SUN DASH RADIUS 252 PN 8000809705. The device is intended to be used for the tanning of human skin.
Class I - DangerousWhat Should You Do?
- Check if you have this product: PN 8000809705; Serial numbers 248-1298
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
β οΈ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- JK Products & Services, Inc
- Reason for Recall:
- The integrity of the wires inside main electrical chord become compromised from excessive flexing (during the door open/close process).
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
SUN DASH RADIUS 252 PN 8000809705. The device is intended to be used for the tanning of human skin.
Product Codes/Lot Numbers:
PN 8000809705; Serial numbers 248-1298
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2635-2014
Related Recalls
SUN|DASH RADIUS 252 tanning system; 52 lamps, Ultraviolet Radiation. The device is intended to be used for tanning of the human skin.
JK Products & Services
The firm has tracked a systematic failure in the door cord on the Sun|Dash Radius 252 product line. In some instances when the door cord is installed with little or no slack outside the upper decor piece, the cord has been known to fail causing a short in the wiring at the pivot point in the cord.
BeautyAngel Health Fitness Natural Collagen Stimulation Anti Aging For human use.
JK Products & Services
The Beauty Angel was marketed without a 510k.