BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, for FILMARRAY Systems, when used in conjunction with Culture Media Bottles: BACT/ALERT FA PLUS, REF: 410851, and BACT/ALERT PF PLUS, REF: 410853
Class I - DangerousWhat Should You Do?
- Check if you have this product: BCID2 Panel REF: RFIT-ASY-0147, UDI-DI: 00815381020338, All unexpired lots. BACT/ALERT REF/UDI-DI: Lot/Expiration: 410851/03573026596057: 0004101638/3/20/2024, 0004101930/7/28/2024, 0004102026/9/4/2024, 0004102408/2/16/2025, 0004102996/9/24/2025, 0004102946/8/29/2025, 0004102956/9/17/2025, 0004102998/9/24/2025. 410853/03573026596095: 0004101718/4/24/2024, 0004101790/6/1/2024, 0004101958/8/3/2024, 0004102452/3/3/2025, 0004102964/9/19/2025
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- BioFire Diagnostics, LLC
- Reason for Recall:
- If blood culture identification panel is used in conjunction with specific lots of culture media bottles, then false positive Serratia marcescens results may occur, due to an increased level of non-viable organism from serratia marcescens targets in culture media bottles, false positive result may lead to an inappropriate change in patient therapy.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, for FILMARRAY Systems, when used in conjunction with Culture Media Bottles: BACT/ALERT FA PLUS, REF: 410851, and BACT/ALERT PF PLUS, REF: 410853
Product Codes/Lot Numbers:
BCID2 Panel REF: RFIT-ASY-0147, UDI-DI: 00815381020338, All unexpired lots. BACT/ALERT REF/UDI-DI: Lot/Expiration: 410851/03573026596057: 0004101638/3/20/2024, 0004101930/7/28/2024, 0004102026/9/4/2024, 0004102408/2/16/2025, 0004102996/9/24/2025, 0004102946/8/29/2025, 0004102956/9/17/2025, 0004102998/9/24/2025. 410853/03573026596095: 0004101718/4/24/2024, 0004101790/6/1/2024, 0004101958/8/3/2024, 0004102452/3/3/2025, 0004102964/9/19/2025
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2456-2024
Related Recalls
BioFire Respiratory Panel 2.1, IVD, REF: 423742 (30 test kit)
BioFire Diagnostics
Due to unintended movement of pouch chemistry, nucleic acid test may result in elevated risk of false negative results and risk of control failures.
BioFire Respiratory Panel 2.1 plus Kit, REF: 423740, For Filmarray systems
BioFire Diagnostics
Due to manufacturing error, respiratory panel may result in elevated control failures and/or false negative results.
FilmArra Pneumonia Panel plus (Pneumoplus), REF: RFIT-LBL-0374
BioFire Diagnostics
Increased risk of control failures and false negative test results with multiplexed nucleic acid test.