Philips Allura System configured with a 1 Phase UPS. Allura family (R1.x, R2.x). Labeled as the following with corresponding model numbers: 1. Allura Xper FD10, Model Number 722003 2. Allura Xper FD10/10, Model Number 722005 3. Allura Xper FD20, Model Number 722006 4. Allura Xper FD20 Biplane, Model Number 722008 5. Allura Xper FD10, Model Number 722010 6. Allura Xper FD10/10, Model Number 722011 7. Allura Xper FD20, Model Number 722012 8. Allura Xper FD20 Biplane, Model Number 722013 9. Allura Xper FD20 OR Table, Model Number 722023 10. Allura Xper FD10, Model Number 722026 11. Allura Xper FD10/10, Model Number 722027 12. Allura Xper FD20, Model Number 722028 13. Allura Xper FD20/10, Model Number 722029 14. Allura Xper FD20 OR Table, Model Number 722035 15. Allura Xper FD20/20, Model Number 722038 16. Allura Xper FD20/15, Model Number 722058 17. Allura Xper FD20/15 OR Table, Model Number 722059.
Class I - DangerousWhat Should You Do?
- Check if you have this product: 1. Model Number/System Code (UDI); Serial Numbers: 722003 (No UDI); 1225, 789, 1138, 445, 740, 152, 1411. 2. Model Number/System Code (UDI); Serial Numbers: 722005 (No UDI); 35. 3. Model Number/System Code (UDI); Serial Numbers: 722006 (No UDI); 168, 2198, 2184, 2606, 195, 161, 1956, 2256, 2070, 1502, 376, 1096, 1731. 4. Model Number/System Code (UDI); Serial Numbers: 722008 (No UDI); 246, 2470. 5. Model Number/System Code (UDI); Serial Numbers: 722010 (00884838059030); 911, 1180, 936, 338, 660, 411, 992, 1210, 873, 1028, 921. 6. Model Number/System Code (UDI); Serial Numbers: 722011 (00884838059047); 308. 7. Model Number/System Code (UDI); Serial Numbers: 722012 (00884838059054); 525, 823, 1141, 1078, 1811, 2092, 1641, 114, 2134, 1589, 1446, 1675, 1625, 1989, 2156, 455, 233, 2929, 2930, 749, 1080, 149, 854. 8. Model Number/System Code (UDI); Serial Numbers: 722013 (00884838059061); 580, 254, 78, 357, 428. 9. Model Number/System Code (UDI); Serial Numbers: 722023 (00884838059085); 164, 66. 10. Model Number/System Code (UDI); Serial Numbers: 722026 (00884838054189); 287, 825, 462, 192, 74, 191, 204, 155, 967, 1132, 205, 1246, 253, 156, 234, 233, 845, 122, 259, 185, 618, 531, 855, 235, 335, 1068, 843, 805, 646, 965, 258, 220, 602, 908, 221, 664, 109, 110, 116, 117, 115, 284, 389, 223, 694, 222, 436, 1192, 102, 914, 396, 704, 370, 201, 774, 696, 838, 344, 93, 1072, 601, 245, 1340, 604, 627, 190, 605, 935, 297, 98, 99, 100, 101, 544, 1129, 1268, 841, 1239, 1293, 648, 1038, 1231, 215, 683, 142, 1050, 224, 73, 608, 1051, 1052, 323, 324, 931, 265, 431, 654, 786, 617, 87, 990, 1101, 738, 1287, 75, 688, 151, 121, 78, 114, 79, 770, 909, 1000, 612, 319, 458, 381, 826, 716, 432, 244, 534, 807, 962, 986, 961, 638, 830, 759, 1115, 707, 271, 296, 257, 86, 926, 448, 536, 525, 682, 1218, 1273, 1156, 146, 343, 406, 722, 247, 970, 672, 824, 903, 1234, 1065, 673, 603, 888, 790, 758, 314, 304, 606, 263, 262, 457, 1222, 251, 441, 686, 1024, 1086, 992, 178, 1178, 435, 710, 592, 1245, 660, 345, 118, 1219, 333, 338, 360, 730, 551, 349, 727, 519, 133, 195, 196, 137, 528, 460, 529, 530, 126, 455, 542, 566, 978, 1201, 1202, 22, 277, 1317, 545, 723, 359, 846, 506, 197, 628, 1001, 119, 120, 857, 789, 481, 1138, 818, 734, 1137, 154, 313, 502. 11. Model Number/System Code (UDI); Serial Numbers: 722027 (00884838054196); 203, 123, 133, 284, 134, 222, 255, 106, 107, 234, 42, 179, 306, 307, 265, 267, 40, 332, 112, 305, 104, 231, 263, 342, 71, 351, 298, 116, 50, 29, 39, 49, 56, 152, 153, 202, 164, 160, 64, 99, 101, 254, 84, 297, 268, 77, 115, 114, 142, 30, 79, 38, 189, 300, 299, 247, 282, 70. 12. Model Number/System Code (UDI); Serial Numbers: 722028 (00884838054202); 596, 819, 1598, 316, 1946, 541, 565, 647, 674, 864, 710, 946, 949, 880, 1041, 447, 696, 166, 618, 313, 1214, 1846, 1978, 1649, 1256, 904, 242, 284, 1397, 1957, 207, 208, 526, 1677, 1756, 431, 2760, 601, 377, 1454, 1629, 275, 2933, 1340, 539, 356, 236, 237, 238, 422, 525, 350, 560, 1407, 272, 494, 80, 324, 282, 262, 1275, 406, 540, 1017, 840, 842, 136, 137, 793, 1130, 260, 636, 1446, 741, 768, 901, 1657, 1828, 1988, 638, 267, 654, 759, 1656, 1736, 1076, 929, 251, 989, 125, 211, 368, 499, 261, 1424, 1261, 1854, 924, 934, 1142, 1651, 774, 2174, 1463, 2846, 779, 1111, 519, 590, 453, 452, 680, 1074, 318, 11, 1191, 1212, 1211, 1269, 340, 481, 468, 317, 1774, 2502, 982, 339, 128, 655, 143, 1482, 1405, 140, 1448, 1413, 464, 2103, 139, 303, 476, 419, 521, 522, 1132, 580, 861, 1014, 256, 1266, 917, 1472, 2007, 2008, 1652, 1396, 1646, 1394, 254, 326, 1824, 343, 424, 435, 437, 543, 130, 987, 1187, 888, 273, 336, 375, 887, 533, 516, 449, 408, 1549, 376, 744, 1034, 420, 553, 110, 2261, 983, 627, 903, 1099, 1481, 363, 1690, 1233, 2206, 2310, 2506, 1331, 365, 203, 204, 1538, 625, 528, 310, 743, 445, 189, 292, 1466, 1200, 1374, 173, 161, 620, 1085, 361, 281, 475, 456, 1310, 315, 908, 672, 1522, 127, 132, 141, 145, 180, 1001, 463, 1484, 1835, 1244, 1022, 1004, 2173, 210, 1350, 1723, 1834, 2061, 2710, 591, 564, 407, 1751, 2284, 697, 364, 520, 2070, 582, 932, 930, 129, 218, 1483, 485, 176, 257, 1920, 962, 1809, 279, 459, 2196, 889, 556, 1149, 1095, 1870, 2253, 347, 909, 1498, 1650, 952, 160, 1569, 966, 306, 234, 213, 1202, 155, 222, 1122, 131, 133, 1855, 224, 490, 1570, 2249, 1869, 1447, 631, 637, 529, 493, 534, 491, 206, 1890, 205, 571, 197, 194, 195, 216, 217, 229, 227, 228, 305, 488, 1822, 1880, 209, 825, 1461, 2709, 2707, 579, 1634, 48, 1783, 498, 462, 1043, 142, 159, 280, 535, 2252, 158, 184, 185, 473, 489, 1826, 466, 671, 978, 402, 1499, 386, 1710, 1708, 1326, 1032, 454, 918, 1705, 914, 698, 1387, 666, 1995, 1889, 1983, 1568, 1980, 1845, 416, 394, 428, 1368, 426, 559, 684, 1333, 547, 550, 190, 973, 1068, 1069, 1630, 346, 1156, 632, 1567, 2407, 2422, 1428, 121, 333, 151, 372, 371, 911, 1633, 717, 1155, 265, 1343, 1439, 1600, 758, 942, 878, 1572, 1039, 763, 231, 663, 822, 1008, 430, 200, 1292, 1046, 728, 1088, 951, 440, 536, 1390, 1645, 827, 1378, 2372, 2373, 503, 2877, 2878, 2720, 853, 415, 1932, 9, 578, 2168, 635, 412, 575, 799, 766, 776, 434, 389, 380, 624, 119, 441, 1937, 105, 111, 687, 2076, 577, 954, 192, 1907, 810, 1128, 899, 1309, 472, 1112, 1307, 1308, 51, 767, 800, 826, 1021, 1235, 1329, 1462, 2314, 1135, 1134, 395, 1210, 2153, 29, 802, 1258, 398, 665, 328, 439, 1632, 322, 2112, 621, 226, 311, 1118, 611, 1120, 1604, 1602, 1601, 1603, 830, 1195, 352, 1268, 1125, 1879, 469, 775, 502, 2274, 113. 13. Model Number/System Code (UDI); Serial Numbers: 722029 (00884838054219); 91, 71, 110, 126, 21, 62, 88, 36, 100, 30, 63, 60, 94, 95, 56, 31, 107, 43. 14. Model Number/System Code (UDI); Serial Numbers: 722035(00884838054240); 85, 24, 101, 117, 21, 38, 177, 121, 183, 88, 57, 125, 119, 100, 165, 47, 52, 131, 124, 169, 46, 74, 30, 60, 59, 196, 216, 187, 34, 222, 72, 111, 128, 1, 96, 53, 160, 104, 191, 70, 154, 142, 134, 129, 65. 15. Model Number/System Code (UDI); Serial Numbers: 722038 (00884838054226); 39, 63, 42, 55, 78, 99, 60, 71, 115, 108, 67, 106, 118, 130, 49, 119, 126, 46, 37, 38, 48, 62, 35, 120, 41, 157, 96, 70, 137, 166, 34, 116, 72, 64. 16. Model Number/System Code (UDI); Serial Numbers: 722058 (00884838059115); 337, 22, 202, 71, 239, 186, 56, 24, 168, 129, 123, 259, 134, 38, 171, 264, 61, 99, 162, 65, 278, 277, 196, 23, 205, 212, 111, 49, 89, 150, 177, 60, 131, 377, 377, 206, 106, 223, 229, 288, 169, 125, 126, 320, 139, 213. 17. Model Number/System Code (UDI); Serial Numbers: 722059 (00884838059122); 7, 8.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Reason for Recall:
- Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Philips Allura System configured with a 1 Phase UPS. Allura family (R1.x, R2.x). Labeled as the following with corresponding model numbers: 1. Allura Xper FD10, Model Number 722003 2. Allura Xper FD10/10, Model Number 722005 3. Allura Xper FD20, Model Number 722006 4. Allura Xper FD20 Biplane, Model Number 722008 5. Allura Xper FD10, Model Number 722010 6. Allura Xper FD10/10, Model Number 722011 7. Allura Xper FD20, Model Number 722012 8. Allura Xper FD20 Biplane, Model Number 722013 9. Allura Xper FD20 OR Table, Model Number 722023 10. Allura Xper FD10, Model Number 722026 11. Allura Xper FD10/10, Model Number 722027 12. Allura Xper FD20, Model Number 722028 13. Allura Xper FD20/10, Model Number 722029 14. Allura Xper FD20 OR Table, Model Number 722035 15. Allura Xper FD20/20, Model Number 722038 16. Allura Xper FD20/15, Model Number 722058 17. Allura Xper FD20/15 OR Table, Model Number 722059.
Product Codes/Lot Numbers:
1. Model Number/System Code (UDI); Serial Numbers: 722003 (No UDI); 1225, 789, 1138, 445, 740, 152, 1411. 2. Model Number/System Code (UDI); Serial Numbers: 722005 (No UDI); 35. 3. Model Number/System Code (UDI); Serial Numbers: 722006 (No UDI); 168, 2198, 2184, 2606, 195, 161, 1956, 2256, 2070, 1502, 376, 1096, 1731. 4. Model Number/System Code (UDI); Serial Numbers: 722008 (No UDI); 246, 2470. 5. Model Number/System Code (UDI); Serial Numbers: 722010 (00884838059030); 911, 1180, 936, 338, 660, 411, 992, 1210, 873, 1028, 921. 6. Model Number/System Code (UDI); Serial Numbers: 722011 (00884838059047); 308. 7. Model Number/System Code (UDI); Serial Numbers: 722012 (00884838059054); 525, 823, 1141, 1078, 1811, 2092, 1641, 114, 2134, 1589, 1446, 1675, 1625, 1989, 2156, 455, 233, 2929, 2930, 749, 1080, 149, 854. 8. Model Number/System Code (UDI); Serial Numbers: 722013 (00884838059061); 580, 254, 78, 357, 428. 9. Model Number/System Code (UDI); Serial Numbers: 722023 (00884838059085); 164, 66. 10. Model Number/System Code (UDI); Serial Numbers: 722026 (00884838054189); 287, 825, 462, 192, 74, 191, 204, 155, 967, 1132, 205, 1246, 253, 156, 234, 233, 845, 122, 259, 185, 618, 531, 855, 235, 335, 1068, 843, 805, 646, 965, 258, 220, 602, 908, 221, 664, 109, 110, 116, 117, 115, 284, 389, 223, 694, 222, 436, 1192, 102, 914, 396, 704, 370, 201, 774, 696, 838, 344, 93, 1072, 601, 245, 1340, 604, 627, 190, 605, 935, 297, 98, 99, 100, 101, 544, 1129, 1268, 841, 1239, 1293, 648, 1038, 1231, 215, 683, 142, 1050, 224, 73, 608, 1051, 1052, 323, 324, 931, 265, 431, 654, 786, 617, 87, 990, 1101, 738, 1287, 75, 688, 151, 121, 78, 114, 79, 770, 909, 1000, 612, 319, 458, 381, 826, 716, 432, 244, 534, 807, 962, 986, 961, 638, 830, 759, 1115, 707, 271, 296, 257, 86, 926, 448, 536, 525, 682, 1218, 1273, 1156, 146, 343, 406, 722, 247, 970, 672, 824, 903, 1234, 1065, 673, 603, 888, 790, 758, 314, 304, 606, 263, 262, 457, 1222, 251, 441, 686, 1024, 1086, 992, 178, 1178, 435, 710, 592, 1245, 660, 345, 118, 1219, 333, 338, 360, 730, 551, 349, 727, 519, 133, 195, 196, 137, 528, 460, 529, 530, 126, 455, 542, 566, 978, 1201, 1202, 22, 277, 1317, 545, 723, 359, 846, 506, 197, 628, 1001, 119, 120, 857, 789, 481, 1138, 818, 734, 1137, 154, 313, 502. 11. Model Number/System Code (UDI); Serial Numbers: 722027 (00884838054196); 203, 123, 133, 284, 134, 222, 255, 106, 107, 234, 42, 179, 306, 307, 265, 267, 40, 332, 112, 305, 104, 231, 263, 342, 71, 351, 298, 116, 50, 29, 39, 49, 56, 152, 153, 202, 164, 160, 64, 99, 101, 254, 84, 297, 268, 77, 115, 114, 142, 30, 79, 38, 189, 300, 299, 247, 282, 70. 12. Model Number/System Code (UDI); Serial Numbers: 722028 (00884838054202); 596, 819, 1598, 316, 1946, 541, 565, 647, 674, 864, 710, 946, 949, 880, 1041, 447, 696, 166, 618, 313, 1214, 1846, 1978, 1649, 1256, 904, 242, 284, 1397, 1957, 207, 208, 526, 1677, 1756, 431, 2760, 601, 377, 1454, 1629, 275, 2933, 1340, 539, 356, 236, 237, 238, 422, 525, 350, 560, 1407, 272, 494, 80, 324, 282, 262, 1275, 406, 540, 1017, 840, 842, 136, 137, 793, 1130, 260, 636, 1446, 741, 768, 901, 1657, 1828, 1988, 638, 267, 654, 759, 1656, 1736, 1076, 929, 251, 989, 125, 211, 368, 499, 261, 1424, 1261, 1854, 924, 934, 1142, 1651, 774, 2174, 1463, 2846, 779, 1111, 519, 590, 453, 452, 680, 1074, 318, 11, 1191, 1212, 1211, 1269, 340, 481, 468, 317, 1774, 2502, 982, 339, 128, 655, 143, 1482, 1405, 140, 1448, 1413, 464, 2103, 139, 303, 476, 419, 521, 522, 1132, 580, 861, 1014, 256, 1266, 917, 1472, 2007, 2008, 1652, 1396, 1646, 1394, 254, 326, 1824, 343, 424, 435, 437, 543, 130, 987, 1187, 888, 273, 336, 375, 887, 533, 516, 449, 408, 1549, 376, 744, 1034, 420, 553, 110, 2261, 983, 627, 903, 1099, 1481, 363, 1690, 1233, 2206, 2310, 2506, 1331, 365, 203, 204, 1538, 625, 528, 310, 743, 445, 189, 292, 1466, 1200, 1374, 173, 161, 620, 1085, 361, 281, 475, 456, 1310, 315, 908, 672, 1522, 127, 132, 141, 145, 180, 1001, 463, 1484, 1835, 1244, 1022, 1004, 2173, 210, 1350, 1723, 1834, 2061, 2710, 591, 564, 407, 1751, 2284, 697, 364, 520, 2070, 582, 932, 930, 129, 218, 1483, 485, 176, 257, 1920, 962, 1809, 279, 459, 2196, 889, 556, 1149, 1095, 1870, 2253, 347, 909, 1498, 1650, 952, 160, 1569, 966, 306, 234, 213, 1202, 155, 222, 1122, 131, 133, 1855, 224, 490, 1570, 2249, 1869, 1447, 631, 637, 529, 493, 534, 491, 206, 1890, 205, 571, 197, 194, 195, 216, 217, 229, 227, 228, 305, 488, 1822, 1880, 209, 825, 1461, 2709, 2707, 579, 1634, 48, 1783, 498, 462, 1043, 142, 159, 280, 535, 2252, 158, 184, 185, 473, 489, 1826, 466, 671, 978, 402, 1499, 386, 1710, 1708, 1326, 1032, 454, 918, 1705, 914, 698, 1387, 666, 1995, 1889, 1983, 1568, 1980, 1845, 416, 394, 428, 1368, 426, 559, 684, 1333, 547, 550, 190, 973, 1068, 1069, 1630, 346, 1156, 632, 1567, 2407, 2422, 1428, 121, 333, 151, 372, 371, 911, 1633, 717, 1155, 265, 1343, 1439, 1600, 758, 942, 878, 1572, 1039, 763, 231, 663, 822, 1008, 430, 200, 1292, 1046, 728, 1088, 951, 440, 536, 1390, 1645, 827, 1378, 2372, 2373, 503, 2877, 2878, 2720, 853, 415, 1932, 9, 578, 2168, 635, 412, 575, 799, 766, 776, 434, 389, 380, 624, 119, 441, 1937, 105, 111, 687, 2076, 577, 954, 192, 1907, 810, 1128, 899, 1309, 472, 1112, 1307, 1308, 51, 767, 800, 826, 1021, 1235, 1329, 1462, 2314, 1135, 1134, 395, 1210, 2153, 29, 802, 1258, 398, 665, 328, 439, 1632, 322, 2112, 621, 226, 311, 1118, 611, 1120, 1604, 1602, 1601, 1603, 830, 1195, 352, 1268, 1125, 1879, 469, 775, 502, 2274, 113. 13. Model Number/System Code (UDI); Serial Numbers: 722029 (00884838054219); 91, 71, 110, 126, 21, 62, 88, 36, 100, 30, 63, 60, 94, 95, 56, 31, 107, 43. 14. Model Number/System Code (UDI); Serial Numbers: 722035(00884838054240); 85, 24, 101, 117, 21, 38, 177, 121, 183, 88, 57, 125, 119, 100, 165, 47, 52, 131, 124, 169, 46, 74, 30, 60, 59, 196, 216, 187, 34, 222, 72, 111, 128, 1, 96, 53, 160, 104, 191, 70, 154, 142, 134, 129, 65. 15. Model Number/System Code (UDI); Serial Numbers: 722038 (00884838054226); 39, 63, 42, 55, 78, 99, 60, 71, 115, 108, 67, 106, 118, 130, 49, 119, 126, 46, 37, 38, 48, 62, 35, 120, 41, 157, 96, 70, 137, 166, 34, 116, 72, 64. 16. Model Number/System Code (UDI); Serial Numbers: 722058 (00884838059115); 337, 22, 202, 71, 239, 186, 56, 24, 168, 129, 123, 259, 134, 38, 171, 264, 61, 99, 162, 65, 278, 277, 196, 23, 205, 212, 111, 49, 89, 150, 177, 60, 131, 377, 377, 206, 106, 223, 229, 288, 169, 125, 126, 320, 139, 213. 17. Model Number/System Code (UDI); Serial Numbers: 722059 (00884838059122); 7, 8.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2503-2025
Related Recalls
Philips Azurion 3M15, Model Numbers: 722064, 722222 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Philips Azurion 3M12, Model Numbers: 722063, 722221 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Philips Azurion 7B12/12, Model Numbers: 722067, 722225 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.