uMI 550 System - Product Usage: intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model: 88000057 GTIN/UDI Codes: 06971576832026 / 01 06971576832026 240 88000057 21 200017 06971576832026 / 01 06971576832026 240 88000057 21 200023 06971576832026 / 01 06971576832026 240 88000057 21 200024 06971576832026 / 01 06971576832026 240 88000057 21 200036 06971576832026 / 01 06971576832026 240 88000057 21 200045 06971576832026 / 01 06971576832026 240 88000057 21 200047 06971576832057 / 01 06971576832057 240 88000057 21 230002 06971576832057 / 01 06971576832057 240 88000057 21 230003 06971576832057 / 01 06971576832057 240 88000057 21 230004 06971576832057 / 01 06971576832057 240 88000057 21 200059 06971576832057 / 01 06971576832057 240 88000057 21 230005 06971576832057 / 01 06971576832057 240 88000057 21 230006 06971576832057 / 01 06971576832057 240 88000057 21 230007 06971576832057 / 01 06971576832057 240 88000057 21 230008 06971576832057 / 01 06971576832057 240 88000057 21 230009 06971576832057 / 01 06971576832057 240 88000057 21 230010 06971576832057 / 01 06971576832057 240 88000057 21 230011 06971576832057 / 01 06971576832057 240 88000057 21 230012 06971576832057 / 01 06971576832057 240 88000057 21 230013 06971576832057 / 01 06971576832057 240 88000057 21 230014 06971576832057 / 01 06971576832057 240 88000057 21 230015 06971576832057 / 01 06971576832057 240 88000057 21 230016 06971576832057 / 01 06971576832057 240 88000057 21 230017 Serial Numbers: 200017; 200023; 200024; 200036; 200045; 200047; 230002; 230003; 230004; 200059; 230005; 230006; 230007; 230008; 230009; 230010; 230011; 230012; 230013; 230014; 230015; 230016; 230017;
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Shanghai United Imaging Healthcare Co., Ltd.
- Reason for Recall:
- The EXAM acquisition workflow could occasionally freeze after canceling the PET scan.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
uMI 550 System - Product Usage: intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.
Product Codes/Lot Numbers:
Model: 88000057 GTIN/UDI Codes: 06971576832026 / 01 06971576832026 240 88000057 21 200017 06971576832026 / 01 06971576832026 240 88000057 21 200023 06971576832026 / 01 06971576832026 240 88000057 21 200024 06971576832026 / 01 06971576832026 240 88000057 21 200036 06971576832026 / 01 06971576832026 240 88000057 21 200045 06971576832026 / 01 06971576832026 240 88000057 21 200047 06971576832057 / 01 06971576832057 240 88000057 21 230002 06971576832057 / 01 06971576832057 240 88000057 21 230003 06971576832057 / 01 06971576832057 240 88000057 21 230004 06971576832057 / 01 06971576832057 240 88000057 21 200059 06971576832057 / 01 06971576832057 240 88000057 21 230005 06971576832057 / 01 06971576832057 240 88000057 21 230006 06971576832057 / 01 06971576832057 240 88000057 21 230007 06971576832057 / 01 06971576832057 240 88000057 21 230008 06971576832057 / 01 06971576832057 240 88000057 21 230009 06971576832057 / 01 06971576832057 240 88000057 21 230010 06971576832057 / 01 06971576832057 240 88000057 21 230011 06971576832057 / 01 06971576832057 240 88000057 21 230012 06971576832057 / 01 06971576832057 240 88000057 21 230013 06971576832057 / 01 06971576832057 240 88000057 21 230014 06971576832057 / 01 06971576832057 240 88000057 21 230015 06971576832057 / 01 06971576832057 240 88000057 21 230016 06971576832057 / 01 06971576832057 240 88000057 21 230017 Serial Numbers: 200017; 200023; 200024; 200036; 200045; 200047; 230002; 230003; 230004; 200059; 230005; 230006; 230007; 230008; 230009; 230010; 230011; 230012; 230013; 230014; 230015; 230016; 230017;
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2593-2021
Related Recalls
Shanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography (PET) and Computed Tomography (CT), Model uMI 550 Part Number: 88000657 88000057
Shanghai United Imaging Healthcare Co.
Due to a software issue where the process of patient scanning, the scatter correction may occasionally fail with will potentially cause a failure of the PET image reconstruction and generation of PET images.
Positron Emission Tomography and Computed Tomography System, Model: uMI 550
Shanghai United Imaging Healthcare Co.
The wireless VSM module of a mobile PET/CT system, operating in an environment with strong Wi-Fi signals, may experience ECG signal and respiratory signal loss due to Wi-Fi interference. ECG and respiratory signal loss during acquisition can result in the failure of ECG and respiratory-gated reconstruction of the PET scan, which may require rescanning of the patient.
Digital Medical X-ray Imaging System, Model: uDR 596i, uDR 592h
Shanghai United Imaging Healthcare Co.
X-ray imaging system positioning image and protocol label is reversed for Flexion and Extension on C-Spine and L-Spine, this will cause the image to be incorrectly labeled and may cause the patient to have a repeat exposure.