DABRA Laser (RA-308 Excimer Laser). The DABRA Laser and DABRA Catheter is an excimer laser ultraviolet light source and delivery system, collectively the DABRA Laser System". The system is designed to deliver the energy to the surgical site to ablate unwanted material. The optical conduit is used in conjunction with existing endovascular accessories and tools to complete a surgical procedure. The software is programmed onto an EEPROM chip, which is a component of the DABRA Excimer Laser RA-308 system. Catheter for crossing total occlusions in vascular system.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Serial Number(s): RA00027 through RA00176;all DABRA excimer laser RA-308 product is impacted.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Ra Medical Systems, Inc.
Reason for Recall:
A software issue was identified which could result in user or patient injury, or may adversely impact laser performance. It is possible that an unintended release of laser light (radiation) may occur, injuring the user or patient.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

DABRA Laser (RA-308 Excimer Laser). The DABRA Laser and DABRA Catheter is an excimer laser ultraviolet light source and delivery system, collectively the DABRA Laser System". The system is designed to deliver the energy to the surgical site to ablate unwanted material. The optical conduit is used in conjunction with existing endovascular accessories and tools to complete a surgical procedure. The software is programmed onto an EEPROM chip, which is a component of the DABRA Excimer Laser RA-308 system. Catheter for crossing total occlusions in vascular system.

Product Codes/Lot Numbers:

Serial Number(s): RA00027 through RA00176;all DABRA excimer laser RA-308 product is impacted.

Distribution:

Nationwide distribution

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2497-2020

Related Recalls

Due to footswitch not meeting required specification for protection from ingress of solids or liquids. If spilling of liquids near or on the footswitch, it may fail in the 'on' position and allow the laser to continue to lase.

Oct 16, 2020 Infusion Pumps View Details →

The firm has become aware there is a potential problem with its DABRA catheters resulting in higher catheter calibration failure rate due to the product shelf life is currently less than the labeled shelf life.

Sep 19, 2019 Surgical Instruments Nationwide View Details →