Trauma Tubes and Plates: TUBE & SCP PLT 90DX150LG TUBE & SCP PLT 90DX205LG TUBE & SCP PLT 90DX255LG TUBE & SCP PLT 95DX150LG TUBE & SCP PLT 95DX205LG TUBE & SCP PLT 95DX255LG TUBE&SCP PLT 130DX8H TUBE&SCP PLT 130DX10H TUBE&SCP PLT 130DX12H TUBE&SCP PLT 130DX14H TUBE&SCP PLT 135DX8H TUBE&SCP PLT 135DX10H TUBE&SCP PLT 135DX12H TUBE&SCP PLT 135DX14H TUBE&SCP PLT 140DX8H TUBE&SCP PLT 140DX10H TUBE&SCP PLT 140DX12H TUBE&SCP PLT 140DX14H TUBE&SCP PLT 145DX8H TUBE&SCP PLT 145DX10H TUBE&SCP PLT 145DX12H TUBE&SCP PLT 145DX14H TUBE&SCP PLT 150DX6H TUBE&SCP PLT 150DX8H TUBE&SCP PLT 150DX10H TUBE&SCP PLT 150DX12H TUBE&SCP PLT 150DX14H VERSA-FX II SC TUBE PLT 90DX6H VERSA-FX II SC TUBE PLT 90DX8H VERSA-FXII SC TUBE PLT 90DX10H VERSA-FXII SC TUBE PLT 90DX12H VERSA-FXII SC TUBE PLT 90DX14H VERSA-FXII SC TUBE PLT 95DX6H VERSA-FXII SC TUBE PLT 95DX8H VERSA-FXII SC TUBE PLT 95DX10H VERSA-FXII SC TUBE PLT 95DX12H VERSA-FXII SC TUBE PLT 95DX14H VERSA-FX KEYLESS 90D X 8HOLE VERSA-FX KEYLESS 90D X 11HOLE VERSA-FX KEYLESS 95D X 8HOLE VERSA-FX KEYLESS 95D X 11HOLE VERSA-FX KEYLESS 95D X 14HOLE VERSA-FX II STD TUBE 130DX8H VERSA-FX II STD TUBE 130DX10H VERSA-FX II STD TUBE 130DX12H VERSA-FX II STD TUBE 130DX14H VERSA-FX II STD TUBE 135DX8H VERSA-FX II STD TUBE 135DX10H VERSA-FX II STD TUBE 135DX12H VERSA-FX II STD TUBE 135DX14H VERSA-FX II STD TUBE 140DX8H VERSA-FX II STD TUBE 140DX10H VERSA-FX II STD TUBE 140DX12H VERSA-FX II STD TUBE 140DX14H VERSA-FX II STD TUBE 145DX8H VERSA-FX II STD TUBE 145DX10H VERSA-FX II STD TUBE 145DX12H VERSA-FX II STD TUBE 145DX14H VERSA-FX II STD TUBE 150DX8H VERSA-FX II STD TUBE 150DX10H VERSA-FX II STD TUBE 150DX12H VERSA-FX II STD TUBE 150DX14H VERSA-FX II KEYLESS TB 130DX8H VERSA-FX II KEYLESS TB 135DX8H VERSA-FX II KEYLESS TB 140DX8H VERSA-FX II KEYLESS TB 145DX8H VERSA-FX II KEYLESS TB 150DX8H VERSA-FX KEYLESS 90D X 8HOLE VERSA-FX KEYLESS 90D X 11HOLE VERSA-FX KEYLESS 95D X 8HOLE VERSA-FX KEYLESS 95D X 11HOLE VERSA-FX KEYLESS 95D X 14HOLE
Class I - DangerousWhat Should You Do?
- Check if you have this product: Item Numbers: 00118109006, 00118109008, 00118109010, 00118109506, 00118109508, 00118109510, 00118113008, 00118113010, 00118113012, 00118113014, 00118113508, 00118113510, 00118113512, 00118113514, 00118114008, 00118114010, 00118114012, 00118114014, 00118114508, 00118114510, 00118114512, 00118114514, 00118115006, 00118115008, 00118115010, 00118115012, 00118115014, 00119309006, 00119309008, 00119309010, 00119309012, 00119309014, 00119309506, 00119309508, 00119309510, 00119309512, 00119309514, 00119809008, 00119809011, 00119809508, 00119809511, 00119809514, 00119313008, 00119313010, 00119313012, 00119313014, 00119313508, 00119313510, 00119313512, 00119313514, 00119314008, 00119314010, 00119314012, 00119314014, 00119314508, 00119314510, 00119314512, 00119314514, 00119315008, 00119315010, 00119315012, 00119315014, 00119413008, 00119413508, 00119414008, 00119414508, 00119415008, 00119809008, 00119809011, 00119809508, 00119809511, 00119809514 EXPIRATION DATE PRIOR TO 2024-07
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Zimmer, Inc.
- Reason for Recall:
- Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Trauma Tubes and Plates: TUBE & SCP PLT 90DX150LG TUBE & SCP PLT 90DX205LG TUBE & SCP PLT 90DX255LG TUBE & SCP PLT 95DX150LG TUBE & SCP PLT 95DX205LG TUBE & SCP PLT 95DX255LG TUBE&SCP PLT 130DX8H TUBE&SCP PLT 130DX10H TUBE&SCP PLT 130DX12H TUBE&SCP PLT 130DX14H TUBE&SCP PLT 135DX8H TUBE&SCP PLT 135DX10H TUBE&SCP PLT 135DX12H TUBE&SCP PLT 135DX14H TUBE&SCP PLT 140DX8H TUBE&SCP PLT 140DX10H TUBE&SCP PLT 140DX12H TUBE&SCP PLT 140DX14H TUBE&SCP PLT 145DX8H TUBE&SCP PLT 145DX10H TUBE&SCP PLT 145DX12H TUBE&SCP PLT 145DX14H TUBE&SCP PLT 150DX6H TUBE&SCP PLT 150DX8H TUBE&SCP PLT 150DX10H TUBE&SCP PLT 150DX12H TUBE&SCP PLT 150DX14H VERSA-FX II SC TUBE PLT 90DX6H VERSA-FX II SC TUBE PLT 90DX8H VERSA-FXII SC TUBE PLT 90DX10H VERSA-FXII SC TUBE PLT 90DX12H VERSA-FXII SC TUBE PLT 90DX14H VERSA-FXII SC TUBE PLT 95DX6H VERSA-FXII SC TUBE PLT 95DX8H VERSA-FXII SC TUBE PLT 95DX10H VERSA-FXII SC TUBE PLT 95DX12H VERSA-FXII SC TUBE PLT 95DX14H VERSA-FX KEYLESS 90D X 8HOLE VERSA-FX KEYLESS 90D X 11HOLE VERSA-FX KEYLESS 95D X 8HOLE VERSA-FX KEYLESS 95D X 11HOLE VERSA-FX KEYLESS 95D X 14HOLE VERSA-FX II STD TUBE 130DX8H VERSA-FX II STD TUBE 130DX10H VERSA-FX II STD TUBE 130DX12H VERSA-FX II STD TUBE 130DX14H VERSA-FX II STD TUBE 135DX8H VERSA-FX II STD TUBE 135DX10H VERSA-FX II STD TUBE 135DX12H VERSA-FX II STD TUBE 135DX14H VERSA-FX II STD TUBE 140DX8H VERSA-FX II STD TUBE 140DX10H VERSA-FX II STD TUBE 140DX12H VERSA-FX II STD TUBE 140DX14H VERSA-FX II STD TUBE 145DX8H VERSA-FX II STD TUBE 145DX10H VERSA-FX II STD TUBE 145DX12H VERSA-FX II STD TUBE 145DX14H VERSA-FX II STD TUBE 150DX8H VERSA-FX II STD TUBE 150DX10H VERSA-FX II STD TUBE 150DX12H VERSA-FX II STD TUBE 150DX14H VERSA-FX II KEYLESS TB 130DX8H VERSA-FX II KEYLESS TB 135DX8H VERSA-FX II KEYLESS TB 140DX8H VERSA-FX II KEYLESS TB 145DX8H VERSA-FX II KEYLESS TB 150DX8H VERSA-FX KEYLESS 90D X 8HOLE VERSA-FX KEYLESS 90D X 11HOLE VERSA-FX KEYLESS 95D X 8HOLE VERSA-FX KEYLESS 95D X 11HOLE VERSA-FX KEYLESS 95D X 14HOLE
Product Codes/Lot Numbers:
Item Numbers: 00118109006, 00118109008, 00118109010, 00118109506, 00118109508, 00118109510, 00118113008, 00118113010, 00118113012, 00118113014, 00118113508, 00118113510, 00118113512, 00118113514, 00118114008, 00118114010, 00118114012, 00118114014, 00118114508, 00118114510, 00118114512, 00118114514, 00118115006, 00118115008, 00118115010, 00118115012, 00118115014, 00119309006, 00119309008, 00119309010, 00119309012, 00119309014, 00119309506, 00119309508, 00119309510, 00119309512, 00119309514, 00119809008, 00119809011, 00119809508, 00119809511, 00119809514, 00119313008, 00119313010, 00119313012, 00119313014, 00119313508, 00119313510, 00119313512, 00119313514, 00119314008, 00119314010, 00119314012, 00119314014, 00119314508, 00119314510, 00119314512, 00119314514, 00119315008, 00119315010, 00119315012, 00119315014, 00119413008, 00119413508, 00119414008, 00119414508, 00119415008, 00119809008, 00119809011, 00119809508, 00119809511, 00119809514 EXPIRATION DATE PRIOR TO 2024-07
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2550-2014
Related Recalls
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.