ReShape Integrated Dual Balloon System, Gen 1, Catalog No. RSM110, Model No. 01-0013-001 The ReShape Integrated Dual Balloon System is indicated for weight reduction when used in conjunction with diet and exercise, in obese patients with a Body Mass Index (BMI) of 30 40 kg/m2 and one or more obesity related comorbid conditions. It is indicated for use in adult patients who have failed weight reduction with diet and exercise alone.
Class I - DangerousWhat Should You Do?
- Check if you have this product: 160317-001, 160371-002, 160317-003, 160324-010.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Reshape Medical Inc
- Reason for Recall:
- Due to a suspected misalignment of the connection between the proximal balloon cap of the implant and the delivery catheter. The possible misalignment is the result of an assembly error during manufacturing and may lead to leakage of saline solution during the balloon inflation procedure.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
ReShape Integrated Dual Balloon System, Gen 1, Catalog No. RSM110, Model No. 01-0013-001 The ReShape Integrated Dual Balloon System is indicated for weight reduction when used in conjunction with diet and exercise, in obese patients with a Body Mass Index (BMI) of 30 40 kg/m2 and one or more obesity related comorbid conditions. It is indicated for use in adult patients who have failed weight reduction with diet and exercise alone.
Product Codes/Lot Numbers:
160317-001, 160371-002, 160317-003, 160324-010.
Distribution:
Distributed in: US, NY, OH, CA
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2552-2016
Related Recalls
ReShape Medical is recalling the Integrated Dual Balloon System Gen 1 due to the potential leak of distillation fluid during balloon inflation.