VITEK¿ 2 Gram Negative Susceptibility card (AST-GN93), REF 414985, 20 cards per carton.
Class I - DangerousWhat Should You Do?
- Check if you have this product: 6830196203 10/Jun/18 6830245203 29/Jul/18
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Biomerieux Inc
- Reason for Recall:
- The integrity of the product container may be compromised thereby exposing product to moisture which could lead to antibiotic degradation (loss of potency).
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
VITEK¿ 2 Gram Negative Susceptibility card (AST-GN93), REF 414985, 20 cards per carton.
Product Codes/Lot Numbers:
6830196203 10/Jun/18 6830245203 29/Jul/18
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2463-2017
Related Recalls
Potential for false resistant results for Colistin (cs02n) with multi-drug-resistant Pseudomonas aeruginosa and Acinetobacter baumannii complex used in specific cards
Ceftriaxone (cro02n) concentration errors associated with specific recently manufactured test kits which would have the potential for for false susceptible results. The risk is present for isolates that have a result of MIC=0.5, 1, 2.