ProVue Instrument Driver for JResultNet (ProVue.jrm) v. 1.5.3 and below JResultNet is a stand-alone software package that provides connectivity between in-vitro diagnostic instruments (IVDs) and Laboratory Information Systems (LIS) aiding the flow of laboratory information, including test orders and results, specimen information, and patient information.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Version #s: 1.0.0 Date Released 6/17/2004, 1.0.1 Date Released 8/18/2004, 1.1.1 Date Released 9/9/2004, 1.2.0 Date Released 10/11/2004, 1.3.0 Date Released 11/1/2004, 1.3.1 Date Released 1/20/2005, 1.3.2 Date Released 2/18/2005, 1.4.0 Date Released 3/30/2005, 1.4.1 Date Released 8/31/2005, 1.4.2 Date Released 9/12/2005, 1.4.2 Date Released 10/5/2005, 1.4.3 Date Released 2/8/2006, 1.4.4 Date Released 2/20/2006, 1.4.5 Date Released 3/3/2006, 1.4.5 Date Released 3/23/2006, 1.4.5 Date Released 5/17/2006, 1.5.0 Date Released 6/26/2007, 1.5.1 Date Released 9/17/2007, 1.5.1 Date Released 12/13/2007, 1.5.1 Date Released 11/7/2008, 1.5.2 Date Released 4/13/2009, 1.5.3 Date Released 1/12/2011.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Data Innovations, LLC
Reason for Recall:
Potential issues with the ProVue driver have been identified which may cause data to be misassociated with patient results under a given set of circumstances.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

ProVue Instrument Driver for JResultNet (ProVue.jrm) v. 1.5.3 and below JResultNet is a stand-alone software package that provides connectivity between in-vitro diagnostic instruments (IVDs) and Laboratory Information Systems (LIS) aiding the flow of laboratory information, including test orders and results, specimen information, and patient information.

Product Codes/Lot Numbers:

Version #s: 1.0.0 Date Released 6/17/2004, 1.0.1 Date Released 8/18/2004, 1.1.1 Date Released 9/9/2004, 1.2.0 Date Released 10/11/2004, 1.3.0 Date Released 11/1/2004, 1.3.1 Date Released 1/20/2005, 1.3.2 Date Released 2/18/2005, 1.4.0 Date Released 3/30/2005, 1.4.1 Date Released 8/31/2005, 1.4.2 Date Released 9/12/2005, 1.4.2 Date Released 10/5/2005, 1.4.3 Date Released 2/8/2006, 1.4.4 Date Released 2/20/2006, 1.4.5 Date Released 3/3/2006, 1.4.5 Date Released 3/23/2006, 1.4.5 Date Released 5/17/2006, 1.5.0 Date Released 6/26/2007, 1.5.1 Date Released 9/17/2007, 1.5.1 Date Released 12/13/2007, 1.5.1 Date Released 11/7/2008, 1.5.2 Date Released 4/13/2009, 1.5.3 Date Released 1/12/2011.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2637-2014

Related Recalls

Quality control results are not processed for Abbott Diagnostics Alinity ci series (abdicqii) instrument. This has been determined to be a driver malfunction. When an instrument transmits results from a quality control (QC) sample ID ("Q^Control" in SPM segment), and the option Hold Results Until Complete is enabled, the results do not parse into IM. The communication Trace shows the sample being transmitted, but no information is logged in the Specimen Event Log (SEL). There is a potential for harm if QC results are not processed, downstream auto verification rules may misfire based on missing QC results leading to a delay in, or incorrect treatment. QC results will not be available for evaluation by the lab. There have been no reported instances of patient harm from this malfunction.

Aug 29, 2018 Infusion Pumps View Details →

Software issue resulting in the results of one patient potentially being associated with another.

Oct 23, 2014 Diagnostic Equipment Nationwide View Details →

When connected to the Sysmex UF-100, the Instrument Manager driver sysu100i driver software versions 7.00.0001, 7.00.0002, 7.00.0003, 7.00.0004, 7.00.0005, 8.00.0001, 8.00.0002, 8.00.0003, and 8.00.0004, and is configured with option "Strip Zeros from Result" set to "Yes" may result in sample error

Nov 6, 2013 Surgical Instruments Nationwide View Details →