3M Attest Super Rapid 5 Steam-Plus Steam Challenge Pack, REF 41482VF; designed to qualify and monitor dynamic-air removal steam sterilization processes
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI/DI: 50707387787283( (shipper); Lot Codes: 41482VF, exp. 33P4ND; 33P4YN, exp. 01/02/2025
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- 3M Company - Health Care Business
- Reason for Recall:
- Specific lots of 3M Attest Super Rapid Readout Biological Indicators (1492V), 3M Attest Super Rapid 5 Steam-Plus Challenge (41482V, 41482VF), and 3M Attest Super Rapid Readout Steam Challenge (1496V) may contain biological indicators with a cap that melts, deforms and/or cracks following a steam sterilization cycle of 132¿C and 135¿C.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
3M Attest Super Rapid 5 Steam-Plus Steam Challenge Pack, REF 41482VF; designed to qualify and monitor dynamic-air removal steam sterilization processes
Product Codes/Lot Numbers:
UDI/DI: 50707387787283( (shipper); Lot Codes: 41482VF, exp. 33P4ND; 33P4YN, exp. 01/02/2025
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2640-2023
Related Recalls
Clarity Precision Grip Attachments, for use in orthodontic treatment. Part numbers/SKUs: 359-830 359-831 359-832 359-833 359-834 359-835 359-836 359-837
3M Company - Health Care Business
Attachments may not mate with the tooth as intended in the digital treatment design.
3M Tegaderm CHG, Chlorhexidine Gluconate I.V. Securement Dressing, REF 1658R
3M Company - Health Care Business
Contamination with foreign substance matter
3M Durapore Surgical Tape, Catalog Number 1538-118
3M Company - Health Care Business
Select lots of the 3M Durapore Surgical Tape, Catalog Number 1538-118, were incorrectly labeled with a shelf life of 5 years, rather than 3 years.