Calibrate (unexposed) nanoDot D2DNS, Model Numbers: a) 18140-000; b) 18140-KIT; c) 18140-SET; radiation monitoring dosimeter used with the microSTAR readers

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI/DI 0860003399903, all batch numbers
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Landauer
Reason for Recall:
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
completed

Product Information

Full Description:

Calibrate (unexposed) nanoDot D2DNS, Model Numbers: a) 18140-000; b) 18140-KIT; c) 18140-SET; radiation monitoring dosimeter used with the microSTAR readers

Product Codes/Lot Numbers:

UDI/DI 0860003399903, all batch numbers

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2570-2023

Related Recalls

Permanent discontinuance of the microSTARii medical dosimetry system. Landauer has initiated this action to remove all reference to use of the microStar readers with the nanoDot or any otherbdosimeter for medical applications from the product s user manual and software and to remove microStar readers from customers who use them only for medical applications.

Feb 8, 2024 Other Medical Devices Nationwide View Details →

MicroSTARii readers may be contributing to the nanodot dosimeter measurement inaccuracy in two possible ways. The first is related to the interaction of the reader s LED beam profile with the nanoDot dosimeter. The second is debris formation due to drawer actuation, resulting in a change of reader response that may not be detectable unless known dose QCs are being run regularly.

Nov 15, 2023 Other Medical Devices View Details →

LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.

Jul 12, 2023 Diagnostic Equipment View Details →