Tiger Cannulated Tray Lid Part # 210-00-001-1 NON-STERILE; Combined Tiger/Headless Screw Part # 210-99-101 NON-STERILE; These are components of the Tiger Cannulated Screw System and the Tiger Headless Cannulated Screw System The Tiger Cannulated Screw Fixation System and the Tiger Headless Cannulated Screw Fixation System implants are intended for fixation of fractures, non-unions, arthrodesis and osteotomies of the small bones in the hand and foot. The implants and guide wires are intended for single use only.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot Numbers ALU120514552473 and 45479
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Trilliant Surgical Ltd.
- Reason for Recall:
- Sterilization tray lids for the Tiger and Tiger Headless Cannulated Screw Systems were reported to occasionally be difficult to remove from the sterilization tray.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Tiger Cannulated Tray Lid Part # 210-00-001-1 NON-STERILE; Combined Tiger/Headless Screw Part # 210-99-101 NON-STERILE; These are components of the Tiger Cannulated Screw System and the Tiger Headless Cannulated Screw System The Tiger Cannulated Screw Fixation System and the Tiger Headless Cannulated Screw Fixation System implants are intended for fixation of fractures, non-unions, arthrodesis and osteotomies of the small bones in the hand and foot. The implants and guide wires are intended for single use only.
Product Codes/Lot Numbers:
Lot Numbers ALU120514552473 and 45479
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2464-2015
Related Recalls
3.5mm Aresenal Screw Drill Bit , Ref: 330-35-001, Qty. 70, Rx Only, Non-Sterile, UDI: (01)00812926029696 - Product Usage: A drill bit facilitates implant insertion
Trilliant Surgical
Labeling Error; specifically, drill bits from lots # TSL008495A are marked with the lot # of TSL008495. This could lead to parts being misidentified and impact device traceability.
8mm HTR Sterile Hammer Toe Reaming Kit, Material: Stainless Steel, Sterile: R, PX only, UDI: (01)00812926022345(17)231225 (10)TSL007202
Trilliant Surgical
The firm has identified that parts from the kit, lot: TSL007202, have the potential to contain the incorrect contents of the 8mm HTR Sterile Hammer Toe Reaming Kit. The kit should contain (2) reamers total, (1) Concave Reamer and (1) Convex Reamer; there is the potential that kits from the identified lot may have (2) of the same reamer type as opposed to one (1) of each type.
Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.