Gallant VR Implantable Cardioverter Defibrillator: VR ICD REF CDVRA500Q Gallant DR Implantable Cardioverter Defibrillator: DR ICD REF CDDRA500Q Neutrino VR Implantable Cardioverter Defibrillator: NxT VR ICD REF CDVRA600Q Neutrino DR Implantable Cardioverter Defibrillator: NxT DR ICD REF CDDRA600Q Entrant VR Implantable Cardioverter Defibrillator: VR ICD REF CDVRA300Q Entrant DR Implantable Cardioverter Defibrillator: DR ICD REF CDDRA300Q The primary function of ICDs and CRT-Ds are to treat life-threatening ventricular tachyarrhythmias via anti-tachycardia pacing (ATP), or cardioversion/defibrillation. Implantable Cardioverter Defibrillators (ICDs) can either be a dual chamber (also known as DR if rate responsive) device that is connected to a right atrial (RA) pacing lead and right ventricular defibrillation lead or single chamber ICD (also known as VR if rate responsive) with only a defibrillation lead in the right ventricle (RV). The ICD, more generally known as an implantable pulse generator (IPG), is typically implanted in a subcutaneous device pocket located in the pectoral region and is attached to the pacing/defibrillator leads that are implanted through the transvenous technique with the distal tips positioned in the RA (dual chamber) and RV. Cardiac Resynchronization Therapy Defibrillator (CRT-D) systems consist of a right atrial (RA) pacing lead, a right ventricular (RV) defibrillation lead, as well as a third lead placed in the coronary venous system overlying the left ventricle (LV) to improve ventricular resynchronization through biventricular pacing for treatment of heart failure. CRT-D devices, more generally known as an implantable pulse generator (IPG), are typically implanted in a subcutaneous device pocket located in the pectoral region and are attached to the pacing/defibrillator leads that are implanted through transvenous technique.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Gallant: Model Number: VR ICD REF CDVRA500Q UDI-DI Code: 05415067031990 Serial Numbers: 111033498 810000696 810002524 111031208 111025117 111030959 111006327 111028481 111014296 111033775 810021414 810001946 810001996 810002011 810001942 810007459 111036536 810019608 810019393 810019581 111042257 810026722 810028531 810028585 810027003 111011936 810021423 111025530 111015227 111036960 810026790 111036957 111036933 111036964 111007506 810021413 810021445 810021449 111033604 111000553 111000558 111000567 111000472 111000706 111000667 111006793 111000664 111000678 111018664 111018644 111018802 111012460 111006783 111006791 111006789 111000786 111000582 111000550 111000721 111001035 111001025 111014950 111021456 111021461 111021703 111021722 111021450 111036417 111012272 111011597 810001013 810001011 810006759 810001046 810001021 810019014 111024417 810002439 111024415 810006774 111024454 810022318 111036914 810013066 810015963 111002002 810021662 810013192 810011364 111003380 810011481 111020960 111011436 111020992 810015820 810015794 111009835 111004271 111003046 111024175 111022969 111009840 810019313 810018954 111021002 111009844 810008620 111010818 810018999 111021710 810016110 810019437 810008498 111024550 810008497 810019426 810011359 810019424 810011511 810011353 810011328 111011181 111011372 111021007 111021712 111024183 810013436 810019054 810019546 810022117 810027515 810013071 810007469 111020996 111009733 111003861 111009782 810011335 810016103 111021851 810004570 111008584 111009245 111003267 810008499 111021988 111012318 111003040 111010800 111003332 111022768 810016526 810027041 111009747 810013248 111011899 111010457 810021351 810021705 810015780 111018656 111010771 111021437 810005318 111010761 111022974 111021994 810007454 810008483 810011285 810011458 810005317 111008582 111003173 111008588 111003555 111018654 810027111 810019552 810016459 111009820 810008557 810026959 111002712 111021707 111006776 111006779 111010814 111017230 810021576 810016112 111003506 810027040 810027014 810003350 111009781 111018131 810021674 111011908 810008527 810019917 111008583 810008575 111004282 810016339 810027099 810015815 810016102 111002437 111011229 111018270 810021707 810013068 810013216 810015840 810018976 810021355 810011459 111003500 111019459 810003430 111009834 111010787 111000447 111004193 111009841 111013996 111011233 111011254 111021837 111019423 810011451 111018786 810016113 810016649 810013446 810013250 810013442 810027039 810026943 810026944 810027020 810027165 111009759 111006801 111021440 111010437 111019431 810011457 810016665 810016449 810019012 810021618 810008573 810021560 810011535 111009788 810013454 810008650 810015824 111006778 810005319 111024379 810013349 810004013 111009809 810008519 111024368 810003553 111024378 111006784 810016140 810008486 111006797 810013266 111018109 111018795 111022789 111011910 810021606 810007463 810007421 810003677 111011216 810027514 111018640 810019556 810008612 111018784 111024402 810003437 810011343 810008526 111004721 111021995 810027548 810018995 810016138 810019427 810005706 111022585 810011387 111018291 111011963 111009708 111010450 111017239 111022090 111024180 111011906 111021723 111021436 810022090 810015784 111009794 111021433 810003343 810013241 111010980 810027103 111011915 111002410 810027161 111001995 111009741 111006541 111002011 810016121 111004259 111003960 810021559 111024204 810013232 111016740 810027168 810015796 111001027 111012290 111002434 111010962 111001969 810015822 111016745 111024202 111012258 111010952 111024184 111013804 111022788 111010465 810019529 810011303 111013808 810011549 810011538 111021812 810005366 810007473 111023732 111014948 111013807 111001583 111023720 111022587 111011938 810026979 111012462 111012307 111003869 111024186 810016277 111004242 810005286 111022781 111021845 810004846 810015783 111019446 810022064 111010799 111011214 810022363 111009828 810008597 111021790 111003559 810016560 111011185 810013225 111013993 810013206 111011922 111003276 810018982 111003858 810013077 810011449 810011550 810016303 111011237 111019415 111003865 111006785 111003863 111004182 111021466 810004849 111012256 111006775 810022137 810019296 810019297 111018120 111018679 810003615 810008594 111003065 111003037 810022370 810022129 810019042 111004513 111003074 111022971 111003957 810005356 111004518 111023531 111003371 111014919 111009837 111003539 111002743 810019250 810011523 810019254 111021846 111022077 810021666 810022122 810021671 810019272 810019301 810019290 810019278 810019065 810019325 810019408 810019310 810019293 810019009 111009737 111011215 111009766 111009777 111003521 111010981 111011968 111009261 111003686 111017247 111017245 111003567 111003505 810015772 810015773 111004497 111017234 810013185 810019050 810011339 810022145 810022376 810022079 810022078 810021530 810021687 810021563 810018946 810019017 810016151 810027033 810013203 810011349 810011369 810008563 810004696 111021010 111024381 111024199 111020974 111019443 111022093 111022955 810016344 810016656 810016653 810016532 111011363 111008706 111003365 111003283 111000464 111003963 810019260 111004523 111004267 810013197 810022087 810011461 810022126 810016310 111003691 111003350 810007437 111021445 111003695 810013245 111000473 111011969 111012312 810013196 810011305 810015962 111017246 810016105 810011298 810027536 810011487 111012295 810027150 111012279 810016644 111024214 111018263 111018111 111000666 111000560 111011981 111000658 810027036 111020994 111009262 810005321 111011380 111003688 111003510 810021545 810005705 810005253 810005296 111022085 111024394 810003454 111018643 810018983 111011603 111011367 111000462 111021429 810013477 810022094 810019059 111010780 810013451 111021843 810013458 810019282 111000541 111021815 810011470 810008640 111000757 810011337 810016363 810021520 111000792 111003514 111000441 111021811 810011358 810027159 810016146 810019063 810019539 810022082 111021717 810011295 111024398 111000440 810015797 810015800 810013448 810016145 810016322 810019268 810021602 810021664 111001076 111004164 810007406 111011247 111011227 111011226 111010454 111010436 111010773 111010448 810022110 111024549 111024380 111021824 810022096 810008474 810007403 810007413 810007434 111021000 810026542 810016270 810018953 810026999 810007427 810008536 810008637 810019443 810021702 111008602 810005709 111021724 810005279 111012296 111000803 111003508 111004172 111000996 111020952 111014955 810018966 111001082 810016155 810018955 810018996 111003680 810019057 810016524 810016334 810019019 810018980 111009773 810021369 810019524 810019525 810021376 111022089 111011375 111024197 810019441 810019442 810007414 810008607 810021578 810008595 810022141 810008591 810008603 810021616 810007416 111000150 810019010 810011503 810022107 810007410 810016323 810013062 111020968 111024569 810019007 810016321 111019441 111009829 810019550 810011314 810008632 810022091 810027144 111021438 111023543 111023723 111023725 810011347 810018958 111022778 810022359 810021356 810015787 111011957 111011365 111011377 810019332 111014945 810019531 111000268 111024367 111003325 810016307 111000159 111010469 810015779 111003548 810016384 810022140 111013983 111003326 111008599 810026973 810011492 810011506 810027535 111005533 111011212 810011317 810003947 111024205 810021505 810011377 810016306 111014012 810027029 810016539 111000674 111000608 111001084 810016650 111001585 111001034 111011432 111011612 810016543 810011394 810003434 111018288 111018261 111012467 810022371 810019324 111014938 111001997 810027128 111011905 810027016 810027000 111011974 810026968 111009789 111013980 111021842 111022088 111010776 111014009 111021982 111021850 810011368 111024192 810011524 111021537 111024181 810027025 810027163 111016837 111024082 111022073 111022977 810016148 810016305 810004758 111022762 810021537 810007405 111022589 810011505 111011360 111024193 810016356 111018785 810019410 810016534 810016547 111003083 810013336 810013094 810011548 810007432 810008525 111018648 111018276 111022080 810004816 111006552 111018278 810015821 810021541 111021318 810011322 810005324 111020982 111010785 111009753 111000214 111000657 111011453 810015961 810008615 111009749 810015958 810003344 111015637 111024213 810007430 111002399 810026950 111018676 111002401 111001026 810016648 810011496 111022963 111008597 810011334 810011531 111011245 810019421 810011468 111009767 111003968 111006551 111009248 810027526 111008605 111008579 111024194 810021585 111006546 810019323 111016728 111022979 111000213 810019920 810016326 810016373 111022590 810013183 810004568 810015827 810011381 810022065 810015802 810026988 810016330 810027030 810027043 810004847 810016357 810015786 111022596 810015832 111019435 111016733 111021795 111004270 810011365 111019408 111003677 111011221 111010439 111011912 111014932 810018997 810011508 810013069 810013200 810019425 810022383 111016738 111013998 111023545 111014003 111011633 111003335 111003042 111011973 111009744 111003328 111003348 111004643 111004281 810016355 810007417 810021620 810016150 810016657 810015943 111021714 810007441 111003554 810008609 810027160 111011190 111018117 111003502 810026937 810013076 111003287 111009723 111009345 111018639 111009772 111011206 111021711 111012266 810021503 111011207 111020970 111015650 810021605 810019062 810019308 111008590 111009731 111003512 810013047 810021722 111008609 111002397 810018952 111003052 111009270 111024382 810016378 810021733 810021717 810018963 810005282 810016555 111021821 810008490 111021984 810003459 111013969 810019011 810013227 111003674 810013231 111011900 810019048 111018652 111012261 810003734 111003278 111012475 111021016 810011320 810021729 810011306 810021597 111003682 810011513 810013078 111008598 111018660 810019258 111011954 810013239 810019435 111021803 111004713 810019319 111023736 810026952 810026987 111011990 810019068 111002741 111005530 111011246 810013348 111011626 111004522 111004514 111018659 810007426 111004716 810018988 810008599 111003136 810022071 810019033 810013240 810003428 810008518 810019330 111011255 810016165 111014928 111011442 111003354 810016438 111009814 111000261 810016157 111016830 111003290 111012305 810007423 111003488 111004508 111021825 111003363 111011934 810008638 810016458 810016457 111014930 111003347 111021462 111006548 111008604 810013091 111024390 111001880 111024176 111009819 810021354 810021710 810021697 810019291 810018977 810016521 810016638 810015788 810013224 810027013 111003072 810016098 111009725 111019447 810011499 111012281 111011236 111016726 111019462 111023519 810011441 111021697 111009732 810015814 810013198 111020956 111024547 810011301 111009768 111021029 111002404 111011200 111022777 111006774 111003689 111021987 111021431 111021538 111020959 810013287 810021612 111013972 810026953 111014005 111017224 111018806 810026928 810011373 111003651 810005298 810021518 810015789 810019528 810005248 810013205 810013447 810013188 111021849 111000247 111006554 111017243 810022073 111016725 111014010 111003652 111022582 111022772 111022581 810011356 810007458 810021373 810026962 810011378 111010790 111002392 810016116 111024393 111021979 111021991 111022595 111024376 810026992 810003674 111021696 810016550 111003359 111010443 111002417 111021720 111023542 810011522 111009696 111009697 810016568 111009796 111009246 111009807 810027023 810011448 810021589 111010464 810016337 810021538 111018800 810027155 111021003 810003427 111020958 111006787 111010772 810019438 111022945 810005307 111022959 810008556 810013430 111001024 111001580 111012464 810022092 810021699 810004815 111023551 111011444 111018136 111001014 111011224 111003345 111001080 111002453 111011985 111001577 111003289 111003954 810007433 111011217 111012302 810005245 111009260 810027545 111003285 810016579 810022136 111009779 810027142 111009780 810022118 111009771 810003342 810005290 810022127 111011627 810019280 111008699 111018796 111008705 111008611 111003566 810005237 810022386 111010442 111002378 111014917 111021441 111022901 111021458 111010435 111023562 111014929 810019429 111011225 810026989 810021601 810016328 111020979 111022076 111020948 810004828 810003370 810003348 111003134 111004192 810027547 810013478 111012457 111011929 111021464 810005349 111021802 810026997 810011276 111009798 810019423 111010441 810011504 810003452 810016647 111013985 111009726 111009810 111011947 810021712 810027553 111008577 810013463 111011346 810019000 810016124 810016132 810005292 810016313 111017233 111001962 810021519 810008592 810012608 111023739 810013223 111011920 810016663 810022115 810011316 111017231 810011302 810013361 111017226 810005323 810015804 111022083 810013456 111011614 111009240 810026938 111000991 111009267 111004190 111012321 111019422 111021783 810013272 810011530 111005540 111004250 810026942 111020986 111006792 810011283 810011296 111021434 111011211 810013075 810013357 111009740 111024198 111009751 111011179 810005291 810019522 111009268 810007470 111024083 111003654 111003653 810003458 810021510 111002420 111011955 810003466 810013433 810003461 810013099 111004245 111011618 111004272 810003456 810016578 111021708 111009743 111009718 810027024 111002400 111002393 810026957 111001884 111011253 111022961 810008604 111012470 111018133 111019407 111004178 111019421 111004718 810011477 111004720 111000151 111000995 101000034 111001539 111011248 810016361 111012317 111001961 111000446 810003617 810012603 810013338 111010781 111019456 810019527 111021986 101000038 810013067 111001994 111001540 810019013 111005531 810003374 111015615 810013074 810019041 111000535 810019043 810027519 111018798 810019322 111018801 111013970 810026956 111000575 810021592 111013975 810021586 810021610 111011916 111001023 810018993 111021001 111001581 111023734 111001553 810015826 810021348 810021352 111010961 111010812 111018293 111024392 810012607 111018127 111018268 111018119 111018128 111001857 810016274 111018130 111024377 111000315 810018943 810021719 111020987 111001554 810021737 111018804 111001007 810021374 810021349 810021564 810021593 810021583 810021607 810021555 111016769 111016841 810019289 111021006 111020963 111022966 111022936 111022770 111002414 111002390 111003965 810013053 810016557 111022071 111021841 810019044 810021567 810008553 111008589 810019016 111018810 111003870 111022975 810021580 810027522 810016153 810022385 111021983 810027531 810026545 810026965 810027006 810021591 810019922 810021667 810021595 111000805 810027046 810022128 810021362 810021570 810021367 810021571 810021357 810021380 810021368 810021378 111015645 111019450 111008610 810019264 810019305 810021551 111005537 810003886 810021574 111021831 111003176 810013339 101000031 111024720 111010969 810027120 111013810 810022103 810015828 111018790 111005527 111001859 111003272 111001072 101000029 810021539 111003090 810018991 810016324 111006556 810016327 810027534 810027149 810027162 810019318 810022381 810007452 810007449 810007411 810007468 810007447 111000764 111003073 111011610 111021974 810019037 810021594 810021608 810021691 810021542 810021725 810021721 810018939 810019558 810015811 111004501 111000662 111024548 111003513 111000438 111001081 111003538 111000552 111019451 810016137 810015775 111000557 111003687 111003678 111000451 111000708 810013070 810013045 111000449 111000455 111000263 111000154 111000527 111000675 111001566 111000548 111018791 111000562 111000531 111000141 111000458 111000988 111001016 111000547 101000030 101000036 111003321 111001542 111000554 101000042 111000319 111001866 111000576 111000581 111000476 111000316 111000470 111000538 111000566 111000452 111009830 111009754 111009811 111000474 111003668 810018985 111000525 810019535 111021797 111013987 111013979 111008701 111006782 810027104 810011293 810011478 810013218 111011925 111003515 111024088 111014914 810022364 810019255 810016444 810003616 810003338 111010974 111015638 111017228 810003672 111011201 111003257 810013335 111003048 111000672 810013269 810027146 810021661 810021575 810026955 810026954 111011230 111011228 810021549 810018969 111021699 111013809 810022106 111021853 810016135 111008603 111021314 111002389 111002398 111021716 810011326 810011462 111022937 111003536 111016731 111016730 111016727 810016445 111023561 111023547 810019407 810021658 810005238 810016666 810016346 810016122 810021516 810013346 810008559 111019436 810022086 810026982 810021523 810021718 810022114 810016641 810019304 810026546 810008587 810007444 810007415 111000456 810018972 810027157 810021735 810027166 810027044 810019439 810027119 810018986 810019004 810016267 810015945 810005325 810026931 810027004 810019271 810027097 810021681 810011385 111011381 111018115 111011374 111024399 810027169 111003673 810016662 810016542 111002730 111010957 810026961 810007442 111009799 810016317 810016097 810016354 810019038 111011970 111018110 111021820 111022767 111020993 810022113 810027034 810011338 111011935 111023540 810021379 810013083 810016573 810007460 810026964 810003678 810019515 111024719 810011290 810016529 810015816 810016156 810015836 810027529 810027530 810013230 810013190 810013432 810013445 810011319 810011533 810011519 810011329 810019284 810019331 810022116 810022100 810013073 810021532 810021531 810021527 810021535 810021522 810021621 810027100 810027026 810027037 810027027 810016563 810016642 810018989 810019252 810013238 810011318 810007448 810016659 111024084 810013464 810011547 810003340 111000461 810011297 111009813 111024185 111011946 810027047 111021828 810022083 111004161 111023564 111003336 111020962 111020977 111021013 810005313 810026966 810011315 810011527 810026969 810011475 111021008 111021009 111018292 111003277 111000147 111000158 111001548 111001889 111001541 111012293 111011628 111003039 101000041 111003533 810021613 111021806 111006780 111001966 810027135 111000272 111000105 111000139 111000260 111000160 111000246 111000579 810016280 111011948 111011197 111016833 111024178 111017237 810011308 111009734 810016463 111009822 111024171 810016301 111012294 810015812 810016163 111023534 111002419 810021526 810011310 810016111 810022088 810003614 810013341 111000673 810011331 810016676 810016533 810011327 810027140 111023520 810019018 810027035 111004255 810021689 810011287 810016460 111004173 810003346 810007474 111002436 111000990 111004180 810013340 810011383 810011372 111024208 810015837 810027170 810026990 810011526 111004714 810016123 111016716 111003519 111024395 111002465 111003523 111022771 111009795 810011355 810018942 810019925 810019256 111003967 111000610 111000702 111024388 111004175 111003284 111001854 111016839 810022120 111002446 111014918 111000763 111021976 111015618 111015634 111015640 111015611 111015616 111015613 111036294 111036299 111036292 111014921 111014926 111014936 111014942 111036288 111015646 111015643 111015631 111015627 111015626 111015623 111015641 111015620 111021990 111022094 111021985 111015649 111015639 111021782 111021817 111021792 111021818 111021852 111023718 111024089 111014925 111015630 111021856 111021719 111001584 111001578 111000986 111001571 111000989 111001010 111001570 111000984 111001003 111000987 111001017 111000985 111001576 111001886 111001867 111001860 111001885 111001868 111001865 111001562 111009778 111016825 111016828 111017229 111016823 111016826 111016838 111016831 111016842 111016834 111017223 111017232 111017248 111016822 111016761 111018134 111004504 111003059 111018295 111018123 111009755 111018107 111000671 111000709 111000797 111018126 111000704 111018284 111000775 111000753 111000723 111036304 111000777 111000752 111000782 111000768 111000703 111000762 111000784 111000705 111000766 111000799 111000767 111000780 111000809 111000795 111000783 111000754 111036424 111013990 111013989 111013971 111013995 111013966 111013960 111013964 111013957 111024170 111036429 111036426 111036425 111014954 111014920 111014941 111014939 111003648 111003525 111003546 111009703 111009757 810013261 810022382 810013364 810027131 810027129 810013354 810027118 810019053 810015838 810013208 810013044 810013032 810022125 810022132 111000442 111001550 111000269 810013443 810013429 111001090 111001582 111001568 111004509 111004517 111001419 111001543 111001018 810011367 810011485 810011332 111000454 111001418 111001569 111000524 810016465 810016367 810016558 111009801 810021678 810021553 810021677 810021659 810021683 810011325 810011352 810011311 810011345 810011512 810019025 810019049 810018944 810019052 810021528 810021514 810021582 810021728 111021822 111018281 111018259 111003268 810016154 111021992 810019926 810016273 810021732 810019002 810015835 810007402 810019055 111021819 111009838 810027541 810003381 810013079 111024373 111020988 810019540 111011977 111011914 810013213 111022956 810022357 810021599 111018809 810011400 111010760 810027005 810016340 810027127 810004569 111024383 810027096 810022384 111013811 810016643 810027045 111011219 810019541 810003537 810008480 111013954 810003612 111021030 111013955 111013962 111013959 111004260 810016645 111019439 111018685 111009730 810013262 810021554 111015632 810005699 810011536 810008500 810018974 810019034 111022953 111016764 810019262 810018970 810019412 810013441 810015817 810015801 810022109 810022379 810013226 810027158 111022938 810016101 810016269 111002010 111024404 111024371 810021680 111024189 810016382 810013465 810013089 810019031 111022588 111014004 111004249 810021709 111004268 111022967 810005236 111011440 810027521 111009739 810005701 111021844 111014002 810022139 111004247 111009742 111021978 810011397 810011480 810015944 111009770 111009265 111009256 111009251 111009264 810003949 810003950 810003951 810004012 111021018 111021014 111021015 111020980 111012301 111011971 111011455 111011909 111011993 810027017 111011352 111011261 111011172 810003504 111020964 111021031 111020950 111020999 111009722 111009802 111009746 111009758 111009716 111001083 111001068 111001579 111000471 111000578 111001855 111000469 111001882 111001005 111000208 111000580 111000466 111000209 111013953 810013087 111011353 111024207 111022075 111000677 111000533 111000728 111000725 111000720 111001088 810015948 810011507 111018793 111019467 810013233 810026945 111011608 810013187 810005247 111015642 810027125 810007401 810019015 810005294 111018271 810005235 810013186 810008479 810005316 111016835 111011956 111010972 111010795 111014946 810003620 111010794 111011182 111010958 111010821 111010809 111010973 111010968 111011348 111011371 111011609 111011454 111011258 810004813 111012313 111011641 111011637 111011949 111011630 111011621 111010811 111012268 810013029 810022104 111000148 810003511 111022079 111011356 111001887 111001078 111001069 111001012 111001011 111000994 111000998 111000997 810016464 111015636 810013452 111012278 810021736 111000755 111000155 111003063 111003035 111018661 111009842 810021600 810016446 111003351 810005357 111016723 111013984 810016272 111024187 810007422 111003374 111005526 111003370 810018981 810022102 111003147 111002721 810016347 810016637 111009843 810008602 111003649 810016679 810016279 810015808 810016368 111003368 111009252 111002732 111011450 111004256 111021705 111004197 111004507 810004821 810003369 810019251 810013035 810027022 810013058 810019026 111011364 111004500 810019326 111011357 810022066 810016664 111000164 810027552 810019294 111022766 111001590 111010819 111016732 111016743 111016715 111009815 111009728 111009825 810008516 111021452 810007450 810008562 810008487 810003619 810003613 111009804 111009806 810008610 810008524 810027090 810027093 810027107 810003560 810003507 810003554 810003508 111006794 810027092 810027102 810027098 810026948 810027095 810013251 810021730 810021511 111016766 111018789 111018787 111018662 111018682 111018267 810011495 810011541 810011542 810011456 810011444 111019457 111019432 810011363 111022594 111022591 111022900 111022946 111022962 111018282 111018112 111018290 810004817 810004823 810004810 111018269 111018277 111021428 810011471 810011279 111003188 111022082 111003175 111003352 111006795 810022362 111010805 111009745 111011451 111018649 111018665 810011466 111010793 810022380 810021534 810022067 810022063 810026984 810027001 111022933 810013050 810013258 810013222 810013257 810013259 810013252 810013228 810013234 810013260 111014927 111014916 111014931 111010806 810011515 810011286 111022763 810013280 810026977 810026933 111022593 111011354 111011243 810018941 810016266 810019314 810021603 810022377 810016567 810016308 810008598 810019916 810027527 810016129 111017249 111005521 810019411 111012461 111009709 111004530 111011937 111021463 810008584 810021695 810019061 111009705 111012297 111010758 810022080 810018994 810013435 810015809 810008572 111005528 810005246 810019536 111020947 111006544 810021670 810015952 810008641 111009761 810015957 111011959 111002736 111001077 111001000 810013243 810021676 810005707 810019417 111021830 810008601 111024200 111002438 111004712 810019263 810019549 111003958 810015805 810005305 810019534 111009253 111011355 810022081 111011220 111018122 111018108 111009258 810021584 810012605 810013437 810013211 111024169 810016576 810016677 111022973 810011392 810011537 810016114 810019265 810012606 810012602 111000450 810016680 810005283 111022086 111021316 810005284 111021700 810015777 111013958 111013963 810004825 111021439 111002381 810015776 810016451 810008568 810019414 810019542 111018264 111022764 810027009 810007445 111021312 810016667 810019517 810027021 810013182 810013469 810021614 810003371 810019521 111022583 810013363 810013470 111006786 111011921 111000532 810003431 111002441 111023530 111011267 111010817 111021467 810003465 810011284 111009785 111006543 111006536 810018964 111011605 111011256 111011897 111010959 111012264 111009797 810026983 810021693 810011488 111022785 810015818 111012469 111017242 111011980 111003675 111011625 111003655 810022134 810013286 810004814 111013803 810022097 810016343 810013366 810013273 810021507 810013263 111003191 111008593 810003425 810003432 111000146 810021366 111000761 111000606 111000101 111012298 101000035 111012319 111004277 111000463 810016312 810021350 111024174 810013281 810019554 111024190 111024173 111024172 810021359 111000663 810013353 810003436 810016544 111002386 111000140 111024401 111010762 810016358 111003541 111009823 810007464 810011374 810021384 810008600 810013194 111023549 810016134 111022584 810013191 111018667 111003266 810007451 111024203 810016133 111024568 111023529 810027549 111001085 810019912 111022950 810013040 810013041 111009765 810016437 111003322 810013475 111021442 111003498 111003270 810003559 111011240 111001032 810003618 111020975 810021572 810005704 810013246 810021726 810008623 810019064 111019455 810003679 111012311 111014949 111018124 810027110 111012280 810008616 810008560 111011193 111020951 111011975 111018788 111011234 111010820 810021529 810018960 810013249 810022130 810007466 810008577 810016349 111024403 111002740 111016734 111023524 810013082 810008482 111013991 810018979 111004186 810005320 810005368 810005250 111000103 111024384 111003658 111016843 111003494 810019036 111003564 810013365 810011348 111013967 810018959 111022760 111022773 810013081 111024372 111024400 111011441 111013978 810027147 810019051 111018672 810021731 810004809 810005702 810021363 111022087 111016767 810019046 810027546 810013034 810013059 810027148 810027012 111004198 810027007 111000143 810027550 810016553 111003341 111002405 111023548 111010456 810019295 111024201 111021011 111013994 810003349 810021590 810015781 111000211 111000212 810019253 810004567 810013483 810019523 810013450 810004566 810008580 810015823 810007431 111009255 111013974 810018940 810007404 111019414 810011509 111013812 810011497 810011309 111014011 111021813 111000710 111001075 111003053 810003558 810021569 810015947 111011953 111009717 111003959 111000676 810015955 111000607 111014008 111024188 111011382 111009839 810016125 810005365 810005306 111001028 111021446 810011393 111002394 810013334 111001004 111011950 810016348 111018647 111009808 810005367 810019431 111021799 810027108 111021977 810011342 810021375 810013201 810013482 810011362 810011384 111021788 111023538 810004752 111019427 111010774 111011349 810011292 810011517 810005280 810015774 810021517 111022578 111018294 111021315 111021443 810016144 111003127 111011178 111012459 810027525 810013209 810015792 111002402 810016319 111016844 111019463 111006555 111022940 810021509 810008583 111000665 111036453 111036561 810008405 111028682 810028576 810028578 810028556 111042323 810022327 810022123 810021704 810022124 810011498 810011521 810011446 111000271 111018132 111003366 111011595 810007438 810013350 111024182 111024196 111021981 810021734 810019022 810013473 810021562 810016100 810016107 810016128 810016106 810021692 810021694 810016362 810016142 810022085 810011376 810013480 810019267 810019024 810011321 810013037 810013039 810013063 810005352 111021805 111021833 810021588 810011354 111024397 810013096 810022072 111021435 810011366 810021696 810011455 111010978 810011472 810007408 111003862 111003662 111023552 111018280 111002746 810011370 111001863 111003263 111001030 111002449 111012479 111012270 111002738 111002734 810018957 111011196 111016736 111011988 111003342 111020985 810011467 111019417 111024396 810011494 810011336 111011257 111010434 111016719 810019914 111024085 810019259 810019919 810021552 111020978 111012257 111010455 111010788 111003281 810003367 111011239 810016455 111011186 111021787 111021789 810008576 810005281 111012456 111011199 111003038 111002006 810019286 111003962 111003381 810021544 810021365 111011251 111011962 111022761 111015635 111011218 810016141 810011278 111015614 810027122 111011176 810027540 810018990 111021801 810007446 111013806 810011324 810011330 111021809 810021573 810021617 111002705 111003355 810003341 111002747 111024389 111003045 111003146 111003183 111004168 111024086 111008703 810027164 810027151 810027523 111023532 111018642 810011445 111004286 111004177 111000701 810027008 810018951 111022776 111000157 111018113 111018807 111020957 810004643 111020953 111018797 810005241 810005315 810005289 810013036 810015778 810019533 810019316 810019281 810015791 810015829 810015810 810018956 111018808 810027019 810013103 810013056 810015825 111003128 111009763 111016722 111008702 111011592 810016574 111009787 111018803 810021716 810016314 810027533 111003323 111004515 111021808 810005285 810016377 810008629 810016551 111009752 111009711 111001070 111004527 810003438 111018812 810027042 111004520 810021619 111018683 111012269 111022579 111021835 810015793 810003351 810022144 810003552 111009805 810021515 111010451 111015624 111012480 810027543 111009706 111009816 111021786 111003952 111014933 810011540 810016152 810016099 810011389 810019520 111001021 111020965 810008586 111010767 111021005 810021550 810027538 810016127 111024217 111000106 810016375 111023525 810019045 810021675 111012282 810015953 810008492 111010768 111019458 810008551 111004176 810019283 810016448 810026996 111009827 111003951 810027532 111012306 810013055 111031417 810026998 111003353 810011518 111011952 111011904 111022939 810005252 810021525 111012320 111004187 810021598 810022369 810027528 111010766 111012304 111010963 111024366 111010791 810003368 810019006 810018968 111004524 111000568 111010756 111004640 111004715 111000153 111000448 810005703 111008586 810003337 111000526 111003180 111003657 111000561 111011594 111002742 111011992 111000145 111000324 810015960 810015813 810019405 810016569 810013449 111021798 111022775 810004807 810016376 810019269 810019311 810016545 810026929 810026939 810021708 111017235 111003174 810003433 810027516 810016441 810019005 111009762 111009239 111000321 810016268 810027152 111023541 111003144 111020971 111020989 111024216 111024206 810007457 810007424 810026936 810013444 810016443 111018668 810011452 810019279 810003372 111012291 810016143 810027520 810027517 111013986 111011361 111013997 111012253 810003464 111002403 111011264 810016658 810011299 810019333 111023568 111010792 810019915 810018978 111010784 111011965 810021358 810011282 810013031 111018273 111018286 111018279 111015858 111000259 111015612 810016352 810021685 810003735 111015857 810013283 111012265 111012465 810013220 111014915 111016762 111016765 111021796 810027018 111016829 111018680 111010765 810027123 810022108 810021377 111021695 111021785 810016130 810004571 810013284 111004642 111042196 810019553 810027011 810016160 810003380 810019924 111020997 111021784 111022765 111021827 111021536 810013184 81
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Abbott Medical
Reason for Recall:
Due to Bluetooth malfunction on a subset of Implantable Cardioverter Defibrillator (ICD) and Cardia Resynchronization Therapy Defibrillator *CRT-D) devices that were manufactured prior to April 2022 as a result of crystal component issue within the Bluetooth Low Energy (BLE) circuit. This results in the device entering an inductive telemetry only mode with loss of Bluetooth communication. In some instances, thought BLE communication is unavailable, the circuit may remain in a state which draws excessive power affecting battery longevity.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Gallant VR Implantable Cardioverter Defibrillator: VR ICD REF CDVRA500Q Gallant DR Implantable Cardioverter Defibrillator: DR ICD REF CDDRA500Q Neutrino VR Implantable Cardioverter Defibrillator: NxT VR ICD REF CDVRA600Q Neutrino DR Implantable Cardioverter Defibrillator: NxT DR ICD REF CDDRA600Q Entrant VR Implantable Cardioverter Defibrillator: VR ICD REF CDVRA300Q Entrant DR Implantable Cardioverter Defibrillator: DR ICD REF CDDRA300Q The primary function of ICDs and CRT-Ds are to treat life-threatening ventricular tachyarrhythmias via anti-tachycardia pacing (ATP), or cardioversion/defibrillation. Implantable Cardioverter Defibrillators (ICDs) can either be a dual chamber (also known as DR if rate responsive) device that is connected to a right atrial (RA) pacing lead and right ventricular defibrillation lead or single chamber ICD (also known as VR if rate responsive) with only a defibrillation lead in the right ventricle (RV). The ICD, more generally known as an implantable pulse generator (IPG), is typically implanted in a subcutaneous device pocket located in the pectoral region and is attached to the pacing/defibrillator leads that are implanted through the transvenous technique with the distal tips positioned in the RA (dual chamber) and RV. Cardiac Resynchronization Therapy Defibrillator (CRT-D) systems consist of a right atrial (RA) pacing lead, a right ventricular (RV) defibrillation lead, as well as a third lead placed in the coronary venous system overlying the left ventricle (LV) to improve ventricular resynchronization through biventricular pacing for treatment of heart failure. CRT-D devices, more generally known as an implantable pulse generator (IPG), are typically implanted in a subcutaneous device pocket located in the pectoral region and are attached to the pacing/defibrillator leads that are implanted through transvenous technique.

Product Codes/Lot Numbers:

Gallant: Model Number: VR ICD REF CDVRA500Q UDI-DI Code: 05415067031990 Serial Numbers: 111033498 810000696 810002524 111031208 111025117 111030959 111006327 111028481 111014296 111033775 810021414 810001946 810001996 810002011 810001942 810007459 111036536 810019608 810019393 810019581 111042257 810026722 810028531 810028585 810027003 111011936 810021423 111025530 111015227 111036960 810026790 111036957 111036933 111036964 111007506 810021413 810021445 810021449 111033604 111000553 111000558 111000567 111000472 111000706 111000667 111006793 111000664 111000678 111018664 111018644 111018802 111012460 111006783 111006791 111006789 111000786 111000582 111000550 111000721 111001035 111001025 111014950 111021456 111021461 111021703 111021722 111021450 111036417 111012272 111011597 810001013 810001011 810006759 810001046 810001021 810019014 111024417 810002439 111024415 810006774 111024454 810022318 111036914 810013066 810015963 111002002 810021662 810013192 810011364 111003380 810011481 111020960 111011436 111020992 810015820 810015794 111009835 111004271 111003046 111024175 111022969 111009840 810019313 810018954 111021002 111009844 810008620 111010818 810018999 111021710 810016110 810019437 810008498 111024550 810008497 810019426 810011359 810019424 810011511 810011353 810011328 111011181 111011372 111021007 111021712 111024183 810013436 810019054 810019546 810022117 810027515 810013071 810007469 111020996 111009733 111003861 111009782 810011335 810016103 111021851 810004570 111008584 111009245 111003267 810008499 111021988 111012318 111003040 111010800 111003332 111022768 810016526 810027041 111009747 810013248 111011899 111010457 810021351 810021705 810015780 111018656 111010771 111021437 810005318 111010761 111022974 111021994 810007454 810008483 810011285 810011458 810005317 111008582 111003173 111008588 111003555 111018654 810027111 810019552 810016459 111009820 810008557 810026959 111002712 111021707 111006776 111006779 111010814 111017230 810021576 810016112 111003506 810027040 810027014 810003350 111009781 111018131 810021674 111011908 810008527 810019917 111008583 810008575 111004282 810016339 810027099 810015815 810016102 111002437 111011229 111018270 810021707 810013068 810013216 810015840 810018976 810021355 810011459 111003500 111019459 810003430 111009834 111010787 111000447 111004193 111009841 111013996 111011233 111011254 111021837 111019423 810011451 111018786 810016113 810016649 810013446 810013250 810013442 810027039 810026943 810026944 810027020 810027165 111009759 111006801 111021440 111010437 111019431 810011457 810016665 810016449 810019012 810021618 810008573 810021560 810011535 111009788 810013454 810008650 810015824 111006778 810005319 111024379 810013349 810004013 111009809 810008519 111024368 810003553 111024378 111006784 810016140 810008486 111006797 810013266 111018109 111018795 111022789 111011910 810021606 810007463 810007421 810003677 111011216 810027514 111018640 810019556 810008612 111018784 111024402 810003437 810011343 810008526 111004721 111021995 810027548 810018995 810016138 810019427 810005706 111022585 810011387 111018291 111011963 111009708 111010450 111017239 111022090 111024180 111011906 111021723 111021436 810022090 810015784 111009794 111021433 810003343 810013241 111010980 810027103 111011915 111002410 810027161 111001995 111009741 111006541 111002011 810016121 111004259 111003960 810021559 111024204 810013232 111016740 810027168 810015796 111001027 111012290 111002434 111010962 111001969 810015822 111016745 111024202 111012258 111010952 111024184 111013804 111022788 111010465 810019529 810011303 111013808 810011549 810011538 111021812 810005366 810007473 111023732 111014948 111013807 111001583 111023720 111022587 111011938 810026979 111012462 111012307 111003869 111024186 810016277 111004242 810005286 111022781 111021845 810004846 810015783 111019446 810022064 111010799 111011214 810022363 111009828 810008597 111021790 111003559 810016560 111011185 810013225 111013993 810013206 111011922 111003276 810018982 111003858 810013077 810011449 810011550 810016303 111011237 111019415 111003865 111006785 111003863 111004182 111021466 810004849 111012256 111006775 810022137 810019296 810019297 111018120 111018679 810003615 810008594 111003065 111003037 810022370 810022129 810019042 111004513 111003074 111022971 111003957 810005356 111004518 111023531 111003371 111014919 111009837 111003539 111002743 810019250 810011523 810019254 111021846 111022077 810021666 810022122 810021671 810019272 810019301 810019290 810019278 810019065 810019325 810019408 810019310 810019293 810019009 111009737 111011215 111009766 111009777 111003521 111010981 111011968 111009261 111003686 111017247 111017245 111003567 111003505 810015772 810015773 111004497 111017234 810013185 810019050 810011339 810022145 810022376 810022079 810022078 810021530 810021687 810021563 810018946 810019017 810016151 810027033 810013203 810011349 810011369 810008563 810004696 111021010 111024381 111024199 111020974 111019443 111022093 111022955 810016344 810016656 810016653 810016532 111011363 111008706 111003365 111003283 111000464 111003963 810019260 111004523 111004267 810013197 810022087 810011461 810022126 810016310 111003691 111003350 810007437 111021445 111003695 810013245 111000473 111011969 111012312 810013196 810011305 810015962 111017246 810016105 810011298 810027536 810011487 111012295 810027150 111012279 810016644 111024214 111018263 111018111 111000666 111000560 111011981 111000658 810027036 111020994 111009262 810005321 111011380 111003688 111003510 810021545 810005705 810005253 810005296 111022085 111024394 810003454 111018643 810018983 111011603 111011367 111000462 111021429 810013477 810022094 810019059 111010780 810013451 111021843 810013458 810019282 111000541 111021815 810011470 810008640 111000757 810011337 810016363 810021520 111000792 111003514 111000441 111021811 810011358 810027159 810016146 810019063 810019539 810022082 111021717 810011295 111024398 111000440 810015797 810015800 810013448 810016145 810016322 810019268 810021602 810021664 111001076 111004164 810007406 111011247 111011227 111011226 111010454 111010436 111010773 111010448 810022110 111024549 111024380 111021824 810022096 810008474 810007403 810007413 810007434 111021000 810026542 810016270 810018953 810026999 810007427 810008536 810008637 810019443 810021702 111008602 810005709 111021724 810005279 111012296 111000803 111003508 111004172 111000996 111020952 111014955 810018966 111001082 810016155 810018955 810018996 111003680 810019057 810016524 810016334 810019019 810018980 111009773 810021369 810019524 810019525 810021376 111022089 111011375 111024197 810019441 810019442 810007414 810008607 810021578 810008595 810022141 810008591 810008603 810021616 810007416 111000150 810019010 810011503 810022107 810007410 810016323 810013062 111020968 111024569 810019007 810016321 111019441 111009829 810019550 810011314 810008632 810022091 810027144 111021438 111023543 111023723 111023725 810011347 810018958 111022778 810022359 810021356 810015787 111011957 111011365 111011377 810019332 111014945 810019531 111000268 111024367 111003325 810016307 111000159 111010469 810015779 111003548 810016384 810022140 111013983 111003326 111008599 810026973 810011492 810011506 810027535 111005533 111011212 810011317 810003947 111024205 810021505 810011377 810016306 111014012 810027029 810016539 111000674 111000608 111001084 810016650 111001585 111001034 111011432 111011612 810016543 810011394 810003434 111018288 111018261 111012467 810022371 810019324 111014938 111001997 810027128 111011905 810027016 810027000 111011974 810026968 111009789 111013980 111021842 111022088 111010776 111014009 111021982 111021850 810011368 111024192 810011524 111021537 111024181 810027025 810027163 111016837 111024082 111022073 111022977 810016148 810016305 810004758 111022762 810021537 810007405 111022589 810011505 111011360 111024193 810016356 111018785 810019410 810016534 810016547 111003083 810013336 810013094 810011548 810007432 810008525 111018648 111018276 111022080 810004816 111006552 111018278 810015821 810021541 111021318 810011322 810005324 111020982 111010785 111009753 111000214 111000657 111011453 810015961 810008615 111009749 810015958 810003344 111015637 111024213 810007430 111002399 810026950 111018676 111002401 111001026 810016648 810011496 111022963 111008597 810011334 810011531 111011245 810019421 810011468 111009767 111003968 111006551 111009248 810027526 111008605 111008579 111024194 810021585 111006546 810019323 111016728 111022979 111000213 810019920 810016326 810016373 111022590 810013183 810004568 810015827 810011381 810022065 810015802 810026988 810016330 810027030 810027043 810004847 810016357 810015786 111022596 810015832 111019435 111016733 111021795 111004270 810011365 111019408 111003677 111011221 111010439 111011912 111014932 810018997 810011508 810013069 810013200 810019425 810022383 111016738 111013998 111023545 111014003 111011633 111003335 111003042 111011973 111009744 111003328 111003348 111004643 111004281 810016355 810007417 810021620 810016150 810016657 810015943 111021714 810007441 111003554 810008609 810027160 111011190 111018117 111003502 810026937 810013076 111003287 111009723 111009345 111018639 111009772 111011206 111021711 111012266 810021503 111011207 111020970 111015650 810021605 810019062 810019308 111008590 111009731 111003512 810013047 810021722 111008609 111002397 810018952 111003052 111009270 111024382 810016378 810021733 810021717 810018963 810005282 810016555 111021821 810008490 111021984 810003459 111013969 810019011 810013227 111003674 810013231 111011900 810019048 111018652 111012261 810003734 111003278 111012475 111021016 810011320 810021729 810011306 810021597 111003682 810011513 810013078 111008598 111018660 810019258 111011954 810013239 810019435 111021803 111004713 810019319 111023736 810026952 810026987 111011990 810019068 111002741 111005530 111011246 810013348 111011626 111004522 111004514 111018659 810007426 111004716 810018988 810008599 111003136 810022071 810019033 810013240 810003428 810008518 810019330 111011255 810016165 111014928 111011442 111003354 810016438 111009814 111000261 810016157 111016830 111003290 111012305 810007423 111003488 111004508 111021825 111003363 111011934 810008638 810016458 810016457 111014930 111003347 111021462 111006548 111008604 810013091 111024390 111001880 111024176 111009819 810021354 810021710 810021697 810019291 810018977 810016521 810016638 810015788 810013224 810027013 111003072 810016098 111009725 111019447 810011499 111012281 111011236 111016726 111019462 111023519 810011441 111021697 111009732 810015814 810013198 111020956 111024547 810011301 111009768 111021029 111002404 111011200 111022777 111006774 111003689 111021987 111021431 111021538 111020959 810013287 810021612 111013972 810026953 111014005 111017224 111018806 810026928 810011373 111003651 810005298 810021518 810015789 810019528 810005248 810013205 810013447 810013188 111021849 111000247 111006554 111017243 810022073 111016725 111014010 111003652 111022582 111022772 111022581 810011356 810007458 810021373 810026962 810011378 111010790 111002392 810016116 111024393 111021979 111021991 111022595 111024376 810026992 810003674 111021696 810016550 111003359 111010443 111002417 111021720 111023542 810011522 111009696 111009697 810016568 111009796 111009246 111009807 810027023 810011448 810021589 111010464 810016337 810021538 111018800 810027155 111021003 810003427 111020958 111006787 111010772 810019438 111022945 810005307 111022959 810008556 810013430 111001024 111001580 111012464 810022092 810021699 810004815 111023551 111011444 111018136 111001014 111011224 111003345 111001080 111002453 111011985 111001577 111003289 111003954 810007433 111011217 111012302 810005245 111009260 810027545 111003285 810016579 810022136 111009779 810027142 111009780 810022118 111009771 810003342 810005290 810022127 111011627 810019280 111008699 111018796 111008705 111008611 111003566 810005237 810022386 111010442 111002378 111014917 111021441 111022901 111021458 111010435 111023562 111014929 810019429 111011225 810026989 810021601 810016328 111020979 111022076 111020948 810004828 810003370 810003348 111003134 111004192 810027547 810013478 111012457 111011929 111021464 810005349 111021802 810026997 810011276 111009798 810019423 111010441 810011504 810003452 810016647 111013985 111009726 111009810 111011947 810021712 810027553 111008577 810013463 111011346 810019000 810016124 810016132 810005292 810016313 111017233 111001962 810021519 810008592 810012608 111023739 810013223 111011920 810016663 810022115 810011316 111017231 810011302 810013361 111017226 810005323 810015804 111022083 810013456 111011614 111009240 810026938 111000991 111009267 111004190 111012321 111019422 111021783 810013272 810011530 111005540 111004250 810026942 111020986 111006792 810011283 810011296 111021434 111011211 810013075 810013357 111009740 111024198 111009751 111011179 810005291 810019522 111009268 810007470 111024083 111003654 111003653 810003458 810021510 111002420 111011955 810003466 810013433 810003461 810013099 111004245 111011618 111004272 810003456 810016578 111021708 111009743 111009718 810027024 111002400 111002393 810026957 111001884 111011253 111022961 810008604 111012470 111018133 111019407 111004178 111019421 111004718 810011477 111004720 111000151 111000995 101000034 111001539 111011248 810016361 111012317 111001961 111000446 810003617 810012603 810013338 111010781 111019456 810019527 111021986 101000038 810013067 111001994 111001540 810019013 111005531 810003374 111015615 810013074 810019041 111000535 810019043 810027519 111018798 810019322 111018801 111013970 810026956 111000575 810021592 111013975 810021586 810021610 111011916 111001023 810018993 111021001 111001581 111023734 111001553 810015826 810021348 810021352 111010961 111010812 111018293 111024392 810012607 111018127 111018268 111018119 111018128 111001857 810016274 111018130 111024377 111000315 810018943 810021719 111020987 111001554 810021737 111018804 111001007 810021374 810021349 810021564 810021593 810021583 810021607 810021555 111016769 111016841 810019289 111021006 111020963 111022966 111022936 111022770 111002414 111002390 111003965 810013053 810016557 111022071 111021841 810019044 810021567 810008553 111008589 810019016 111018810 111003870 111022975 810021580 810027522 810016153 810022385 111021983 810027531 810026545 810026965 810027006 810021591 810019922 810021667 810021595 111000805 810027046 810022128 810021362 810021570 810021367 810021571 810021357 810021380 810021368 810021378 111015645 111019450 111008610 810019264 810019305 810021551 111005537 810003886 810021574 111021831 111003176 810013339 101000031 111024720 111010969 810027120 111013810 810022103 810015828 111018790 111005527 111001859 111003272 111001072 101000029 810021539 111003090 810018991 810016324 111006556 810016327 810027534 810027149 810027162 810019318 810022381 810007452 810007449 810007411 810007468 810007447 111000764 111003073 111011610 111021974 810019037 810021594 810021608 810021691 810021542 810021725 810021721 810018939 810019558 810015811 111004501 111000662 111024548 111003513 111000438 111001081 111003538 111000552 111019451 810016137 810015775 111000557 111003687 111003678 111000451 111000708 810013070 810013045 111000449 111000455 111000263 111000154 111000527 111000675 111001566 111000548 111018791 111000562 111000531 111000141 111000458 111000988 111001016 111000547 101000030 101000036 111003321 111001542 111000554 101000042 111000319 111001866 111000576 111000581 111000476 111000316 111000470 111000538 111000566 111000452 111009830 111009754 111009811 111000474 111003668 810018985 111000525 810019535 111021797 111013987 111013979 111008701 111006782 810027104 810011293 810011478 810013218 111011925 111003515 111024088 111014914 810022364 810019255 810016444 810003616 810003338 111010974 111015638 111017228 810003672 111011201 111003257 810013335 111003048 111000672 810013269 810027146 810021661 810021575 810026955 810026954 111011230 111011228 810021549 810018969 111021699 111013809 810022106 111021853 810016135 111008603 111021314 111002389 111002398 111021716 810011326 810011462 111022937 111003536 111016731 111016730 111016727 810016445 111023561 111023547 810019407 810021658 810005238 810016666 810016346 810016122 810021516 810013346 810008559 111019436 810022086 810026982 810021523 810021718 810022114 810016641 810019304 810026546 810008587 810007444 810007415 111000456 810018972 810027157 810021735 810027166 810027044 810019439 810027119 810018986 810019004 810016267 810015945 810005325 810026931 810027004 810019271 810027097 810021681 810011385 111011381 111018115 111011374 111024399 810027169 111003673 810016662 810016542 111002730 111010957 810026961 810007442 111009799 810016317 810016097 810016354 810019038 111011970 111018110 111021820 111022767 111020993 810022113 810027034 810011338 111011935 111023540 810021379 810013083 810016573 810007460 810026964 810003678 810019515 111024719 810011290 810016529 810015816 810016156 810015836 810027529 810027530 810013230 810013190 810013432 810013445 810011319 810011533 810011519 810011329 810019284 810019331 810022116 810022100 810013073 810021532 810021531 810021527 810021535 810021522 810021621 810027100 810027026 810027037 810027027 810016563 810016642 810018989 810019252 810013238 810011318 810007448 810016659 111024084 810013464 810011547 810003340 111000461 810011297 111009813 111024185 111011946 810027047 111021828 810022083 111004161 111023564 111003336 111020962 111020977 111021013 810005313 810026966 810011315 810011527 810026969 810011475 111021008 111021009 111018292 111003277 111000147 111000158 111001548 111001889 111001541 111012293 111011628 111003039 101000041 111003533 810021613 111021806 111006780 111001966 810027135 111000272 111000105 111000139 111000260 111000160 111000246 111000579 810016280 111011948 111011197 111016833 111024178 111017237 810011308 111009734 810016463 111009822 111024171 810016301 111012294 810015812 810016163 111023534 111002419 810021526 810011310 810016111 810022088 810003614 810013341 111000673 810011331 810016676 810016533 810011327 810027140 111023520 810019018 810027035 111004255 810021689 810011287 810016460 111004173 810003346 810007474 111002436 111000990 111004180 810013340 810011383 810011372 111024208 810015837 810027170 810026990 810011526 111004714 810016123 111016716 111003519 111024395 111002465 111003523 111022771 111009795 810011355 810018942 810019925 810019256 111003967 111000610 111000702 111024388 111004175 111003284 111001854 111016839 810022120 111002446 111014918 111000763 111021976 111015618 111015634 111015640 111015611 111015616 111015613 111036294 111036299 111036292 111014921 111014926 111014936 111014942 111036288 111015646 111015643 111015631 111015627 111015626 111015623 111015641 111015620 111021990 111022094 111021985 111015649 111015639 111021782 111021817 111021792 111021818 111021852 111023718 111024089 111014925 111015630 111021856 111021719 111001584 111001578 111000986 111001571 111000989 111001010 111001570 111000984 111001003 111000987 111001017 111000985 111001576 111001886 111001867 111001860 111001885 111001868 111001865 111001562 111009778 111016825 111016828 111017229 111016823 111016826 111016838 111016831 111016842 111016834 111017223 111017232 111017248 111016822 111016761 111018134 111004504 111003059 111018295 111018123 111009755 111018107 111000671 111000709 111000797 111018126 111000704 111018284 111000775 111000753 111000723 111036304 111000777 111000752 111000782 111000768 111000703 111000762 111000784 111000705 111000766 111000799 111000767 111000780 111000809 111000795 111000783 111000754 111036424 111013990 111013989 111013971 111013995 111013966 111013960 111013964 111013957 111024170 111036429 111036426 111036425 111014954 111014920 111014941 111014939 111003648 111003525 111003546 111009703 111009757 810013261 810022382 810013364 810027131 810027129 810013354 810027118 810019053 810015838 810013208 810013044 810013032 810022125 810022132 111000442 111001550 111000269 810013443 810013429 111001090 111001582 111001568 111004509 111004517 111001419 111001543 111001018 810011367 810011485 810011332 111000454 111001418 111001569 111000524 810016465 810016367 810016558 111009801 810021678 810021553 810021677 810021659 810021683 810011325 810011352 810011311 810011345 810011512 810019025 810019049 810018944 810019052 810021528 810021514 810021582 810021728 111021822 111018281 111018259 111003268 810016154 111021992 810019926 810016273 810021732 810019002 810015835 810007402 810019055 111021819 111009838 810027541 810003381 810013079 111024373 111020988 810019540 111011977 111011914 810013213 111022956 810022357 810021599 111018809 810011400 111010760 810027005 810016340 810027127 810004569 111024383 810027096 810022384 111013811 810016643 810027045 111011219 810019541 810003537 810008480 111013954 810003612 111021030 111013955 111013962 111013959 111004260 810016645 111019439 111018685 111009730 810013262 810021554 111015632 810005699 810011536 810008500 810018974 810019034 111022953 111016764 810019262 810018970 810019412 810013441 810015817 810015801 810022109 810022379 810013226 810027158 111022938 810016101 810016269 111002010 111024404 111024371 810021680 111024189 810016382 810013465 810013089 810019031 111022588 111014004 111004249 810021709 111004268 111022967 810005236 111011440 810027521 111009739 810005701 111021844 111014002 810022139 111004247 111009742 111021978 810011397 810011480 810015944 111009770 111009265 111009256 111009251 111009264 810003949 810003950 810003951 810004012 111021018 111021014 111021015 111020980 111012301 111011971 111011455 111011909 111011993 810027017 111011352 111011261 111011172 810003504 111020964 111021031 111020950 111020999 111009722 111009802 111009746 111009758 111009716 111001083 111001068 111001579 111000471 111000578 111001855 111000469 111001882 111001005 111000208 111000580 111000466 111000209 111013953 810013087 111011353 111024207 111022075 111000677 111000533 111000728 111000725 111000720 111001088 810015948 810011507 111018793 111019467 810013233 810026945 111011608 810013187 810005247 111015642 810027125 810007401 810019015 810005294 111018271 810005235 810013186 810008479 810005316 111016835 111011956 111010972 111010795 111014946 810003620 111010794 111011182 111010958 111010821 111010809 111010973 111010968 111011348 111011371 111011609 111011454 111011258 810004813 111012313 111011641 111011637 111011949 111011630 111011621 111010811 111012268 810013029 810022104 111000148 810003511 111022079 111011356 111001887 111001078 111001069 111001012 111001011 111000994 111000998 111000997 810016464 111015636 810013452 111012278 810021736 111000755 111000155 111003063 111003035 111018661 111009842 810021600 810016446 111003351 810005357 111016723 111013984 810016272 111024187 810007422 111003374 111005526 111003370 810018981 810022102 111003147 111002721 810016347 810016637 111009843 810008602 111003649 810016679 810016279 810015808 810016368 111003368 111009252 111002732 111011450 111004256 111021705 111004197 111004507 810004821 810003369 810019251 810013035 810027022 810013058 810019026 111011364 111004500 810019326 111011357 810022066 810016664 111000164 810027552 810019294 111022766 111001590 111010819 111016732 111016743 111016715 111009815 111009728 111009825 810008516 111021452 810007450 810008562 810008487 810003619 810003613 111009804 111009806 810008610 810008524 810027090 810027093 810027107 810003560 810003507 810003554 810003508 111006794 810027092 810027102 810027098 810026948 810027095 810013251 810021730 810021511 111016766 111018789 111018787 111018662 111018682 111018267 810011495 810011541 810011542 810011456 810011444 111019457 111019432 810011363 111022594 111022591 111022900 111022946 111022962 111018282 111018112 111018290 810004817 810004823 810004810 111018269 111018277 111021428 810011471 810011279 111003188 111022082 111003175 111003352 111006795 810022362 111010805 111009745 111011451 111018649 111018665 810011466 111010793 810022380 810021534 810022067 810022063 810026984 810027001 111022933 810013050 810013258 810013222 810013257 810013259 810013252 810013228 810013234 810013260 111014927 111014916 111014931 111010806 810011515 810011286 111022763 810013280 810026977 810026933 111022593 111011354 111011243 810018941 810016266 810019314 810021603 810022377 810016567 810016308 810008598 810019916 810027527 810016129 111017249 111005521 810019411 111012461 111009709 111004530 111011937 111021463 810008584 810021695 810019061 111009705 111012297 111010758 810022080 810018994 810013435 810015809 810008572 111005528 810005246 810019536 111020947 111006544 810021670 810015952 810008641 111009761 810015957 111011959 111002736 111001077 111001000 810013243 810021676 810005707 810019417 111021830 810008601 111024200 111002438 111004712 810019263 810019549 111003958 810015805 810005305 810019534 111009253 111011355 810022081 111011220 111018122 111018108 111009258 810021584 810012605 810013437 810013211 111024169 810016576 810016677 111022973 810011392 810011537 810016114 810019265 810012606 810012602 111000450 810016680 810005283 111022086 111021316 810005284 111021700 810015777 111013958 111013963 810004825 111021439 111002381 810015776 810016451 810008568 810019414 810019542 111018264 111022764 810027009 810007445 111021312 810016667 810019517 810027021 810013182 810013469 810021614 810003371 810019521 111022583 810013363 810013470 111006786 111011921 111000532 810003431 111002441 111023530 111011267 111010817 111021467 810003465 810011284 111009785 111006543 111006536 810018964 111011605 111011256 111011897 111010959 111012264 111009797 810026983 810021693 810011488 111022785 810015818 111012469 111017242 111011980 111003675 111011625 111003655 810022134 810013286 810004814 111013803 810022097 810016343 810013366 810013273 810021507 810013263 111003191 111008593 810003425 810003432 111000146 810021366 111000761 111000606 111000101 111012298 101000035 111012319 111004277 111000463 810016312 810021350 111024174 810013281 810019554 111024190 111024173 111024172 810021359 111000663 810013353 810003436 810016544 111002386 111000140 111024401 111010762 810016358 111003541 111009823 810007464 810011374 810021384 810008600 810013194 111023549 810016134 111022584 810013191 111018667 111003266 810007451 111024203 810016133 111024568 111023529 810027549 111001085 810019912 111022950 810013040 810013041 111009765 810016437 111003322 810013475 111021442 111003498 111003270 810003559 111011240 111001032 810003618 111020975 810021572 810005704 810013246 810021726 810008623 810019064 111019455 810003679 111012311 111014949 111018124 810027110 111012280 810008616 810008560 111011193 111020951 111011975 111018788 111011234 111010820 810021529 810018960 810013249 810022130 810007466 810008577 810016349 111024403 111002740 111016734 111023524 810013082 810008482 111013991 810018979 111004186 810005320 810005368 810005250 111000103 111024384 111003658 111016843 111003494 810019036 111003564 810013365 810011348 111013967 810018959 111022760 111022773 810013081 111024372 111024400 111011441 111013978 810027147 810019051 111018672 810021731 810004809 810005702 810021363 111022087 111016767 810019046 810027546 810013034 810013059 810027148 810027012 111004198 810027007 111000143 810027550 810016553 111003341 111002405 111023548 111010456 810019295 111024201 111021011 111013994 810003349 810021590 810015781 111000211 111000212 810019253 810004567 810013483 810019523 810013450 810004566 810008580 810015823 810007431 111009255 111013974 810018940 810007404 111019414 810011509 111013812 810011497 810011309 111014011 111021813 111000710 111001075 111003053 810003558 810021569 810015947 111011953 111009717 111003959 111000676 810015955 111000607 111014008 111024188 111011382 111009839 810016125 810005365 810005306 111001028 111021446 810011393 111002394 810013334 111001004 111011950 810016348 111018647 111009808 810005367 810019431 111021799 810027108 111021977 810011342 810021375 810013201 810013482 810011362 810011384 111021788 111023538 810004752 111019427 111010774 111011349 810011292 810011517 810005280 810015774 810021517 111022578 111018294 111021315 111021443 810016144 111003127 111011178 111012459 810027525 810013209 810015792 111002402 810016319 111016844 111019463 111006555 111022940 810021509 810008583 111000665 111036453 111036561 810008405 111028682 810028576 810028578 810028556 111042323 810022327 810022123 810021704 810022124 810011498 810011521 810011446 111000271 111018132 111003366 111011595 810007438 810013350 111024182 111024196 111021981 810021734 810019022 810013473 810021562 810016100 810016107 810016128 810016106 810021692 810021694 810016362 810016142 810022085 810011376 810013480 810019267 810019024 810011321 810013037 810013039 810013063 810005352 111021805 111021833 810021588 810011354 111024397 810013096 810022072 111021435 810011366 810021696 810011455 111010978 810011472 810007408 111003862 111003662 111023552 111018280 111002746 810011370 111001863 111003263 111001030 111002449 111012479 111012270 111002738 111002734 810018957 111011196 111016736 111011988 111003342 111020985 810011467 111019417 111024396 810011494 810011336 111011257 111010434 111016719 810019914 111024085 810019259 810019919 810021552 111020978 111012257 111010455 111010788 111003281 810003367 111011239 810016455 111011186 111021787 111021789 810008576 810005281 111012456 111011199 111003038 111002006 810019286 111003962 111003381 810021544 810021365 111011251 111011962 111022761 111015635 111011218 810016141 810011278 111015614 810027122 111011176 810027540 810018990 111021801 810007446 111013806 810011324 810011330 111021809 810021573 810021617 111002705 111003355 810003341 111002747 111024389 111003045 111003146 111003183 111004168 111024086 111008703 810027164 810027151 810027523 111023532 111018642 810011445 111004286 111004177 111000701 810027008 810018951 111022776 111000157 111018113 111018807 111020957 810004643 111020953 111018797 810005241 810005315 810005289 810013036 810015778 810019533 810019316 810019281 810015791 810015829 810015810 810018956 111018808 810027019 810013103 810013056 810015825 111003128 111009763 111016722 111008702 111011592 810016574 111009787 111018803 810021716 810016314 810027533 111003323 111004515 111021808 810005285 810016377 810008629 810016551 111009752 111009711 111001070 111004527 810003438 111018812 810027042 111004520 810021619 111018683 111012269 111022579 111021835 810015793 810003351 810022144 810003552 111009805 810021515 111010451 111015624 111012480 810027543 111009706 111009816 111021786 111003952 111014933 810011540 810016152 810016099 810011389 810019520 111001021 111020965 810008586 111010767 111021005 810021550 810027538 810016127 111024217 111000106 810016375 111023525 810019045 810021675 111012282 810015953 810008492 111010768 111019458 810008551 111004176 810019283 810016448 810026996 111009827 111003951 810027532 111012306 810013055 111031417 810026998 111003353 810011518 111011952 111011904 111022939 810005252 810021525 111012320 111004187 810021598 810022369 810027528 111010766 111012304 111010963 111024366 111010791 810003368 810019006 810018968 111004524 111000568 111010756 111004640 111004715 111000153 111000448 810005703 111008586 810003337 111000526 111003180 111003657 111000561 111011594 111002742 111011992 111000145 111000324 810015960 810015813 810019405 810016569 810013449 111021798 111022775 810004807 810016376 810019269 810019311 810016545 810026929 810026939 810021708 111017235 111003174 810003433 810027516 810016441 810019005 111009762 111009239 111000321 810016268 810027152 111023541 111003144 111020971 111020989 111024216 111024206 810007457 810007424 810026936 810013444 810016443 111018668 810011452 810019279 810003372 111012291 810016143 810027520 810027517 111013986 111011361 111013997 111012253 810003464 111002403 111011264 810016658 810011299 810019333 111023568 111010792 810019915 810018978 111010784 111011965 810021358 810011282 810013031 111018273 111018286 111018279 111015858 111000259 111015612 810016352 810021685 810003735 111015857 810013283 111012265 111012465 810013220 111014915 111016762 111016765 111021796 810027018 111016829 111018680 111010765 810027123 810022108 810021377 111021695 111021785 810016130 810004571 810013284 111004642 111042196 810019553 810027011 810016160 810003380 810019924 111020997 111021784 111022765 111021827 111021536 810013184 81

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2614-2023

Related Recalls

There is a potential issue affecting Eterna Spinal Cord Stimulation (SCS) Implantable Pulse Generators (IPGs), Model 32400 and Liberta RC Deep Brain Stimulation (DBS) Implantable Pulse Generators (IPGs), Model 62400 . These devicess may experience a loss of communication with the Clinician Programmer (CP) and/or the Patient Controller (PC).

Oct 6, 2025 Implants & Prosthetics View Details β†’

There is a potential issue affecting Eterna Spinal Cord Stimulation (SCS) Implantable Pulse Generators (IPGs), Model 32400 and Liberta RC Deep Brain Stimulation (DBS) Implantable Pulse Generators (IPGs), Model 62400 . These devicess may experience a loss of communication with the Clinician Programmer (CP) and/or the Patient Controller (PC).

Oct 6, 2025 Implants & Prosthetics View Details β†’

Due to a manufacturing issue, leadless pacemakers may have been exposed to higher than normal temperatures which may result in premature battery failure.

Feb 15, 2025 Implants & Prosthetics Nationwide View Details β†’