CPT Hip Joint Metal/Polymer Semi-Constrained Cemented Prosthesis, used in Hip Arthroplasty; Item Numbers 00811400000 00811400010 00811400100 00811400110 00811400200 00811400210 00811400218 00811400230 00811400300 00811400310 00811400318 00811400330 00811400400 00811400410 00811400420 00811400423 00811400426 00811400430 00811400500 00811400510 00811400530 00811401218 00811401318 00811404000 00811405000

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Item Number / UDI-DI 00811400000 / 00889024145733 00811400010 / 00889024145740 00811400100 / 00889024145757 00811400110 / 00889024145764 00811400200 / 00889024145771 00811400210 / 00889024145788 00811400218 / 00889024145795 00811400230 / 00889024145801 00811400300 / 00889024145818 00811400310 / 00889024145825 00811400318 / 00889024145832 00811400330 / 00889024145849 00811400400 / 00889024145856 00811400410 / 00889024145863 00811400420 / 00889024145870 00811400423 / 00889024145887 00811400426 / 00889024145894 00811400430 / 00889024145900 00811400500 / 00889024145917 00811400510 / 00889024145924 00811400530 / 00889024145931 00811401218 / 00889024145948 00811401318 / 00889024145955 00811404000 / 00889024145962 00811405000 / 00889024145979
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Zimmer, Inc.
Reason for Recall:
Affected product has an increased risk of postoperative perisprosthetic femoral fracture (PFF). The IFU is being updated to reflect the risk of PFF.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

CPT Hip Joint Metal/Polymer Semi-Constrained Cemented Prosthesis, used in Hip Arthroplasty; Item Numbers 00811400000 00811400010 00811400100 00811400110 00811400200 00811400210 00811400218 00811400230 00811400300 00811400310 00811400318 00811400330 00811400400 00811400410 00811400420 00811400423 00811400426 00811400430 00811400500 00811400510 00811400530 00811401218 00811401318 00811404000 00811405000

Product Codes/Lot Numbers:

Item Number / UDI-DI 00811400000 / 00889024145733 00811400010 / 00889024145740 00811400100 / 00889024145757 00811400110 / 00889024145764 00811400200 / 00889024145771 00811400210 / 00889024145788 00811400218 / 00889024145795 00811400230 / 00889024145801 00811400300 / 00889024145818 00811400310 / 00889024145825 00811400318 / 00889024145832 00811400330 / 00889024145849 00811400400 / 00889024145856 00811400410 / 00889024145863 00811400420 / 00889024145870 00811400423 / 00889024145887 00811400426 / 00889024145894 00811400430 / 00889024145900 00811400500 / 00889024145917 00811400510 / 00889024145924 00811400530 / 00889024145931 00811401218 / 00889024145948 00811401318 / 00889024145955 00811404000 / 00889024145962 00811405000 / 00889024145979

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2450-2024

Related Recalls

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.

Dec 2, 2025 Other Medical Devices Nationwide View Details →

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.

Dec 2, 2025 Other Medical Devices Nationwide View Details →

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.

Dec 2, 2025 Other Medical Devices Nationwide View Details →