The Plum A+ Infusion Systems are a multipurpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care. The Plum A+ family infusion pumps are self contained, microprocessor based infusion devices. The pumps are available in single channel and in triple channel pumping mechanism types.

Class I - Dangerous
🏥 Medical Devices Recalled: February 1, 2013 ICU Infusion Pumps Nationwide

What Should You Do?

  1. Check if you have this product:
    List number: 20792 Serial Number 16061001 16061002 16061003 16061004 16061005 16061006 16061007 16061008 16061009 16061010 16061011 16061012 16061013 16061014 16061015 16061016 16061017 16061018 16061019 16061020 16061021 16061022 16061023 16061025 16061026 16061027 16061028 16061029 16061030 16061031 16061032 16061033 16061034 16061035 16061036 16061037 16061038 16061039 16061040 16061041 16061042 16061043 16061044 16061045 16061046 16061047 16061048 16061049 16061050 16061051 16061052 16061053 16061054 16061055 16061056 16061057 16061058 16061059 16061060 16061061 16061062 16061063 16061064 16061065 16061066 16061067 16061068 16061069 16061070 16061071 16061072 16061073 16061074 16061075 16061076 16061077 16061078 16061079 16061080 16061081 16061082 16061083 16061084 16061085 16061086 16061087 16061088 16061089 16061090 16061091 16061092 16061093 16061094 16061095 16061096 16061097 16061098 16061099 16061100 16061101 16061102 16061103 16061104 16061105 16061106 16061107 16061108 16061109 16061110 16061111 16061112 16061113 16061114 16061115 16061116 16061117 16061118 16061119 16061120 16061121 16061122 16061123 16061124 16061125 16061126 16061127 16061128 16061129 16061130 16061131 16061132 16061133 16061134 16061135 16061136 16061137 16061138 16061139 16061140 16061141 16061142 16061143 16061144 16061145 16061146 16061147 16061148 16061149 16061150 16061151 16061152 16061153 16061154 16061155 16061156 16061157 16061158 16061159 16061160 16061161 16061162 16061163 16061164 16061165 16061166 16061167 16061168 16061169 16061170 16061171 16061172 16061173 16061174 16061175 16061176 16061177 16061178 16061179 16061180 16061181 16061182 16061183 16061184 16061185 16061186 16061187 16061188 16061189 16061190 16061191 16061192 16061193 16061194 16061195 16061196 16061197 16061198 16061199 16061200 16061201 16061202 16061203 16061204 16061205 16061206 16061207 16061208 16061209 16061210 16061211 16061212 16061213 16061214 16061215 16061216 16061217 16061218 16061219 16061220 16061221 16061222 16061223 16061224 16061225 16061226 16061227 16061228 16061229 16061230 16061231 16061232 16061233 16061234 16061235 16061236 16061237 16061238 16061239 16061240 16061241 16061242 16061243 16061244 16061245 16061246 16061248 16061249 16061250 16061251 16061252 16061253 16061254 16061255 16061256 16061257 16061258 16061259 16061260 16061261 16061262 16061263 16061264 16061265 16061266 16061267 16061268 16061269 16061270 16061271 16061272 16061273 16061274 16061275 16061276 16061277 16061278 16061279 16061280 16061281 16061282 16061283 16061284 16061285 16061286 16061287 16061288 16061289 16061290 16061291 16061292 16061293 16061294 16061295 16061296 16061297 16061298 16061299 16061300 16061301 16061302 16061303 16061304 16061305 16061306 16061307 16061308 16061309 16061310 16061311 16061312 16061313 16061314 16061315 16061316 16061317 16061318 16061319 16061320 16061321 16061322 16061323 16061324 16061325 16061326 16061327 16061328 16061329 16061330 16061331 16061332 16061333 16061334 16061335 16061336 16061337 16061338 16061339 16061340 16061341 16061342 16061343 16061344 16061345 16061346 16061347 16061348 16061349 16061350 16061351 16061352 16061353 16061354 16061355 16061356 16061357 16061358 16061359 16061360 16061361 16061362 16061363 16061364 16061365 16061366 16061367 16061368 16061369 16061370 16061371 16061372 16061373 16061374 16061375 16061376 16061377 16061378 16061379 16061380 16061381 16061382 16061383 16061384 16061385 16061386 16061387 16061388 16061389 16061390 16061391 16061392 16061393 16061394 16061395 16061396 16061397 16061398 16061399 16061400 16061401 16061402 16061403 16061404 16061405 16061406 16061407 16061408 16061409 16061410 16061411 16061412 16061413 16061414 16061415 16061416 16061417 16061418 16061419 16061420 16061421 16061422 16061423 16061424 16061425 16061426 16061427 16061428 16061429 16061430 16061431 16061432 16061433 16061434 16061435 16061436 16061437 16061438 16061439 16061440 16061441 16061442 16061443 16061444 16061445 16061446 16061447 16061448 16061449 16061450 16061451 16061452 16061453 16061454 16061455 16061456 16061457 16061458 16061459 16061460 16061461 16061462 16061463 16061464 16061465 16061466 16061467 16061468 16061469 16061470 16061471 16061472 16061473 16061474 16061475 16061476 16061477 16061478 16061479 16061480 16061481 16061482 16061483 16061484 16061485 16061486 16061487 16061488 16061489 16061490 16061491 16061492 16061493 16061494 16061495 16061496 16061497 16061498 16061499 16061500 16061501 16061502 16061503 16061504 16061505 16061506 16061507 16061508 16061509 16061510 16061511 16061512 16061513 16061514 16061515 16061516 16061517 16061518 16061519 16061520 16061521 16061522 16061523 16061524 16061525 16061526 16061527 16061528 16061529 16061530 16061531 16061532 16061533 16061534 16061535 16061536 16061537 16061538 16061539 16061540 16061541 16061542 16061543 16061544 16061545 16061546 16061547 16061548 16061549 16061550 16061551 16061552 16061553 16061554 16061556 16061557 16061558 16061559 16061560 16061561 16061562 16061563 16061564 16061566 16061567 16061568 16061569 16061570 16061571 16061572 16061573 16061574 16061575 16061576 16061577 16061578 16061579 16061580 16061581 16061582 16061583 16061584 16061585 16061586 16061587 16061588 16061589 16061591 16061592 16061593 16061594 16061596 16061597 16061598 16061599 16061600 16061601 16061602 16061603 16061604 16061605 16061606 16061607 16061608 16061609 16061610 16061611 16061612 16061613 16061614 16061615 16061616 16061617 16061618 16061619 16061620 16061621 16061622 16061623 16061624 16061625 16061626 16061627 16061628 16061630 16061631 16061632 16061633 16061634 16061635 16061636 16061637 16061638 16061639 16061640 16061641 16061642 16061643 16061644 16061645 16061646 16061647 16061648 16061649 16061650 16061651 16061652 16061653 16061654 16061655 16061656 16061657 16061658 16061659 16061660 16061661 16061662 16061663 16061664 16061665 16061666 16061667 16061668 16061669 16061670 16061671 16061672 16061673 16061674 16061675 16061676 16061677 16061678 16061679 16061680 16061681 16061682 16061683 16061684 16061685 16061687 16061688 16061689 16061690 16061691 16061692 16061693 16061694 16061695 16061696 16061698 16061699 16061700 16061701 16061702 16061703 16061704 16061705 16061706 16061707 16061708 16061709 16061710 16061711 16061712 16061713 16061714 16061715 16061716 16061717 16061718 16061719 16061720 16061721 16061722 16061723 16061724 16061725 16061726 16061727 16061728 16061729 16061730 16061731 16061732 16061733 16061734 16061735 16061736 16061737 16061738 16061739 16061740 16061741 16061742 16061743 16061744 16061745 16061746 16061747 16061748 16061749 16061750 16061751 16061752 16061753 16061754 16061755 16061756 16061757 16061758 16061759 16061760 16061761 16061762 16061763 16061764 16061765 16061766 16061767 16061768 16061769 16061770 16061771 16061772 16061773 16061774 16061775 16061776 16061777 16061778 16061779 16061780 16061781 16061782 16061783 16061784 16061785 16061786 16061787 16061788 16061789 16061790 16061791 16061792 16061793 16061794 16061795 16061796 16061797 16061798 16061799 16061800 16061801 16061802 16061803 16061804 16061805 16061806 16061808 16061809 16061810 16061811 16061812 16061813 16061814 16061815 16061816 16061817 16061818 16061819 16061820 16061821 16061822 16061823 16061824 16061825 16061826 16061827 16061828 16061829 16061830 16061831 16061832 16061833 16061834 16061835 16061836 16061837 16061838 16061839 16061840 16061841 16061842 16061843 16061844 16061845 16061846 16061847 16061848 16061849 16061850 16061851 16061852 16061853 16061854 16061855 16061856 16061857 16061858 16061859 16061860 16061861 16061862 16061863 16061864 16061865 16061866 16061867 16061868 16061869 16061870 16061871 16061872 16061873 16061874 16061875 16061877 16061878 16061879 16061880 16061881 16061882 16061883 16061884 16061885 16061886 16061887 16061888 16061889 16061890 16061891 16061892 16061894 16061895 16061896 16061897 16061898 16061899 16061900 16061901 16061902 16061903 16061904 16061905 16061906 16061907 16061908 16061909 16061910 16061911 16061912 16061913 16061914 16061915 16061916 16061917 16061918 16061919 16061920 16061921 16061922 16061923 16061924 16061925 16061926 16061927 16061928 16061929 16061930 16061931 16061932 16061933 16061934 16061935 16061936 16061937 16061938 16061939 16061940 16061941 16061942 16061943 16061944 16061945 16061946 16061947 16061948 16061949 16061951 16061952 16061953 16061954 16061955 16061956 16061957 16061958 16061959 16061960 16061961 16061962 16061963 16061964 16061965 16061966 16061967 16061968 16061969 16061970 16061971 16061972 16061973 16061974 16061975 16061976 16061977 16061978 16061979 16061980 16061981 16061982 16061983 16061984 16061985 16061986 16061987 16061988 16061989 16061990 16061991 16061992 16061993 16061994 16061995 16061996 16061997 16061999 16062000 16062001 16062002 16062003 16062004 16062005 16062006 16062007 16062008 16062009 16062010 16062011 16062012 16062014 16062015 16062016 16062017 16062018 16062019 16062020 16062021 16062022 16062023 16062024 16062025 16062026 16062027 16062029 16062030 16062031 16062032 16062033 16062034 16062035 16062036 16062037 16062038 16062039 16062041 16062042 16062043 16062044 16062045 16062046 16062047 16062048 16062050 16062051 16062052 16062053 16062054 16062055 16062056 16062057 16062058 16062059 16062060 16062061 16062062 16062063 16062064 16062065 16062066 16062067 16062068 16062069 16062070 16062071 16062072 16062073 16062074 16062075 16062076 16062077 16062078 16062079 16062080 16062081 16062082 16062083 16062084 16062085 16062086 16062087 16062088 16062089 16062090 16062091 16062092 16062093 16062094 16062095 16062096 16062097 16062098 16062099 16062100 16062101 16062102 16062103 16062104 16062105 16062106 16062107 16062108 16062109 16062110 16062111 16062112 16062113 16062114 16062116 16062117 16062118 16062119 16062120 16062121 16062122 16062123 16062124 16062125 16062126 16062127 16062128 16062129 16062130 16062131 16062132 16062133 16062134 16062135 16062136 16062137 16062138 16062139 16062140 16062141 16062142 16062143 16062144 16062145 16062146 16062147 16062148 16062149 16062150 16062151 16062152 16062153 16062154 16062155 16062156 16062157 16062158 16062159 16062160 16062161 16062162 16062163 16062164 16062165 16062166 16062167 16062168 16062169 16062170 16062171 16062172 16062173 16062174 16062175 16062176 16062177 16062178 16062179 16062180 16062181 16062183 16062184 16062185 16062186 16062187 16062188 16062189 16062190 16062191 16062192 16062193 16062194 16062195 16062196 16062197 16062198 16062199 16062200 16062201 16062202 16062203 16062204 16062205 16062206 16062211 16062212 16062213 16062214 16062215 16062216 16062217 16062218 16062219 16062220 16062221 16062222 16062223 16062224 16062225 16062226 16062227 16062228 16062229 16062230 16062356 16062357 16062358 16062359 16062360 16062361 16062362 16062363 16062364 16062365 16062366 16062367 16062368 16062370 16062371 16062372 16062373 16062374 16062375 16062376 16062377 16062378 16062379 16062380 16062381 16062382 16062383 16062384 16062385 16062387 16062388 16062389 16062390 16062391 16062392 16062393 16062394 16062395 16062396 16062397 16062398 16062399 16062400 16062401 16062402 16062403 16062404 16062405 16062406 16062407 16062408 16062409 16062410 16062411 16062412 16062413 16062414 16062415 16062416 16062417 16062418 16062419 16062420 16062421 16062422 16062423 16062424 16062425 16062426 16062427 16062428 16062429 16062430 16062431 16062432 16062433 16062434 16062435 16062441 16062442 16062443 16062444 16062445 16062446 16062447 16062448 16062449 16062450 16062451 16062452 16062453 16062454 16062455 16062456 16062457 16062458 16062459 16062460 16062461 16062462 16062463 16062464 16062465 16062466 16062467 16062468 16062469 16062470 16062481 16062482 16062483 16062484 16062485 16062486 16062487 16062488 16062489 16062490 16062491 16062492 16062493 16062494 16062495 16062496 16062497 16062498 16062499 16062500 16062501 16062502 16062503 16062504 16062505 16062506 16062507 16062508 16062509 16062510 16062511 16062512 16062513 16062514 16062515 16062516 16062517 16062518 16062519 16062520 16062521 16062522 16062523 16062524 16062525 16062526 16062527 16062528 16062529 16062530 16062531 16062532 16062533 16062534 16062535 16062536 16062537 16062538 16062539 16062540 16062541 16062542 16062543 16062544 16062545 16062546 16062547 16062548 16062549 16062550 16062551 16062552 16062553 16062554 16062555 16062556 16062557 16062558 16062559 16062560 16062561 16062562 16062563 16062564 16062565 16062566 16062567 16062568 16062569 16062570 16062571 16062572 16062573 16062574 16062575 16062576 16062577 16062578 16062579 16062580 16062581 16062582 16062583 16062584 16062585 16062586 16062587 16062588 16062589 16062590 16062591 16062592 16062593 16062594 16062595 16062596 16062597 16062598 16062599 16062600 16062601 16062602 16062604 16062605 16062606 16062607 16062608 16062609 16062610 16062611 16062612 16062613 16062614 16062615 16062616 16062617 16062618 16062619 16062620 16062621 16062622 16062623 16062624 16062625 16062626 16062627 16062628 16062629 16062630 16062631 16062632 16062633 16062634 16062635 16062636 16062637 16062638 16062639 16062640 16062641 16062642 16062643 16062644 16062645 16062646 16062647 16062648 16062649 16062650 16062651 16062652 16062653 16062654 16062655 16062656 16062657 16062658 16062659 16062660 16062661 16062662 16062663 16062664 16062665 16062666 16062667 16062668 16062669 16062670 16062671 16062672 16062673 16062674 16062675 16062676 16062677 16062678 16062679 16062680 16062681 16062682 16062683 16062684 16062685 16062686 16062687 16062688 16062689 16062690 16062691 16062692 16062693 16062694 16062695 16062696 16062697 16062698 16062699 16062700 16062701 16062702 16062703 16062704 16062705 16062707 16062708 16062709 16062710 16062711 16062712 16062713 16062714 16062715 16062716 16062717 16062718 16062719 16062720 16062721 16062722 16062723 16062724 16062725 16062726 16062727 16062728 16062729 16062730 16062731 16062732 16062733 16062734 16062735 16062736 16062738 16062739 16062740 16062741 16062742 16062743 16062744 16062745 16062746 16062747 16062748 16062749 16062750 16062751 16062752 16062753 16062754 16062755 16062756 16062757 16062758 16062759 16062760 16062761 16062762 16062763 16062764 16062765 16062766 16062767 16062768 16062769 16062770 16062771 16062772 16062773 16062774 16062775 16062776 16062777 16062778 16062779 16062780 16062781 16062782 16062783 16062784 16062785 16062786 16062787 16062788 16062789 16062790 16062791 16062792 16062793 16062794 16062795 16062796 16062797 16062798 16062799 16062800 16062801 16062802 16062803 16062804 16062805 16062806 16062807 16062808 16062809 16062810 16062811 16062812 16062813 16062814 16062815 16062816 16062817 16062818 16062819 16062820 16062821 16062822 16062823 16062824 16062825 16062826 16062827 16062828 16062829 16062830 16062831 16062832 16062833 16062834 16062835 16062836 16062837 16062838 16062839 16062840 16062841 16062842 16062843 16062844 16062845 16062846 16062847 16062848 16062849 16062850 16062851 16062852 16062853 16062854 16062855 16062856 16062857 16062858 16062859 16062860 16062861 16062862 16062863 16062864 16062865 16062866 16062867 16062868 16062869 16062870 16062871 16062872 16062873 16062874 16062875 16062876 16062877 16062878 16062879 16062880 16062881 16062882 16062883 16062884 16062885 16062886 16062887 16062888 16062889 16062890 16062891 16062892 16062893 16062894 16062895 16062896 16062897 16062898 16062899 16062900 16062901 16062902 16062903 16062904 16062905 16062906 16062907 16062908 16062909 16062910 16062911 16062912 16062913 16062914 16062916 16062917 16062918 16062919 16062920 16062921 16062922 16062923 16062925 16062926 16062927 16062928 16062929 16062930 16062931 16062932 16062933 16062934 16062935 16062936 16062937 16062938 16062939 16062940 16062941 16062942 16062943 16062944 16062945 16062946 16062947 16062948 16062949 16062950 16062951 16062952 16062953 16062954 16062955 16062956 16062957 16062958 16062959 16062960 16062961 16062962 16062963 16062964 16062965 16062966 16062967 16062968 16062969 16062970 16062971 16062972 16062973 16062974 16062975 16062976 16062977 16062978 16062979 16062980 16062981 16062982 16062983 16062984 16062985 16062986 16062987 16062988 16062989 16062990 16062991 16062992 16062993 16062994 16062995 16062996 16062997 16062998 16062999 16063000 16063001 16063002 16063003 16063004 16063005 16063006 16063007 16063008 16063009 16063010 16063011 16063012 16063013 16063014 16063015 16063016 16063017 16063018 16063019 16063020 16063021 16063022 16063024 16063025 16063026 16063027 16063028 16063029 16063030 16063031 16063032 16063033 16063034 16063035 16063036 16063037 16063038 16063039 16063040 16063041 16063042 16063043 16063044 16063045 16063046 16063047 16063048 16063049 16063050 16063051 16063052 16063053 16063054 16063055 16063056 16063057 16063058 16063059 16063060 16063061 16063062 16063063 16063064 16063065 16063066 16063067 16063068 16063069 16063070 16063071 16063072 16063073 16063074 16063075 16063076 16063077 16063078 16063079 16063081 16063082 16063083 16063084 16063085 16063086 16063087 16063088 16063089 16063090 16063091 16063092 16063093 16063094 16063095 16063096 16063097 16063098 16063099 16063100 16063101 16063102 16063103 16063104 16063105 16063106 16063107 16063108 16063109 16063110 16063111 16063112 16063113 16063114 16063115 16063116 16063117 16063118 16063119 16063120 16063121 16063122 16063123 16063124 16063125 16063126 16063127 16063128 16063129 16063130 16063131 16063132 16063133 16063134 16063135 16063136 16063137 16063138 16063139 16063140 16063141 16063142 16063143 16063144 16063145 16063146 16063147 16063148 16063149 16063150 16063151 16063152 16063153 16063154 16063155 16063156 16063157 16063158 16063159 16063160 16063161 16063162 16063163 16063164 16063165 16063166 16063167 16063168 16063169 16063170 16063171 16063172 16063173 16063174 16063175 16063176 16063177 16063178 16063179 16063180 16063181 16063182 16063183 16063184 16063185 16063186 16063187 16063188 16063189 16063190 16063191 16063192 16063193 16063194 16063195 16063196 16063197 16063198 16063199 16063200 16063201 16063202 16063203 16063204 16063205 16063206 16063207 16063208 16063209 16063210 16063211 16063212 16063213 16063214 16063215 16063216 16063217 16063218 16063219 16063220 16063221 16063222 16063223 16063224 16063225 16063226 16063227 16063228 16063229 16063230 16063231 16063232 16063233 16063234 16063235 16063236 16063237 16063239 16063240 16063241 16063242 16063243 16063244 16063245 16063246 16063247 16063248 16063249 16063250 16063251 16063252 16063253 16063254 16063255 16063256 16063257 16063258 16063259 16063260 16063261 16063262 16063263 16063264 16063265 16063266 16063267 16063268 16063269 16063270 16063271 16063272 16063273 16063274 16063275 16063276 16063277 16063278 16063279 16063281 16063282 16063283 16063284 16063285 16063286 16063287 16063288 16063289 16063290 16063291 16063293 16063294 16063295 16063296 16063297 16063298 16063299 16063300 16063301 16063302 16063303 16063304 16063305 16063306 16063307 16063308 16063309 16063310 16063311 16063312 16063313 16063314 16063315 16063316 16063317 16063318 16063319 16063320 16063321 16063322 16063323 16063324 16063325 16063326 16063327 16063328 16063329 16063330 16063331 16063332 16063333 16063334 16063335 16063336 16063337 16063338 16063339 16063340 16063341 16063342 16063343 16063344 16063345 16063346 16063347 16063348 16063349 16063350 16063351 16063352 16063353 16063354 16063355 16063356 16063357 16063358 16063359 16063360 16063361 16063362 16063363 16063364 16063365 16063366 16063367 16063368 16063369 16063370 16063371 16063372 16063373 16063374 16063375 16063376 16063377 16063378 16063379 16063380 16063381 16063382 16063383 16063384 16063385 16063386 16063387 16063388 16063389 16063390 16063391 16063392 16063393 16063394 16063395 16063396 16063397 16063398 16063399 16063400 16063401 16063402 16063403 16063404 16063405 16063406 16063407 16063408 16063409 16063410 16063411 16063412 16063413 16063414 16063415 16063416 16063417 16063418 16063419 16063420 16063421 16063422 16063423 16063425 16063426 16063427 16063428 16063429 16063430 16063431 16063432 16063433 16063434 16063435 16063436 16063437 16063438 16063439 16063440 16063441 16063442 16063443 16063444 16063445 16063446 16063447 16063448 16063449 16063450 16063451 16063452 16063453 16063454 16063455 16063456 16063457 16063458 16063459 16063460 16063461 16063462 16063463 16063464 16063465 16063466 16063467 16063468 16063469 16063470 16063471 16063472 16063473 16063474 16063475 16063476 16063477 16063478 16063479 16063480 16063481 16063482 16063483 16063484 16063485 16063486 16063487 16063488 16063489 16063490 16063491 16063492 16063493 16063494 16063495 16063496 16063497 16063498 16063499 16063500 16063501 16063502 16063503 16063504 16063505 16063506 16063507 16063508 16063509 16063510 16063511 16063512 16063513 16063514 16063515 16063516 16063517 16063518 16063519 16063520 16063521 16063522 16063523 16063524 16063525 16063526 16063527 16063528 16063529 16063530 16063531 16063532 16063533 16063534 16063535 16063536 16063537 16063538 16063539 16063540 16063541 16063542 16063543 16063544 16063545 16063546 16063547 16063548 16063549 16063550 16063551 16063552 16063553 16063554 16063555 16063556 16063557 16063558 16063559 16063560 16063561 16063562 16063563 16063564 16063565 16063566 16063567 16063568 16063569 16063570 16063571 16063572 16063573 16063574 16063575 16063576 16063577 16063578 16063579 16063580 16063581 16063582 16063583 16063584 16063585 16063586 16063587 16063588 16063589 16063590 16063591 16063592 16063593 16063594 16063595 16063596 16063597 16063598 16063599 16063600 16063601 16063602 16063603 16063604 16063605 16063606 16063607 16063608 16063609 16063610 16063611 16063612 16063613 16063614 16063615 16063616 16063617 16063618 16063619 16063620 16063621 16063622 16063623 16063624 16063625 16063626 16063627 16063628 16063629 16063630 16063631 16063632 16063633 16063634 16063635 16063636 16063637 16063638 16063639 16063640 16063641 16063642 16063643 16063644 16063645 16063646 16063647 16063648 16063649 16063650 16063651 16063652 16063653 16063654 16063655 16063656 16063657 16063658 16063659 16063660 16063662 16063663 16063664 16063665 16063666 16063667 16063668 16063669 16063670 16063671 16063672 16063673 16063674 16063675 16063676 16063677 16063678 16063679 16063680 16063681 16063682 16063683 16063684 16063685 16063686 16063687 16063688 16063689 16063690 16063691 16063692 16063693 16063694 16063695 16063696 16063697 16063698 16063699 16063700 16063701 16063702 16063703 16063704 16063705 16063706 16063707 16063708 16063709 16063710 16063711 16063712 16063713 16063714 16063715 16063716 16063717 16063718 16063719 16063720 16063721 16063722 16063723 16063724 16063725 16063726 16063727 16063728 16063729 16063730 16063731 16063732 16063733 16063734 16063735 16063736 16063737 16063738 16063739 16063740 16063741 16063742 16063743 16063744 16063745 16063746 16063747 16063748 16063749 16063750 16063751 16063752 16063753 16063754 16063755 16063756 16063757 16063758 16063759 16063760 16063761 16063762 16063763 16063764 16063765 16063766 16063767 16063768 16063769 16063770 16063771 16063772 16063773 16063774 16063775 16063776 16063777 16063778 16063779 16063780 16063781 16063782 16063783 16063784 16063785 16063786 16063787 16063788 16063789 16063790 16063791 16063792 16063793 16063794 16063795 16063796 16063797 16063798 16063800 16063801 16063802 16063803 16063804 16063805 16063806 16063807 16063808 16063809 16063810 16063811 16063812 16063813 16063814 16063815 16063816 16063817 16063818 16063819 16063820 16063821 16063822 16063823 16063824 16063825 16063826 16063827 16063828 16063829 16063830 16063831 16063832 16063833 16063834 16063835 16063836 16063837 16063838 16063840 16063841 16063842 16063843 16063844 16063845 16063846 16063847 16063848 16063849 16063850 16063851 16063852 16063853 16063854 16063855 16063856 16063857 16063858 16063859 16063860 16063861 16063862 16063863 16063864 16063865 16063866 16063867 16063868 16063869 16063870 16063871 16063872 16063873 16063874 16063875 16063876 16063877 16063878 16063879 16063880 16063881 16063882 16063883 16063884 16063885 16063886 16063887 16063888 16063889 16063890 16063891 16063892 16063893 16063894 16063895 16063896 16063897 16063899 16063900 16063901 16063902 16063903 16063904 16063905 16063906 16063907 16063908 16063909 16063910 16063911 16063912 16063913 16063914 16063915 16063916 16063917 16063918 16063919 16063920 16063921 16063922 16063923 16063924 16063925 16063927 16063928 16063929 16063930 16063931 16063932 16063933 16063934 16063935 16063936 16063937 16063938 16063939 16063940 16063941 16063942 16063943 16063944 16063945 16063946 16063947 16063948 16063949 16063950 16063951 16063952 16063953 16063954 16063955 16063956 16063957 16063958 16063959 16063960 16063962 16063963 16063964 16063965 16063966 16063967 16063968 16063969 16063970 16063971 16063972 16063973 16063974 16063975 16063976 16063977 16063978 16063979 16063980 16063982 16063983 16063984 16063985 16063986 16063987 16063988 16063989 16063990 16063991 16063992 16063993 16063994 16063995 16063996 16063997 16063998 16063999 16064000 16064001 16064002 16064003 16064004 16064005 16064006 16064007 16064008 16064009 16064010 16064011 16064014 16064015 16064016 16064017 16064018 16064019 16064020 16064021 16064022 16064023 16064024 16064025 16064026 16064027 16064028 16064029 16064030 16064031 16064032 16064033 16064034 16064035 16064037 16064038 16064039 16064040 16064041 16064042 16064043 16064044 16064045 16064046 16064047 16064048 16064049 16064050 16064051 16064052 16064053 16064054 16064055 16064056 16064057 16064058 16064059 16064060 16064061 16064062 16064063 16064064 16064065 16064066 16064067 16064068 16064069 16064070 16064071 16064072 16064073 16064074 16064075 16064076 16064077 16064078 16064079 16064080 16064081 16064082 16064083 16064084 16064085 16064086 16064087 16064088 16064089 16064090 16064091 16064092 16064093 16064094 16064095 16064096 16064097 16064098 16064099 16064101 16064102 16064103 16064104 16064105 16064106 16064107 16064108 16064109 16064110 16064111 16064112 16064113 16064114 16064115 16064116 16064117 16064118 16064119 16064120 16064121 16064122 16064123 16064124 16064125 16064126 16064127 16064128 16064129 16064130 16064131 16064132 16064134 16064135 16064136 16064138 16064139 16064141 16064142 16064143 16064144 16064145 16064146 16064148 16064149 16064150 16064151 16064152 16064153 16064154 16064155 16064156 16064157 16064158 16064159 16064160 16064161 16064163 16064164 16064165 16064166 16064167 16064168 16064169 16064170 16064171 16064172 16064173 16064174 16064175 16064176 16064177 16064178 16064179 16064180 16064181 16064182 16064183 16064184 16064185 16064186 16064187 16064188 16064189 16064190 16064191 16064192 16064193 16064195 16064196 16064197 16064198 16064199 16064200 16064201 16064202 16064203 16064204 16064205 16064206 16064207 16064208 16064209 16064210 16064211 16064212 16064213 16064214 16064215 16064216 16064217 16064218 16064219 16064220 16064221 16064222 16064223 16064224 16064225 16064226 16064227 16064228 16064229 16064230 16064231 16064232 16064233 16064234 16064235 16064236 16064238 16064239 16064240 16064241 16064242 16064243 16064244 16064245 16064246 16064247 16064248 16064249 16064250 16064251 16064252 16064253 16064254 16064255 16064256 16064257 16064258 16064259 16064260 16064261 16064262 16064263 16064264 16064265 16064266 16064267 16064268 16064269 16064270 16064271 16064272 16064273 16064274 16064275 16064276 16064277 16064278 16064279 16064280 16064281 16064282 16064283 16064284 16064286 16064287 16064288 16064289 16064290 16064291 16064292 16064293 16064294 16064295 16064296 16064297 16064298 16064301 16064302 16064303 16064304 16064305 16064306 16064307 16064308 16064309 16064310 16064311 16064312 16064314 16064315 16064316 16064317 16064319 16064320 16064321 16064322 16064323 16064324 16064325 16064326 16064327 16064328 16064329 16064330 16064331 16064332 16064333 16064334 16064335 16064336 16064337 16064338 16064339 16064340 16064341 16064342 16064343 16064344 16064345 16064346 16064347 16064348 16064349 16064350 16064351 16064352 16064353 16064354 16064355 16064356 16064357 16064358 16064359 16064360 16064361 16064362 16064363 16064364 16064365 16064366 16064367 16064368 16064369 16064370 16064372 16064373 16064374 16064375 16064376 16064377 16064378 16064379 16064380 16064381 16064382 16064383 16064384 16064385 16064387 16064388 16064389 16064390 16064391 16064392 16064393 16064394 16064395 16064396 16064397 16064398 16064399 16064400 16064401 16064402 16064403 16064404 16064405 16064406 16064407 16064408 16064409 16064410 16064411 16064412 16064413 16064414 16064415 16064416 16064417 16064418 16064419 16064420 16064421 16064422 16064424 16064425 16064426 16064427 16064428 16064429 16064430 16064431 16064432 16064433 16064434 16064435 16064436 16064437 16064438 16064439 16064440 16064441 16064442 16064443 16064444 16064445 16064446 16064447 16064448 16064449 16064450 16064452 16064453 16064454 16064455 16064456 16064457 16064458 16064459 16064460 16064
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
ICU
Reason for Recall:
A possible interruption of therapy when an E321 error code occurs while using the infuser, which is caused by battery issues.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

The Plum A+ Infusion Systems are a multipurpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care. The Plum A+ family infusion pumps are self contained, microprocessor based infusion devices. The pumps are available in single channel and in triple channel pumping mechanism types.

Product Codes/Lot Numbers:

List number: 20792 Serial Number 16061001 16061002 16061003 16061004 16061005 16061006 16061007 16061008 16061009 16061010 16061011 16061012 16061013 16061014 16061015 16061016 16061017 16061018 16061019 16061020 16061021 16061022 16061023 16061025 16061026 16061027 16061028 16061029 16061030 16061031 16061032 16061033 16061034 16061035 16061036 16061037 16061038 16061039 16061040 16061041 16061042 16061043 16061044 16061045 16061046 16061047 16061048 16061049 16061050 16061051 16061052 16061053 16061054 16061055 16061056 16061057 16061058 16061059 16061060 16061061 16061062 16061063 16061064 16061065 16061066 16061067 16061068 16061069 16061070 16061071 16061072 16061073 16061074 16061075 16061076 16061077 16061078 16061079 16061080 16061081 16061082 16061083 16061084 16061085 16061086 16061087 16061088 16061089 16061090 16061091 16061092 16061093 16061094 16061095 16061096 16061097 16061098 16061099 16061100 16061101 16061102 16061103 16061104 16061105 16061106 16061107 16061108 16061109 16061110 16061111 16061112 16061113 16061114 16061115 16061116 16061117 16061118 16061119 16061120 16061121 16061122 16061123 16061124 16061125 16061126 16061127 16061128 16061129 16061130 16061131 16061132 16061133 16061134 16061135 16061136 16061137 16061138 16061139 16061140 16061141 16061142 16061143 16061144 16061145 16061146 16061147 16061148 16061149 16061150 16061151 16061152 16061153 16061154 16061155 16061156 16061157 16061158 16061159 16061160 16061161 16061162 16061163 16061164 16061165 16061166 16061167 16061168 16061169 16061170 16061171 16061172 16061173 16061174 16061175 16061176 16061177 16061178 16061179 16061180 16061181 16061182 16061183 16061184 16061185 16061186 16061187 16061188 16061189 16061190 16061191 16061192 16061193 16061194 16061195 16061196 16061197 16061198 16061199 16061200 16061201 16061202 16061203 16061204 16061205 16061206 16061207 16061208 16061209 16061210 16061211 16061212 16061213 16061214 16061215 16061216 16061217 16061218 16061219 16061220 16061221 16061222 16061223 16061224 16061225 16061226 16061227 16061228 16061229 16061230 16061231 16061232 16061233 16061234 16061235 16061236 16061237 16061238 16061239 16061240 16061241 16061242 16061243 16061244 16061245 16061246 16061248 16061249 16061250 16061251 16061252 16061253 16061254 16061255 16061256 16061257 16061258 16061259 16061260 16061261 16061262 16061263 16061264 16061265 16061266 16061267 16061268 16061269 16061270 16061271 16061272 16061273 16061274 16061275 16061276 16061277 16061278 16061279 16061280 16061281 16061282 16061283 16061284 16061285 16061286 16061287 16061288 16061289 16061290 16061291 16061292 16061293 16061294 16061295 16061296 16061297 16061298 16061299 16061300 16061301 16061302 16061303 16061304 16061305 16061306 16061307 16061308 16061309 16061310 16061311 16061312 16061313 16061314 16061315 16061316 16061317 16061318 16061319 16061320 16061321 16061322 16061323 16061324 16061325 16061326 16061327 16061328 16061329 16061330 16061331 16061332 16061333 16061334 16061335 16061336 16061337 16061338 16061339 16061340 16061341 16061342 16061343 16061344 16061345 16061346 16061347 16061348 16061349 16061350 16061351 16061352 16061353 16061354 16061355 16061356 16061357 16061358 16061359 16061360 16061361 16061362 16061363 16061364 16061365 16061366 16061367 16061368 16061369 16061370 16061371 16061372 16061373 16061374 16061375 16061376 16061377 16061378 16061379 16061380 16061381 16061382 16061383 16061384 16061385 16061386 16061387 16061388 16061389 16061390 16061391 16061392 16061393 16061394 16061395 16061396 16061397 16061398 16061399 16061400 16061401 16061402 16061403 16061404 16061405 16061406 16061407 16061408 16061409 16061410 16061411 16061412 16061413 16061414 16061415 16061416 16061417 16061418 16061419 16061420 16061421 16061422 16061423 16061424 16061425 16061426 16061427 16061428 16061429 16061430 16061431 16061432 16061433 16061434 16061435 16061436 16061437 16061438 16061439 16061440 16061441 16061442 16061443 16061444 16061445 16061446 16061447 16061448 16061449 16061450 16061451 16061452 16061453 16061454 16061455 16061456 16061457 16061458 16061459 16061460 16061461 16061462 16061463 16061464 16061465 16061466 16061467 16061468 16061469 16061470 16061471 16061472 16061473 16061474 16061475 16061476 16061477 16061478 16061479 16061480 16061481 16061482 16061483 16061484 16061485 16061486 16061487 16061488 16061489 16061490 16061491 16061492 16061493 16061494 16061495 16061496 16061497 16061498 16061499 16061500 16061501 16061502 16061503 16061504 16061505 16061506 16061507 16061508 16061509 16061510 16061511 16061512 16061513 16061514 16061515 16061516 16061517 16061518 16061519 16061520 16061521 16061522 16061523 16061524 16061525 16061526 16061527 16061528 16061529 16061530 16061531 16061532 16061533 16061534 16061535 16061536 16061537 16061538 16061539 16061540 16061541 16061542 16061543 16061544 16061545 16061546 16061547 16061548 16061549 16061550 16061551 16061552 16061553 16061554 16061556 16061557 16061558 16061559 16061560 16061561 16061562 16061563 16061564 16061566 16061567 16061568 16061569 16061570 16061571 16061572 16061573 16061574 16061575 16061576 16061577 16061578 16061579 16061580 16061581 16061582 16061583 16061584 16061585 16061586 16061587 16061588 16061589 16061591 16061592 16061593 16061594 16061596 16061597 16061598 16061599 16061600 16061601 16061602 16061603 16061604 16061605 16061606 16061607 16061608 16061609 16061610 16061611 16061612 16061613 16061614 16061615 16061616 16061617 16061618 16061619 16061620 16061621 16061622 16061623 16061624 16061625 16061626 16061627 16061628 16061630 16061631 16061632 16061633 16061634 16061635 16061636 16061637 16061638 16061639 16061640 16061641 16061642 16061643 16061644 16061645 16061646 16061647 16061648 16061649 16061650 16061651 16061652 16061653 16061654 16061655 16061656 16061657 16061658 16061659 16061660 16061661 16061662 16061663 16061664 16061665 16061666 16061667 16061668 16061669 16061670 16061671 16061672 16061673 16061674 16061675 16061676 16061677 16061678 16061679 16061680 16061681 16061682 16061683 16061684 16061685 16061687 16061688 16061689 16061690 16061691 16061692 16061693 16061694 16061695 16061696 16061698 16061699 16061700 16061701 16061702 16061703 16061704 16061705 16061706 16061707 16061708 16061709 16061710 16061711 16061712 16061713 16061714 16061715 16061716 16061717 16061718 16061719 16061720 16061721 16061722 16061723 16061724 16061725 16061726 16061727 16061728 16061729 16061730 16061731 16061732 16061733 16061734 16061735 16061736 16061737 16061738 16061739 16061740 16061741 16061742 16061743 16061744 16061745 16061746 16061747 16061748 16061749 16061750 16061751 16061752 16061753 16061754 16061755 16061756 16061757 16061758 16061759 16061760 16061761 16061762 16061763 16061764 16061765 16061766 16061767 16061768 16061769 16061770 16061771 16061772 16061773 16061774 16061775 16061776 16061777 16061778 16061779 16061780 16061781 16061782 16061783 16061784 16061785 16061786 16061787 16061788 16061789 16061790 16061791 16061792 16061793 16061794 16061795 16061796 16061797 16061798 16061799 16061800 16061801 16061802 16061803 16061804 16061805 16061806 16061808 16061809 16061810 16061811 16061812 16061813 16061814 16061815 16061816 16061817 16061818 16061819 16061820 16061821 16061822 16061823 16061824 16061825 16061826 16061827 16061828 16061829 16061830 16061831 16061832 16061833 16061834 16061835 16061836 16061837 16061838 16061839 16061840 16061841 16061842 16061843 16061844 16061845 16061846 16061847 16061848 16061849 16061850 16061851 16061852 16061853 16061854 16061855 16061856 16061857 16061858 16061859 16061860 16061861 16061862 16061863 16061864 16061865 16061866 16061867 16061868 16061869 16061870 16061871 16061872 16061873 16061874 16061875 16061877 16061878 16061879 16061880 16061881 16061882 16061883 16061884 16061885 16061886 16061887 16061888 16061889 16061890 16061891 16061892 16061894 16061895 16061896 16061897 16061898 16061899 16061900 16061901 16061902 16061903 16061904 16061905 16061906 16061907 16061908 16061909 16061910 16061911 16061912 16061913 16061914 16061915 16061916 16061917 16061918 16061919 16061920 16061921 16061922 16061923 16061924 16061925 16061926 16061927 16061928 16061929 16061930 16061931 16061932 16061933 16061934 16061935 16061936 16061937 16061938 16061939 16061940 16061941 16061942 16061943 16061944 16061945 16061946 16061947 16061948 16061949 16061951 16061952 16061953 16061954 16061955 16061956 16061957 16061958 16061959 16061960 16061961 16061962 16061963 16061964 16061965 16061966 16061967 16061968 16061969 16061970 16061971 16061972 16061973 16061974 16061975 16061976 16061977 16061978 16061979 16061980 16061981 16061982 16061983 16061984 16061985 16061986 16061987 16061988 16061989 16061990 16061991 16061992 16061993 16061994 16061995 16061996 16061997 16061999 16062000 16062001 16062002 16062003 16062004 16062005 16062006 16062007 16062008 16062009 16062010 16062011 16062012 16062014 16062015 16062016 16062017 16062018 16062019 16062020 16062021 16062022 16062023 16062024 16062025 16062026 16062027 16062029 16062030 16062031 16062032 16062033 16062034 16062035 16062036 16062037 16062038 16062039 16062041 16062042 16062043 16062044 16062045 16062046 16062047 16062048 16062050 16062051 16062052 16062053 16062054 16062055 16062056 16062057 16062058 16062059 16062060 16062061 16062062 16062063 16062064 16062065 16062066 16062067 16062068 16062069 16062070 16062071 16062072 16062073 16062074 16062075 16062076 16062077 16062078 16062079 16062080 16062081 16062082 16062083 16062084 16062085 16062086 16062087 16062088 16062089 16062090 16062091 16062092 16062093 16062094 16062095 16062096 16062097 16062098 16062099 16062100 16062101 16062102 16062103 16062104 16062105 16062106 16062107 16062108 16062109 16062110 16062111 16062112 16062113 16062114 16062116 16062117 16062118 16062119 16062120 16062121 16062122 16062123 16062124 16062125 16062126 16062127 16062128 16062129 16062130 16062131 16062132 16062133 16062134 16062135 16062136 16062137 16062138 16062139 16062140 16062141 16062142 16062143 16062144 16062145 16062146 16062147 16062148 16062149 16062150 16062151 16062152 16062153 16062154 16062155 16062156 16062157 16062158 16062159 16062160 16062161 16062162 16062163 16062164 16062165 16062166 16062167 16062168 16062169 16062170 16062171 16062172 16062173 16062174 16062175 16062176 16062177 16062178 16062179 16062180 16062181 16062183 16062184 16062185 16062186 16062187 16062188 16062189 16062190 16062191 16062192 16062193 16062194 16062195 16062196 16062197 16062198 16062199 16062200 16062201 16062202 16062203 16062204 16062205 16062206 16062211 16062212 16062213 16062214 16062215 16062216 16062217 16062218 16062219 16062220 16062221 16062222 16062223 16062224 16062225 16062226 16062227 16062228 16062229 16062230 16062356 16062357 16062358 16062359 16062360 16062361 16062362 16062363 16062364 16062365 16062366 16062367 16062368 16062370 16062371 16062372 16062373 16062374 16062375 16062376 16062377 16062378 16062379 16062380 16062381 16062382 16062383 16062384 16062385 16062387 16062388 16062389 16062390 16062391 16062392 16062393 16062394 16062395 16062396 16062397 16062398 16062399 16062400 16062401 16062402 16062403 16062404 16062405 16062406 16062407 16062408 16062409 16062410 16062411 16062412 16062413 16062414 16062415 16062416 16062417 16062418 16062419 16062420 16062421 16062422 16062423 16062424 16062425 16062426 16062427 16062428 16062429 16062430 16062431 16062432 16062433 16062434 16062435 16062441 16062442 16062443 16062444 16062445 16062446 16062447 16062448 16062449 16062450 16062451 16062452 16062453 16062454 16062455 16062456 16062457 16062458 16062459 16062460 16062461 16062462 16062463 16062464 16062465 16062466 16062467 16062468 16062469 16062470 16062481 16062482 16062483 16062484 16062485 16062486 16062487 16062488 16062489 16062490 16062491 16062492 16062493 16062494 16062495 16062496 16062497 16062498 16062499 16062500 16062501 16062502 16062503 16062504 16062505 16062506 16062507 16062508 16062509 16062510 16062511 16062512 16062513 16062514 16062515 16062516 16062517 16062518 16062519 16062520 16062521 16062522 16062523 16062524 16062525 16062526 16062527 16062528 16062529 16062530 16062531 16062532 16062533 16062534 16062535 16062536 16062537 16062538 16062539 16062540 16062541 16062542 16062543 16062544 16062545 16062546 16062547 16062548 16062549 16062550 16062551 16062552 16062553 16062554 16062555 16062556 16062557 16062558 16062559 16062560 16062561 16062562 16062563 16062564 16062565 16062566 16062567 16062568 16062569 16062570 16062571 16062572 16062573 16062574 16062575 16062576 16062577 16062578 16062579 16062580 16062581 16062582 16062583 16062584 16062585 16062586 16062587 16062588 16062589 16062590 16062591 16062592 16062593 16062594 16062595 16062596 16062597 16062598 16062599 16062600 16062601 16062602 16062604 16062605 16062606 16062607 16062608 16062609 16062610 16062611 16062612 16062613 16062614 16062615 16062616 16062617 16062618 16062619 16062620 16062621 16062622 16062623 16062624 16062625 16062626 16062627 16062628 16062629 16062630 16062631 16062632 16062633 16062634 16062635 16062636 16062637 16062638 16062639 16062640 16062641 16062642 16062643 16062644 16062645 16062646 16062647 16062648 16062649 16062650 16062651 16062652 16062653 16062654 16062655 16062656 16062657 16062658 16062659 16062660 16062661 16062662 16062663 16062664 16062665 16062666 16062667 16062668 16062669 16062670 16062671 16062672 16062673 16062674 16062675 16062676 16062677 16062678 16062679 16062680 16062681 16062682 16062683 16062684 16062685 16062686 16062687 16062688 16062689 16062690 16062691 16062692 16062693 16062694 16062695 16062696 16062697 16062698 16062699 16062700 16062701 16062702 16062703 16062704 16062705 16062707 16062708 16062709 16062710 16062711 16062712 16062713 16062714 16062715 16062716 16062717 16062718 16062719 16062720 16062721 16062722 16062723 16062724 16062725 16062726 16062727 16062728 16062729 16062730 16062731 16062732 16062733 16062734 16062735 16062736 16062738 16062739 16062740 16062741 16062742 16062743 16062744 16062745 16062746 16062747 16062748 16062749 16062750 16062751 16062752 16062753 16062754 16062755 16062756 16062757 16062758 16062759 16062760 16062761 16062762 16062763 16062764 16062765 16062766 16062767 16062768 16062769 16062770 16062771 16062772 16062773 16062774 16062775 16062776 16062777 16062778 16062779 16062780 16062781 16062782 16062783 16062784 16062785 16062786 16062787 16062788 16062789 16062790 16062791 16062792 16062793 16062794 16062795 16062796 16062797 16062798 16062799 16062800 16062801 16062802 16062803 16062804 16062805 16062806 16062807 16062808 16062809 16062810 16062811 16062812 16062813 16062814 16062815 16062816 16062817 16062818 16062819 16062820 16062821 16062822 16062823 16062824 16062825 16062826 16062827 16062828 16062829 16062830 16062831 16062832 16062833 16062834 16062835 16062836 16062837 16062838 16062839 16062840 16062841 16062842 16062843 16062844 16062845 16062846 16062847 16062848 16062849 16062850 16062851 16062852 16062853 16062854 16062855 16062856 16062857 16062858 16062859 16062860 16062861 16062862 16062863 16062864 16062865 16062866 16062867 16062868 16062869 16062870 16062871 16062872 16062873 16062874 16062875 16062876 16062877 16062878 16062879 16062880 16062881 16062882 16062883 16062884 16062885 16062886 16062887 16062888 16062889 16062890 16062891 16062892 16062893 16062894 16062895 16062896 16062897 16062898 16062899 16062900 16062901 16062902 16062903 16062904 16062905 16062906 16062907 16062908 16062909 16062910 16062911 16062912 16062913 16062914 16062916 16062917 16062918 16062919 16062920 16062921 16062922 16062923 16062925 16062926 16062927 16062928 16062929 16062930 16062931 16062932 16062933 16062934 16062935 16062936 16062937 16062938 16062939 16062940 16062941 16062942 16062943 16062944 16062945 16062946 16062947 16062948 16062949 16062950 16062951 16062952 16062953 16062954 16062955 16062956 16062957 16062958 16062959 16062960 16062961 16062962 16062963 16062964 16062965 16062966 16062967 16062968 16062969 16062970 16062971 16062972 16062973 16062974 16062975 16062976 16062977 16062978 16062979 16062980 16062981 16062982 16062983 16062984 16062985 16062986 16062987 16062988 16062989 16062990 16062991 16062992 16062993 16062994 16062995 16062996 16062997 16062998 16062999 16063000 16063001 16063002 16063003 16063004 16063005 16063006 16063007 16063008 16063009 16063010 16063011 16063012 16063013 16063014 16063015 16063016 16063017 16063018 16063019 16063020 16063021 16063022 16063024 16063025 16063026 16063027 16063028 16063029 16063030 16063031 16063032 16063033 16063034 16063035 16063036 16063037 16063038 16063039 16063040 16063041 16063042 16063043 16063044 16063045 16063046 16063047 16063048 16063049 16063050 16063051 16063052 16063053 16063054 16063055 16063056 16063057 16063058 16063059 16063060 16063061 16063062 16063063 16063064 16063065 16063066 16063067 16063068 16063069 16063070 16063071 16063072 16063073 16063074 16063075 16063076 16063077 16063078 16063079 16063081 16063082 16063083 16063084 16063085 16063086 16063087 16063088 16063089 16063090 16063091 16063092 16063093 16063094 16063095 16063096 16063097 16063098 16063099 16063100 16063101 16063102 16063103 16063104 16063105 16063106 16063107 16063108 16063109 16063110 16063111 16063112 16063113 16063114 16063115 16063116 16063117 16063118 16063119 16063120 16063121 16063122 16063123 16063124 16063125 16063126 16063127 16063128 16063129 16063130 16063131 16063132 16063133 16063134 16063135 16063136 16063137 16063138 16063139 16063140 16063141 16063142 16063143 16063144 16063145 16063146 16063147 16063148 16063149 16063150 16063151 16063152 16063153 16063154 16063155 16063156 16063157 16063158 16063159 16063160 16063161 16063162 16063163 16063164 16063165 16063166 16063167 16063168 16063169 16063170 16063171 16063172 16063173 16063174 16063175 16063176 16063177 16063178 16063179 16063180 16063181 16063182 16063183 16063184 16063185 16063186 16063187 16063188 16063189 16063190 16063191 16063192 16063193 16063194 16063195 16063196 16063197 16063198 16063199 16063200 16063201 16063202 16063203 16063204 16063205 16063206 16063207 16063208 16063209 16063210 16063211 16063212 16063213 16063214 16063215 16063216 16063217 16063218 16063219 16063220 16063221 16063222 16063223 16063224 16063225 16063226 16063227 16063228 16063229 16063230 16063231 16063232 16063233 16063234 16063235 16063236 16063237 16063239 16063240 16063241 16063242 16063243 16063244 16063245 16063246 16063247 16063248 16063249 16063250 16063251 16063252 16063253 16063254 16063255 16063256 16063257 16063258 16063259 16063260 16063261 16063262 16063263 16063264 16063265 16063266 16063267 16063268 16063269 16063270 16063271 16063272 16063273 16063274 16063275 16063276 16063277 16063278 16063279 16063281 16063282 16063283 16063284 16063285 16063286 16063287 16063288 16063289 16063290 16063291 16063293 16063294 16063295 16063296 16063297 16063298 16063299 16063300 16063301 16063302 16063303 16063304 16063305 16063306 16063307 16063308 16063309 16063310 16063311 16063312 16063313 16063314 16063315 16063316 16063317 16063318 16063319 16063320 16063321 16063322 16063323 16063324 16063325 16063326 16063327 16063328 16063329 16063330 16063331 16063332 16063333 16063334 16063335 16063336 16063337 16063338 16063339 16063340 16063341 16063342 16063343 16063344 16063345 16063346 16063347 16063348 16063349 16063350 16063351 16063352 16063353 16063354 16063355 16063356 16063357 16063358 16063359 16063360 16063361 16063362 16063363 16063364 16063365 16063366 16063367 16063368 16063369 16063370 16063371 16063372 16063373 16063374 16063375 16063376 16063377 16063378 16063379 16063380 16063381 16063382 16063383 16063384 16063385 16063386 16063387 16063388 16063389 16063390 16063391 16063392 16063393 16063394 16063395 16063396 16063397 16063398 16063399 16063400 16063401 16063402 16063403 16063404 16063405 16063406 16063407 16063408 16063409 16063410 16063411 16063412 16063413 16063414 16063415 16063416 16063417 16063418 16063419 16063420 16063421 16063422 16063423 16063425 16063426 16063427 16063428 16063429 16063430 16063431 16063432 16063433 16063434 16063435 16063436 16063437 16063438 16063439 16063440 16063441 16063442 16063443 16063444 16063445 16063446 16063447 16063448 16063449 16063450 16063451 16063452 16063453 16063454 16063455 16063456 16063457 16063458 16063459 16063460 16063461 16063462 16063463 16063464 16063465 16063466 16063467 16063468 16063469 16063470 16063471 16063472 16063473 16063474 16063475 16063476 16063477 16063478 16063479 16063480 16063481 16063482 16063483 16063484 16063485 16063486 16063487 16063488 16063489 16063490 16063491 16063492 16063493 16063494 16063495 16063496 16063497 16063498 16063499 16063500 16063501 16063502 16063503 16063504 16063505 16063506 16063507 16063508 16063509 16063510 16063511 16063512 16063513 16063514 16063515 16063516 16063517 16063518 16063519 16063520 16063521 16063522 16063523 16063524 16063525 16063526 16063527 16063528 16063529 16063530 16063531 16063532 16063533 16063534 16063535 16063536 16063537 16063538 16063539 16063540 16063541 16063542 16063543 16063544 16063545 16063546 16063547 16063548 16063549 16063550 16063551 16063552 16063553 16063554 16063555 16063556 16063557 16063558 16063559 16063560 16063561 16063562 16063563 16063564 16063565 16063566 16063567 16063568 16063569 16063570 16063571 16063572 16063573 16063574 16063575 16063576 16063577 16063578 16063579 16063580 16063581 16063582 16063583 16063584 16063585 16063586 16063587 16063588 16063589 16063590 16063591 16063592 16063593 16063594 16063595 16063596 16063597 16063598 16063599 16063600 16063601 16063602 16063603 16063604 16063605 16063606 16063607 16063608 16063609 16063610 16063611 16063612 16063613 16063614 16063615 16063616 16063617 16063618 16063619 16063620 16063621 16063622 16063623 16063624 16063625 16063626 16063627 16063628 16063629 16063630 16063631 16063632 16063633 16063634 16063635 16063636 16063637 16063638 16063639 16063640 16063641 16063642 16063643 16063644 16063645 16063646 16063647 16063648 16063649 16063650 16063651 16063652 16063653 16063654 16063655 16063656 16063657 16063658 16063659 16063660 16063662 16063663 16063664 16063665 16063666 16063667 16063668 16063669 16063670 16063671 16063672 16063673 16063674 16063675 16063676 16063677 16063678 16063679 16063680 16063681 16063682 16063683 16063684 16063685 16063686 16063687 16063688 16063689 16063690 16063691 16063692 16063693 16063694 16063695 16063696 16063697 16063698 16063699 16063700 16063701 16063702 16063703 16063704 16063705 16063706 16063707 16063708 16063709 16063710 16063711 16063712 16063713 16063714 16063715 16063716 16063717 16063718 16063719 16063720 16063721 16063722 16063723 16063724 16063725 16063726 16063727 16063728 16063729 16063730 16063731 16063732 16063733 16063734 16063735 16063736 16063737 16063738 16063739 16063740 16063741 16063742 16063743 16063744 16063745 16063746 16063747 16063748 16063749 16063750 16063751 16063752 16063753 16063754 16063755 16063756 16063757 16063758 16063759 16063760 16063761 16063762 16063763 16063764 16063765 16063766 16063767 16063768 16063769 16063770 16063771 16063772 16063773 16063774 16063775 16063776 16063777 16063778 16063779 16063780 16063781 16063782 16063783 16063784 16063785 16063786 16063787 16063788 16063789 16063790 16063791 16063792 16063793 16063794 16063795 16063796 16063797 16063798 16063800 16063801 16063802 16063803 16063804 16063805 16063806 16063807 16063808 16063809 16063810 16063811 16063812 16063813 16063814 16063815 16063816 16063817 16063818 16063819 16063820 16063821 16063822 16063823 16063824 16063825 16063826 16063827 16063828 16063829 16063830 16063831 16063832 16063833 16063834 16063835 16063836 16063837 16063838 16063840 16063841 16063842 16063843 16063844 16063845 16063846 16063847 16063848 16063849 16063850 16063851 16063852 16063853 16063854 16063855 16063856 16063857 16063858 16063859 16063860 16063861 16063862 16063863 16063864 16063865 16063866 16063867 16063868 16063869 16063870 16063871 16063872 16063873 16063874 16063875 16063876 16063877 16063878 16063879 16063880 16063881 16063882 16063883 16063884 16063885 16063886 16063887 16063888 16063889 16063890 16063891 16063892 16063893 16063894 16063895 16063896 16063897 16063899 16063900 16063901 16063902 16063903 16063904 16063905 16063906 16063907 16063908 16063909 16063910 16063911 16063912 16063913 16063914 16063915 16063916 16063917 16063918 16063919 16063920 16063921 16063922 16063923 16063924 16063925 16063927 16063928 16063929 16063930 16063931 16063932 16063933 16063934 16063935 16063936 16063937 16063938 16063939 16063940 16063941 16063942 16063943 16063944 16063945 16063946 16063947 16063948 16063949 16063950 16063951 16063952 16063953 16063954 16063955 16063956 16063957 16063958 16063959 16063960 16063962 16063963 16063964 16063965 16063966 16063967 16063968 16063969 16063970 16063971 16063972 16063973 16063974 16063975 16063976 16063977 16063978 16063979 16063980 16063982 16063983 16063984 16063985 16063986 16063987 16063988 16063989 16063990 16063991 16063992 16063993 16063994 16063995 16063996 16063997 16063998 16063999 16064000 16064001 16064002 16064003 16064004 16064005 16064006 16064007 16064008 16064009 16064010 16064011 16064014 16064015 16064016 16064017 16064018 16064019 16064020 16064021 16064022 16064023 16064024 16064025 16064026 16064027 16064028 16064029 16064030 16064031 16064032 16064033 16064034 16064035 16064037 16064038 16064039 16064040 16064041 16064042 16064043 16064044 16064045 16064046 16064047 16064048 16064049 16064050 16064051 16064052 16064053 16064054 16064055 16064056 16064057 16064058 16064059 16064060 16064061 16064062 16064063 16064064 16064065 16064066 16064067 16064068 16064069 16064070 16064071 16064072 16064073 16064074 16064075 16064076 16064077 16064078 16064079 16064080 16064081 16064082 16064083 16064084 16064085 16064086 16064087 16064088 16064089 16064090 16064091 16064092 16064093 16064094 16064095 16064096 16064097 16064098 16064099 16064101 16064102 16064103 16064104 16064105 16064106 16064107 16064108 16064109 16064110 16064111 16064112 16064113 16064114 16064115 16064116 16064117 16064118 16064119 16064120 16064121 16064122 16064123 16064124 16064125 16064126 16064127 16064128 16064129 16064130 16064131 16064132 16064134 16064135 16064136 16064138 16064139 16064141 16064142 16064143 16064144 16064145 16064146 16064148 16064149 16064150 16064151 16064152 16064153 16064154 16064155 16064156 16064157 16064158 16064159 16064160 16064161 16064163 16064164 16064165 16064166 16064167 16064168 16064169 16064170 16064171 16064172 16064173 16064174 16064175 16064176 16064177 16064178 16064179 16064180 16064181 16064182 16064183 16064184 16064185 16064186 16064187 16064188 16064189 16064190 16064191 16064192 16064193 16064195 16064196 16064197 16064198 16064199 16064200 16064201 16064202 16064203 16064204 16064205 16064206 16064207 16064208 16064209 16064210 16064211 16064212 16064213 16064214 16064215 16064216 16064217 16064218 16064219 16064220 16064221 16064222 16064223 16064224 16064225 16064226 16064227 16064228 16064229 16064230 16064231 16064232 16064233 16064234 16064235 16064236 16064238 16064239 16064240 16064241 16064242 16064243 16064244 16064245 16064246 16064247 16064248 16064249 16064250 16064251 16064252 16064253 16064254 16064255 16064256 16064257 16064258 16064259 16064260 16064261 16064262 16064263 16064264 16064265 16064266 16064267 16064268 16064269 16064270 16064271 16064272 16064273 16064274 16064275 16064276 16064277 16064278 16064279 16064280 16064281 16064282 16064283 16064284 16064286 16064287 16064288 16064289 16064290 16064291 16064292 16064293 16064294 16064295 16064296 16064297 16064298 16064301 16064302 16064303 16064304 16064305 16064306 16064307 16064308 16064309 16064310 16064311 16064312 16064314 16064315 16064316 16064317 16064319 16064320 16064321 16064322 16064323 16064324 16064325 16064326 16064327 16064328 16064329 16064330 16064331 16064332 16064333 16064334 16064335 16064336 16064337 16064338 16064339 16064340 16064341 16064342 16064343 16064344 16064345 16064346 16064347 16064348 16064349 16064350 16064351 16064352 16064353 16064354 16064355 16064356 16064357 16064358 16064359 16064360 16064361 16064362 16064363 16064364 16064365 16064366 16064367 16064368 16064369 16064370 16064372 16064373 16064374 16064375 16064376 16064377 16064378 16064379 16064380 16064381 16064382 16064383 16064384 16064385 16064387 16064388 16064389 16064390 16064391 16064392 16064393 16064394 16064395 16064396 16064397 16064398 16064399 16064400 16064401 16064402 16064403 16064404 16064405 16064406 16064407 16064408 16064409 16064410 16064411 16064412 16064413 16064414 16064415 16064416 16064417 16064418 16064419 16064420 16064421 16064422 16064424 16064425 16064426 16064427 16064428 16064429 16064430 16064431 16064432 16064433 16064434 16064435 16064436 16064437 16064438 16064439 16064440 16064441 16064442 16064443 16064444 16064445 16064446 16064447 16064448 16064449 16064450 16064452 16064453 16064454 16064455 16064456 16064457 16064458 16064459 16064460 16064

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2683-2017

Related Recalls

A possible interruption of therapy when an E321 error code occurs while using the infuser, which is caused by battery issues.

Feb 1, 2013 Infusion Pumps Nationwide View Details →

A possible interruption of therapy when an E321 error code occurs while using the infuser, which is caused by battery issues.

Feb 1, 2013 Infusion Pumps Nationwide View Details →

A possible interruption of therapy when an E321 error code occurs while using the infuser, which is caused by battery issues.

Feb 1, 2013 Infusion Pumps Nationwide View Details →