AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) LAP CHOLE PACK, Model/Item Number CILC69L; 2) LAP CHOLE PACK, Model/Item Number HNLC80J; 3) LAP CHOLE PACK, Model/Item Number JKLC55H; 4) LAPAROSCOPY PACK, Model/Item Number LLLA21K; 5) GENERAL LAPAROSCOPY MPH, Model/Item Number MPGL15G;
Class I - DangerousWhat Should You Do?
- Check if you have this product: 1) Model/Item Number CILC69L: UDI/DI 00191072235503, Kit Lot Numbers: 42-7937111, 42-8049311; 2) Model/Item Number HNLC80J: UDI/DI 00191072193650, Kit Lot Numbers: 42-8044511, 42-8118911, 42-7977011, 42-7803211; 3) Model/Item Number JKLC55H: UDI/DI 00191072235138, Kit Lot Numbers: 42-7814311; 4) Model/Item Number LLLA21K: UDI/DI 00191072233486, Kit Lot Numbers: 70-052582, 70-053048, 70-051564, 70-051373, 70-051163, 70-053446, 70-053776; 5) Model/Item Number MPGL15G: UDI/DI 00191072223197, Kit Lot Numbers: 5436916, 5436615
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- American Contract Systems Inc
- Reason for Recall:
- During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) LAP CHOLE PACK, Model/Item Number CILC69L; 2) LAP CHOLE PACK, Model/Item Number HNLC80J; 3) LAP CHOLE PACK, Model/Item Number JKLC55H; 4) LAPAROSCOPY PACK, Model/Item Number LLLA21K; 5) GENERAL LAPAROSCOPY MPH, Model/Item Number MPGL15G;
Product Codes/Lot Numbers:
1) Model/Item Number CILC69L: UDI/DI 00191072235503, Kit Lot Numbers: 42-7937111, 42-8049311; 2) Model/Item Number HNLC80J: UDI/DI 00191072193650, Kit Lot Numbers: 42-8044511, 42-8118911, 42-7977011, 42-7803211; 3) Model/Item Number JKLC55H: UDI/DI 00191072235138, Kit Lot Numbers: 42-7814311; 4) Model/Item Number LLLA21K: UDI/DI 00191072233486, Kit Lot Numbers: 70-052582, 70-053048, 70-051564, 70-051373, 70-051163, 70-053446, 70-053776; 5) Model/Item Number MPGL15G: UDI/DI 00191072223197, Kit Lot Numbers: 5436916, 5436615
Distribution:
Distributed in: US, IL, LA
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2465-2025
Related Recalls
Convenience kits containing Medline Namic Angiographic Control Syringes with Rotating Adaptor CORONARY ANGIO PACK Model numbers ANCA80AP ANCA80AQ
American Contract Systems
Impacted kits contain syringes recalled by Medline Industries. Syringe rotating adaptor may unwind during use, which may result in a loose connection or full disconnection between the syringe and manifold. Potential risk for biohazard exposure, blood loss, infection, or air embolism.
Medical convenience kits Item Number/Description HSAN63N ANGIO PACK IHAN02W ANGIOGRAPHY PACK MHCA98AK CORONARY ANGIO PACK- 205991 RCAN45C ANGIO CUSTOM RCH UDCO10F CVL CORONARY PACK 219224
American Contract Systems
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.