Medfusion¿ Model 4000 Syringe Infusion Pump, with PharmGuard¿ Server Software (PGS). Sold under the following names: MEDFUSION¿ 4000 Pump V1.0; MEDFUSION¿ 4000 Pump V1.1, v1.1.1, v1.1.2; MEDFUSION¿ 4000 Pump V1.1, V1.1.1, V1.1.2; MEDFUSION¿ 4000 Pump V1.5, V1.5.1; MEDFUSION¿ 4000 Pump V1.5, V1.5.1; MEDFUSION¿ 4000 Pump V1.6, V1.6.1 Product Usage: The pumps are indicated for the following uses: 1. In the administration of fluids requiring precisely controlled infusion rates including blood or blood products, lipids, drugs, antibiotics, enteral solutions and other therapeutic fluids; 2. By the following delivery routes: arterial, epidural, intravenous, intrathecal, subcutaneous, and enteral; 3. By the following delivery modes: continuous, volume/time, mass, body weight, intermittent and bolus.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Model Numbers (Serial Numbers): a. 4000-0100-50 (2000245, 2000246, 2000247, 2000248, 2000251, 2000252, 2000254, 2000255, 2000256, 2000257, 2000258, 2000260, 2000261, 2000264, 2000265, 2000267, 2000268, 2000269, 2000270, 2000272, 2000273, 2000274, 2000276,2000277, 2000278, 2000279, 2000280, 2000281, 2000282, 2000283, 2000284, 2000285, 2000286, 2000287, 2000288, 2000290, 2000291, 2000297, 2000298, 2000299, 2000300, 2000302, 2000304, 2000308, 2000309, 2000310, 2000311, 2000312, 2000313, 2000314, 2000315, 2000317, 2000318, 2000320, 2000321, 2000322, 2000324, 2000325, 2000326, 2000329, 2000331, 2000332, 2000336, 2000338, 2000341, 2000343, 2000345, 2000346, 2000348, 2000349, 2000352, 2000353, 2000355, 2000356, 2000358, 2000359, 2000360, 2000363, 2000364, 2000366, 2000370, 2000371, 2000374, 2000375, 2000376, 2000377, 2000379, 2000380, 2000381, 2000383, 2000385, 2000386, 2000387, 2000388, 2000389, 2000390, 2000392, 2000393, 2000395, 2000396, 2000403, 2000408, 2000409, 2000413, 2000415, 2000417, 2000426, 2000433, 2000434, 2000305, 2000330, 2000333, 2000357, 2000361, 2000362, 2000365, 2000369, 2000382, 2000391, 2000397, 2000398, 2000400, 2000402, 2000405, 2000414, 2000419, 2000423, 2000425, 2000430, 2000431, 2000439, 2000440, 2000441, 2000443, 2000316, 2000428, 2000438, 2000262, 2000342, 2000263, 2000327, 2000334, 2000378, 2000303, 2000351, 2000437, 2000301, 2000249, 2000275, 2000292, 2000373, 2000407, 2000394, 2000295, 2000335, 2000435, 2000422, 2000293, 2000418, 2000259, 2000367, 2000253, 2000416, 2000289, 2000344, 2000372, 2000350, 2000266, 2000250, 2000328, 2000401, 2000384, 2000404, 2000319, 2000399, 2000339, 2000412, 2000294, 2000340, 2000296, 2000429, 2000337, 2000368, 2000323, 2000307, 2000442); b. 4000-0101-51 (2002111, 2002121, 2002126, 2002127, 2002156, 2002171, 2002172, 2002221, 2002228, 2002241, 2002251, 2002253, 2002255, 2002263, 2002265, 2002272, 2002283, 2002291, 2002307, 2002318, 2002324, 2002336, 2002337, 2002339, 2002344, 2002353, 2002355, 2002356, 2002359, 2002195, 2002196, 2002197, 2002198, 2002199, 2002378, 2002379, 2004959, 2002407, 2002409, 2002410, 2002411, 2002412, 2002414, 2002415, 2002416, 2002417, 2002418, 2002419, 2002420, 2002421, 2002423, 2002427, 2002428, 2002429, 2002431, 2002432, 2002433, 2002437, 2002438, 2002439, 2002440, 2002441, 2002443, 2002445, 2002447, 2002448, 2002451, 2002452, 2002453, 2002454, 2002455, 2002456, 2002459, 2002460, 2002461, 2002462, 2002463, 2002464, 2002465, 2002466, 2002468, 2002469, 2002470, 2002472, 2002473, 2002474, 2002475, 2002478, 2002480, 2002483, 2002484, 2002485, 2002488, 2002489, 2002490, 2002491, 2002492, 2002494, 2002495, 2002496, 2002497, 2002498, 2002579, 2002600, 2002608, 2002616, 2002619, 2002630, 2002634, 2002643, 2002649, 2002705, 2002719, 2002722, 2002744, 2002771, 2002799, 2002811, 2002824, 2002826, 2002833, 2002835, 2002842, 2002844, 2002846, 2002857, 2004163, 2004164, 2004165, 2004167, 2004168, 2004169, 2004171, 2004172, 2004174, 2004176, 2004177, 2004178, 2004179, 2004180, 2004182, 2004183, 2004184, 2004185, 2004186, 2004187, 2004188, 2004189, 2004190, 2004191, 2004192, 2004193, 2004194, 2004195, 2004196, 2004197, 2004198, 2004199, 2004200, 2004201, 2004202, 2004204, 2004205, 2004206, 2004207, 2004208, 2004209, 2004210, 2004212, 2004215, 2004216, 2004217, 2004218, 2004219, 2004220, 2004221, 2004222, 2004225, 2004226, 2004227, 2004228, 2004229, 2004230, 2004231, 2004232, 2004233, 2004234, 2004235, 2004237, 2004238, 2004239, 2004240, 2004241, 2004242, 2004244, 2004245, 2004246, 2004247, 2004248, 2004250, 2004251, 2004252, 2004254, 2004255, 2004256, 2004258, 2004259, 2004260, 2004261, 2004262, 2004263, 2004265, 2004266, 2004267, 2004268, 2004269, 2004270, 2004273, 2004274, 2004275, 2004276, 2004277, 2004278, 2004279, 2004280, 2004282, 2004283, 2004284, 2004285, 2004286, 2004287, 2004289, 2004291, 2004292, 2004293, 2004296, 2004297, 2004298, 2004299, 2004300, 2004301, 2004303, 2004304, 2004305, 2004306, 2004307, 2004309, 2004310, 2004311, 2004312, 2004313, 2004315, 2004317, 2004319, 2004320, 2004322, 2004323, 2004324, 2004325, 2004326, 2004327, 2004328, 2004330, 2004331, 2004332, 2004333, 2004334, 2004335, 2004336, 2004337, 2004339, 2004340, 2004342, 2004343, 2004344, 2004345, 2004346, 2004347, 2004348, 2004349, 2004350, 2004351, 2004353, 2004354, 2004355, 2004356, 2004357, 2004358, 2004360, 2004361, 2004363, 2004364, 2004366, 2004367, 2004369, 2004370, 2004371, 2004372, 2004373, 2004374, 2004375, 2004376, 2004377, 2004378, 2004379, 2004380, 2004382, 2004383, 2004384, 2004385, 2004386, 2004387, 2004388, 2004389, 2004390, 2004391, 2004394, 2004395, 2004396, 2004397, 2004398, 2004399, 2004400, 2004403, 2004406, 2004407, 2004408, 2004409, 2004410, 2004412, 2004413, 2004414, 2004416, 2004417, 2004418, 2004420, 2004421, 2004422, 2004423, 2004424, 2004426, 2004428, 2004429, 2004430, 2004431, 2004432, 2004433, 2004434, 2004437, 2004438, 2004439, 2004440, 2004442, 2004443, 2004444, 2004445, 2004446, 2004447, 2004448, 2004449, 2004450, 2004452, 2004453, 2004454, 2004456, 2004457, 2004459, 2004461, 2004462, 2004464, 2004466, 2004467, 2004468, 2004469, 2004470, 2004472, 2004473, 2004474, 2004475, 2004476, 2004481, 2004482, 2004483, 2004484, 2004485, 2004486, 2004487, 2004488, 2004489, 2004491, 2004493, 2004494, 2004496, 2004499, 2004500, 2004501, 2004502, 2004515, 2004517, 2004518, 2004521, 2004523, 2004524, 2004525, 2004529, 2004535, 2004539, 2004543, 2004544, 2004545, 2004547, 2004549, 2004552, 2004554, 2004555, 2004556, 2004557, 2004558, 2004559, 2004560, 2004561, 2004562, 2004564, 2004566, 2004567, 2004568, 2004569, 2004570, 2004572, 2004573, 2004574, 2004575, 2004576, 2004577, 2004578, 2004579, 2004580, 2004582, 2004583, 2004584, 2004588, 2004589, 2004591, 2004593, 2004594, 2004595, 2004597, 2004598, 2004599, 2004600, 2004601, 2004602, 2004603, 2004605, 2004608, 2004610, 2004611, 2004612, 2004614, 2004615, 2004616, 2004618, 2004619, 2004620, 2004621, 2004622, 2004624, 2004625, 2004627, 2004119, 2008536, 2008575, 2004833, 2004843, 2004847, 2004854, 2004861, 2004865, 2004868, 2004877, 2004893, 2004894, 2004898, 2004907, 2004934, 2004938, 2005073, 2005077, 2005083, 2005085, 2005086, 2005092, 2005093, 2005097, 2005098, 2005100, 2005112, 2005113, 2005116, 2005118, 2005120, 2005130, 2005132, 2005135, 2005137, 2005138, 2005142, 2005145, 2005147, 2005149, 2005151, 2005153, 2005161, 2005163, 2005166, 2005167, 2005170, 2005173, 2005178, 2005183, 2005286, 2005287, 2005293, 2005294, 2005299, 2005303, 2005309, 2005311, 2005312, 2005314, 2005320, 2005321, 2005325, 2005326, 2005331, 2005333, 2005334, 2005338, 2005339, 2005341, 2005342, 2005349, 2005353, 2005361, 2005367, 2005370, 2005382, 2005386, 2005392, 2005403, 2005410, 2005413, 2005420, 2005428, 2005432, 2005434, 2005441, 2005443, 2005447, 2005458, 2005461, 2005462, 2005465, 2005469, 2005471, 2005475, 2005488, 2005489, 2005497, 2005498, 2005506, 2005515, 2005533, 2005535, 2005543, 2005554, 2005555, 2005580, 2005582, 2005585, 2005586, 2005587, 2005593, 2005596, 2005597, 2005599, 2005606, 2005614, 2005616, 2005620, 2005642, 2005644, 2005656, 2005658, 2005663, 2005664, 2005667, 2005673, 2005689, 2005697, 2005713, 2005725, 2005727, 2005741, 2005745, 2005747, 2005757, 2005762, 2005774, 2005775, 2005782, 2005788, 2005791, 2005804, 2005818, 2005822, 2005829, 2005834, 2005835, 2005842, 2005854, 2005855, 2005856, 2005861, 2005862, 2005869, 2005877, 2005879, 2005884, 2005885, 2005886, 2005890, 2005899, 2005900, 2005903, 2005907, 2005914, 2005915, 2005927, 2006134, 2006135, 2006136, 2006137, 2006138, 2006139, 2006140, 2006142, 2006143, 2006144, 2006145, 2006147, 2006149, 2006150, 2006151, 2006152, 2006153, 2006154, 2006156, 2006158, 2006159, 2006160, 2006161, 2006162, 2006163, 2006164, 2006165, 2006166, 2006167, 2006168, 2006169, 2006170, 2006171, 2006172, 2006173, 2006174, 2006175, 2006176, 2006177, 2006178, 2006179, 2006180, 2006181, 2006182, 2006183, 2006184, 2006185, 2006187, 2006188, 2006189, 2006191, 2006192, 2006193, 2006194, 2006195, 2006196, 2006197, 2006198, 2006199, 2006200, 2006201, 2006202, 2006203, 2006204, 2006205, 2006206, 2006207, 2006208, 2006209, 2006211, 2006212, 2006213, 2006215, 2006216, 2006217, 2006218, 2006219, 2006221, 2006224, 2006225, 2006226, 2006227, 2006228, 2006232, 2006233, 2006234, 2006235, 2006238, 2006239, 2006241, 2006243, 2006245, 2006246, 2006247, 2006248, 2006250, 2006251, 2006252, 2006253, 2006254, 2006255, 2006257, 2006258, 2006260, 2006262, 2006263, 2006265, 2006267, 2006268, 2006270, 2006272, 2006273, 2006274, 2006276, 2006277, 2006280, 2006281, 2006282, 2006283, 2006284, 2006285, 2006286, 2006287, 2006289, 2006290, 2006291, 2006294, 2006295, 2006297, 2006299, 2006300, 2006301, 2006302, 2006303, 2006304, 2006305, 2006307, 2006308, 2006309, 2006310, 2006311, 2006313, 2006314, 2006316, 2006317, 2006319, 2006366, 2007861, 2007899, 2007918, 2006392, 2006393, 2006395, 2006397, 2006399, 2006400, 2006401, 2006403, 2006404, 2006405, 2006406, 2006407, 2006408, 2006409, 2006412, 2006413, 2006414, 2006415, 2006416, 2006417, 2006418, 2006419, 2006420, 2006421, 2006422, 2006426, 2006428, 2006430, 2006431, 2006432, 2006433, 2006434, 2006435, 2006437, 2006439, 2006440, 2006441, 2006442, 2006443, 2006444, 2006445, 2006446, 2006447, 2006448, 2006451, 2006452, 2006453, 2006455, 2006457, 2006458, 2006459, 2006460, 2006461, 2006462, 2006463, 2006464, 2006466, 2006468, 2006472, 2006473, 2006475, 2006477, 2006484, 2006485, 2006488, 2006489, 2006490, 2006491, 2006494, 2006495, 2006496, 2006497, 2006498, 2006500, 2006501, 2006502, 2006503, 2006505, 2006506, 2006507, 2006508, 2006509, 2006511, 2006513, 2006514, 2006515, 2006516, 2006517, 2006518, 2006519, 2006521, 2006522, 2006523, 2006524, 2006526, 2006528, 2006529, 2006530, 2006531, 2006534, 2006536, 2006537, 2006539, 2006542, 2006543, 2006544, 2006546, 2006547, 2006548, 2006549, 2006552, 2006555, 2006558, 2006559, 2006560, 2006561, 2006562, 2006564, 2006566, 2006567, 2006568, 2006569, 2006570, 2006571, 2006576, 2006577, 2006578, 2006579, 2006580, 2006581, 2006584, 2006585, 2006586, 2006588, 2006589, 2006590, 2006591, 2006592, 2006593, 2006595, 2006597, 2006598, 2006599, 2006600, 2006601, 2006603, 2006604, 2006606, 2006607, 2006610, 2006611, 2006612, 2006613, 2006615, 2006616, 2006617, 2006618, 2006619, 2006620, 2006621, 2006623, 2006624, 2006626, 2006629, 2006630, 2006631, 2006633, 2006634, 2006640, 2006641, 2006642, 2006643, 2006644, 2006647, 2006648, 2006649, 2006651, 2006652, 2006653, 2006654, 2006656, 2006658, 2006659, 2006660, 2006661, 2006662, 2006663, 2006665, 2006666, 2006667, 2006669, 2006670, 2006671, 2006768, 2006769, 2006770, 2006772, 2006774, 2006775, 2006776, 2006777, 2006780, 2006781, 2006782, 2006784, 2006785, 2006787, 2006788, 2006790, 2006792, 2006793, 2006795, 2006798, 2006800, 2006801, 2006802, 2006803, 2006804, 2006805, 2006808, 2006809, 2006811, 2006812, 2006814, 2006815, 2006818, 2006820, 2006821, 2006822, 2006823, 2006826, 2006827, 2006828, 2006829, 2006830, 2006832, 2006833, 2006835, 2006836, 2006839, 2006840, 2006841, 2006842, 2006846, 2006847, 2006848, 2006849, 2006853, 2006855, 2006856, 2006857, 2006861, 2006862, 2006863, 2006864, 2006867, 2006869, 2006870, 2006871, 2006872, 2006873, 2006874, 2006875, 2006877, 2006878, 2006879, 2006880, 2006881, 2006882, 2006883, 2006885, 2006886, 2006887, 2006888, 2006890, 2006895, 2006896, 2006897, 2006900, 2006901, 2006902, 2006906, 2006907, 2006908, 2006909, 2006910, 2006916, 2006917, 2006926, 2006927, 2006928, 2006929, 2006930, 2006931, 2006932, 2006934, 2006935, 2006938, 2006939, 2006940, 2006942, 2006944, 2006946, 2006948, 2006949, 2006950, 2006951, 2006952, 2006954, 2006955, 2006956, 2006957, 2006958, 2006959, 2006960, 2006961, 2006963, 2006965, 2006968, 2006969, 2006970, 2006972, 2006975, 2006976, 2006977, 2006979, 2006981, 2006982, 2006983, 2006984, 2006985, 2006987, 2006988, 2006989, 2006991, 2006992, 2006994, 2006995, 2006996, 2006997, 2007001, 2007002, 2007007, 2007008, 2007009, 2007010, 2007011, 2007012, 2007015, 2007019, 2007020, 2007021, 2007022, 2007023, 2007024, 2007026, 2007027, 2007028, 2007029, 2007031, 2007033, 2007036, 2007038, 2007041, 2007042, 2007043, 2007044, 2007046, 2007049, 2007050, 2007054, 2007055, 2007059, 2007061, 2007062, 2007064, 2007067, 2007069, 2007073, 2007075, 2007076, 2007078, 2007079, 2007080, 2007081, 2007082, 2007083, 2007085, 2007087, 2007088, 2007089, 2007090, 2007091, 2007092, 2007093, 2007094, 2007097, 2007098, 2007099, 2007100, 2007101, 2007102, 2007103, 2007105, 2007107, 2007110, 2007112, 2007114, 2007118, 2007119, 2007121, 2007122, 2007127, 2007128, 2007130, 2007134, 2007136, 2007137, 2007138, 2007140, 2007141, 2007142, 2007143, 2007147, 2007149, 2007150, 2007151, 2007153, 2007154, 2007155, 2007159, 2007163, 2007164, 2007170, 2007171, 2007172, 2007173, 2007174, 2007175, 2007176, 2007178, 2007183, 2007184, 2007187, 2007188, 2007191, 2007194, 2007195, 2007196, 2007198, 2007199, 2007200, 2007202, 2007205, 2007206, 2007209, 2007211, 2007212, 2007213, 2007215, 2007216, 2007217, 2007218, 2007219, 2007220, 2007221, 2007223, 2007224, 2007225, 2007226, 2007227, 2007232, 2007233, 2007239, 2007240, 2007241, 2007242, 2007244, 2007246, 2007249, 2007250, 2007252, 2007253, 2007254, 2007256, 2007257, 2007259, 2007260, 2007261, 2007266, 2007269, 2007271, 2007274, 2007275, 2007277, 2007278, 2007279, 2007280, 2007282, 2007283, 2007284, 2007286, 2007287, 2007288, 2007291, 2007292, 2007294, 2007297, 2007298, 2007299, 2007301, 2007302, 2007310, 2007311, 2007314, 2007315, 2007316, 2007317, 2007319, 2007321, 2007323, 2007324, 2007325, 2007326, 2007329, 2007330, 2007333, 2007334, 2007335, 2007336, 2007337, 2007340, 2007341, 2007343, 2007344, 2007685, 2004774, 2007946, 2002220, 2007869, 2004811, 2007947, 2007949, 2007953, 2007956, 2007957, 2007958, 2008036, 2004812, 2004813, 2004816, 2004817, 2004819, 2004821, 2004822, 2004825, 2004826, 2004827, 2004828, 2004829, 2004830, 2004831, 2006646, 2007005, 2007988, 2007993, 2008052, 2008055, 2008056, 2008060, 2008061, 2008062, 2008063, 2008066, 2008067, 2008068, 2008069, 2008072, 2008074, 2008077, 2008079, 2008087, 2008089, 2008090, 2008091, 2008092, 2008094, 2008098, 2008099, 2008101, 2008102, 2008106, 2008107, 2008108, 2008110, 2008111, 2008112, 2008117, 2008120, 2008123, 2008125, 2008126, 2008132, 2008134, 2008135, 2008136, 2008137, 2008138, 2008139, 2008140, 2008141, 2008142, 2008144, 2008145, 2008147, 2008148, 2008149, 2008150, 2008151, 2008154, 2008156, 2008158, 2008159, 2008161, 2008162, 2008163, 2008164, 2008167, 2008168, 2008172, 2008174, 2008176, 2008177, 2008178, 2008181, 2008182, 2008184, 2008185, 2008186, 2008187, 2008189, 2008191, 2008192, 2008193, 2008196, 2008197, 2008198, 2008199, 2008200, 2008202, 2008204, 2008208, 2008210, 2008215, 2008216, 2008220, 2008228, 2008232, 2008236, 2008237, 2008238, 2008240, 2008241, 2008242, 2008243, 2008245, 2008246, 2008247, 2008248, 2008249, 2008250, 2008256, 2008258, 2008260, 2008261, 2008264, 2008265, 2008267, 2008268, 2008270, 2008273, 2008277, 2008279, 2008283, 2008284, 2008285, 2008286, 2008289, 2008290, 2008291, 2008292, 2008294, 2008299, 2008301, 2008304, 2008307, 2008308, 2008309, 2008313, 2008315, 2008316, 2008317, 2008318, 2008319, 2008323, 2008325, 2008326, 2008330, 2008331, 2008333, 2008335, 2008336, 2008338, 2008339, 2008341, 2008342, 2008346, 2008349, 2008352, 2008355, 2008359, 2008004, 2008008, 2008010, 2008360, 2008361, 2008362, 2002693, 2008363, 2008365, 2008031, 2008033, 2008375, 2008654, 2008719, 2008723, 2008817, 2008917, 2007276, 2002729, 2007160, 2009028, 2009042, 2002180, 2009087, 2006819, 2009090, 2009092, 2008378, 2008379, 2009507, 2009508, 2002408, 2009631, 2009634, 2009642, 2009643, 2009644, 2009646, 2009647, 2009648, 2009651, 2009653, 2009654, 2009656, 2009658, 2009659, 2009662, 2009663, 2009665, 2009666, 2009667, 2009671, 2009672, 2009673, 2009674, 2009676, 2009680, 2009681, 2009682, 2009687, 2009688, 2009701, 2009702, 2009704, 2009705, 2009706, 2009707, 2009710, 2009714, 2009720, 2009722, 2009728, 2009730, 2009731, 2009732, 2009735, 2009736, 2009737, 2009739, 2009742, 2009743, 2009746, 2009748, 2009755, 2009757, 2009759, 2009760, 2009764, 2009766, 2009767, 2009769, 2009773, 2009774, 2009775, 2009776, 2009778, 2009780, 2009781, 2009783, 2009784, 2009786, 2009787, 2009789, 2009794, 2009795, 2009796, 2009799, 2009802, 2009803, 2009810, 2009812, 2009819, 2009823, 2009824, 2009827, 2009828, 2009831, 2009833, 2009834, 2009836, 2009837, 2009842, 2009843, 2009845, 2009846, 2009848, 2009849, 2009853, 2009854, 2009855, 2002486, 2002422, 2002581, 2004531, 2002502, 2010054, 2002467, 2002580, 2010143, 2010153, 2010154, 2010156, 2010201, 2010205, 2010256, 2010259, 2010271, 2006852, 2010343, 2010345, 2010346, 2010347, 2010348, 2010349, 2010350, 2010352, 2010356, 2010357, 2010358, 2010359, 2010360, 2010363, 2010365, 2010366, 2010371, 2010373, 2010374, 2010375, 2010376, 2010377, 2010379, 2010380, 2010385, 2010387, 2010390, 2010392, 2010393, 2010394, 2010395, 2010396, 2008171, 2009552, 2009553, 2009554, 2009555, 2009558, 2009560, 2009562, 2009563, 2009565, 2009568, 2009569, 2009570, 2009571, 2009573, 2009574, 2009575, 2009577, 2009578, 2009579, 2009581, 2009582, 2009583, 2009584, 2009585, 2009586, 2009588, 2009589, 2009590, 2009591, 2009592, 2009593, 2009594, 2009595, 2009596, 2009597, 2009599, 2009600, 2009601, 2009602, 2009603, 2002762, 2007228, 2007065, 2010383, 2010398, 2010399, 2010400, 2010401, 2010403, 2010404, 2010405, 2010406, 2010407, 2010410, 2010411, 2010412, 2010414, 2010415, 2010416, 2010417, 2010418, 2010421, 2010423, 2010424, 2010425, 2010426, 2010427, 2010428, 2010429, 2010430, 2010431, 2010434, 2010435, 2010436, 2010437, 2010438, 2010440, 2010441, 2010443, 2010444, 2010445, 2010446, 2010447, 2010448, 2010449, 2010450, 2010451, 2010452, 2010457, 2010459, 2010463, 2010464, 2010465, 2010466, 2010467, 2010468, 2010487, 2010488, 2010490, 2010491, 2010492, 2010493, 2010494, 2010495, 2010496, 2010498, 2010499, 2010500, 2010501, 2010502, 2010503, 2010504, 2010505, 2010506, 2010507, 2010508, 2010509, 2010510, 2010511, 2010512, 2010513, 2010514, 2010515, 2010516, 2010517, 2010518, 2010519, 2010520, 2010522, 2010523, 2010524, 2010525, 2010526, 2010527, 2010528, 2010529, 2010531, 2010532, 2010533, 2010536, 2010539, 2010540, 2010541, 2010543, 2010544, 2010545, 2010546, 2010548, 2010550, 2010551, 2010552, 2010553, 2010554, 2010555, 2010557, 2010558, 2010559, 2010560, 2010561, 2010563, 2010564, 2010565, 2010566, 2010568, 2010569, 2010571, 2010572, 2010573, 2010574, 2010575, 2010576, 2010578, 2010579, 2010580, 2010581, 2010582, 2010583, 2010584, 2010585, 2010586, 2010587, 2010588, 2010589, 2010590, 2010593, 2010594, 2010595, 2010596, 2010597, 2010598, 2010599, 2010600, 2010601, 2010602, 2010604, 2010605, 2010606, 2010607, 2010608, 2010609, 2010612, 2010613, 2010614, 2010615, 2010616, 2010617, 2010620, 2010621, 2010622, 2010623, 2010624, 2010625, 2010626, 2010628, 2010629, 2010631, 2010632, 2010634, 2010636, 2010637, 2010639, 2010640, 2010641, 2010642, 2010643, 2010644, 2010646, 2010648, 2010649, 2010650, 2010651, 2010655, 2010656, 2010658, 2010659, 2010660, 2010662, 2010664, 2010665, 2010666, 2010667, 2010670, 2002479, 2010469, 2010470, 2010471, 2010472, 2010473, 2010474, 2010672, 2010675, 2010676, 2010677, 2010678, 2010679, 2010680, 2010681, 2010682, 2010684, 2010687, 2010692, 2010697, 2010698, 2010704, 2010705, 2010706, 2010707, 2010708, 2010710, 2010711, 2010712, 2010713, 2010714, 2010715, 2010716, 2010717, 2010718, 2010719, 2010722, 2010723, 2010724, 2010726, 2010727, 2010728, 2010729, 2010731, 2010732, 2010733, 2010735, 2010736, 2010737, 2010738, 2010739, 2010744, 2010747, 2010748, 2010749, 2010750, 2010751, 2010753, 2010754, 2010755, 2010756, 2010759, 2010760, 2010761, 2010764, 2010765, 2010767, 2010768, 2010770, 2010771, 2010772, 2010773, 2010775, 2010776, 2010777, 2010778, 2010780, 2010781, 2010783, 2010784, 2010787, 2010791, 2010792, 2010793, 2010794, 2010799, 2010800, 2010801, 2010803, 2010804, 2010806, 2010808, 2010810, 2010811, 2010813, 2010816, 2010818, 2010820, 2010821, 2010824, 2010827, 2010829, 2010830, 2010831, 2010835, 2010836, 2010838, 2010839, 2010840, 2010841, 2010842, 2010844, 2010847, 2010848, 2010849, 2010850, 2010853, 2010856, 2010857, 2010858, 2010859, 2010860, 2010861, 2010862, 2010863, 2010865, 2010866, 2010868, 2010869, 2010870, 2010873, 2010875, 2010876, 2010879, 2010880, 2010881, 2010883, 2010884, 2010885, 2010886, 2010887, 2010890, 2010891, 2010893, 2010896, 2010897, 2010898, 2010899, 2010901, 2010902, 2010903, 2010908, 2010909, 2010911, 2010914, 2010916, 2010917, 2010918, 2010919, 2010920, 2010921, 2010922, 2010925, 2010926, 2010928, 2010929, 2010930, 2010932, 2010933, 2010934, 2010936, 2010938, 2010940, 2010941, 2010944, 2010945, 2010946, 2010948, 2010949, 2010951, 2010952, 2010954, 2010955, 2010956, 2010957, 2010958, 2010959, 2010960, 2010961, 2010962, 2010964, 2010965, 2010967, 2010968, 2010969, 2010970, 2010971, 2010973, 2010974, 2010975, 2010976, 2010977, 2010978, 2010979, 2010980, 2010981, 2010982, 2010985, 2010986, 2010988, 2010990, 2010991, 2010992, 2010996, 2010997, 2010998, 2011000, 2011001, 2011002, 2011004, 2011005, 2011006, 2011008, 2011009, 2011010, 2011012, 2011013, 2011014, 2011016, 2011017, 2011018, 2011019, 2011020, 2011026, 2011028, 2011029, 2011031, 2011032, 2011034, 2011035, 2011037, 2011039, 2011042, 2011043, 2011044, 2011050, 2011051, 2011053, 2011070, 2011071, 2011074, 2011075, 2011076, 2011080, 2011081, 2011082, 2011083, 2011084, 2011085, 2011086, 2011087, 2011088, 2011089, 2011091, 2011095, 2011097, 2011099, 2011100, 2011101, 2011102, 2011105, 2011108, 2011109, 2011110, 2011114, 2011116, 2011118, 2011119, 2011120, 2011122, 2011123, 2011124, 2011125, 2011128, 2011129, 2011130, 2011131, 2011132, 2011133, 2011134, 2011135, 2011137, 2011138, 2011140, 2011141, 2011142, 2011145, 2011147, 2011149, 2011152, 2011153, 2011155, 2011156, 2011157, 2011158, 2011161, 2011162, 2011163, 2011165, 2011166, 2011167, 2011168, 2011172, 2011174, 2011175, 2011176, 2011179, 2011180, 2011181, 2011182, 2011183, 2011184, 2011185, 2011186, 2011187, 2011188, 2011189, 2011191, 2011193, 2011194, 2011195, 2011196, 2011197, 2011200, 2011201, 2011203, 2011204, 2011207, 2011208, 2011209, 2011210, 2011212, 2011214, 2011215, 2011216, 2011218, 2011219, 2011220, 2011221, 2011224, 2011227, 2011228, 2011229, 2011232, 2011237, 2011238, 2011239, 2011241, 2011242, 2011243, 2004532, 2004173, 2007148, 2002113, 2011259, 2011260, 2011261, 2011265, 2011266, 2011268, 2011269, 2011271, 2011273, 2007281, 2009112, 2009115, 2009116, 2009119, 2009120, 2009123, 2009124, 2009126, 2009127, 2009129, 2009130, 2009131, 2009132, 2009133, 2009134, 2009135, 2009136, 2009137, 2009138, 2009139, 2009140, 2009142, 2009143, 2009144, 2009146, 2009147, 2009148, 2009149, 2009151, 2009153, 2009155, 2009156, 2009157, 2009158, 2009160, 2009161, 2009162, 2009164, 2009165, 2009166, 2009168, 2009169, 2009170, 2009171, 2009172, 2009173, 2009174, 2009175, 2006492, 2004546, 2006556, 2002424, 2006423, 2007268, 2009249, 2009250, 2004281, 2004329, 2006868, 2002752, 2002434, 2002854, 2007262, 2008920, 2006860, 2009251, 2009252, 2009254, 2009256, 2009257, 2009258, 2009259, 2009260, 2009262, 2009263, 2009265, 2009266, 2009267, 2009268, 2009269, 2009270, 2009271, 2009272, 2009273, 2009276, 2009277, 2009278, 2009280, 2009281, 2009283, 2009284, 2009285, 2009286, 2009287, 2009288, 2009289, 2009290, 2009291, 2009293, 2009294, 2009295, 2009296, 2009298, 2009301, 2009303, 2009305, 2009308, 2009309, 2009311, 2009312, 2009314, 2009315, 2009316, 2009319, 2009320, 2009322, 2009323, 2009324, 2009325, 2009326, 2009327, 2009329, 2009331, 2009332, 2009334, 2009335, 2009336, 2009337, 2009338, 2009341, 2009344, 2009346, 2009347, 2009348, 2009349, 2009350, 2009353, 2009354, 2009356, 2009357, 2009358, 2009359, 2009360, 2009361, 2009362, 2009363, 2009365, 2009366, 2009367, 2009368, 2009369, 2009374, 2009375, 2009376, 2009377, 2009378, 2009379, 2009380, 2009381, 2009383, 2009384, 2009385, 2009387, 2009389, 2009391, 2009392, 2009394, 2009395, 2009398, 2009399, 2009400, 2009402, 2009403, 2009404, 2009405, 2009406, 2009408, 2009410, 2009412, 2009414, 2009415, 2009416, 2009417, 2009418, 2009419, 2009420, 2009421, 2009424, 2009426, 2009427, 2009428, 2009899, 2009903, 2009917, 2010292, 2010296, 2010305, 2010309, 2010310, 2010316, 2010318, 2010325, 2010331, 2010336, 2011294, 2009292, 2009307, 2009345, 2009355, 2009364, 2009370, 2011299, 2011300, 2011305, 2011311, 2011320, 2011336, 2002501, 2008932, 2002413, 2002481, 2002482, 2002477, 2006813, 2002435, 2002732, 2002471, 2002442, 2005368, 2006791, 2002687, 2010219, 2002764, 2002499, 2009310, 2010635, 2009614, 2009629, 2009865, 2009866, 2009888, 2011367, 2011372, 2011381, 2011384, 2010611, 2011412, 2002436, 2002457, 2011462, 2011465, 2002613, 2010194, 2010213, 2007017, 2011478, 2011479, 2011480, 2004249, 2004479, 2004526, 2004586, 2004592, 2004272, 2004405, 2004534, 2002147, 2002162, 2006214, 2002136, 2002181, 2009689, 2005087, 2006236, 2002319, 2002306, 2002358, 2011600, 2011612, 2011616, 2011617, 2004404, 2004590, 2004617, 2004520, 2004288, 2007229, 2004480, 2004888, 2002342, 2002222, 2011115, 2006512, 2006527, 2006858, 2006269, 2002130, 2002167, 2006551, 2002141, 2005702, 2005624, 2010924, 2014557, 2014564, 2014565, 2014574, 2014577, 2014579, 2014581, 2014582, 2010987, 2010547, 2006635, 2011737, 2011739, 2011745, 2011747, 2011775, 2011782, 2011783, 2011787, 2011788, 2011857, 2011859, 2011900, 2011901, 2011902, 2011920, 2011948, 2002210, 2011570, 2011975, 2011976, 2011977, 2011978, 2011979, 2011980, 2011981, 2011982, 2011983, 2011984, 2011985, 2011986, 2011987, 2011988, 2011989, 2011990, 2011991, 2011992, 2010284, 2010864, 2002230, 2002235, 2006636, 2007111, 2011104, 2002152, 2011699, 2012018, 2012019, 2012020, 2012116, 2012225, 2012226, 2012227, 2012228, 2012229, 2012230, 2012232, 2012233, 2012235, 2012236, 2012237, 2012238, 2012240, 2012241, 2012242, 2012243, 2012245, 2012247, 2012248, 2012249, 2012250, 2012251, 2012252, 2012253, 2012254, 2012255, 2012256, 2012257, 2012258, 2012259, 2012260, 2012261, 2012262, 2012264, 2012265, 2012266, 2012267, 2012268, 2012269, 2012270, 2012272, 2012273, 2012276, 2012277, 2012280, 2012281, 2012282, 2012283, 2012285, 2012286, 2012287, 2012289, 2012290, 2012291, 2012293, 2012294, 2012295, 2012296, 2012297, 2012298, 2012299, 2012300, 2012301, 2012302, 2012303, 2012305, 2012322, 2012323, 2012324, 2012325, 2012326, 2012327, 2012328, 2012330, 2012331, 2012332, 2012333, 2012334, 2012335, 2012336, 2012339, 2012341, 2012342, 2012344, 2012345, 2012346, 2012348, 2012349, 2012350, 2012351, 2012352, 2012354, 2012355, 2012356, 2012357, 2012358, 2012359, 2012360, 2012361, 2012363, 2012364, 2012365, 2012366, 2012367, 2012369, 2012370, 2012371, 2012372, 2012373, 2012374, 2012375, 2012376, 2012378, 2012379, 2012380, 2012381, 2012382, 2012383, 2012385, 2012386, 2012388, 2012389, 2012390, 2012391, 2012392, 2012393, 2012395, 2012396, 2012397, 2012398, 2012399, 2012400, 2012402, 2012403, 2012404, 2012405, 2012406, 2012407, 2012408, 2012409, 2012410, 2012413, 2012415, 2012417, 2012418, 2012420, 2012421, 2012422, 2012424, 2012425, 2012426, 2012427, 2012428, 2012430, 2012431, 2012432, 2012433, 2012434, 2012435, 2012436, 2012437, 2012438, 2012439, 2012440, 2012441, 2012442, 2012443, 2012444, 2012446, 2012447, 2012449, 2012450, 2012451, 2012452, 2012453, 2012454, 2012455, 2012456, 2012457, 2012458, 2012459, 2012460, 2012461, 2012462, 2012463, 2012465, 2012466, 2012467, 2012468, 2012469, 2012470, 2012471, 2012472, 2012473, 2012474, 2012475, 2012476, 2012477, 2012478, 2012479, 2012480, 2012481, 2012482, 2012483, 2012484, 2012485, 2012486, 2012487, 2012488, 2012490, 2012491, 2012493, 2012494, 2012496, 2012497, 2012498, 2012499, 2012501, 2012502, 2012503, 2012504, 2012506, 2012507, 2012509, 2012511, 2012512, 2012514, 2012515, 2012516, 2012517, 2012519, 2012520, 2012521, 2012524, 2012525, 2012526, 2012527, 2012528, 2012529, 2012530, 2012531, 2012532, 2012534, 2012535, 2012536, 2012537, 2012539, 2012540, 2012544, 2012545, 2012546, 2012547, 2012548, 2012549, 2012550, 2012551, 2012552, 2012553, 2012554, 2012559, 2012560, 2012562, 2012563, 2012564, 2012565, 2012566, 2012567, 2012568, 2012569, 2012570, 2012571, 2012572, 2012573, 2012574, 2012576, 2012577, 2012578, 2012579, 2012580, 2012581, 2012582, 2012583, 2012584, 2012586, 2012587, 2012588, 2012589, 2012590, 2012591, 2012592, 2012593, 2012594, 2012595, 2012596, 2012597, 2012598, 2012600, 2012602, 2012603, 2012604, 2012605, 2012606, 2012607, 2012608, 2012610, 2012611, 2012612, 2012613, 2012614, 2012615, 2012618, 2012619, 2012620, 2012621, 2012622, 2012623, 2012624, 2012626, 2012627, 2012628, 2012629, 2012630, 2012631, 2012632, 2012633, 2012634, 2012635, 2012636, 2012637, 2012638, 2012639, 2012640, 2012641, 2012644, 2012646, 2012647, 2012648, 2012649, 2012650, 2012651, 2012652, 2012653, 2012654, 2012655, 2012656, 2012657, 2012658, 2012659, 2012660, 2012661, 2012662, 2012663, 2012664, 2012665, 2012767, 2012769, 2012772, 2012774, 2012775, 2012776, 2012777, 2012779, 2012780, 2012781, 2012783, 2012788, 2007295, 2012041, 2006627, 2007056, 2006563, 2010912, 2002331, 2010369, 2012042, 2013025, 2013100, 2013102, 2013103, 2002114, 2002211, 2011408, 2011410, 2009306, 2012645, 2012735, 2012995, 2013132, 2013202, 2013203, 2013204, 2013205, 2013206, 2013207, 2013208, 2013209, 2013210, 2013211, 2013212, 2013213, 2009274, 2009401, 2013214, 2013215, 2013216, 2009318, 2009790, 2011206, 2005911, 2010354, 2004908, 2013121, 2006157, 2008078, 2005923, 2010388, 2010439, 2005445, 2008217, 2009830, 2013291, 2005409, 2005454, 2005568, 2008296, 2008311, 2011407, 2011409, 2011411, 2005873, 2005136, 2005567, 2005472, 2010364, 2005450, 2008306, 2010344, 2004871, 2005174, 2010685, 2010828, 2010963, 2005298, 2005328, 2005565, 2005655, 2005715, 2005843, 2005916, 2004915, 2005401, 2005904, 2005805, 2005457, 2002724, 2008515, 2013447, 2013449, 2013452, 2013453, 2013455, 2013457, 2013458, 2013459, 2013461, 2013463, 2005395, 2005360, 2005389, 2005430, 2005557, 2005636, 2005770, 2004842, 2004883, 2005108, 2005591, 2005680, 2004839, 2004902, 2005871, 2006609, 2006854, 2005509, 2005608, 2005849, 2007139, 2005128, 2005364, 2005552, 2006465, 2005301, 2005302, 2005455, 2005521, 2005536, 2005583, 2002758, 2004941, 2010207, 2004917, 2004923, 2005296, 2005545, 2005919, 2010432, 2007251, 2004895, 2004912, 2005074, 2005089, 2005159, 2005310, 2005417, 2005501, 2005674, 2005748, 2005824, 2006903, 2004836, 2004876, 2005411, 2005480, 2005558, 2005726, 2005933, 2005323, 2005424, 2005453, 2005544, 2005550, 2005662, 2005692, 2005709, 2005716, 2005719, 2005872, 2006436, 2002717, 2006843, 2013552, 2004928, 2005175, 2005324, 2005517, 2005632, 2005681, 2005720, 2005814, 2005920, 2007158, 2004906, 2005528, 2005569, 2005710, 2005157, 2005329, 2005464, 2005516, 2005577, 2005686, 2005735, 2005799, 2005830, 2005846, 2004858, 2005601, 2005690, 2005717, 2005724, 2005783, 2005837, 2005921, 2004860, 2004882, 2004911, 2005111, 2005369, 2005414, 2005841, 2004903, 2004939, 2005350, 2005358, 2005485, 2005522, 2005562, 2005863, 2009729, 2013587, 2005099, 2005492, 2005612, 2005648, 2005693, 2010537, 2013589, 2013590, 2013596, 2013598, 2013604, 2013606, 2013607, 2013608, 2013609, 2013610, 2004935, 2005387, 2005426, 2005553, 2005666, 2005793, 2005797, 2005806, 2005388, 2005847, 2005859, 2005888, 2006920, 2006978, 2004889, 2004909, 2004919, 2005405, 2005561, 2005588, 2005639, 2005640, 2005801, 2005898, 2005531, 2004891, 2005365, 2005408, 2005538, 2005540, 2005723, 2005864, 2004845, 2004918, 2005646, 2005687, 2005171, 2005357, 2005698, 2005330, 2005463, 2005559, 2005622, 2005653, 2005431, 2005563, 2014001, 2002426, 2005416, 2005527, 2005684, 2005677, 2010433, 2010455, 2005400, 2005435, 2005767, 2005406, 2005850, 2009751, 2009765, 2010460, 2010619, 2010633, 2014214, 2014216, 2014217, 2014219, 2014220, 2014221, 2014222, 2014223, 2014224, 2014225, 2014226, 2014228, 2014229, 2014232, 2014233, 2014234, 2014235, 2005729, 2005816, 2005372, 2005532, 2007263, 2007328, 2005524, 2002209, 2002236, 2005519, 2005896, 2002326, 2002610, 2002692, 2002703, 2002712, 2002798, 2008559, 2014572, 2014612, 2014613, 2014614, 2014615, 2014616, 2014617, 2014618, 2014619, 2014621, 2014623, 2014624, 2014625, 2014626, 2014627, 2014630, 2014631, 2014633, 2014635, 2014636, 2014637, 2014638, 2014639, 2014640, 2014641, 2014644, 2014646, 2014648, 2014649, 2014650, 2014651, 2014653, 2014654, 2014656, 2014658, 2014661, 2014662, 2014663, 2014664, 2014665, 2014668, 2014669, 2014670, 2014671, 2014674, 2014675, 2014676, 2014677, 2014678, 2014681, 2014684, 20146
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Smiths Medical ASD Inc.
Reason for Recall:
Customers who utilize PharmGuard¿ Server Software (PGS) with Medfusion¿ 4000 pumps manufactured prior to November 10, 2017, may experience wireless network connectivity issues on certain pumps due to a potential Radio Module issue on some Medfusion¿ 4000 pumps.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Medfusion¿ Model 4000 Syringe Infusion Pump, with PharmGuard¿ Server Software (PGS). Sold under the following names: MEDFUSION¿ 4000 Pump V1.0; MEDFUSION¿ 4000 Pump V1.1, v1.1.1, v1.1.2; MEDFUSION¿ 4000 Pump V1.1, V1.1.1, V1.1.2; MEDFUSION¿ 4000 Pump V1.5, V1.5.1; MEDFUSION¿ 4000 Pump V1.5, V1.5.1; MEDFUSION¿ 4000 Pump V1.6, V1.6.1 Product Usage: The pumps are indicated for the following uses: 1. In the administration of fluids requiring precisely controlled infusion rates including blood or blood products, lipids, drugs, antibiotics, enteral solutions and other therapeutic fluids; 2. By the following delivery routes: arterial, epidural, intravenous, intrathecal, subcutaneous, and enteral; 3. By the following delivery modes: continuous, volume/time, mass, body weight, intermittent and bolus.

Product Codes/Lot Numbers:

Model Numbers (Serial Numbers): a. 4000-0100-50 (2000245, 2000246, 2000247, 2000248, 2000251, 2000252, 2000254, 2000255, 2000256, 2000257, 2000258, 2000260, 2000261, 2000264, 2000265, 2000267, 2000268, 2000269, 2000270, 2000272, 2000273, 2000274, 2000276,2000277, 2000278, 2000279, 2000280, 2000281, 2000282, 2000283, 2000284, 2000285, 2000286, 2000287, 2000288, 2000290, 2000291, 2000297, 2000298, 2000299, 2000300, 2000302, 2000304, 2000308, 2000309, 2000310, 2000311, 2000312, 2000313, 2000314, 2000315, 2000317, 2000318, 2000320, 2000321, 2000322, 2000324, 2000325, 2000326, 2000329, 2000331, 2000332, 2000336, 2000338, 2000341, 2000343, 2000345, 2000346, 2000348, 2000349, 2000352, 2000353, 2000355, 2000356, 2000358, 2000359, 2000360, 2000363, 2000364, 2000366, 2000370, 2000371, 2000374, 2000375, 2000376, 2000377, 2000379, 2000380, 2000381, 2000383, 2000385, 2000386, 2000387, 2000388, 2000389, 2000390, 2000392, 2000393, 2000395, 2000396, 2000403, 2000408, 2000409, 2000413, 2000415, 2000417, 2000426, 2000433, 2000434, 2000305, 2000330, 2000333, 2000357, 2000361, 2000362, 2000365, 2000369, 2000382, 2000391, 2000397, 2000398, 2000400, 2000402, 2000405, 2000414, 2000419, 2000423, 2000425, 2000430, 2000431, 2000439, 2000440, 2000441, 2000443, 2000316, 2000428, 2000438, 2000262, 2000342, 2000263, 2000327, 2000334, 2000378, 2000303, 2000351, 2000437, 2000301, 2000249, 2000275, 2000292, 2000373, 2000407, 2000394, 2000295, 2000335, 2000435, 2000422, 2000293, 2000418, 2000259, 2000367, 2000253, 2000416, 2000289, 2000344, 2000372, 2000350, 2000266, 2000250, 2000328, 2000401, 2000384, 2000404, 2000319, 2000399, 2000339, 2000412, 2000294, 2000340, 2000296, 2000429, 2000337, 2000368, 2000323, 2000307, 2000442); b. 4000-0101-51 (2002111, 2002121, 2002126, 2002127, 2002156, 2002171, 2002172, 2002221, 2002228, 2002241, 2002251, 2002253, 2002255, 2002263, 2002265, 2002272, 2002283, 2002291, 2002307, 2002318, 2002324, 2002336, 2002337, 2002339, 2002344, 2002353, 2002355, 2002356, 2002359, 2002195, 2002196, 2002197, 2002198, 2002199, 2002378, 2002379, 2004959, 2002407, 2002409, 2002410, 2002411, 2002412, 2002414, 2002415, 2002416, 2002417, 2002418, 2002419, 2002420, 2002421, 2002423, 2002427, 2002428, 2002429, 2002431, 2002432, 2002433, 2002437, 2002438, 2002439, 2002440, 2002441, 2002443, 2002445, 2002447, 2002448, 2002451, 2002452, 2002453, 2002454, 2002455, 2002456, 2002459, 2002460, 2002461, 2002462, 2002463, 2002464, 2002465, 2002466, 2002468, 2002469, 2002470, 2002472, 2002473, 2002474, 2002475, 2002478, 2002480, 2002483, 2002484, 2002485, 2002488, 2002489, 2002490, 2002491, 2002492, 2002494, 2002495, 2002496, 2002497, 2002498, 2002579, 2002600, 2002608, 2002616, 2002619, 2002630, 2002634, 2002643, 2002649, 2002705, 2002719, 2002722, 2002744, 2002771, 2002799, 2002811, 2002824, 2002826, 2002833, 2002835, 2002842, 2002844, 2002846, 2002857, 2004163, 2004164, 2004165, 2004167, 2004168, 2004169, 2004171, 2004172, 2004174, 2004176, 2004177, 2004178, 2004179, 2004180, 2004182, 2004183, 2004184, 2004185, 2004186, 2004187, 2004188, 2004189, 2004190, 2004191, 2004192, 2004193, 2004194, 2004195, 2004196, 2004197, 2004198, 2004199, 2004200, 2004201, 2004202, 2004204, 2004205, 2004206, 2004207, 2004208, 2004209, 2004210, 2004212, 2004215, 2004216, 2004217, 2004218, 2004219, 2004220, 2004221, 2004222, 2004225, 2004226, 2004227, 2004228, 2004229, 2004230, 2004231, 2004232, 2004233, 2004234, 2004235, 2004237, 2004238, 2004239, 2004240, 2004241, 2004242, 2004244, 2004245, 2004246, 2004247, 2004248, 2004250, 2004251, 2004252, 2004254, 2004255, 2004256, 2004258, 2004259, 2004260, 2004261, 2004262, 2004263, 2004265, 2004266, 2004267, 2004268, 2004269, 2004270, 2004273, 2004274, 2004275, 2004276, 2004277, 2004278, 2004279, 2004280, 2004282, 2004283, 2004284, 2004285, 2004286, 2004287, 2004289, 2004291, 2004292, 2004293, 2004296, 2004297, 2004298, 2004299, 2004300, 2004301, 2004303, 2004304, 2004305, 2004306, 2004307, 2004309, 2004310, 2004311, 2004312, 2004313, 2004315, 2004317, 2004319, 2004320, 2004322, 2004323, 2004324, 2004325, 2004326, 2004327, 2004328, 2004330, 2004331, 2004332, 2004333, 2004334, 2004335, 2004336, 2004337, 2004339, 2004340, 2004342, 2004343, 2004344, 2004345, 2004346, 2004347, 2004348, 2004349, 2004350, 2004351, 2004353, 2004354, 2004355, 2004356, 2004357, 2004358, 2004360, 2004361, 2004363, 2004364, 2004366, 2004367, 2004369, 2004370, 2004371, 2004372, 2004373, 2004374, 2004375, 2004376, 2004377, 2004378, 2004379, 2004380, 2004382, 2004383, 2004384, 2004385, 2004386, 2004387, 2004388, 2004389, 2004390, 2004391, 2004394, 2004395, 2004396, 2004397, 2004398, 2004399, 2004400, 2004403, 2004406, 2004407, 2004408, 2004409, 2004410, 2004412, 2004413, 2004414, 2004416, 2004417, 2004418, 2004420, 2004421, 2004422, 2004423, 2004424, 2004426, 2004428, 2004429, 2004430, 2004431, 2004432, 2004433, 2004434, 2004437, 2004438, 2004439, 2004440, 2004442, 2004443, 2004444, 2004445, 2004446, 2004447, 2004448, 2004449, 2004450, 2004452, 2004453, 2004454, 2004456, 2004457, 2004459, 2004461, 2004462, 2004464, 2004466, 2004467, 2004468, 2004469, 2004470, 2004472, 2004473, 2004474, 2004475, 2004476, 2004481, 2004482, 2004483, 2004484, 2004485, 2004486, 2004487, 2004488, 2004489, 2004491, 2004493, 2004494, 2004496, 2004499, 2004500, 2004501, 2004502, 2004515, 2004517, 2004518, 2004521, 2004523, 2004524, 2004525, 2004529, 2004535, 2004539, 2004543, 2004544, 2004545, 2004547, 2004549, 2004552, 2004554, 2004555, 2004556, 2004557, 2004558, 2004559, 2004560, 2004561, 2004562, 2004564, 2004566, 2004567, 2004568, 2004569, 2004570, 2004572, 2004573, 2004574, 2004575, 2004576, 2004577, 2004578, 2004579, 2004580, 2004582, 2004583, 2004584, 2004588, 2004589, 2004591, 2004593, 2004594, 2004595, 2004597, 2004598, 2004599, 2004600, 2004601, 2004602, 2004603, 2004605, 2004608, 2004610, 2004611, 2004612, 2004614, 2004615, 2004616, 2004618, 2004619, 2004620, 2004621, 2004622, 2004624, 2004625, 2004627, 2004119, 2008536, 2008575, 2004833, 2004843, 2004847, 2004854, 2004861, 2004865, 2004868, 2004877, 2004893, 2004894, 2004898, 2004907, 2004934, 2004938, 2005073, 2005077, 2005083, 2005085, 2005086, 2005092, 2005093, 2005097, 2005098, 2005100, 2005112, 2005113, 2005116, 2005118, 2005120, 2005130, 2005132, 2005135, 2005137, 2005138, 2005142, 2005145, 2005147, 2005149, 2005151, 2005153, 2005161, 2005163, 2005166, 2005167, 2005170, 2005173, 2005178, 2005183, 2005286, 2005287, 2005293, 2005294, 2005299, 2005303, 2005309, 2005311, 2005312, 2005314, 2005320, 2005321, 2005325, 2005326, 2005331, 2005333, 2005334, 2005338, 2005339, 2005341, 2005342, 2005349, 2005353, 2005361, 2005367, 2005370, 2005382, 2005386, 2005392, 2005403, 2005410, 2005413, 2005420, 2005428, 2005432, 2005434, 2005441, 2005443, 2005447, 2005458, 2005461, 2005462, 2005465, 2005469, 2005471, 2005475, 2005488, 2005489, 2005497, 2005498, 2005506, 2005515, 2005533, 2005535, 2005543, 2005554, 2005555, 2005580, 2005582, 2005585, 2005586, 2005587, 2005593, 2005596, 2005597, 2005599, 2005606, 2005614, 2005616, 2005620, 2005642, 2005644, 2005656, 2005658, 2005663, 2005664, 2005667, 2005673, 2005689, 2005697, 2005713, 2005725, 2005727, 2005741, 2005745, 2005747, 2005757, 2005762, 2005774, 2005775, 2005782, 2005788, 2005791, 2005804, 2005818, 2005822, 2005829, 2005834, 2005835, 2005842, 2005854, 2005855, 2005856, 2005861, 2005862, 2005869, 2005877, 2005879, 2005884, 2005885, 2005886, 2005890, 2005899, 2005900, 2005903, 2005907, 2005914, 2005915, 2005927, 2006134, 2006135, 2006136, 2006137, 2006138, 2006139, 2006140, 2006142, 2006143, 2006144, 2006145, 2006147, 2006149, 2006150, 2006151, 2006152, 2006153, 2006154, 2006156, 2006158, 2006159, 2006160, 2006161, 2006162, 2006163, 2006164, 2006165, 2006166, 2006167, 2006168, 2006169, 2006170, 2006171, 2006172, 2006173, 2006174, 2006175, 2006176, 2006177, 2006178, 2006179, 2006180, 2006181, 2006182, 2006183, 2006184, 2006185, 2006187, 2006188, 2006189, 2006191, 2006192, 2006193, 2006194, 2006195, 2006196, 2006197, 2006198, 2006199, 2006200, 2006201, 2006202, 2006203, 2006204, 2006205, 2006206, 2006207, 2006208, 2006209, 2006211, 2006212, 2006213, 2006215, 2006216, 2006217, 2006218, 2006219, 2006221, 2006224, 2006225, 2006226, 2006227, 2006228, 2006232, 2006233, 2006234, 2006235, 2006238, 2006239, 2006241, 2006243, 2006245, 2006246, 2006247, 2006248, 2006250, 2006251, 2006252, 2006253, 2006254, 2006255, 2006257, 2006258, 2006260, 2006262, 2006263, 2006265, 2006267, 2006268, 2006270, 2006272, 2006273, 2006274, 2006276, 2006277, 2006280, 2006281, 2006282, 2006283, 2006284, 2006285, 2006286, 2006287, 2006289, 2006290, 2006291, 2006294, 2006295, 2006297, 2006299, 2006300, 2006301, 2006302, 2006303, 2006304, 2006305, 2006307, 2006308, 2006309, 2006310, 2006311, 2006313, 2006314, 2006316, 2006317, 2006319, 2006366, 2007861, 2007899, 2007918, 2006392, 2006393, 2006395, 2006397, 2006399, 2006400, 2006401, 2006403, 2006404, 2006405, 2006406, 2006407, 2006408, 2006409, 2006412, 2006413, 2006414, 2006415, 2006416, 2006417, 2006418, 2006419, 2006420, 2006421, 2006422, 2006426, 2006428, 2006430, 2006431, 2006432, 2006433, 2006434, 2006435, 2006437, 2006439, 2006440, 2006441, 2006442, 2006443, 2006444, 2006445, 2006446, 2006447, 2006448, 2006451, 2006452, 2006453, 2006455, 2006457, 2006458, 2006459, 2006460, 2006461, 2006462, 2006463, 2006464, 2006466, 2006468, 2006472, 2006473, 2006475, 2006477, 2006484, 2006485, 2006488, 2006489, 2006490, 2006491, 2006494, 2006495, 2006496, 2006497, 2006498, 2006500, 2006501, 2006502, 2006503, 2006505, 2006506, 2006507, 2006508, 2006509, 2006511, 2006513, 2006514, 2006515, 2006516, 2006517, 2006518, 2006519, 2006521, 2006522, 2006523, 2006524, 2006526, 2006528, 2006529, 2006530, 2006531, 2006534, 2006536, 2006537, 2006539, 2006542, 2006543, 2006544, 2006546, 2006547, 2006548, 2006549, 2006552, 2006555, 2006558, 2006559, 2006560, 2006561, 2006562, 2006564, 2006566, 2006567, 2006568, 2006569, 2006570, 2006571, 2006576, 2006577, 2006578, 2006579, 2006580, 2006581, 2006584, 2006585, 2006586, 2006588, 2006589, 2006590, 2006591, 2006592, 2006593, 2006595, 2006597, 2006598, 2006599, 2006600, 2006601, 2006603, 2006604, 2006606, 2006607, 2006610, 2006611, 2006612, 2006613, 2006615, 2006616, 2006617, 2006618, 2006619, 2006620, 2006621, 2006623, 2006624, 2006626, 2006629, 2006630, 2006631, 2006633, 2006634, 2006640, 2006641, 2006642, 2006643, 2006644, 2006647, 2006648, 2006649, 2006651, 2006652, 2006653, 2006654, 2006656, 2006658, 2006659, 2006660, 2006661, 2006662, 2006663, 2006665, 2006666, 2006667, 2006669, 2006670, 2006671, 2006768, 2006769, 2006770, 2006772, 2006774, 2006775, 2006776, 2006777, 2006780, 2006781, 2006782, 2006784, 2006785, 2006787, 2006788, 2006790, 2006792, 2006793, 2006795, 2006798, 2006800, 2006801, 2006802, 2006803, 2006804, 2006805, 2006808, 2006809, 2006811, 2006812, 2006814, 2006815, 2006818, 2006820, 2006821, 2006822, 2006823, 2006826, 2006827, 2006828, 2006829, 2006830, 2006832, 2006833, 2006835, 2006836, 2006839, 2006840, 2006841, 2006842, 2006846, 2006847, 2006848, 2006849, 2006853, 2006855, 2006856, 2006857, 2006861, 2006862, 2006863, 2006864, 2006867, 2006869, 2006870, 2006871, 2006872, 2006873, 2006874, 2006875, 2006877, 2006878, 2006879, 2006880, 2006881, 2006882, 2006883, 2006885, 2006886, 2006887, 2006888, 2006890, 2006895, 2006896, 2006897, 2006900, 2006901, 2006902, 2006906, 2006907, 2006908, 2006909, 2006910, 2006916, 2006917, 2006926, 2006927, 2006928, 2006929, 2006930, 2006931, 2006932, 2006934, 2006935, 2006938, 2006939, 2006940, 2006942, 2006944, 2006946, 2006948, 2006949, 2006950, 2006951, 2006952, 2006954, 2006955, 2006956, 2006957, 2006958, 2006959, 2006960, 2006961, 2006963, 2006965, 2006968, 2006969, 2006970, 2006972, 2006975, 2006976, 2006977, 2006979, 2006981, 2006982, 2006983, 2006984, 2006985, 2006987, 2006988, 2006989, 2006991, 2006992, 2006994, 2006995, 2006996, 2006997, 2007001, 2007002, 2007007, 2007008, 2007009, 2007010, 2007011, 2007012, 2007015, 2007019, 2007020, 2007021, 2007022, 2007023, 2007024, 2007026, 2007027, 2007028, 2007029, 2007031, 2007033, 2007036, 2007038, 2007041, 2007042, 2007043, 2007044, 2007046, 2007049, 2007050, 2007054, 2007055, 2007059, 2007061, 2007062, 2007064, 2007067, 2007069, 2007073, 2007075, 2007076, 2007078, 2007079, 2007080, 2007081, 2007082, 2007083, 2007085, 2007087, 2007088, 2007089, 2007090, 2007091, 2007092, 2007093, 2007094, 2007097, 2007098, 2007099, 2007100, 2007101, 2007102, 2007103, 2007105, 2007107, 2007110, 2007112, 2007114, 2007118, 2007119, 2007121, 2007122, 2007127, 2007128, 2007130, 2007134, 2007136, 2007137, 2007138, 2007140, 2007141, 2007142, 2007143, 2007147, 2007149, 2007150, 2007151, 2007153, 2007154, 2007155, 2007159, 2007163, 2007164, 2007170, 2007171, 2007172, 2007173, 2007174, 2007175, 2007176, 2007178, 2007183, 2007184, 2007187, 2007188, 2007191, 2007194, 2007195, 2007196, 2007198, 2007199, 2007200, 2007202, 2007205, 2007206, 2007209, 2007211, 2007212, 2007213, 2007215, 2007216, 2007217, 2007218, 2007219, 2007220, 2007221, 2007223, 2007224, 2007225, 2007226, 2007227, 2007232, 2007233, 2007239, 2007240, 2007241, 2007242, 2007244, 2007246, 2007249, 2007250, 2007252, 2007253, 2007254, 2007256, 2007257, 2007259, 2007260, 2007261, 2007266, 2007269, 2007271, 2007274, 2007275, 2007277, 2007278, 2007279, 2007280, 2007282, 2007283, 2007284, 2007286, 2007287, 2007288, 2007291, 2007292, 2007294, 2007297, 2007298, 2007299, 2007301, 2007302, 2007310, 2007311, 2007314, 2007315, 2007316, 2007317, 2007319, 2007321, 2007323, 2007324, 2007325, 2007326, 2007329, 2007330, 2007333, 2007334, 2007335, 2007336, 2007337, 2007340, 2007341, 2007343, 2007344, 2007685, 2004774, 2007946, 2002220, 2007869, 2004811, 2007947, 2007949, 2007953, 2007956, 2007957, 2007958, 2008036, 2004812, 2004813, 2004816, 2004817, 2004819, 2004821, 2004822, 2004825, 2004826, 2004827, 2004828, 2004829, 2004830, 2004831, 2006646, 2007005, 2007988, 2007993, 2008052, 2008055, 2008056, 2008060, 2008061, 2008062, 2008063, 2008066, 2008067, 2008068, 2008069, 2008072, 2008074, 2008077, 2008079, 2008087, 2008089, 2008090, 2008091, 2008092, 2008094, 2008098, 2008099, 2008101, 2008102, 2008106, 2008107, 2008108, 2008110, 2008111, 2008112, 2008117, 2008120, 2008123, 2008125, 2008126, 2008132, 2008134, 2008135, 2008136, 2008137, 2008138, 2008139, 2008140, 2008141, 2008142, 2008144, 2008145, 2008147, 2008148, 2008149, 2008150, 2008151, 2008154, 2008156, 2008158, 2008159, 2008161, 2008162, 2008163, 2008164, 2008167, 2008168, 2008172, 2008174, 2008176, 2008177, 2008178, 2008181, 2008182, 2008184, 2008185, 2008186, 2008187, 2008189, 2008191, 2008192, 2008193, 2008196, 2008197, 2008198, 2008199, 2008200, 2008202, 2008204, 2008208, 2008210, 2008215, 2008216, 2008220, 2008228, 2008232, 2008236, 2008237, 2008238, 2008240, 2008241, 2008242, 2008243, 2008245, 2008246, 2008247, 2008248, 2008249, 2008250, 2008256, 2008258, 2008260, 2008261, 2008264, 2008265, 2008267, 2008268, 2008270, 2008273, 2008277, 2008279, 2008283, 2008284, 2008285, 2008286, 2008289, 2008290, 2008291, 2008292, 2008294, 2008299, 2008301, 2008304, 2008307, 2008308, 2008309, 2008313, 2008315, 2008316, 2008317, 2008318, 2008319, 2008323, 2008325, 2008326, 2008330, 2008331, 2008333, 2008335, 2008336, 2008338, 2008339, 2008341, 2008342, 2008346, 2008349, 2008352, 2008355, 2008359, 2008004, 2008008, 2008010, 2008360, 2008361, 2008362, 2002693, 2008363, 2008365, 2008031, 2008033, 2008375, 2008654, 2008719, 2008723, 2008817, 2008917, 2007276, 2002729, 2007160, 2009028, 2009042, 2002180, 2009087, 2006819, 2009090, 2009092, 2008378, 2008379, 2009507, 2009508, 2002408, 2009631, 2009634, 2009642, 2009643, 2009644, 2009646, 2009647, 2009648, 2009651, 2009653, 2009654, 2009656, 2009658, 2009659, 2009662, 2009663, 2009665, 2009666, 2009667, 2009671, 2009672, 2009673, 2009674, 2009676, 2009680, 2009681, 2009682, 2009687, 2009688, 2009701, 2009702, 2009704, 2009705, 2009706, 2009707, 2009710, 2009714, 2009720, 2009722, 2009728, 2009730, 2009731, 2009732, 2009735, 2009736, 2009737, 2009739, 2009742, 2009743, 2009746, 2009748, 2009755, 2009757, 2009759, 2009760, 2009764, 2009766, 2009767, 2009769, 2009773, 2009774, 2009775, 2009776, 2009778, 2009780, 2009781, 2009783, 2009784, 2009786, 2009787, 2009789, 2009794, 2009795, 2009796, 2009799, 2009802, 2009803, 2009810, 2009812, 2009819, 2009823, 2009824, 2009827, 2009828, 2009831, 2009833, 2009834, 2009836, 2009837, 2009842, 2009843, 2009845, 2009846, 2009848, 2009849, 2009853, 2009854, 2009855, 2002486, 2002422, 2002581, 2004531, 2002502, 2010054, 2002467, 2002580, 2010143, 2010153, 2010154, 2010156, 2010201, 2010205, 2010256, 2010259, 2010271, 2006852, 2010343, 2010345, 2010346, 2010347, 2010348, 2010349, 2010350, 2010352, 2010356, 2010357, 2010358, 2010359, 2010360, 2010363, 2010365, 2010366, 2010371, 2010373, 2010374, 2010375, 2010376, 2010377, 2010379, 2010380, 2010385, 2010387, 2010390, 2010392, 2010393, 2010394, 2010395, 2010396, 2008171, 2009552, 2009553, 2009554, 2009555, 2009558, 2009560, 2009562, 2009563, 2009565, 2009568, 2009569, 2009570, 2009571, 2009573, 2009574, 2009575, 2009577, 2009578, 2009579, 2009581, 2009582, 2009583, 2009584, 2009585, 2009586, 2009588, 2009589, 2009590, 2009591, 2009592, 2009593, 2009594, 2009595, 2009596, 2009597, 2009599, 2009600, 2009601, 2009602, 2009603, 2002762, 2007228, 2007065, 2010383, 2010398, 2010399, 2010400, 2010401, 2010403, 2010404, 2010405, 2010406, 2010407, 2010410, 2010411, 2010412, 2010414, 2010415, 2010416, 2010417, 2010418, 2010421, 2010423, 2010424, 2010425, 2010426, 2010427, 2010428, 2010429, 2010430, 2010431, 2010434, 2010435, 2010436, 2010437, 2010438, 2010440, 2010441, 2010443, 2010444, 2010445, 2010446, 2010447, 2010448, 2010449, 2010450, 2010451, 2010452, 2010457, 2010459, 2010463, 2010464, 2010465, 2010466, 2010467, 2010468, 2010487, 2010488, 2010490, 2010491, 2010492, 2010493, 2010494, 2010495, 2010496, 2010498, 2010499, 2010500, 2010501, 2010502, 2010503, 2010504, 2010505, 2010506, 2010507, 2010508, 2010509, 2010510, 2010511, 2010512, 2010513, 2010514, 2010515, 2010516, 2010517, 2010518, 2010519, 2010520, 2010522, 2010523, 2010524, 2010525, 2010526, 2010527, 2010528, 2010529, 2010531, 2010532, 2010533, 2010536, 2010539, 2010540, 2010541, 2010543, 2010544, 2010545, 2010546, 2010548, 2010550, 2010551, 2010552, 2010553, 2010554, 2010555, 2010557, 2010558, 2010559, 2010560, 2010561, 2010563, 2010564, 2010565, 2010566, 2010568, 2010569, 2010571, 2010572, 2010573, 2010574, 2010575, 2010576, 2010578, 2010579, 2010580, 2010581, 2010582, 2010583, 2010584, 2010585, 2010586, 2010587, 2010588, 2010589, 2010590, 2010593, 2010594, 2010595, 2010596, 2010597, 2010598, 2010599, 2010600, 2010601, 2010602, 2010604, 2010605, 2010606, 2010607, 2010608, 2010609, 2010612, 2010613, 2010614, 2010615, 2010616, 2010617, 2010620, 2010621, 2010622, 2010623, 2010624, 2010625, 2010626, 2010628, 2010629, 2010631, 2010632, 2010634, 2010636, 2010637, 2010639, 2010640, 2010641, 2010642, 2010643, 2010644, 2010646, 2010648, 2010649, 2010650, 2010651, 2010655, 2010656, 2010658, 2010659, 2010660, 2010662, 2010664, 2010665, 2010666, 2010667, 2010670, 2002479, 2010469, 2010470, 2010471, 2010472, 2010473, 2010474, 2010672, 2010675, 2010676, 2010677, 2010678, 2010679, 2010680, 2010681, 2010682, 2010684, 2010687, 2010692, 2010697, 2010698, 2010704, 2010705, 2010706, 2010707, 2010708, 2010710, 2010711, 2010712, 2010713, 2010714, 2010715, 2010716, 2010717, 2010718, 2010719, 2010722, 2010723, 2010724, 2010726, 2010727, 2010728, 2010729, 2010731, 2010732, 2010733, 2010735, 2010736, 2010737, 2010738, 2010739, 2010744, 2010747, 2010748, 2010749, 2010750, 2010751, 2010753, 2010754, 2010755, 2010756, 2010759, 2010760, 2010761, 2010764, 2010765, 2010767, 2010768, 2010770, 2010771, 2010772, 2010773, 2010775, 2010776, 2010777, 2010778, 2010780, 2010781, 2010783, 2010784, 2010787, 2010791, 2010792, 2010793, 2010794, 2010799, 2010800, 2010801, 2010803, 2010804, 2010806, 2010808, 2010810, 2010811, 2010813, 2010816, 2010818, 2010820, 2010821, 2010824, 2010827, 2010829, 2010830, 2010831, 2010835, 2010836, 2010838, 2010839, 2010840, 2010841, 2010842, 2010844, 2010847, 2010848, 2010849, 2010850, 2010853, 2010856, 2010857, 2010858, 2010859, 2010860, 2010861, 2010862, 2010863, 2010865, 2010866, 2010868, 2010869, 2010870, 2010873, 2010875, 2010876, 2010879, 2010880, 2010881, 2010883, 2010884, 2010885, 2010886, 2010887, 2010890, 2010891, 2010893, 2010896, 2010897, 2010898, 2010899, 2010901, 2010902, 2010903, 2010908, 2010909, 2010911, 2010914, 2010916, 2010917, 2010918, 2010919, 2010920, 2010921, 2010922, 2010925, 2010926, 2010928, 2010929, 2010930, 2010932, 2010933, 2010934, 2010936, 2010938, 2010940, 2010941, 2010944, 2010945, 2010946, 2010948, 2010949, 2010951, 2010952, 2010954, 2010955, 2010956, 2010957, 2010958, 2010959, 2010960, 2010961, 2010962, 2010964, 2010965, 2010967, 2010968, 2010969, 2010970, 2010971, 2010973, 2010974, 2010975, 2010976, 2010977, 2010978, 2010979, 2010980, 2010981, 2010982, 2010985, 2010986, 2010988, 2010990, 2010991, 2010992, 2010996, 2010997, 2010998, 2011000, 2011001, 2011002, 2011004, 2011005, 2011006, 2011008, 2011009, 2011010, 2011012, 2011013, 2011014, 2011016, 2011017, 2011018, 2011019, 2011020, 2011026, 2011028, 2011029, 2011031, 2011032, 2011034, 2011035, 2011037, 2011039, 2011042, 2011043, 2011044, 2011050, 2011051, 2011053, 2011070, 2011071, 2011074, 2011075, 2011076, 2011080, 2011081, 2011082, 2011083, 2011084, 2011085, 2011086, 2011087, 2011088, 2011089, 2011091, 2011095, 2011097, 2011099, 2011100, 2011101, 2011102, 2011105, 2011108, 2011109, 2011110, 2011114, 2011116, 2011118, 2011119, 2011120, 2011122, 2011123, 2011124, 2011125, 2011128, 2011129, 2011130, 2011131, 2011132, 2011133, 2011134, 2011135, 2011137, 2011138, 2011140, 2011141, 2011142, 2011145, 2011147, 2011149, 2011152, 2011153, 2011155, 2011156, 2011157, 2011158, 2011161, 2011162, 2011163, 2011165, 2011166, 2011167, 2011168, 2011172, 2011174, 2011175, 2011176, 2011179, 2011180, 2011181, 2011182, 2011183, 2011184, 2011185, 2011186, 2011187, 2011188, 2011189, 2011191, 2011193, 2011194, 2011195, 2011196, 2011197, 2011200, 2011201, 2011203, 2011204, 2011207, 2011208, 2011209, 2011210, 2011212, 2011214, 2011215, 2011216, 2011218, 2011219, 2011220, 2011221, 2011224, 2011227, 2011228, 2011229, 2011232, 2011237, 2011238, 2011239, 2011241, 2011242, 2011243, 2004532, 2004173, 2007148, 2002113, 2011259, 2011260, 2011261, 2011265, 2011266, 2011268, 2011269, 2011271, 2011273, 2007281, 2009112, 2009115, 2009116, 2009119, 2009120, 2009123, 2009124, 2009126, 2009127, 2009129, 2009130, 2009131, 2009132, 2009133, 2009134, 2009135, 2009136, 2009137, 2009138, 2009139, 2009140, 2009142, 2009143, 2009144, 2009146, 2009147, 2009148, 2009149, 2009151, 2009153, 2009155, 2009156, 2009157, 2009158, 2009160, 2009161, 2009162, 2009164, 2009165, 2009166, 2009168, 2009169, 2009170, 2009171, 2009172, 2009173, 2009174, 2009175, 2006492, 2004546, 2006556, 2002424, 2006423, 2007268, 2009249, 2009250, 2004281, 2004329, 2006868, 2002752, 2002434, 2002854, 2007262, 2008920, 2006860, 2009251, 2009252, 2009254, 2009256, 2009257, 2009258, 2009259, 2009260, 2009262, 2009263, 2009265, 2009266, 2009267, 2009268, 2009269, 2009270, 2009271, 2009272, 2009273, 2009276, 2009277, 2009278, 2009280, 2009281, 2009283, 2009284, 2009285, 2009286, 2009287, 2009288, 2009289, 2009290, 2009291, 2009293, 2009294, 2009295, 2009296, 2009298, 2009301, 2009303, 2009305, 2009308, 2009309, 2009311, 2009312, 2009314, 2009315, 2009316, 2009319, 2009320, 2009322, 2009323, 2009324, 2009325, 2009326, 2009327, 2009329, 2009331, 2009332, 2009334, 2009335, 2009336, 2009337, 2009338, 2009341, 2009344, 2009346, 2009347, 2009348, 2009349, 2009350, 2009353, 2009354, 2009356, 2009357, 2009358, 2009359, 2009360, 2009361, 2009362, 2009363, 2009365, 2009366, 2009367, 2009368, 2009369, 2009374, 2009375, 2009376, 2009377, 2009378, 2009379, 2009380, 2009381, 2009383, 2009384, 2009385, 2009387, 2009389, 2009391, 2009392, 2009394, 2009395, 2009398, 2009399, 2009400, 2009402, 2009403, 2009404, 2009405, 2009406, 2009408, 2009410, 2009412, 2009414, 2009415, 2009416, 2009417, 2009418, 2009419, 2009420, 2009421, 2009424, 2009426, 2009427, 2009428, 2009899, 2009903, 2009917, 2010292, 2010296, 2010305, 2010309, 2010310, 2010316, 2010318, 2010325, 2010331, 2010336, 2011294, 2009292, 2009307, 2009345, 2009355, 2009364, 2009370, 2011299, 2011300, 2011305, 2011311, 2011320, 2011336, 2002501, 2008932, 2002413, 2002481, 2002482, 2002477, 2006813, 2002435, 2002732, 2002471, 2002442, 2005368, 2006791, 2002687, 2010219, 2002764, 2002499, 2009310, 2010635, 2009614, 2009629, 2009865, 2009866, 2009888, 2011367, 2011372, 2011381, 2011384, 2010611, 2011412, 2002436, 2002457, 2011462, 2011465, 2002613, 2010194, 2010213, 2007017, 2011478, 2011479, 2011480, 2004249, 2004479, 2004526, 2004586, 2004592, 2004272, 2004405, 2004534, 2002147, 2002162, 2006214, 2002136, 2002181, 2009689, 2005087, 2006236, 2002319, 2002306, 2002358, 2011600, 2011612, 2011616, 2011617, 2004404, 2004590, 2004617, 2004520, 2004288, 2007229, 2004480, 2004888, 2002342, 2002222, 2011115, 2006512, 2006527, 2006858, 2006269, 2002130, 2002167, 2006551, 2002141, 2005702, 2005624, 2010924, 2014557, 2014564, 2014565, 2014574, 2014577, 2014579, 2014581, 2014582, 2010987, 2010547, 2006635, 2011737, 2011739, 2011745, 2011747, 2011775, 2011782, 2011783, 2011787, 2011788, 2011857, 2011859, 2011900, 2011901, 2011902, 2011920, 2011948, 2002210, 2011570, 2011975, 2011976, 2011977, 2011978, 2011979, 2011980, 2011981, 2011982, 2011983, 2011984, 2011985, 2011986, 2011987, 2011988, 2011989, 2011990, 2011991, 2011992, 2010284, 2010864, 2002230, 2002235, 2006636, 2007111, 2011104, 2002152, 2011699, 2012018, 2012019, 2012020, 2012116, 2012225, 2012226, 2012227, 2012228, 2012229, 2012230, 2012232, 2012233, 2012235, 2012236, 2012237, 2012238, 2012240, 2012241, 2012242, 2012243, 2012245, 2012247, 2012248, 2012249, 2012250, 2012251, 2012252, 2012253, 2012254, 2012255, 2012256, 2012257, 2012258, 2012259, 2012260, 2012261, 2012262, 2012264, 2012265, 2012266, 2012267, 2012268, 2012269, 2012270, 2012272, 2012273, 2012276, 2012277, 2012280, 2012281, 2012282, 2012283, 2012285, 2012286, 2012287, 2012289, 2012290, 2012291, 2012293, 2012294, 2012295, 2012296, 2012297, 2012298, 2012299, 2012300, 2012301, 2012302, 2012303, 2012305, 2012322, 2012323, 2012324, 2012325, 2012326, 2012327, 2012328, 2012330, 2012331, 2012332, 2012333, 2012334, 2012335, 2012336, 2012339, 2012341, 2012342, 2012344, 2012345, 2012346, 2012348, 2012349, 2012350, 2012351, 2012352, 2012354, 2012355, 2012356, 2012357, 2012358, 2012359, 2012360, 2012361, 2012363, 2012364, 2012365, 2012366, 2012367, 2012369, 2012370, 2012371, 2012372, 2012373, 2012374, 2012375, 2012376, 2012378, 2012379, 2012380, 2012381, 2012382, 2012383, 2012385, 2012386, 2012388, 2012389, 2012390, 2012391, 2012392, 2012393, 2012395, 2012396, 2012397, 2012398, 2012399, 2012400, 2012402, 2012403, 2012404, 2012405, 2012406, 2012407, 2012408, 2012409, 2012410, 2012413, 2012415, 2012417, 2012418, 2012420, 2012421, 2012422, 2012424, 2012425, 2012426, 2012427, 2012428, 2012430, 2012431, 2012432, 2012433, 2012434, 2012435, 2012436, 2012437, 2012438, 2012439, 2012440, 2012441, 2012442, 2012443, 2012444, 2012446, 2012447, 2012449, 2012450, 2012451, 2012452, 2012453, 2012454, 2012455, 2012456, 2012457, 2012458, 2012459, 2012460, 2012461, 2012462, 2012463, 2012465, 2012466, 2012467, 2012468, 2012469, 2012470, 2012471, 2012472, 2012473, 2012474, 2012475, 2012476, 2012477, 2012478, 2012479, 2012480, 2012481, 2012482, 2012483, 2012484, 2012485, 2012486, 2012487, 2012488, 2012490, 2012491, 2012493, 2012494, 2012496, 2012497, 2012498, 2012499, 2012501, 2012502, 2012503, 2012504, 2012506, 2012507, 2012509, 2012511, 2012512, 2012514, 2012515, 2012516, 2012517, 2012519, 2012520, 2012521, 2012524, 2012525, 2012526, 2012527, 2012528, 2012529, 2012530, 2012531, 2012532, 2012534, 2012535, 2012536, 2012537, 2012539, 2012540, 2012544, 2012545, 2012546, 2012547, 2012548, 2012549, 2012550, 2012551, 2012552, 2012553, 2012554, 2012559, 2012560, 2012562, 2012563, 2012564, 2012565, 2012566, 2012567, 2012568, 2012569, 2012570, 2012571, 2012572, 2012573, 2012574, 2012576, 2012577, 2012578, 2012579, 2012580, 2012581, 2012582, 2012583, 2012584, 2012586, 2012587, 2012588, 2012589, 2012590, 2012591, 2012592, 2012593, 2012594, 2012595, 2012596, 2012597, 2012598, 2012600, 2012602, 2012603, 2012604, 2012605, 2012606, 2012607, 2012608, 2012610, 2012611, 2012612, 2012613, 2012614, 2012615, 2012618, 2012619, 2012620, 2012621, 2012622, 2012623, 2012624, 2012626, 2012627, 2012628, 2012629, 2012630, 2012631, 2012632, 2012633, 2012634, 2012635, 2012636, 2012637, 2012638, 2012639, 2012640, 2012641, 2012644, 2012646, 2012647, 2012648, 2012649, 2012650, 2012651, 2012652, 2012653, 2012654, 2012655, 2012656, 2012657, 2012658, 2012659, 2012660, 2012661, 2012662, 2012663, 2012664, 2012665, 2012767, 2012769, 2012772, 2012774, 2012775, 2012776, 2012777, 2012779, 2012780, 2012781, 2012783, 2012788, 2007295, 2012041, 2006627, 2007056, 2006563, 2010912, 2002331, 2010369, 2012042, 2013025, 2013100, 2013102, 2013103, 2002114, 2002211, 2011408, 2011410, 2009306, 2012645, 2012735, 2012995, 2013132, 2013202, 2013203, 2013204, 2013205, 2013206, 2013207, 2013208, 2013209, 2013210, 2013211, 2013212, 2013213, 2009274, 2009401, 2013214, 2013215, 2013216, 2009318, 2009790, 2011206, 2005911, 2010354, 2004908, 2013121, 2006157, 2008078, 2005923, 2010388, 2010439, 2005445, 2008217, 2009830, 2013291, 2005409, 2005454, 2005568, 2008296, 2008311, 2011407, 2011409, 2011411, 2005873, 2005136, 2005567, 2005472, 2010364, 2005450, 2008306, 2010344, 2004871, 2005174, 2010685, 2010828, 2010963, 2005298, 2005328, 2005565, 2005655, 2005715, 2005843, 2005916, 2004915, 2005401, 2005904, 2005805, 2005457, 2002724, 2008515, 2013447, 2013449, 2013452, 2013453, 2013455, 2013457, 2013458, 2013459, 2013461, 2013463, 2005395, 2005360, 2005389, 2005430, 2005557, 2005636, 2005770, 2004842, 2004883, 2005108, 2005591, 2005680, 2004839, 2004902, 2005871, 2006609, 2006854, 2005509, 2005608, 2005849, 2007139, 2005128, 2005364, 2005552, 2006465, 2005301, 2005302, 2005455, 2005521, 2005536, 2005583, 2002758, 2004941, 2010207, 2004917, 2004923, 2005296, 2005545, 2005919, 2010432, 2007251, 2004895, 2004912, 2005074, 2005089, 2005159, 2005310, 2005417, 2005501, 2005674, 2005748, 2005824, 2006903, 2004836, 2004876, 2005411, 2005480, 2005558, 2005726, 2005933, 2005323, 2005424, 2005453, 2005544, 2005550, 2005662, 2005692, 2005709, 2005716, 2005719, 2005872, 2006436, 2002717, 2006843, 2013552, 2004928, 2005175, 2005324, 2005517, 2005632, 2005681, 2005720, 2005814, 2005920, 2007158, 2004906, 2005528, 2005569, 2005710, 2005157, 2005329, 2005464, 2005516, 2005577, 2005686, 2005735, 2005799, 2005830, 2005846, 2004858, 2005601, 2005690, 2005717, 2005724, 2005783, 2005837, 2005921, 2004860, 2004882, 2004911, 2005111, 2005369, 2005414, 2005841, 2004903, 2004939, 2005350, 2005358, 2005485, 2005522, 2005562, 2005863, 2009729, 2013587, 2005099, 2005492, 2005612, 2005648, 2005693, 2010537, 2013589, 2013590, 2013596, 2013598, 2013604, 2013606, 2013607, 2013608, 2013609, 2013610, 2004935, 2005387, 2005426, 2005553, 2005666, 2005793, 2005797, 2005806, 2005388, 2005847, 2005859, 2005888, 2006920, 2006978, 2004889, 2004909, 2004919, 2005405, 2005561, 2005588, 2005639, 2005640, 2005801, 2005898, 2005531, 2004891, 2005365, 2005408, 2005538, 2005540, 2005723, 2005864, 2004845, 2004918, 2005646, 2005687, 2005171, 2005357, 2005698, 2005330, 2005463, 2005559, 2005622, 2005653, 2005431, 2005563, 2014001, 2002426, 2005416, 2005527, 2005684, 2005677, 2010433, 2010455, 2005400, 2005435, 2005767, 2005406, 2005850, 2009751, 2009765, 2010460, 2010619, 2010633, 2014214, 2014216, 2014217, 2014219, 2014220, 2014221, 2014222, 2014223, 2014224, 2014225, 2014226, 2014228, 2014229, 2014232, 2014233, 2014234, 2014235, 2005729, 2005816, 2005372, 2005532, 2007263, 2007328, 2005524, 2002209, 2002236, 2005519, 2005896, 2002326, 2002610, 2002692, 2002703, 2002712, 2002798, 2008559, 2014572, 2014612, 2014613, 2014614, 2014615, 2014616, 2014617, 2014618, 2014619, 2014621, 2014623, 2014624, 2014625, 2014626, 2014627, 2014630, 2014631, 2014633, 2014635, 2014636, 2014637, 2014638, 2014639, 2014640, 2014641, 2014644, 2014646, 2014648, 2014649, 2014650, 2014651, 2014653, 2014654, 2014656, 2014658, 2014661, 2014662, 2014663, 2014664, 2014665, 2014668, 2014669, 2014670, 2014671, 2014674, 2014675, 2014676, 2014677, 2014678, 2014681, 2014684, 20146

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2554-2019

Related Recalls

CADD-Solis HPCA Ambulatory Infusion Pump; Model/Number: 21-2111-0402-09-KYO 21-2111-0402-09-KYO CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP 21-2111-0300-09_LOAN CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR CMSA 21-2111-0300-09_CMSA CADD Solis PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR FUKUOKA 21-2111-0300-09_FUK CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR KYOTO 21-2111-0300-09_KYO CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR OKAYAMA 21-2111-0300-09_OKA CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR OOMIYA 21-2111-0300-09_OMI CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR OSAKA 21-2111-0300-09_OSA CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR SAPPORO 21-2111-0300-09_SAP CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR TOKYO 21-2111-0300-09_TOK CADD_SOLIS_MODEL 2110_ENGLISH_1/EA 21-2111-0200-50JP CADD-Solis VIP CADD-SOLIS V4.2 LOAN FUK 21-2111-0402-09-FUK CADD-SOLIS V4.2 LOAN NAG 21-2111-0402-09-NAG CADD-SOLIS V4.2 LOAN OKA 21-2111-0402-09-OKA CADD-SOLIS V4.2 LOAN OSA 21-2111-0402-09-OSA CADD-SOLIS V4.2 LOAN SAI 21-2111-0402-09-SAI CADD-SOLIS V4.2 LOAN SAP 21-2111-0402-09-SAP CADD-SOLIS V4.2 LOAN TOKE 21-2111-0402-09-TOKE CADD-SOLIS V4.2 LOAN TOKW 21-2111-0402-09-TOKW CADD-SOLIS V4.2 LOAN YELLOW KYO 21-2112-0402-09-KYO CADD-SOLIS V4.2 LOAN YELLOW NAG 21-2112-0402-09-NAG CADD-SOLIS V4.2 LOAN YELLOW YOK ¿ 21-2112-0402-09-YOK CADD-SOLIS V4.2 LOAN YOK 21-2111-0402-09-YOK Number:21-2111-0403-02Part Type:List Number/Finished GoodLifecycle Phase:03-ReleasedDescription:PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, FRENCH, 1/EA 21-2111-0403-02 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, ENGLISH, 1/EA 21-2111-0403-51 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, TRIAL, LOANER, ENGLISH, 1/EA 21-2111-0403-78 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, TRIAL, LOANER, YELLOW, ENGLISH, 1/EA 21-2112-0403-78 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, YELLOW, ENGLISH, 1/EA 21-2112-0403-51 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, YELLOW, FRENCH, 1/EA 21-2112-0403-02 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110 TRIAL/LOANER, NA ENGLISH 1/EA 21-2111-0300-00 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, DANISH 1/EA ¿ 21-2111-0300-12 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, DUTCH 1/EA ¿ 21-2111-0300-06 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, FRENCH 1/EA ¿ 21-2111-0300-02 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, GERMAN 1/EA ¿ 21-2111-0300-03 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, ITALIAN 1/EA ¿ 21-2111-0300-07 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, JAPANESE 21-2111-0300-09JP PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NA ENGLISH 1/EA ¿ 21-2111-0300-01 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NFHU, FRENCH 1/EA ¿ 21-2111-0300-232 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NFHU, NA ENGLISH 1/EA ¿ 21-2111-0300-231 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NFHU, SWEDISH 1/EA 21-2111-0300-244 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, OUS ENGLISH 1/EA ¿ 21-2111-0300-50 Not PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, PORTUGUESE 1/EA ¿ 21-2111-0300-17 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, SPANISH 1/EA ¿ 21-2111-0300-08 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, SWEDISH 1/EA ¿ 21-2111-0300-14 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW KEYPAD, MODEL 2110, NA ENGLISH 1/EA ¿ 21-2112-0300-01 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, DUTCH 1/EA 21-2112-0300-06 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, FRENCH 1/EA ¿ 21-2112-0300-02 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, GERMAN 1/EA 21-2112-0300-03 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, OUS ENGLISH 1/EA 21-2112-0300-50 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110,TRIAL/LOANER, NA ENGLISH 1/EA 21-2112-0300-00 PUMP KIT, CADD-SOLIS HPCA, MDL 2100, DANISH 1/EA ¿ 21-2101-0200-12 PUMP KIT, CADD-SOLIS HPCA, MDL 2100, FRENCH 1/EA ¿ 21-2101-0200-02 PUMP KIT, CADD-SOLIS HPCA, MDL 2100, GERMAN 1/EA ¿ 21-2101-0200-03 PUMP KIT, CADD-SOLIS HPCA, MDL 2100, ITALIAN 1/EA ¿ 21-2101-02

Smiths Medical ASD

Class I - Dangerous

Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.

Apr 10, 2025 Infusion Pumps Nationwide View Details →

Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.

Apr 10, 2025 Infusion Pumps Nationwide View Details →

CADD Solis VIP Ambulatory Infusion Pump Models: 1) PUMP KIT, CADD-SOLIS VIP, MDL 2120, NA ENGLISH, MANUAL MODE, 1/EA, Model Number: 21-2127-0105-01; 2) PUMP KIT, CADD-SOLIS VIP, MDL 2120, NA ENGLISH, PHARMGUARD ENABLED, 1/EA, Model Number: 21-2120-0105-01; 3) PUMP KIT, CADD-SOLIS VIP, MDL 2120, NA ENGLISH 1/EA, Model Number: 21-2120-0103-01; 4) PUMP, CADD-SOLIS VIP, MDL 2120, ENGLISH 1/EA, Model Number: 21-2120-0102-51; 5) PUMP KIT, CADD-SOLIS VIP, MDL 2120, NA ENGLISH, PHARMGUARD ENABLED 1/EA, Model Number: 21-2120-0104-01; 6) PUMP KIT, CADD-SOLIS VIP, MDL 2120, CE GLOBAL ENGLISH 1/EA, Model Number: 21-2120-0103-51; 7) PUMP KIT, CADD-SOLIS VIP, MDL 2120, NA ENGLISH, MANUAL MODE 1/EA, Model Number: 21-2127-0104-01; 8) PUMP KIT, CADD-SOLIS VIP, MDL 2120, TRIAL/LOANER, NA ENGLISH, PHARMGUARD, 1/EA, Model Number: 21-2120-0105-00; 9) PUMP, CADD-SOLIS VIP, MDL 2120, TRIAL/LOANER CE GLOBAL ENGLISH 1/EA, Model Number: 21-2120-0103-78; 10) PUMP, CADD-SOLIS VIP, MDL 2120, ENGLISH, TRIAL/LOANER 1/EA, Model Number: 21-2120-0102-78; 11) PUMP, CADD-SOLIS VIP, MDL 2120, NFHU, NA ENGLISH, PHARMGUARD ENABLED 1/EA, Model Number: 21-2120-0104-231; 12) PUMP, CADD-SOLIS VIP, MDL 2120, ENGLISH, NFHU 1/EA, Model Number: 21-2120-0102-249; 13) PUMP, CADD-SOLIS VIP, MDL 2120, NFHU CE GLOBAL ENGLISH 1/EA, Model Number: 21-2120-0103-249; 14) PUMP KIT, CADD-SOLIS VIP, MDL 2120, NA ENGLISH, FACTORY LIBRARY 1/EA, Model Number: 21-2125-0104-01; 15) PUMP, CADD-SOLIS VIP, MDL 2120, ENGLISH, NFHU 1/EA, Model Number: 21-2120-0102-249.

Smiths Medical ASD

Class I - Dangerous

There is a potential for thermal damage in CADD-Solis and CADD- Solis VIP infusion pumps.

Apr 10, 2025 Infusion Pumps Nationwide View Details →