The O-arm O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging for adult and pediatric patients weighing 60 lbs. or greater and having an abdominal thickness greater than 16 cm and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Model No. BI70002000; UDI-DI: 00763000074128, 00763000081041, 00763000355555, 00763000496784, 00763000542801, 00763000616434, 00763000616526, 00763000616564; Serial No. C2786, C2841, C2781, C2624, C2659, C2660, C2665, C2672, C2708, C2723, C2735, C2760, C2774, C2775, C2779, C2798, C2814, C2823, C2873, C2918, C2968, C3290, C3363, C3188, C3322, C3387, C3404, and C3418.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Medtronic Navigation, Inc.-Littleton
Reason for Recall:
Ground cable installed incorrectly.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

The O-arm O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging for adult and pediatric patients weighing 60 lbs. or greater and having an abdominal thickness greater than 16 cm and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects.

Product Codes/Lot Numbers:

Model No. BI70002000; UDI-DI: 00763000074128, 00763000081041, 00763000355555, 00763000496784, 00763000542801, 00763000616434, 00763000616526, 00763000616564; Serial No. C2786, C2841, C2781, C2624, C2659, C2660, C2665, C2672, C2708, C2723, C2735, C2760, C2774, C2775, C2779, C2798, C2814, C2823, C2873, C2918, C2968, C3290, C3363, C3188, C3322, C3387, C3404, and C3418.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2449-2023

Related Recalls

Potential for an electrical component of the O-arm" O2 Imaging System to intermittently shut off,failure can cause the potential for surgical delay, additional imaging, cancellation and rescheduling of surgery, or completion of surgery without the benefit of image guidance.

Apr 26, 2024 Diagnostic Equipment Nationwide View Details →
Class I - Dangerous

Due to a manufacturing issue affecting particular O-arm" O2 Imaging Systems, Medtronic has determined that the gantry tractor motor drive belt may loosen over an extended number of uses at a higher frequency in these systems. If the belt loosens, it may result in rotation of 3D images about the gantry isocenter. While the 3D image is anatomically accurate within the image itself and may be used to confirm therapy, its electronic registered location may be rotated relative to the actual physical position of the patient.

Nov 15, 2019 Diagnostic Equipment Nationwide View Details →

The 0-arm 1000 Imaging System uses energy from the batteries to generate X-rays and move the system. If a charger board is not functioning properly, the batteries will not receive a full recharge of the potential capacity and results in the inability of the system to take X-rays, open the gantry, and move the system from one place to another, which may affect the ability to continue use within the operating room until the charger board and/or the batteries are replaced

May 1, 2019 Diagnostic Equipment Nationwide View Details →