SHIMADZU FDR Visionary Catalog Number: FDR Visionary Suite Stationary x-ray system for taking general radiographic and tomographic images of the whole body. The device is intended to be used in hospitals, clinics, imaging centers, and/or other healthcare facilities by qualified/trained professionals
Class I - DangerousWhat Should You Do?
- Check if you have this product: Serial No. 0162M4501 thru 0562M45103, 3ZC5C2A0C002 thru 3ZC5C1734001, D362M42902 thru 0562M41101, 41E733E69003, MP95A8F5A001 thru MP95A9F6B001
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Shimadzu Medical Systems Usa Com
- Reason for Recall:
- Cracks may occur over time on the tube holding shaft near the base of the tube mounting flange.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
SHIMADZU FDR Visionary Catalog Number: FDR Visionary Suite Stationary x-ray system for taking general radiographic and tomographic images of the whole body. The device is intended to be used in hospitals, clinics, imaging centers, and/or other healthcare facilities by qualified/trained professionals
Product Codes/Lot Numbers:
Serial No. 0162M4501 thru 0562M45103, 3ZC5C2A0C002 thru 3ZC5C1734001, D362M42902 thru 0562M41101, 41E733E69003, MP95A8F5A001 thru MP95A9F6B001
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2655-2017
Related Recalls
In these units, the brake is usually released by gripping the drive handle and applied by letting go of the handle. However, for systems within a specified manufacture period, we have identified that even when letting go of the handle, the systems brakes are not applied. In this event, if the Operator releases the handle when the device is moving, the unit will continue to coast.
Normal operation of the device is to power up the device in the morning, register the first patient, and then perform a fluoroscopy. If, however, the user powers up the device in the morning and makes an error by starting a fluoroscopy while the first patient is still being registered, the device application will abnormally terminate and require service intervention before it can be used again. This event will not occur after the first patient procedure.
TRINIAS for diagnostic imaging and interventional procedures in cardiac angiography, neurovascular angiography, abdominal angiography, peripheral angiography, rotational angiography, multi-purpose angiography and whole body radiographic/fluoroscopic procedures.
Shimadzu Medical Systems Usa Com
Two issues: Event 1: Normal operation of the device is to power up the device in the morning, register the first patient, and then perform a fluoroscopy. If, however, the user powers up the device in the morning and makes an error by starting a fluoroscopy while the first patient is still being registered, the device application will abnormally terminate and require service intervention before it can be used again. This event will not occur after the first patient procedure. Event 2: In rare cases a synchronization error may occur in the data transmission circuit in the device due to external noise. The visibility of the displayed image may be corrupted (images become bit-shifted abnormal images). If this event occurs, it may become difficult to see the object, which may hinder the examination and treatment.