Irix-C Forceps Inserter; Product Code/UDI: T066-0059/ M697T06600591
Class I - DangerousWhat Should You Do?
- Check if you have this product: LOT: XS1920458A
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- XTANT Medical
- Reason for Recall:
- The strike plate fractured off from the forceps inserter devices with minimal impaction or force due to unsatisfactory weld.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Irix-C Forceps Inserter; Product Code/UDI: T066-0059/ M697T06600591
Product Codes/Lot Numbers:
LOT: XS1920458A
Distribution:
Distributed in: AZ, TX, WV
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2500-2020
Related Recalls
The proximal opening of the inserts exhibit an out of specification condition which would prevent the crossbar plates from appropriately engaging with the insert.
Cervical Implants contain a label that incorrectly identifies the grade of titanium used.
The firm received one customer complaint concerning three OsteoSelect Demineralized Bone Matrix (DBM) devices which contained 5.0cc of putty, instead of 10.00cc of putty as labeled.