Brand Name: Perfusor Space Product Name: PERFUSOR SPACE ACM-KIT Model/Catalog Number: 8713040U Product Description: Perfusor Space, Infusion Pump Kit, 1 each Component: N/A
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot Code: Model No 8713040U; UDI-DI 4046963687182; All Serial Numbers
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- B Braun Medical Inc
- Reason for Recall:
- Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Brand Name: Perfusor Space Product Name: PERFUSOR SPACE ACM-KIT Model/Catalog Number: 8713040U Product Description: Perfusor Space, Infusion Pump Kit, 1 each Component: N/A
Product Codes/Lot Numbers:
Lot Code: Model No 8713040U; UDI-DI 4046963687182; All Serial Numbers
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2436-2024
Related Recalls
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).