AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) ANTERIOR HIP PACK, MODEL/ITEM NUMBER AMAH40N; 2)TOTAL KNEE DAN MYER, MODEL/ITEM NUMBER CCKM67F; 3)HIP PROTHESIS PACK, MODEL/ITEM NUMBER HSTH03N; 4)DTC ANTERIOR HIP PACK (PS 209459), MODEL/ITEM NUMBER MIAH65B; 5)DR WHIDD0N TOTAL HIP CPP, MODEL/ITEM NUMBER MPWH01B; 6)KNEE ARTHROSCOPY, MODEL/ITEM NUMBER SPKA57T; 7)ANTERIOR HIP PACK, MODEL/ITEM NUMBER WVAH40H.
Class I - DangerousWhat Should You Do?
- Check if you have this product: 1) MODEL/ITEM NUMBER AMAH40N: UDI/DI 00191072235473, Kit Lot Numbers: 42-7886011, 42-7990911; 2)MODEL/ITEM NUMBER CCKM67F: UDI/DI 00191072233578, Kit Lot Numbers: 70-053862, 70-053457, 70-053042, 70-052715, 70-052358, 70-051850, 70-051243; 3)MODEL/ITEM NUMBER HSTH03N: UDI/DI 00191072189615, Kit Lot Numbers: 42-7115711; 4)MODEL/ITEM NUMBER MIAH65B: UDI/DI 00191072235411, Kit Lot Numbers: 42-8134311; 5)MODEL/ITEM NUMBER MPWH01B: UDI/DI 00191072233271, Kit Lot Numbers: 5436801; 6)MODEL/ITEM NUMBER SPKA57T: UDI/DI 00191072235848, Kit Lot Numbers: 12-7400011, 12-7381311, 12-7358211, 12-7082111, 12-7157111, 12-7309911; 7)MODEL/ITEM NUMBER WVAH40H: UDI/DI 00191072233004, Kit Lot Numbers: 42-7897411, 42-7966211.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- American Contract Systems Inc
- Reason for Recall:
- During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) ANTERIOR HIP PACK, MODEL/ITEM NUMBER AMAH40N; 2)TOTAL KNEE DAN MYER, MODEL/ITEM NUMBER CCKM67F; 3)HIP PROTHESIS PACK, MODEL/ITEM NUMBER HSTH03N; 4)DTC ANTERIOR HIP PACK (PS 209459), MODEL/ITEM NUMBER MIAH65B; 5)DR WHIDD0N TOTAL HIP CPP, MODEL/ITEM NUMBER MPWH01B; 6)KNEE ARTHROSCOPY, MODEL/ITEM NUMBER SPKA57T; 7)ANTERIOR HIP PACK, MODEL/ITEM NUMBER WVAH40H.
Product Codes/Lot Numbers:
1) MODEL/ITEM NUMBER AMAH40N: UDI/DI 00191072235473, Kit Lot Numbers: 42-7886011, 42-7990911; 2)MODEL/ITEM NUMBER CCKM67F: UDI/DI 00191072233578, Kit Lot Numbers: 70-053862, 70-053457, 70-053042, 70-052715, 70-052358, 70-051850, 70-051243; 3)MODEL/ITEM NUMBER HSTH03N: UDI/DI 00191072189615, Kit Lot Numbers: 42-7115711; 4)MODEL/ITEM NUMBER MIAH65B: UDI/DI 00191072235411, Kit Lot Numbers: 42-8134311; 5)MODEL/ITEM NUMBER MPWH01B: UDI/DI 00191072233271, Kit Lot Numbers: 5436801; 6)MODEL/ITEM NUMBER SPKA57T: UDI/DI 00191072235848, Kit Lot Numbers: 12-7400011, 12-7381311, 12-7358211, 12-7082111, 12-7157111, 12-7309911; 7)MODEL/ITEM NUMBER WVAH40H: UDI/DI 00191072233004, Kit Lot Numbers: 42-7897411, 42-7966211.
Distribution:
Distributed in: US, IL, LA
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2472-2025
Related Recalls
Medical convenience kits Item Number/Description HSAN63N ANGIO PACK IHAN02W ANGIOGRAPHY PACK MHCA98AK CORONARY ANGIO PACK- 205991 RCAN45C ANGIO CUSTOM RCH UDCO10F CVL CORONARY PACK 219224
American Contract Systems
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Medical convenience kits Item Number/Description BUCY78E CYSTO PACK CMCP10X CYSTO PACK - 205972 IHCY10Y CYSTO PACK MHCY35AE CYSTO PACK- 212487 PLBY52E CYSTO PACK UDCY31Z CYSTO PACK - 206047 UICY28L MOR CYSTO PACK
American Contract Systems
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.