The Plum A+ Infusion Systems are a multipurpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care. The Plum A+ family infusion pumps are self contained, microprocessor based infusion devices. The pumps are available in single channel and in triple channel pumping mechanism types.
Class I - DangerousWhat Should You Do?
- Check if you have this product: List number: 20679 Serial Number 15441019 15441020 15441037 15441044 15441082 15441093 15441118 15441226 15441229 15441230 15441240 15441241 15441244 15441246 15441247 15441249 15441253 15441256 15441258 15441260 15441265 15441266 15441270 15441271 15441272 15441275 15441285 15441303 15441304 15441315 15441318 15441321 15441323 15441325 15441349 15441350 15441367 15441369 15441371 15441372 15441378 15441395 15441399 15441405 15441409 15441411 15441437 15441458 15441489 15441492 15441493 15441502 15441506 15441508 15441513 15441515 15441516 15441520 15441522 15441524 15441526 15441528 15441532 15441533 15441534 15441535 15441536 15441554 15441601 15441604 15441605 15441610 15441611 15441615 15441633 15441637 15441638 15441649 15441704 15441765 15441803 15441816 15441839 15441942 15441960 15441966 15442037 15442062 15442090 15442124 15442126 15442157 15442193 15442197 15442207 15442235 15442280 15442304 15442306 15442327 15442370 15442401 15442408 15442417 15442448 15442457 15442512 15442544 15442663 15442831 15442838 15442841 15442893 15442905 15442920 15442927 15442961 15442974 15442979 15442998 15443001 15443003 15443005 15443006 15443007 15443008 15443010 15443018 15443023 15443025 15443027 15443028 15443030 15443033 15443034 15443039 15443040 15443062 15443067 15443069 15443103 15443119 15443123 15443126 15443130 15443142 15443148 15443154 15443228 15443248 15443317 15443342 15443346 15443365 15443401 15443403 15443426 15443427 15443445 15443474 15443475 15443478 15443518 15443529 15443533 15443535 15443537 15443542 15443589 15443607 15443609 15443614 15443618 15443619 15443626 15443629 15443635 15443637 15443640 15443642 15443650 15443668 15443671 15443683 15443688 15443719 15443744 15443756 15443757 15443759 15443760 15443771 15443785 15443787 15443789 15443799 15443801 15443808 15443820 15443825 15443830 15443833 15443837 15443859 15443865 15443868 15443879 15443880 15443895 15443910 15443919 15443923 15443927 15443928 15443929 15443934 15443935 15443936 15443942 15443943 15443944 15443946 15443947 15443949 15443950 15443951 15443959 15443961 15443962 15443963 15443965 15443970 15443976 15443977 15443978 15443986 15443987 15443988 15443992 15443993 15443996 15443999 15444000 15444002 15444003 15444005 15444008 15444037 15444040 15444041 15444060 15444065 15444067 15444069 15444072 15444084 15444100 15444107 15444109 15444111 15444115 15444116 15444118 15444119 15444120 15444122 15444123 15444124 15444125 15444127 15444128 15444129 15444132 15444133 15444135 15444136 15444137 15444139 15444140 15444141 15444142 15444143 15444144 15444145 15444146 15444147 15444150 15444152 15444153 15444154 15444156 15444157 15444164 15444166 15444167 15444168 15444171 15444173 15444175 15444176 15444184 15444186 15444188 15444190 15444192 15444193 15444196 15444200 15444206 15444211 15444212 15444213 15444215 15444216 15444217 15444218 15444219 15444221 15444224 15444226 15444227 15444228 15444229 15444230 15444231 15444232 15444233 15444235 15444236 15444238 15444241 15444243 15444245 15444247 15444248 15444249 15444250 15444251 15444252 15444254 15444256 15444261 15444262 15444264 15444269 15444272 15444274 15444276 15444277 15444279 15444280 15444281 15444284 15444285 15444286 15444287 15444288 15444289 15444291 15444292 15444293 15444296 15444298 15444299 15444300 15444301 15444303 15444304 15444308 15444311 15444313 15444314 15444317 15444321 15444322 15444325 15444329 15444332 15444333 15444335 15444337 15444338 15444342 15444346 15444349 15444351 15444354 15444361 15444362 15444364 15444366 15444368 15444370 15444373 15444374 15444376 15444383 15444385 15444386 15444388 15444389 15444394 15444395 15444399 15444400 15444402 15444403 15444412 15444418 15444430 15444431 15444449 15444450 15444452 15444455 15444456 15444459 15444462 15444463 15444465 15444472 15444473 15444474 15444475 15444477 15444482 15444483 15444485 15444486 15444488 15444489 15444491 15444492 15444493 15444496 15444497 15444498 15444500 15444503 15444504 15444511 15444514 15444515 15444516 15444517 15444518 15444519 15444524 15444526 15444528 15444531 15444532 15444537 15444540 15444541 15444543 15444546 15444551 15444553 15444554 15444555 15444556 15444557 15444558 15444559 15444560 15444565 15444566 15444569 15444570 15444571 15444573 15444574 15444575 15444576 15444577 15444578 15444581 15444583 15444584 15444585 15444586 15444587 15444590 15444591 15444592 15444594 15444595 15444597 15444600 15444601 15444604 15444606 15444607 15444608 15444609 15444610 15444613 15444615 15444616 15444620 15444621 15444625 15444627 15444628 15444629 15444630 15444637 15444639 15444642 15444644 15444647 15444648 15444649 15444651 15444652 15444653 15444654 15444655 15444656 15444657 15444658 15444659 15444660 15444662 15444663 15444666 15444667 15444671 15444673 15444674 15444676 15444678 15444679 15444680 15444681 15444682 15444690 15444692 15444694 15444695 15444697 15444699 15444701 15444704 15444705 15444706 15444708 15444709 15444710 15444711 15444713 15444714 15444715 15444717 15444718 15444719 15444721 15444722 15444723 15444724 15444725 15444726 15444727 15444729 15444730 15444731 15444733 15444734 15444735 15444738 15444739 15444740 15444743 15444744 15444746 15444747 15444748 15444750 15444752 15444753 15444754 15444755 15444756 15444757 15444761 15444762 15444764 15444767 15444768 15444771 15444775 15444776 15444777 15444778 15444780 15444782 15444783 15444785 15444786 15444787 15444788 15444790 15444791 15444792 15444794 15444796 15444799 15444801 15444802 15444803 15444805 15444808 15444810 15444811 15444815 15444816 15444817 15444819 15444820 15444821 15444822 15444823 15444825 15444827 15444829 15444830 15444831 15444832 15444833 15444834 15444844 15444848 15444851 15444854 15444857 15444858 15444859 15444862 15444863 15444865 15444866 15444868 15444869 15444870 15444871 15444872 15444874 15444875 15444876 15444877 15444878 15444879 15444880 15444881 15444882 15444883 15444884 15444886 15444887 15444888 15444889 15444890 15444891 15444894 15444896 15444898 15444899 15444901 15444902 15444903 15444904 15444908 15444909 15444910 15444914 15444920 15444921 15444925 15444926 15444927 15444928 15444929 15444930 15444931 15444932 15444933 15444937 15444938 15444939 15444940 15444941 15444942 15444943 15444944 15444945 15444948 15444949 15444950 15444951 15444952 15444953 15444957 15444958 15444959 15444960 15444963 15444964 15444965 15444966 15444967 15444969 15444971 15444976 15444980 15444983 15444984 15444986 15444987 15444988 15444990 15444991 15444992 15444994 15444995 15444996 15444999 15445000 15445003 15445006 15445007 15445008 15445009 15445010 15445012 15445013 15445014 15445017 15445018 15445021 15445022 15445023 15445024 15445025 15445026 15445027 15445028 15445029 15445031 15445032 15445033 15445034 15445035 15445036 15445037 15445038 15445039 15445043 15445044 15445045 15445046 15445048 15445050 15445051 15445052 15445053 15445055 15445057 15445060 15445062 15445063 15445064 15445066 15445067 15445068 15445071 15445072 15445073 15445076 15445077 15445081 15445082 15445085 15445087 15445090 15445091 15445092 15445093 15445094 15445098 15445099 15445100 15445101 15445102 15445103 15445106 15445107 15445108 15445110 15445112 15445113 15445114 15445117 15445118 15445119 15445123 15445124 15445125 15445126 15445128 15445129 15445131 15445132 15445133 15445135 15445136 15445138 15445143 15445144 15445145 15445146 15445147 15445150 15445151 15445153 15445155 15445156 15445157 15445158 15445159 15445168 15445169 15445170 15445171 15445174 15445175 15445176 15445177 15445178 15445179 15445184 15445187 15445188 15445189 15445190 15445192 15445193 15445195 15445196 15445197 15445199 15445200 15445201 15445203 15445206 15445207 15445208 15445210 15445211 15445212 15445213 15445214 15445215 15445217 15445218 15445219 15445222 15445223 15445224 15445226 15445228 15445229 15445230 15445232 15445234 15445235 15445238 15445240 15445241 15445242 15445243 15445244 15445245 15445247 15445250 15445251 15445252 15445253 15445254 15445256 15445257 15445260 15445264 15445266 15445268 15445269 15445270 15445271 15445272 15445273 15445275 15445277 15445283 15445284 15445285 15445287 15445288 15445289 15445294 15445297 15445303 15445305 15445306 15445307 15445308 15445309 15445310 15445311 15445312 15445314 15445317 15445318 15445319 15445320 15445322 15445323 15445324 15445325 15445327 15445328 15445332 15445333 15445334 15445337 15445338 15445339 15445340 15445344 15445347 15445350 15445351 15445352 15445353 15445354 15445355 15445357 15445358 15445359 15445360 15445361 15445362 15445364 15445365 15445366 15445367 15445368 15445369 15445370 15445371 15445372 15445373 15445374 15445375 15445376 15445377 15445379 15445385 15445389 15445390 15445391 15445393 15445394 15445397 15445400 15445406 15445407 15445408 15445409 15445411 15445412 15445414 15445415 15445416 15445417 15445419 15445420 15445425 15445427 15445428 15445429 15445430 15445432 15445433 15445434 15445436 15445437 15445438 15445440 15445441 15445444 15445445 15445446 15445448 15445449 15445450 15445451 15445454 15445455 15445456 15445457 15445458 15445459 15445460 15445461 15445462 15445464 15445466 15445467 15445468 15445469 15445471 15445472 15445473 15445474 15445475 15445477 15445478 15445479 15445480 15445481 15445482 15445483 15445484 15445485 15445487 15445488 15445489 15445490 15445491 15445492 15445494 15445495 15445496 15445497 15445498 15445499 15445501 15445503 15445504 15445506 15445507 15445508 15445509 15445510 15445511 15445512 15445513 15445514 15445516 15445517 15445518 15445519 15445520 15445522 15445524 15445525 15445526 15445527 15445528 15445530 15445531 15445532 15445533 15445534 15445535 15445537 15445538 15445539 15445541 15445542 15445543 15445544 15445546 15445547 15445549 15445552 15445553 15445554 15445555 15445556 15445557 15445559 15445561 15445563 15445564 15445565 15445568 15445569 15445571 15445572 15445573 15445574 15445576 15445577 15445579 15445580 15445582 15445583 15445585 15445586 15445588 15445591 15445592 15445593 15445595 15445596 15445597 15445598 15445599 15445600 15445601 15445602 15445603 15445604 15445605 15445606 15445607 15445608 15445609 15445610 15445611 15445612 15445613 15445614 15445615 15445619 15445620 15445622 15445623 15445624 15445625 15445626 15445630 15445631 15445632 15445633 15445634 15445635 15445636 15445637 15445638 15445640 15445641 15445642 15445643 15445644 15445646 15445647 15445648 15445649 15445650 15445651 15445652 15445653 15445654 15445655 15445656 15445657 15445659 15445660 15445661 15445662 15445663 15445667 15445669 15445670 15445671 15445672 15445673 15445674 15445675 15445677 15445678 15445679 15445680 15445681 15445682 15445684 15445685 15445689 15445690 15445692 15445694 15445695 15445696 15445697 15445698 15445699 15445701 15445702 15445703 15445704 15445705 15445706 15445707 15445709 15445710 15445711 15445712 15445714 15445715 15445716 15445717 15445718 15445721 15445722 15445724 15445725 15445726 15445727 15445728 15445729 15445730 15445731 15445732 15445733 15445734 15445735 15445736 15445737 15445738 15445739 15445740 15445741 15445743 15445744 15445745 15445749 15445750 15445751 15445752 15445753 15445754 15445755 15445756 15445757 15445759 15445761 15445762 15445763 15445766 15445767 15445768 15445773 15445774 15445775 15445776 15445777 15445778 15445779 15445780 15445781 15445783 15445784 15445785 15445786 15445788 15445789 15445790 15445791 15445792 15445793 15445795 15445796 15445797 15445798 15445799 15445803 15445804 15445806 15445807 15445808 15445809 15445810 15445811 15445812 15445814 15445815 15445817 15445819 15445820 15445821 15445822 15445823 15445824 15445825 15445826 15445827 15445829 15445830 15445831 15445834 15445835 15445836 15445837 15445838 15445839 15445841 15445842 15445844 15445845 15445847 15445848 15445849 15445851 15445852 15445853 15445855 15445857 15445858 15445859 15445860 15445861 15445863 15445866 15445867 15445868 15445869 15445871 15445872 15445873 15445875 15445876 15445877 15445878 15445879 15445880 15445881 15445884 15445885 15445887 15445889 15445891 15445892 15445896 15445897 15445898 15445900 15445901 15445902 15445903 15445904 15445906 15445907 15445908 15445909 15445910 15445911 15445913 15445914 15445915 15445916 15445917 15445919 15445920 15445921 15445922 15445923 15445924 15445925 15445928 15445929 15445932 15445936 15445938 15445939 15445940 15445941 15445943 15445945 15445946 15445948 15445949 15445950 15445951 15445952 15445954 15445955 15445957 15445958 15445959 15445960 15445961 15445962 15445967 15445970 15445971 15445972 15445973 15445974 15445975 15445976 15445977 15445978 15445980 15445983 15445984 15445986 15445987 15445988 15445989 15445991 15445993 15445994 15445995 15445996 15445997 15445999 15446000 15446001 15446002 15446003 15446004 15446005 15446006 15446008 15446009 15446010 15446011 15446012 15446014 15446016 15446017 15446018 15446019 15446021 15446022 15446024 15446025 15446026 15446027 15446028 15446029 15446030 15446031 15446034 15446037 15446038 15446039 15446040 15446041 15446042 15446043 15446044 15446045 15446046 15446047 15446048 15446049 15446050 15446051 15446052 15446053 15446054 15446055 15446057 15446060 15446064 15446065 15446066 15446067 15446068 15446069 15446070 15446072 15446073 15446074 15446079 15446080 15446083 15446085 15446094 15446095 15446096 15446097 15446099 15446100 15446101 15446102 15446105 15446106 15446107 15446108 15446109 15446111 15446113 15446114 15446115 15446116 15446117 15446119 15446120 15446121 15446122 15446125 15446126 15446128 15446130 15446132 15446133 15446134 15446135 15446137 15446138 15446139 15446140 15446141 15446142 15446145 15446147 15446148 15446149 15446151 15446154 15446155 15446156 15446159 15446160 15446161 15446162 15446163 15446166 15446167 15446168 15446170 15446171 15446172 15446173 15446175 15446176 15446178 15446180 15446181 15446184 15446186 15446189 15446190 15446194 15446195 15446196 15446197 15446198 15446200 15446201 15446203 15446205 15446206 15446207 15446208 15446210 15446211 15446212 15446215 15446216 15446218 15446219 15446220 15446222 15446223 15446224 15446225 15446226 15446227 15446228 15446229 15446230 15446231 15446232 15446233 15446234 15446235 15446236 15446237 15446239 15446240 15446241 15446243 15446244 15446245 15446246 15446247 15446248 15446250 15446251 15446252 15446253 15446254 15446255 15446256 15446258 15446259 15446261 15446262 15446264 15446265 15446266 15446267 15446268 15446271 15446273 15446274 15446275 15446276 15446279 15446281 15446282 15446283 15446286 15446290 15446296 15446297 15446298 15446299 15446300 15446302 15446303 15446304 15446305 15446306 15446307 15446311 15446312 15446313 15446338 15446340 15446341 15446344 15446347 15446348 15446349 15446350 15446351 15446352 15446402 15446456 15446457 15446460 15446462 15446464 15446465 15446466 15446467 15446471 15446472 15446473 15446475 15446476 15446477 15446478 15446479 15446480 15446481 15446482 15446483 15446488 15446489 15446490 15446491 15446494 15446495 15446496 15446498 15446506 15446507 15446508 15446509 15446510 15446511 15446512 15446513 15446514 15446515 15446517 15446518 15446519 15446520 15446521 15446522 15446523 15446524 15446525 15446526 15446529 15446530 15446531 15446532 15446533 15446536 15446537 15446538 15446539 15446540 15446541 15446542 15446543 15446544 15446546 15446547 15446548 15446550 15446551 15446552 15446553 15446554 15446555 15446556 15446557 15446558 15446560 15446561 15446563 15446564 15446565 15446566 15446567 15446568 15446569 15446570 15446571 15446572 15446573 15446574 15446575 15446576 15446577 15446579 15446580 15446581 15446582 15446583 15446584 15446586 15446588 15446589 15446590 15446591 15446592 15446593 15446594 15446595 15446596 15446597 15446598 15446599 15446600 15446601 15446602 15446603 15446604 15446605 15446606 15446607 15446608 15446609 15446610 15446612 15446613 15446614 15446615 15446617 15446618 15446619 15446620 15446621 15446622 15446623 15446624 15446625 15446626 15446627 15446628 15446629 15446630 15446631 15446632 15446633 15446634 15446635 15446636 15446637 15446639 15446640 15446641 15446643 15446644 15446646 15446647 15446648 15446650 15446651 15446652 15446653 15446654 15446655 15446656 15446657 15446658 15446659 15446660 15446661 15446662 15446663 15446664 15446665 15446666 15446667 15446668 15446671 15446672 15446673 15446674 15446675 15446676 15446677 15446678 15446679 15446680 15446681 15446682 15446683 15446685 15446686 15446687 15446688 15446690 15446692 15446694 15446695 15446697 15446701 15446702 15446703 15446704 15446705 15446706 15446707 15446708 15446709 15446710 15446712 15446714 15446716 15446717 15446719 15446720 15446721 15446722 15446723 15446724 15446725 15446726 15446727 15446728 15446729 15446731 15446732 15446733 15446734 15446735 15446736 15446737 15446739 15446740 15446741 15446742 15446744 15446745 15446746 15446748 15446749 15446750 15446751 15446752 15446754 15446755 15446756 15446758 15446759 15446760 15446761 15446762 15446763 15446764 15446767 15446768 15446769 15446770 15446771 15446772 15446773 15446775 15446776 15446778 15446779 15446781 15446782 15446783 15446784 15446785 15446786 15446787 15446788 15446789 15446790 15446791 15446792 15446795 15446796 15446797 15446798 15446799 15446800 15446801 15446802 15446803 15446804 15446805 15446808 15446809 15446810 15446811 15446812 15446813 15446815 15446816 15446817 15446818 15446819 15446820 15446821 15446822 15446823 15446824 15446825 15446826 15446827 15446828 15446829 15446830 15446831 15446832 15446833 15446834 15446835 15446836 15446837 15446838 15446839 15446840 15446841 15446842 15446843 15446844 15446845 15446846 15446847 15446848 15446849 15446850 15446851 15446852 15446853 15446854 15446855 15446856 15446857 15446858 15446859 15446860 15446861 15446862 15446863 15446864 15446865 15446866 15446867 15446868 15446869 15446870 15446872 15446873 15446874 15446875 15446876 15446877 15446879 15446880 15446881 15446883 15446885 15446886 15446887 15446889 15446890 15446893 15446894 15446895 15446896 15446898 15446900 15446901 15446902 15446903 15446904 15446906 15446907 15446909 15446910 15446911 15446913 15446914 15446915 15446916 15446917 15446918 15446919 15446920 15446923 15446924 15446925 15446926 15446927 15446930 15446931 15446932 15446933 15446934 15446936 15446937 15446938 15446939 15446940 15446941 15446942 15446943 15446945 15446946 15446947 15446948 15446949 15446950 15446951 15446952 15446953 15446954 15446955 15446956 15446957 15446959 15446960 15446961 15446963 15446964 15446965 15446966 15446967 15446968 15446969 15446970 15446971 15446972 15446974 15446976 15446977 15446978 15446979 15446981 15446982 15446983 15446984 15446985 15446986 15446991 15446992 15446993 15446994 15446995 15446996 15446997 15446998 15446999 15447000 15447001 15447002 15447003 15447004 15447007 15447008 15447010 15447011 15447012 15447013 15447014 15447015 15447016 15447017 15447018 15447019 15447020 15447021 15447022 15447023 15447025 15447027 15447028 15447029 15447030 15447031 15447032 15447033 15447034 15447037 15447038 15447039 15447041 15447042 15447043 15447044 15447045 15447047 15447048 15447049 15447051 15447052 15447055 15447056 15447059 15447060 15447061 15447063 15447064 15447065 15447066 15447067 15447068 15447069 15447070 15447071 15447072 15447073 15447074 15447075 15447076 15447077 15447078 15447081 15447082 15447084 15447085 15447086 15447087 15447088 15447089 15447090 15447092 15447093 15447094 15447096 15447098 15447100 15447101 15447102 15447103 15447105 15447106 15447107 15447108 15447109 15447110 15447111 15447116 15447117 15447121 15447122 15447124 15447126 15447127 15447128 15447129 15447130 15447131 15447133 15447134 15447137 15447140 15447141 15447142 15447143 15447146 15447147 15447149 15447150 15447151 15447152 15447153 15447154 15447155 15447156 15447158 15447159 15447160 15447161 15447162 15447163 15447164 15447165 15447166 15447167 15447169 15447170 15447171 15447172 15447173 15447174 15447175 15447177 15447178 15447179 15447180 15447181 15447182 15447183 15447184 15447185 15447186 15447187 15447188 15447189 15447190 15447191 15447192 15447193 15447194 15447195 15447196 15447197 15447199 15447200 15447201 15447202 15447203 15447204 15447205 15447206 15447207 15447210 15447211 15447212 15447213 15447214 15447216 15447217 15447219 15447220 15447221 15447222 15447223 15447224 15447225 15447226 15447227 15447228 15447230 15447232 15447233 15447234 15447236 15447237 15447240 15447241 15447245 15447246 15447247 15447249 15447250 15447251 15447252 15447253 15447254 15447255 15447257 15447258 15447259 15447261 15447262 15447263 15447264 15447265 15447266 15447268 15447270 15447271 15447272 15447273 15447275 15447276 15447278 15447279 15447280 15447283 15447284 15447285 15447286 15447287 15447290 15447291 15447292 15447293 15447294 15447295 15447296 15447297 15447298 15447300 15447301 15447302 15447303 15447304 15447305 15447306 15447307 15447308 15447309 15447310 15447311 15447312 15447313 15447314 15447315 15447316 15447317 15447318 15447319 15447320 15447321 15447322 15447323 15447324 15447325 15447326 15447327 15447328 15447329 15447330 15447331 15447332 15447333 15447334 15447335 15447336 15447337 15447339 15447340 15447341 15447342 15447344 15447345 15447346 15447347 15447348 15447349 15447350 15447351 15447356 15447358 15447359 15447360 15447361 15447362 15447363 15447365 15447366 15447367 15447368 15447369 15447370 15447372 15447374 15447375 15447376 15447379 15447381 15447382 15447384 15447385 15447386 15447387 15447388 15447389 15447390 15447391 15447392 15447393 15447395 15447396 15447397 15447398 15447399 15447400 15447401 15447402 15447403 15447404 15447406 15447408 15447410 15447411 15447412 15447413 15447414 15447418 15447419 15447420 15447421 15447422 15447423 15447424 15447425 15447426 15447427 15447428 15447430 15447431 15447432 15447433 15447434 15447436 15447437 15447438 15447440 15447441 15447442 15447443 15447444 15447445 15447447 15447448 15447449 15447451 15447452 15447453 15447454 15447455 15447457 15447458 15447459 15447460 15447461 15447462 15447463 15447464 15447465 15447466 15447467 15447468 15447469 15447470 15447471 15447472 15447473 15447474 15447476 15447477 15447480 15447481 15447482 15447483 15447484 15447485 15447486 15447487 15447488 15447489 15447490 15447492 15447501 15447506 15447507 15447508 15447510 15447511 15447512 15447513 15447514 15447515 15447517 15447518 15447519 15447521 15447522 15447523 15447524 15447526 15447527 15447528 15447529 15447530 15447531 15447532 15447533 15447534 15447535 15447536 15447537 15447540 15447542 15447543 15447544 15447545 15447548 15447550 15447552 15447553 15447554 15447555 15447559 15447560 15447561 15447563 15447564 15447565 15447566 15447567 15447568 15447569 15447570 15447571 15447573 15447574 15447576 15447577 15447579 15447580 15447581 15447582 15447583 15447584 15447585 15447587 15447588 15447589 15447590 15447591 15447592 15447593 15447594 15447595 15447596 15447597 15447598 15447599 15447601 15447602 15447603 15447606 15447609 15447610 15447611 15447612 15447613 15447614 15447616 15447617 15447618 15447619 15447620 15447621 15447622 15447623 15447624 15447625 15447626 15447627 15447628 15447630 15447631 15447632 15447633 15447634 15447636 15447638 15447640 15447641 15447642 15447643 15447644 15447645 15447646 15447647 15447648 15447649 15447650 15447651 15447652 15447653 15447654 15447655 15447657 15447658 15447659 15447660 15447661 15447663 15447665 15447666 15447667 15447668 15447669 15447670 15447671 15447672 15447673 15447674 15447675 15447676 15447677 15447678 15447679 15447680 15447681 15447682 15447683 15447684 15447685 15447686 15447687 15447689 15447690 15447691 15447692 15447693 15447694 15447695 15447696 15447697 15447698 15447699 15447700 15447701 15447702 15447704 15447705 15447706 15447707 15447708 15447709 15447710 15447712 15447713 15447714 15447715 15447716 15447717 15447718 15447719 15447720 15447722 15447723 15447725 15447726 15447727 15447728 15447729 15447730 15447731 15447732 15447733 15447734 15447735 15447736 15447737 15447738 15447739 15447740 15447741 15447742 15447744 15447745 15447746 15447748 15447749 15447750 15447751 15447752 15447753 15447754 15447755 15447757 15447758 15447759 15447760 15447761 15447762 15447763 15447764 15447765 15447766 15447767 15447768 15447769 15447770 15447771 15447772 15447773 15447774 15447775 15447776 15447777 15447778 15447779 15447780 15447782 15447783 15447784 15447785 15447786 15447787 15447788 15447789 15447790 15447791 15447793 15447794 15447795 15447796 15447797 15447798 15447799 15447800 15447802 15447803 15447804 15447806 15447807 15447808 15447809 15447810 15447812 15447814 15447815 15447816 15447817 15447818 15447820 15447821 15447823 15447824 15447826 15447827 15447828 15447829 15447831 15447832 15447833 15447834 15447835 15447836 15447837 15447838 15447839 15447840 15447841 15447842 15447843 15447845 15447846 15447847 15447848 15447849 15447850 15447851 15447852 15447853 15447854 15447855 15447856 15447857 15447858 15447859 15447860 15447861 15447862 15447863 15447864 15447865 15447866 15447868 15447869 15447870 15447871 15447872 15447873 15447874 15447875 15447876 15447877 15447878 15447879 15447880 15447881 15447882 15447883 15447884 15447885 15447886 15447887 15447888 15447889 15447890 15447891 15447893 15447894 15447895 15447896 15447897 15447898 15447899 15447900 15447901 15447902 15447903 15447904 15447905 15447906 15447907 15447909 15447910 15447911 15447913 15447914 15447915 15447916 15447917 15447918 15447919 15447920 15447921 15447923 15447924 15447926 15447927 15447928 15447929 15447930 15447931 15447932 15447933 15447934 15447935 15447936 15447937 15447938 15447940 15447941 15447942 15447943 15447944 15447945 15447946 15447947 15447948 15447949 15447950 15447951 15447952 15447953 15447955 15447956 15447957 15447959 15447960 15447961 15447962 15447963 15447964 15447965 15447966 15447967 15447968 15447969 15447970 15447971 15447973 15447974 15447975 15447976 15447977 15447978 15447979 15447980 15447981 15447983 15447984 15447985 15447986 15447987 15447988 15447989 15447990 15447991 15447992 15447993 15447994 15447995 15447997 15447998 15447999 15448000 15448001 15448002 15448003 15448004 15448005 15448007 15448008 15448009 15448010 15448011 15448013 15448014 15448015 15448016 15448017 15448018 15448019 15448020 15448021 15448022 15448024 15448025 15448026 15448027 15448029 15448030 15448031 15448033 15448035 15448036 15448037 15448038 15448039 15448040 15448041 15448042 15448043 15448046 15448047 15448048 15448049 15448050 15448051 15448053 15448054 15448055 15448056 15448057 15448058 15448059 15448060 15448061 15448062 15448063 15448064 15448065 15448066 15448067 15448068 15448069 15448070 15448071 15448072 15448073 15448074 15448075 15448076 15448077 15448078 15448079 15448080 15448081 15448082 15448083 15448084 15448085 15448087 15448088 15448089 15448090 15448091 15448092 15448093 15448094 15448095 15448097 15448101 15448103 15448104 15448105 15448106 15448107 15448108 15448109 15448110 15448111 15448112 15448113 15448114 15448115 15448117 15448118 15448119 15448120 15448121 15448122 15448123 15448124 15448125 15448126 15448127 15448128 15448130 15448131 15448132 15448133 15448134 15448135 15448136 15448137 15448138 15448142 15448143 15448144 15448145 15448146 15448147 15448149 15448150 15448151 15448152 15448153 15448154 15448155 15448156 15448157 15448158 15448159 15448160 15448161 15448162 15448165 15448167 15448168 15448169 15448170 15448171 15448172 15448173 15448174 15448175 15448176 15448177 15448179 15448180 15448181 15448182 15448183 15448184 15448185 15448187 15448188 15448189 15448190 15448191 15448192 15448193 15448194 15448195 15448196 15448197 15448198 15448199 15448200 15448201 15448202 15448203 15448205 15448206 15448207 15448208 15448210 15448211 15448212 15448213 15448214 15448215 15448217 15448220 15448221 15448222 15448223 15448224 15448225 15448226 15448227 15448229 15448230 15448231 15448232 15448234 15448236 15448237 15448238 15448239 15448240 15448241 15448242 15448243 15448244 15448246 15448247 15448252 15448253 15448254 15448255 15448256 15448257 15448263 15448264 15448265 15448269 15448271 15448272 15448273 15448274 15448275 15448276 15448277 15448278 15448280 15448281 15448283 15448284 15448285 15448286 15448287 15448288 15448289 15448291 15448292 15448293 15448295 15448296 15448297 15448298 15448299 15448302 15448303 15448304 15448305 15448306 15448307 15448309 15448310 15448311 15448312 15448314 15448315 15448316 15448317 15448320 15448321 15448322 15448324 15448325 15448326 15448328 15448329 15448330 15448332 15448333 15448334 15448336 15448337 15448338 15448339 15448340 15448341 15448343 15448344 15448345 15448346 15448347 15448348 15448349 15448350 15448351 15448356 15448359 15448362 15448363 15448364 15448365 15448366 15448367 15448370 15448371 15448372 15448373 15448374 15448375 15448376 15448377 15448378 15448379 15448380 15448381 15448382 15448383 15448384 15448385 15448386 15448387 15448389 15448390 15448391 15448392 15448393 15448395 15448396 15448397 15448398 15448400 15448401 15448402 15448403 15448404 15448406 15448407 15448409 15448411 15448412 15448413 15448414 15448415 15448416 15448417 15448
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- ICU
- Reason for Recall:
- A possible interruption of therapy when an E321 error code occurs while using the infuser, which is caused by battery issues.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
The Plum A+ Infusion Systems are a multipurpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care. The Plum A+ family infusion pumps are self contained, microprocessor based infusion devices. The pumps are available in single channel and in triple channel pumping mechanism types.
Product Codes/Lot Numbers:
List number: 20679 Serial Number 15441019 15441020 15441037 15441044 15441082 15441093 15441118 15441226 15441229 15441230 15441240 15441241 15441244 15441246 15441247 15441249 15441253 15441256 15441258 15441260 15441265 15441266 15441270 15441271 15441272 15441275 15441285 15441303 15441304 15441315 15441318 15441321 15441323 15441325 15441349 15441350 15441367 15441369 15441371 15441372 15441378 15441395 15441399 15441405 15441409 15441411 15441437 15441458 15441489 15441492 15441493 15441502 15441506 15441508 15441513 15441515 15441516 15441520 15441522 15441524 15441526 15441528 15441532 15441533 15441534 15441535 15441536 15441554 15441601 15441604 15441605 15441610 15441611 15441615 15441633 15441637 15441638 15441649 15441704 15441765 15441803 15441816 15441839 15441942 15441960 15441966 15442037 15442062 15442090 15442124 15442126 15442157 15442193 15442197 15442207 15442235 15442280 15442304 15442306 15442327 15442370 15442401 15442408 15442417 15442448 15442457 15442512 15442544 15442663 15442831 15442838 15442841 15442893 15442905 15442920 15442927 15442961 15442974 15442979 15442998 15443001 15443003 15443005 15443006 15443007 15443008 15443010 15443018 15443023 15443025 15443027 15443028 15443030 15443033 15443034 15443039 15443040 15443062 15443067 15443069 15443103 15443119 15443123 15443126 15443130 15443142 15443148 15443154 15443228 15443248 15443317 15443342 15443346 15443365 15443401 15443403 15443426 15443427 15443445 15443474 15443475 15443478 15443518 15443529 15443533 15443535 15443537 15443542 15443589 15443607 15443609 15443614 15443618 15443619 15443626 15443629 15443635 15443637 15443640 15443642 15443650 15443668 15443671 15443683 15443688 15443719 15443744 15443756 15443757 15443759 15443760 15443771 15443785 15443787 15443789 15443799 15443801 15443808 15443820 15443825 15443830 15443833 15443837 15443859 15443865 15443868 15443879 15443880 15443895 15443910 15443919 15443923 15443927 15443928 15443929 15443934 15443935 15443936 15443942 15443943 15443944 15443946 15443947 15443949 15443950 15443951 15443959 15443961 15443962 15443963 15443965 15443970 15443976 15443977 15443978 15443986 15443987 15443988 15443992 15443993 15443996 15443999 15444000 15444002 15444003 15444005 15444008 15444037 15444040 15444041 15444060 15444065 15444067 15444069 15444072 15444084 15444100 15444107 15444109 15444111 15444115 15444116 15444118 15444119 15444120 15444122 15444123 15444124 15444125 15444127 15444128 15444129 15444132 15444133 15444135 15444136 15444137 15444139 15444140 15444141 15444142 15444143 15444144 15444145 15444146 15444147 15444150 15444152 15444153 15444154 15444156 15444157 15444164 15444166 15444167 15444168 15444171 15444173 15444175 15444176 15444184 15444186 15444188 15444190 15444192 15444193 15444196 15444200 15444206 15444211 15444212 15444213 15444215 15444216 15444217 15444218 15444219 15444221 15444224 15444226 15444227 15444228 15444229 15444230 15444231 15444232 15444233 15444235 15444236 15444238 15444241 15444243 15444245 15444247 15444248 15444249 15444250 15444251 15444252 15444254 15444256 15444261 15444262 15444264 15444269 15444272 15444274 15444276 15444277 15444279 15444280 15444281 15444284 15444285 15444286 15444287 15444288 15444289 15444291 15444292 15444293 15444296 15444298 15444299 15444300 15444301 15444303 15444304 15444308 15444311 15444313 15444314 15444317 15444321 15444322 15444325 15444329 15444332 15444333 15444335 15444337 15444338 15444342 15444346 15444349 15444351 15444354 15444361 15444362 15444364 15444366 15444368 15444370 15444373 15444374 15444376 15444383 15444385 15444386 15444388 15444389 15444394 15444395 15444399 15444400 15444402 15444403 15444412 15444418 15444430 15444431 15444449 15444450 15444452 15444455 15444456 15444459 15444462 15444463 15444465 15444472 15444473 15444474 15444475 15444477 15444482 15444483 15444485 15444486 15444488 15444489 15444491 15444492 15444493 15444496 15444497 15444498 15444500 15444503 15444504 15444511 15444514 15444515 15444516 15444517 15444518 15444519 15444524 15444526 15444528 15444531 15444532 15444537 15444540 15444541 15444543 15444546 15444551 15444553 15444554 15444555 15444556 15444557 15444558 15444559 15444560 15444565 15444566 15444569 15444570 15444571 15444573 15444574 15444575 15444576 15444577 15444578 15444581 15444583 15444584 15444585 15444586 15444587 15444590 15444591 15444592 15444594 15444595 15444597 15444600 15444601 15444604 15444606 15444607 15444608 15444609 15444610 15444613 15444615 15444616 15444620 15444621 15444625 15444627 15444628 15444629 15444630 15444637 15444639 15444642 15444644 15444647 15444648 15444649 15444651 15444652 15444653 15444654 15444655 15444656 15444657 15444658 15444659 15444660 15444662 15444663 15444666 15444667 15444671 15444673 15444674 15444676 15444678 15444679 15444680 15444681 15444682 15444690 15444692 15444694 15444695 15444697 15444699 15444701 15444704 15444705 15444706 15444708 15444709 15444710 15444711 15444713 15444714 15444715 15444717 15444718 15444719 15444721 15444722 15444723 15444724 15444725 15444726 15444727 15444729 15444730 15444731 15444733 15444734 15444735 15444738 15444739 15444740 15444743 15444744 15444746 15444747 15444748 15444750 15444752 15444753 15444754 15444755 15444756 15444757 15444761 15444762 15444764 15444767 15444768 15444771 15444775 15444776 15444777 15444778 15444780 15444782 15444783 15444785 15444786 15444787 15444788 15444790 15444791 15444792 15444794 15444796 15444799 15444801 15444802 15444803 15444805 15444808 15444810 15444811 15444815 15444816 15444817 15444819 15444820 15444821 15444822 15444823 15444825 15444827 15444829 15444830 15444831 15444832 15444833 15444834 15444844 15444848 15444851 15444854 15444857 15444858 15444859 15444862 15444863 15444865 15444866 15444868 15444869 15444870 15444871 15444872 15444874 15444875 15444876 15444877 15444878 15444879 15444880 15444881 15444882 15444883 15444884 15444886 15444887 15444888 15444889 15444890 15444891 15444894 15444896 15444898 15444899 15444901 15444902 15444903 15444904 15444908 15444909 15444910 15444914 15444920 15444921 15444925 15444926 15444927 15444928 15444929 15444930 15444931 15444932 15444933 15444937 15444938 15444939 15444940 15444941 15444942 15444943 15444944 15444945 15444948 15444949 15444950 15444951 15444952 15444953 15444957 15444958 15444959 15444960 15444963 15444964 15444965 15444966 15444967 15444969 15444971 15444976 15444980 15444983 15444984 15444986 15444987 15444988 15444990 15444991 15444992 15444994 15444995 15444996 15444999 15445000 15445003 15445006 15445007 15445008 15445009 15445010 15445012 15445013 15445014 15445017 15445018 15445021 15445022 15445023 15445024 15445025 15445026 15445027 15445028 15445029 15445031 15445032 15445033 15445034 15445035 15445036 15445037 15445038 15445039 15445043 15445044 15445045 15445046 15445048 15445050 15445051 15445052 15445053 15445055 15445057 15445060 15445062 15445063 15445064 15445066 15445067 15445068 15445071 15445072 15445073 15445076 15445077 15445081 15445082 15445085 15445087 15445090 15445091 15445092 15445093 15445094 15445098 15445099 15445100 15445101 15445102 15445103 15445106 15445107 15445108 15445110 15445112 15445113 15445114 15445117 15445118 15445119 15445123 15445124 15445125 15445126 15445128 15445129 15445131 15445132 15445133 15445135 15445136 15445138 15445143 15445144 15445145 15445146 15445147 15445150 15445151 15445153 15445155 15445156 15445157 15445158 15445159 15445168 15445169 15445170 15445171 15445174 15445175 15445176 15445177 15445178 15445179 15445184 15445187 15445188 15445189 15445190 15445192 15445193 15445195 15445196 15445197 15445199 15445200 15445201 15445203 15445206 15445207 15445208 15445210 15445211 15445212 15445213 15445214 15445215 15445217 15445218 15445219 15445222 15445223 15445224 15445226 15445228 15445229 15445230 15445232 15445234 15445235 15445238 15445240 15445241 15445242 15445243 15445244 15445245 15445247 15445250 15445251 15445252 15445253 15445254 15445256 15445257 15445260 15445264 15445266 15445268 15445269 15445270 15445271 15445272 15445273 15445275 15445277 15445283 15445284 15445285 15445287 15445288 15445289 15445294 15445297 15445303 15445305 15445306 15445307 15445308 15445309 15445310 15445311 15445312 15445314 15445317 15445318 15445319 15445320 15445322 15445323 15445324 15445325 15445327 15445328 15445332 15445333 15445334 15445337 15445338 15445339 15445340 15445344 15445347 15445350 15445351 15445352 15445353 15445354 15445355 15445357 15445358 15445359 15445360 15445361 15445362 15445364 15445365 15445366 15445367 15445368 15445369 15445370 15445371 15445372 15445373 15445374 15445375 15445376 15445377 15445379 15445385 15445389 15445390 15445391 15445393 15445394 15445397 15445400 15445406 15445407 15445408 15445409 15445411 15445412 15445414 15445415 15445416 15445417 15445419 15445420 15445425 15445427 15445428 15445429 15445430 15445432 15445433 15445434 15445436 15445437 15445438 15445440 15445441 15445444 15445445 15445446 15445448 15445449 15445450 15445451 15445454 15445455 15445456 15445457 15445458 15445459 15445460 15445461 15445462 15445464 15445466 15445467 15445468 15445469 15445471 15445472 15445473 15445474 15445475 15445477 15445478 15445479 15445480 15445481 15445482 15445483 15445484 15445485 15445487 15445488 15445489 15445490 15445491 15445492 15445494 15445495 15445496 15445497 15445498 15445499 15445501 15445503 15445504 15445506 15445507 15445508 15445509 15445510 15445511 15445512 15445513 15445514 15445516 15445517 15445518 15445519 15445520 15445522 15445524 15445525 15445526 15445527 15445528 15445530 15445531 15445532 15445533 15445534 15445535 15445537 15445538 15445539 15445541 15445542 15445543 15445544 15445546 15445547 15445549 15445552 15445553 15445554 15445555 15445556 15445557 15445559 15445561 15445563 15445564 15445565 15445568 15445569 15445571 15445572 15445573 15445574 15445576 15445577 15445579 15445580 15445582 15445583 15445585 15445586 15445588 15445591 15445592 15445593 15445595 15445596 15445597 15445598 15445599 15445600 15445601 15445602 15445603 15445604 15445605 15445606 15445607 15445608 15445609 15445610 15445611 15445612 15445613 15445614 15445615 15445619 15445620 15445622 15445623 15445624 15445625 15445626 15445630 15445631 15445632 15445633 15445634 15445635 15445636 15445637 15445638 15445640 15445641 15445642 15445643 15445644 15445646 15445647 15445648 15445649 15445650 15445651 15445652 15445653 15445654 15445655 15445656 15445657 15445659 15445660 15445661 15445662 15445663 15445667 15445669 15445670 15445671 15445672 15445673 15445674 15445675 15445677 15445678 15445679 15445680 15445681 15445682 15445684 15445685 15445689 15445690 15445692 15445694 15445695 15445696 15445697 15445698 15445699 15445701 15445702 15445703 15445704 15445705 15445706 15445707 15445709 15445710 15445711 15445712 15445714 15445715 15445716 15445717 15445718 15445721 15445722 15445724 15445725 15445726 15445727 15445728 15445729 15445730 15445731 15445732 15445733 15445734 15445735 15445736 15445737 15445738 15445739 15445740 15445741 15445743 15445744 15445745 15445749 15445750 15445751 15445752 15445753 15445754 15445755 15445756 15445757 15445759 15445761 15445762 15445763 15445766 15445767 15445768 15445773 15445774 15445775 15445776 15445777 15445778 15445779 15445780 15445781 15445783 15445784 15445785 15445786 15445788 15445789 15445790 15445791 15445792 15445793 15445795 15445796 15445797 15445798 15445799 15445803 15445804 15445806 15445807 15445808 15445809 15445810 15445811 15445812 15445814 15445815 15445817 15445819 15445820 15445821 15445822 15445823 15445824 15445825 15445826 15445827 15445829 15445830 15445831 15445834 15445835 15445836 15445837 15445838 15445839 15445841 15445842 15445844 15445845 15445847 15445848 15445849 15445851 15445852 15445853 15445855 15445857 15445858 15445859 15445860 15445861 15445863 15445866 15445867 15445868 15445869 15445871 15445872 15445873 15445875 15445876 15445877 15445878 15445879 15445880 15445881 15445884 15445885 15445887 15445889 15445891 15445892 15445896 15445897 15445898 15445900 15445901 15445902 15445903 15445904 15445906 15445907 15445908 15445909 15445910 15445911 15445913 15445914 15445915 15445916 15445917 15445919 15445920 15445921 15445922 15445923 15445924 15445925 15445928 15445929 15445932 15445936 15445938 15445939 15445940 15445941 15445943 15445945 15445946 15445948 15445949 15445950 15445951 15445952 15445954 15445955 15445957 15445958 15445959 15445960 15445961 15445962 15445967 15445970 15445971 15445972 15445973 15445974 15445975 15445976 15445977 15445978 15445980 15445983 15445984 15445986 15445987 15445988 15445989 15445991 15445993 15445994 15445995 15445996 15445997 15445999 15446000 15446001 15446002 15446003 15446004 15446005 15446006 15446008 15446009 15446010 15446011 15446012 15446014 15446016 15446017 15446018 15446019 15446021 15446022 15446024 15446025 15446026 15446027 15446028 15446029 15446030 15446031 15446034 15446037 15446038 15446039 15446040 15446041 15446042 15446043 15446044 15446045 15446046 15446047 15446048 15446049 15446050 15446051 15446052 15446053 15446054 15446055 15446057 15446060 15446064 15446065 15446066 15446067 15446068 15446069 15446070 15446072 15446073 15446074 15446079 15446080 15446083 15446085 15446094 15446095 15446096 15446097 15446099 15446100 15446101 15446102 15446105 15446106 15446107 15446108 15446109 15446111 15446113 15446114 15446115 15446116 15446117 15446119 15446120 15446121 15446122 15446125 15446126 15446128 15446130 15446132 15446133 15446134 15446135 15446137 15446138 15446139 15446140 15446141 15446142 15446145 15446147 15446148 15446149 15446151 15446154 15446155 15446156 15446159 15446160 15446161 15446162 15446163 15446166 15446167 15446168 15446170 15446171 15446172 15446173 15446175 15446176 15446178 15446180 15446181 15446184 15446186 15446189 15446190 15446194 15446195 15446196 15446197 15446198 15446200 15446201 15446203 15446205 15446206 15446207 15446208 15446210 15446211 15446212 15446215 15446216 15446218 15446219 15446220 15446222 15446223 15446224 15446225 15446226 15446227 15446228 15446229 15446230 15446231 15446232 15446233 15446234 15446235 15446236 15446237 15446239 15446240 15446241 15446243 15446244 15446245 15446246 15446247 15446248 15446250 15446251 15446252 15446253 15446254 15446255 15446256 15446258 15446259 15446261 15446262 15446264 15446265 15446266 15446267 15446268 15446271 15446273 15446274 15446275 15446276 15446279 15446281 15446282 15446283 15446286 15446290 15446296 15446297 15446298 15446299 15446300 15446302 15446303 15446304 15446305 15446306 15446307 15446311 15446312 15446313 15446338 15446340 15446341 15446344 15446347 15446348 15446349 15446350 15446351 15446352 15446402 15446456 15446457 15446460 15446462 15446464 15446465 15446466 15446467 15446471 15446472 15446473 15446475 15446476 15446477 15446478 15446479 15446480 15446481 15446482 15446483 15446488 15446489 15446490 15446491 15446494 15446495 15446496 15446498 15446506 15446507 15446508 15446509 15446510 15446511 15446512 15446513 15446514 15446515 15446517 15446518 15446519 15446520 15446521 15446522 15446523 15446524 15446525 15446526 15446529 15446530 15446531 15446532 15446533 15446536 15446537 15446538 15446539 15446540 15446541 15446542 15446543 15446544 15446546 15446547 15446548 15446550 15446551 15446552 15446553 15446554 15446555 15446556 15446557 15446558 15446560 15446561 15446563 15446564 15446565 15446566 15446567 15446568 15446569 15446570 15446571 15446572 15446573 15446574 15446575 15446576 15446577 15446579 15446580 15446581 15446582 15446583 15446584 15446586 15446588 15446589 15446590 15446591 15446592 15446593 15446594 15446595 15446596 15446597 15446598 15446599 15446600 15446601 15446602 15446603 15446604 15446605 15446606 15446607 15446608 15446609 15446610 15446612 15446613 15446614 15446615 15446617 15446618 15446619 15446620 15446621 15446622 15446623 15446624 15446625 15446626 15446627 15446628 15446629 15446630 15446631 15446632 15446633 15446634 15446635 15446636 15446637 15446639 15446640 15446641 15446643 15446644 15446646 15446647 15446648 15446650 15446651 15446652 15446653 15446654 15446655 15446656 15446657 15446658 15446659 15446660 15446661 15446662 15446663 15446664 15446665 15446666 15446667 15446668 15446671 15446672 15446673 15446674 15446675 15446676 15446677 15446678 15446679 15446680 15446681 15446682 15446683 15446685 15446686 15446687 15446688 15446690 15446692 15446694 15446695 15446697 15446701 15446702 15446703 15446704 15446705 15446706 15446707 15446708 15446709 15446710 15446712 15446714 15446716 15446717 15446719 15446720 15446721 15446722 15446723 15446724 15446725 15446726 15446727 15446728 15446729 15446731 15446732 15446733 15446734 15446735 15446736 15446737 15446739 15446740 15446741 15446742 15446744 15446745 15446746 15446748 15446749 15446750 15446751 15446752 15446754 15446755 15446756 15446758 15446759 15446760 15446761 15446762 15446763 15446764 15446767 15446768 15446769 15446770 15446771 15446772 15446773 15446775 15446776 15446778 15446779 15446781 15446782 15446783 15446784 15446785 15446786 15446787 15446788 15446789 15446790 15446791 15446792 15446795 15446796 15446797 15446798 15446799 15446800 15446801 15446802 15446803 15446804 15446805 15446808 15446809 15446810 15446811 15446812 15446813 15446815 15446816 15446817 15446818 15446819 15446820 15446821 15446822 15446823 15446824 15446825 15446826 15446827 15446828 15446829 15446830 15446831 15446832 15446833 15446834 15446835 15446836 15446837 15446838 15446839 15446840 15446841 15446842 15446843 15446844 15446845 15446846 15446847 15446848 15446849 15446850 15446851 15446852 15446853 15446854 15446855 15446856 15446857 15446858 15446859 15446860 15446861 15446862 15446863 15446864 15446865 15446866 15446867 15446868 15446869 15446870 15446872 15446873 15446874 15446875 15446876 15446877 15446879 15446880 15446881 15446883 15446885 15446886 15446887 15446889 15446890 15446893 15446894 15446895 15446896 15446898 15446900 15446901 15446902 15446903 15446904 15446906 15446907 15446909 15446910 15446911 15446913 15446914 15446915 15446916 15446917 15446918 15446919 15446920 15446923 15446924 15446925 15446926 15446927 15446930 15446931 15446932 15446933 15446934 15446936 15446937 15446938 15446939 15446940 15446941 15446942 15446943 15446945 15446946 15446947 15446948 15446949 15446950 15446951 15446952 15446953 15446954 15446955 15446956 15446957 15446959 15446960 15446961 15446963 15446964 15446965 15446966 15446967 15446968 15446969 15446970 15446971 15446972 15446974 15446976 15446977 15446978 15446979 15446981 15446982 15446983 15446984 15446985 15446986 15446991 15446992 15446993 15446994 15446995 15446996 15446997 15446998 15446999 15447000 15447001 15447002 15447003 15447004 15447007 15447008 15447010 15447011 15447012 15447013 15447014 15447015 15447016 15447017 15447018 15447019 15447020 15447021 15447022 15447023 15447025 15447027 15447028 15447029 15447030 15447031 15447032 15447033 15447034 15447037 15447038 15447039 15447041 15447042 15447043 15447044 15447045 15447047 15447048 15447049 15447051 15447052 15447055 15447056 15447059 15447060 15447061 15447063 15447064 15447065 15447066 15447067 15447068 15447069 15447070 15447071 15447072 15447073 15447074 15447075 15447076 15447077 15447078 15447081 15447082 15447084 15447085 15447086 15447087 15447088 15447089 15447090 15447092 15447093 15447094 15447096 15447098 15447100 15447101 15447102 15447103 15447105 15447106 15447107 15447108 15447109 15447110 15447111 15447116 15447117 15447121 15447122 15447124 15447126 15447127 15447128 15447129 15447130 15447131 15447133 15447134 15447137 15447140 15447141 15447142 15447143 15447146 15447147 15447149 15447150 15447151 15447152 15447153 15447154 15447155 15447156 15447158 15447159 15447160 15447161 15447162 15447163 15447164 15447165 15447166 15447167 15447169 15447170 15447171 15447172 15447173 15447174 15447175 15447177 15447178 15447179 15447180 15447181 15447182 15447183 15447184 15447185 15447186 15447187 15447188 15447189 15447190 15447191 15447192 15447193 15447194 15447195 15447196 15447197 15447199 15447200 15447201 15447202 15447203 15447204 15447205 15447206 15447207 15447210 15447211 15447212 15447213 15447214 15447216 15447217 15447219 15447220 15447221 15447222 15447223 15447224 15447225 15447226 15447227 15447228 15447230 15447232 15447233 15447234 15447236 15447237 15447240 15447241 15447245 15447246 15447247 15447249 15447250 15447251 15447252 15447253 15447254 15447255 15447257 15447258 15447259 15447261 15447262 15447263 15447264 15447265 15447266 15447268 15447270 15447271 15447272 15447273 15447275 15447276 15447278 15447279 15447280 15447283 15447284 15447285 15447286 15447287 15447290 15447291 15447292 15447293 15447294 15447295 15447296 15447297 15447298 15447300 15447301 15447302 15447303 15447304 15447305 15447306 15447307 15447308 15447309 15447310 15447311 15447312 15447313 15447314 15447315 15447316 15447317 15447318 15447319 15447320 15447321 15447322 15447323 15447324 15447325 15447326 15447327 15447328 15447329 15447330 15447331 15447332 15447333 15447334 15447335 15447336 15447337 15447339 15447340 15447341 15447342 15447344 15447345 15447346 15447347 15447348 15447349 15447350 15447351 15447356 15447358 15447359 15447360 15447361 15447362 15447363 15447365 15447366 15447367 15447368 15447369 15447370 15447372 15447374 15447375 15447376 15447379 15447381 15447382 15447384 15447385 15447386 15447387 15447388 15447389 15447390 15447391 15447392 15447393 15447395 15447396 15447397 15447398 15447399 15447400 15447401 15447402 15447403 15447404 15447406 15447408 15447410 15447411 15447412 15447413 15447414 15447418 15447419 15447420 15447421 15447422 15447423 15447424 15447425 15447426 15447427 15447428 15447430 15447431 15447432 15447433 15447434 15447436 15447437 15447438 15447440 15447441 15447442 15447443 15447444 15447445 15447447 15447448 15447449 15447451 15447452 15447453 15447454 15447455 15447457 15447458 15447459 15447460 15447461 15447462 15447463 15447464 15447465 15447466 15447467 15447468 15447469 15447470 15447471 15447472 15447473 15447474 15447476 15447477 15447480 15447481 15447482 15447483 15447484 15447485 15447486 15447487 15447488 15447489 15447490 15447492 15447501 15447506 15447507 15447508 15447510 15447511 15447512 15447513 15447514 15447515 15447517 15447518 15447519 15447521 15447522 15447523 15447524 15447526 15447527 15447528 15447529 15447530 15447531 15447532 15447533 15447534 15447535 15447536 15447537 15447540 15447542 15447543 15447544 15447545 15447548 15447550 15447552 15447553 15447554 15447555 15447559 15447560 15447561 15447563 15447564 15447565 15447566 15447567 15447568 15447569 15447570 15447571 15447573 15447574 15447576 15447577 15447579 15447580 15447581 15447582 15447583 15447584 15447585 15447587 15447588 15447589 15447590 15447591 15447592 15447593 15447594 15447595 15447596 15447597 15447598 15447599 15447601 15447602 15447603 15447606 15447609 15447610 15447611 15447612 15447613 15447614 15447616 15447617 15447618 15447619 15447620 15447621 15447622 15447623 15447624 15447625 15447626 15447627 15447628 15447630 15447631 15447632 15447633 15447634 15447636 15447638 15447640 15447641 15447642 15447643 15447644 15447645 15447646 15447647 15447648 15447649 15447650 15447651 15447652 15447653 15447654 15447655 15447657 15447658 15447659 15447660 15447661 15447663 15447665 15447666 15447667 15447668 15447669 15447670 15447671 15447672 15447673 15447674 15447675 15447676 15447677 15447678 15447679 15447680 15447681 15447682 15447683 15447684 15447685 15447686 15447687 15447689 15447690 15447691 15447692 15447693 15447694 15447695 15447696 15447697 15447698 15447699 15447700 15447701 15447702 15447704 15447705 15447706 15447707 15447708 15447709 15447710 15447712 15447713 15447714 15447715 15447716 15447717 15447718 15447719 15447720 15447722 15447723 15447725 15447726 15447727 15447728 15447729 15447730 15447731 15447732 15447733 15447734 15447735 15447736 15447737 15447738 15447739 15447740 15447741 15447742 15447744 15447745 15447746 15447748 15447749 15447750 15447751 15447752 15447753 15447754 15447755 15447757 15447758 15447759 15447760 15447761 15447762 15447763 15447764 15447765 15447766 15447767 15447768 15447769 15447770 15447771 15447772 15447773 15447774 15447775 15447776 15447777 15447778 15447779 15447780 15447782 15447783 15447784 15447785 15447786 15447787 15447788 15447789 15447790 15447791 15447793 15447794 15447795 15447796 15447797 15447798 15447799 15447800 15447802 15447803 15447804 15447806 15447807 15447808 15447809 15447810 15447812 15447814 15447815 15447816 15447817 15447818 15447820 15447821 15447823 15447824 15447826 15447827 15447828 15447829 15447831 15447832 15447833 15447834 15447835 15447836 15447837 15447838 15447839 15447840 15447841 15447842 15447843 15447845 15447846 15447847 15447848 15447849 15447850 15447851 15447852 15447853 15447854 15447855 15447856 15447857 15447858 15447859 15447860 15447861 15447862 15447863 15447864 15447865 15447866 15447868 15447869 15447870 15447871 15447872 15447873 15447874 15447875 15447876 15447877 15447878 15447879 15447880 15447881 15447882 15447883 15447884 15447885 15447886 15447887 15447888 15447889 15447890 15447891 15447893 15447894 15447895 15447896 15447897 15447898 15447899 15447900 15447901 15447902 15447903 15447904 15447905 15447906 15447907 15447909 15447910 15447911 15447913 15447914 15447915 15447916 15447917 15447918 15447919 15447920 15447921 15447923 15447924 15447926 15447927 15447928 15447929 15447930 15447931 15447932 15447933 15447934 15447935 15447936 15447937 15447938 15447940 15447941 15447942 15447943 15447944 15447945 15447946 15447947 15447948 15447949 15447950 15447951 15447952 15447953 15447955 15447956 15447957 15447959 15447960 15447961 15447962 15447963 15447964 15447965 15447966 15447967 15447968 15447969 15447970 15447971 15447973 15447974 15447975 15447976 15447977 15447978 15447979 15447980 15447981 15447983 15447984 15447985 15447986 15447987 15447988 15447989 15447990 15447991 15447992 15447993 15447994 15447995 15447997 15447998 15447999 15448000 15448001 15448002 15448003 15448004 15448005 15448007 15448008 15448009 15448010 15448011 15448013 15448014 15448015 15448016 15448017 15448018 15448019 15448020 15448021 15448022 15448024 15448025 15448026 15448027 15448029 15448030 15448031 15448033 15448035 15448036 15448037 15448038 15448039 15448040 15448041 15448042 15448043 15448046 15448047 15448048 15448049 15448050 15448051 15448053 15448054 15448055 15448056 15448057 15448058 15448059 15448060 15448061 15448062 15448063 15448064 15448065 15448066 15448067 15448068 15448069 15448070 15448071 15448072 15448073 15448074 15448075 15448076 15448077 15448078 15448079 15448080 15448081 15448082 15448083 15448084 15448085 15448087 15448088 15448089 15448090 15448091 15448092 15448093 15448094 15448095 15448097 15448101 15448103 15448104 15448105 15448106 15448107 15448108 15448109 15448110 15448111 15448112 15448113 15448114 15448115 15448117 15448118 15448119 15448120 15448121 15448122 15448123 15448124 15448125 15448126 15448127 15448128 15448130 15448131 15448132 15448133 15448134 15448135 15448136 15448137 15448138 15448142 15448143 15448144 15448145 15448146 15448147 15448149 15448150 15448151 15448152 15448153 15448154 15448155 15448156 15448157 15448158 15448159 15448160 15448161 15448162 15448165 15448167 15448168 15448169 15448170 15448171 15448172 15448173 15448174 15448175 15448176 15448177 15448179 15448180 15448181 15448182 15448183 15448184 15448185 15448187 15448188 15448189 15448190 15448191 15448192 15448193 15448194 15448195 15448196 15448197 15448198 15448199 15448200 15448201 15448202 15448203 15448205 15448206 15448207 15448208 15448210 15448211 15448212 15448213 15448214 15448215 15448217 15448220 15448221 15448222 15448223 15448224 15448225 15448226 15448227 15448229 15448230 15448231 15448232 15448234 15448236 15448237 15448238 15448239 15448240 15448241 15448242 15448243 15448244 15448246 15448247 15448252 15448253 15448254 15448255 15448256 15448257 15448263 15448264 15448265 15448269 15448271 15448272 15448273 15448274 15448275 15448276 15448277 15448278 15448280 15448281 15448283 15448284 15448285 15448286 15448287 15448288 15448289 15448291 15448292 15448293 15448295 15448296 15448297 15448298 15448299 15448302 15448303 15448304 15448305 15448306 15448307 15448309 15448310 15448311 15448312 15448314 15448315 15448316 15448317 15448320 15448321 15448322 15448324 15448325 15448326 15448328 15448329 15448330 15448332 15448333 15448334 15448336 15448337 15448338 15448339 15448340 15448341 15448343 15448344 15448345 15448346 15448347 15448348 15448349 15448350 15448351 15448356 15448359 15448362 15448363 15448364 15448365 15448366 15448367 15448370 15448371 15448372 15448373 15448374 15448375 15448376 15448377 15448378 15448379 15448380 15448381 15448382 15448383 15448384 15448385 15448386 15448387 15448389 15448390 15448391 15448392 15448393 15448395 15448396 15448397 15448398 15448400 15448401 15448402 15448403 15448404 15448406 15448407 15448409 15448411 15448412 15448413 15448414 15448415 15448416 15448417 15448
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2682-2017
Related Recalls
A possible interruption of therapy when an E321 error code occurs while using the infuser, which is caused by battery issues.
A possible interruption of therapy when an E321 error code occurs while using the infuser, which is caused by battery issues.
A possible interruption of therapy when an E321 error code occurs while using the infuser, which is caused by battery issues.