ABL80-FLEX CO-OX 393-841 Box Label, REF 393-841, IVD, CE, UDI: 05700693938417 - Product Usage: It is intended for use in a laboratory environment, near patient or point-of-care setting.
Class I - DangerousWhat Should You Do?
- Check if you have this product: All lots
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Radiometer Medical ApS
- Reason for Recall:
- Barcode readers on ALB900 and ABL800 analyzers used for measuring pH, blood gases, electrolytes, glucose, lactate, hematocrit and oximetry may misinterpret the contents of some barcode types not using a check digit for patient ID or accession number. Misinterpretation of a barcode, including e.g. patient ID or accession number, may result in patient mix up and/or delayed medical treatment.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
ABL80-FLEX CO-OX 393-841 Box Label, REF 393-841, IVD, CE, UDI: 05700693938417 - Product Usage: It is intended for use in a laboratory environment, near patient or point-of-care setting.
Product Codes/Lot Numbers:
All lots
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2468-2020
Related Recalls
ABL90 FLEX and ABL90 FLEX PLUS analyzers running with Radiometer Windows 10 versions RWIN10 1.0 and RWIN10 1.1 may unexpectedly freeze (the analyzer ceases to operate) during a sample measurement. The analyzer software includes a feature (watchdog) that detects a freeze and automatically restarts the software to continue operating the analyzer. However, even if a restart occurs, any sample being processed at that time will be lost, potential to lead to delay in test results.
Issue with analyzer when the pH of the calibration solution decreases during the in-use period potentially because of bacterial growth in the calibration solution bottles. This may result in a probability of reporting biased out-of-specification pH results on blood samples.
Due to potential software issue that may result in patient mix-up information.