Brilliance 16, Model number 728246 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    728246 -- Brilliance 16 3010 3017 3031 3040 3043 3070 3072 3085 3090 3117 3149 3216 3247 3293 3314 5006 5012 5018 5020 5024 5026 5028 5030 5031 5035 5044 5045 5052 5053 5054 5055 5058 5059 5060 5064 5069 5070 5081 5086 5088 5091 5093 5097 5099 5103 5107 5108 5110 5111 5115 5116 5118 5120 5123 5126 5127 5128 5131 5138 5140 5155 5160 5161 5163 5166 5184 5186 5188 5190 5191 5192 5193 5195 5197 5208 5209 5216 5219 5235 5236 5238A 5270 5272 5279 5282 5285 5287 5294 5295 5315 5316 5350 5354 5357 5360 5368 5394 5433 5434 5443 5453 5457 5480 5487 5501 5502 5505 5519 5523 5526 5554 5572 5573 5576 5609 5615 5633 5647 5653 5683 5723 5758 5762 5764 5771 5811 5897 5929 5937 5956 5974 5995 6021 6022 6031 6033 6048 6053 6072 6220 6229 6243 6280 6283 6298 6309 6321 6360 6375 6405 6426 6441 6465 6474 6484 6503 6511 6517 6519 6523 6549 6559 30049 32287 50011 50053 50087 50180 3123 5674 6555 32366 3465 50181 5135 5237 3179 3411 5009 5048 5077 5412 6154 30011 30110 5101 5181 3324 5405 3050 3441 6387
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Philips Medical Systems (Cleveland) Inc
Reason for Recall:
After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at iso-center (0,0). If the region of interest is not visualized in the images, a CT rescan may be performed.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Brilliance 16, Model number 728246 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.

Product Codes/Lot Numbers:

728246 -- Brilliance 16 3010 3017 3031 3040 3043 3070 3072 3085 3090 3117 3149 3216 3247 3293 3314 5006 5012 5018 5020 5024 5026 5028 5030 5031 5035 5044 5045 5052 5053 5054 5055 5058 5059 5060 5064 5069 5070 5081 5086 5088 5091 5093 5097 5099 5103 5107 5108 5110 5111 5115 5116 5118 5120 5123 5126 5127 5128 5131 5138 5140 5155 5160 5161 5163 5166 5184 5186 5188 5190 5191 5192 5193 5195 5197 5208 5209 5216 5219 5235 5236 5238A 5270 5272 5279 5282 5285 5287 5294 5295 5315 5316 5350 5354 5357 5360 5368 5394 5433 5434 5443 5453 5457 5480 5487 5501 5502 5505 5519 5523 5526 5554 5572 5573 5576 5609 5615 5633 5647 5653 5683 5723 5758 5762 5764 5771 5811 5897 5929 5937 5956 5974 5995 6021 6022 6031 6033 6048 6053 6072 6220 6229 6243 6280 6283 6298 6309 6321 6360 6375 6405 6426 6441 6465 6474 6484 6503 6511 6517 6519 6523 6549 6559 30049 32287 50011 50053 50087 50180 3123 5674 6555 32366 3465 50181 5135 5237 3179 3411 5009 5048 5077 5412 6154 30011 30110 5101 5181 3324 5405 3050 3441 6387

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2655-2016

Related Recalls

Due to a software issue, there is a potential image error of the Region of Interest for expansion/contraction for HFP (Head First Prone), FFS (Feet First Supine) and FFP (Feet First Prone) orientations.

Aug 5, 2025 Other Medical Devices Nationwide View Details →