Access SARS CoV-2 IgG (1st IS) Assay: Access Immunoassay Systems REAGENT PACK, REF C74339 Access SARS-CoV-2 IgG (1st IS) Calibrators, REF C74340 Access SARS-CoV-2 (1st IS) Quality Control (QC), REF C74341

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Part Number: C74339 UDI Code: (01)15099590746919(17)210930(11)210402(10)124466 Lot Number: 124466; Part Number: C74340 UDI Codes: (01)15099590746926(17)210930(11)210402(10)124467; (01)15099590746926(17)211231(11)210703(10)124758; Lot Numbers: 124467; 124758; Part Number: C74341 UDI Codes: (01)15099590746933(17)210930(11)210402(10)124468; (01)15099590746933(17)211130(11)210602(10)124708; Lot Numbers: 124468; 124708;
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Beckman Coulter Inc.
Reason for Recall:
Firm's submission for Emergency Use Authorization (EUA) contained insufficient evidence to be granted EUA Authorization to distribute
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Access SARS CoV-2 IgG (1st IS) Assay: Access Immunoassay Systems REAGENT PACK, REF C74339 Access SARS-CoV-2 IgG (1st IS) Calibrators, REF C74340 Access SARS-CoV-2 (1st IS) Quality Control (QC), REF C74341

Product Codes/Lot Numbers:

Part Number: C74339 UDI Code: (01)15099590746919(17)210930(11)210402(10)124466 Lot Number: 124466; Part Number: C74340 UDI Codes: (01)15099590746926(17)210930(11)210402(10)124467; (01)15099590746926(17)211231(11)210703(10)124758; Lot Numbers: 124467; 124758; Part Number: C74341 UDI Codes: (01)15099590746933(17)210930(11)210402(10)124468; (01)15099590746933(17)211130(11)210602(10)124708; Lot Numbers: 124468; 124708;

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2508-2021

Related Recalls

Class I - Dangerous

Beckman Coulter has identified that Access Total T4 Calibrator lots (PN 33805) exhibit a negative bias on DxI 600/800 instruments, with slopes outside product specifications. A false low patient result could cause a patient to undergo unnecessary additional diagnostic testing. A patient may be treated for a disorder not present or receive delayed treatment for hyperthyroidism. For a pregnant patient being monitored for thyroid hormone supplementation, incorrect treatment or dose may be used. A delayed patient results could cause repeat blood draw with associated complications.

Mar 16, 2026 Other Medical Devices Nationwide View Details →

Due to specific lots of diluents contributing to elevated platelet Daily Checks background counts cycle with use with analyzers.

Nov 10, 2025 Other Medical Devices Nationwide View Details →