Siemens Magnetom Skyra, Verio, Spectra and Biograph MMR Nuclear Magnetic Resonance Imaging System and Tomographic Imager

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Model numbers: 10276755 (Verio), 10684333 (Verio Dot), 10684334 (Verio Dot upgrade), 10432915 (Skyra), 10655588 (Spectra), and 10433372 (Biograph mMR)
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Siemens Medical Solutions USA, Inc
Reason for Recall:
When positioning the Flex Large 4 Coil (part No. 8625761) off-center in left-right direction as it is used for example for hip and shoulder imaging and, in addition at an off-center position with respect to the head-foot direction, the coil can heat up in the area of the electronic housing (white plastic lids).
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Siemens Magnetom Skyra, Verio, Spectra and Biograph MMR Nuclear Magnetic Resonance Imaging System and Tomographic Imager

Product Codes/Lot Numbers:

Model numbers: 10276755 (Verio), 10684333 (Verio Dot), 10684334 (Verio Dot upgrade), 10432915 (Skyra), 10655588 (Spectra), and 10433372 (Biograph mMR)

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2464-2012

Related Recalls

During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.

Mar 9, 2026 Other Medical Devices View Details →
Class I - Dangerous

During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.

Mar 9, 2026 Other Medical Devices View Details →

ARTIS Pheno VE30A and VE40A, Model 10849000

Siemens Medical Solutions USA

Class I - Dangerous

During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.

Mar 12, 2026 Other Medical Devices View Details →