t:slim G4 Insulin Pump with Dexcom G4 Platinum CGM Software version: 4.3.4.3, Firmware version: 004722, Firmware, Fuel Gauge, Binary Rev B.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Serial Numbers: 483731, 483930, 483992, 484013, 492055, 492368, 501157, 501177, 501178, 501187, 501191, 501200, 501202, 501205, 501207, 501268, 501285, 501364, 501368, 501370, 501376, 501385, 501386, 501387, 501432, 501435, 501440, 501443, 501444, 501445, 501466, 501479, 501488, 501490, 501491, 501495, 501503, 501508, 501509, 501519, 501522, 501523, 501526, 501535, 501560, 504522, 504531, 504535, 504540, 504753, 504972, 504990, 505025, 90472671, 90488584
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Tandem Diabetes Care Inc
- Reason for Recall:
- The fuel gauge, the component that reads and reports the battery parameters to the pump, could provide inaccurate readings, which present to the user in one of two ways: A) by triggering a Malfunction 4 Alarm, or B) by triggering a succession of notifications prior to the pump shutting off, including Alert 2 (Battery Low), Alert 3 (Battery Very Low) and Alarm 12 (Battery Very Low). The alarms in both scenarios notify the user that the pump has stopped delivering insulin.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
t:slim G4 Insulin Pump with Dexcom G4 Platinum CGM Software version: 4.3.4.3, Firmware version: 004722, Firmware, Fuel Gauge, Binary Rev B.
Product Codes/Lot Numbers:
Serial Numbers: 483731, 483930, 483992, 484013, 492055, 492368, 501157, 501177, 501178, 501187, 501191, 501200, 501202, 501205, 501207, 501268, 501285, 501364, 501368, 501370, 501376, 501385, 501386, 501387, 501432, 501435, 501440, 501443, 501444, 501445, 501466, 501479, 501488, 501490, 501491, 501495, 501503, 501508, 501509, 501519, 501522, 501523, 501526, 501535, 501560, 504522, 504531, 504535, 504540, 504753, 504972, 504990, 505025, 90472671, 90488584
Distribution:
Nationwide distribution
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2471-2018
Related Recalls
Insulin pump includes a vibration motor that gives tactile feedback for any alerts, alarms, or malfunctions. Pump may exhibit false vibration motor failure due to a software issue causing Malfunction 12: "Pump cannot operate, the mobile app can no longer receive data from the pump. Insulin delivery and any active CGM Sessions have been stopped", which could result in hyperglycemia.
A software defect in versions 7.9.0.1 and 7.10.1 of the pump software for Tandem t:slim X2 and version 7.9.0.1 of the Tandem Mobi pump with Control-IQ+ technology that are paired with a G7 sensor may result in an unexpected automatic insulin correction bolus (autobolus ) which could result in hypoglycemia.
t:slim X2 Insulin Pump with Interoperable Technology (with Tandem t:slim mobile app)
Tandem Diabetes Care
an app defect that occurs when the phone is set to a right-to-left language, which causes app-pump pairing issues and graphical defects, which can potentially lead to incorrect therapy decisions and subsequent hypoglycemia or hyperglycemia.