syngo.CT VA20A, VA30A and VA40A in: syngo.via RT Image Suite / Sim&Go SOMATOM go.Now SOMATOM go.Up SOMATOM go.All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro and SOMATOM X.cite Material Numbers: 11061620 11061628 11061630 11061640 11061660 11061670
Class I - DangerousWhat Should You Do?
- Check if you have this product: Serial Numbers: 128051 130051 128065 128035 130057 128086 128079 128041 128019 128078 128060 128068 112045 128085 128075 128022 128033 130012 128016 130044 130048 130049 128018 119216 108085 128015 130019 130055 130054 128020 128039 130052 130036 130061 128049 117756 128029 130021 130045 130039 128063 128057 128045 128046 ***UPDATED 12/9/21*** 128114 128113 130058 ***Updated 1/31/22: Serial Number 128123 130015 130081 112261 112071 128124 128132 128125 130079 130077 Expanded Recall 3/12/22: 11061620 S/N: 111805 11061660 S/N: 128144, 128135, 128142, 128136, 128133 11061670 S/N: 130088, 130094, 130090, 130087, 130091, 130092, 130089
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Siemens Medical Solutions USA, Inc
- Reason for Recall:
- Potential for data loss when using the contouring or patient marking workflow when editing a structure set after saving and reopening a study, resulting in an incomplete structure set sent to the treatment planning system (TPS).
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
syngo.CT VA20A, VA30A and VA40A in: syngo.via RT Image Suite / Sim&Go SOMATOM go.Now SOMATOM go.Up SOMATOM go.All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro and SOMATOM X.cite Material Numbers: 11061620 11061628 11061630 11061640 11061660 11061670
Product Codes/Lot Numbers:
Serial Numbers: 128051 130051 128065 128035 130057 128086 128079 128041 128019 128078 128060 128068 112045 128085 128075 128022 128033 130012 128016 130044 130048 130049 128018 119216 108085 128015 130019 130055 130054 128020 128039 130052 130036 130061 128049 117756 128029 130021 130045 130039 128063 128057 128045 128046 ***UPDATED 12/9/21*** 128114 128113 130058 ***Updated 1/31/22: Serial Number 128123 130015 130081 112261 112071 128124 128132 128125 130079 130077 Expanded Recall 3/12/22: 11061620 S/N: 111805 11061660 S/N: 128144, 128135, 128142, 128136, 128133 11061670 S/N: 130088, 130094, 130090, 130087, 130091, 130092, 130089
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2446-2021
Related Recalls
ARTIS Icono biplane, floor and ceiling.: ARTIS icono biplane Model 11327600. ARTIS icono ceiling Model 11328100. ARTIS icono floor Model 11327700. interventional fluoroscopic x-ray system
Siemens Medical Solutions USA
During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.
ARTIS pheno, Model 10849000. interventional fluoroscopic x-ray system
Siemens Medical Solutions USA
During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.
ARTIS Pheno VE30A and VE40A, Model 10849000
Siemens Medical Solutions USA
During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.