AB SCIEX API 3200MD" LC/MS/MS System with software: MultiQuantMD 3.0In vitro diagnostic to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens Part Number 5024543.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Serial Numbers: BV20031308, BV20081311, BV20051310, BV20071311, BV20101401, BV20131401
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Ab Sciex
- Reason for Recall:
- Software- AB SCIEX API 3200MD LC/MS/MS SYSTEM with MultiQuantMD 3.0 software may provide incorrect data for patient diagnosis
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
AB SCIEX API 3200MD" LC/MS/MS System with software: MultiQuantMD 3.0In vitro diagnostic to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens Part Number 5024543.
Product Codes/Lot Numbers:
Serial Numbers: BV20031308, BV20081311, BV20051310, BV20071311, BV20101401, BV20131401
Distribution:
Distributed in: FL, NC, OK, PA, TX
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2609-2014
Related Recalls
In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.