Zimmer Gender Solution Patello-Femoral Joint (PFJ) System Patello-Femoral Trochlea Component, Precoat, Size 4, Left. Knee prosthesis.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    00-5926-014-01; Lot 62455332
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

āš ļø Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Zimmer, Inc.
Reason for Recall:
Internal review found that five (5) PFJ implant forgings were identified as have forged laps during receiving inspection process for raw material. These five forgings were not segregated and were subsequently issued to final manufacturing under Part 00-5926-014-01; lot 62455332. A forged lap is a defect that forms when metal folds over itself during forging.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Zimmer Gender Solution Patello-Femoral Joint (PFJ) System Patello-Femoral Trochlea Component, Precoat, Size 4, Left. Knee prosthesis.

Product Codes/Lot Numbers:

00-5926-014-01; Lot 62455332

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2545-2014

Related Recalls

Zimmer Biomet is conducting a medical device recall for two (2) lots of the Dual Mobility Vivacit-EĀæ Bearing due to a commingle. The outer package is labeled as a Size F 44 mm, however, the implant inside the package is a Size G 46 mm, and vice versa.

Sep 26, 2025 Implants & Prosthetics Nationwide View Details →

Zimmer Biomet is conducting a medical device recall for two (2) lots of the Dual Mobility Vivacit-EĀæ Bearing due to a commingle. The outer package is labeled as a Size F 44 mm, however, the implant inside the package is a Size G 46 mm, and vice versa.

Sep 26, 2025 Implants & Prosthetics Nationwide View Details →