CADD-Solis VIP ambulatory Infusion Pumps, Model 2110. Indicated for the following uses: Intra-arterial, subcutaneous, intraperitoneal, in close proximity to nerves, into an intraoperative site, epidural space or subarachnoid space. Therapies that require a continuous rate of infusion, intermittent bolus, and/or with patient-controlled demand doses.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    REF List Number 21-2111-0100-00, UDI/DI 10610586039195, Serial Numbers: 10610586039195; REF List Number 21-2111-0100-51, UDI/DI 30610586039205, Serial Numbers: 1008871, 1011314, 1011695, 1012140, 1012157, 1012387, 1012393, 1012430, 1012607, 1012611, 1012749, 1012751, 1012752, 1012758, 1012765, 1012767, 1012769, 1012773, 1012815, 1012988, 1013182, 1013185, 1013186, 1013187, 1013219, 1013221, 1013223, 1013230, 1013235, 1013237, 1013239, 1013251, 1013256, 1013264, 1013265, 1013278, 1013282, 1013291, 1013295, 1013301, 1013562, 1013817, 1013824, 1013853, 1013866, 1013953, 1013979, 1014000, 1014064, 1014103, 1014117, 1014279, 1014282, 1014288, 1014302, 1014308, 1014447, 1014463, 1016862, 1016876, 1016881, 1014804, 1014806, 1014811, 1014831, 1014844, 1014861, 1014880, 1014895, 1015034, 1015375, 1015382, 1016292, 1016522, 1016709, 1038360, 1016879, 1016927, 1016961, 1023905, 1016989, 1016990, 1017003, 1017005, 1017006, 1017049, 1017051, 1017075, 1017096, 1017143, 1017146, 1017159, 1017161, 1017163, 1017172, 1017173, 1017174, 1017175, 1017180, 1017189, 1017200, 1017202, 1017221, 1017230, 1017237, 1017241, 1017244, 1017252, 1017261, 1017264, 1017267, 1017276, 1017284, 1017286, 1017290, 1017293, 1017294, 1017295, 1017302, 1017325, 1017327, 1017332, 1017333, 1017339, 1017347, 1017359, 1017407, 1017431, 1017433, 1017435, 1017436, 1017444, 1017447, 1017448, 1017455, 1017463, 1017471, 1017486, 1017488, 1017491, 1017493, 1017508, 1017509, 1017511, 1017512, 1017514, 1017516, 1017518, 1017522, 1017525, 1017532, 1017788, 1017970, 1018001, 1018005, 1018018, 1018048, 1018057, 1018062, 1018095, 1018115, 1018222, 1018286, 1018303, 1018339, 1018376, 1018387, 1018393, 1018415, 1018434, 1018528, 1018538, 1018561, 1018620, 1019029, 1019103, 1019208, 1019262, 1019051, 1020066, 1020067, 1020108, 1020109, 1020110, 1020111, 1020125, 1020128, 1020181, 1020185, 1020197, 1020205, 1020210, 1020221, 1026697, 1026698, 1026702, 1026709, 1026828, 1020836, 1020866, 1021919, 1021943, 1021978, 1022324, 1022346, 1022354, 1022368, 1022396, 1022414, 1022449, 1022467, 1022483, 1022502, 1022511, 1022532, 1022541, 1022751, 1022852, 1022864, 1022971, 1020421, 1023190, 1023192, 1023203, 1023206, 1023213, 1023364, 1023409, 1023449, 1023458, 1023460, 1023476, 1023630, 1023633, 1023673, 1023712, 1023763, 1022468, 1022471, 1023830, 1023839, 1023884, 1023912, 1024043, 1024050, 1024062, 1024070, 1024092, 1024099, 1024113, 1024120, 1024126, 1024146, 1024163, 1024460, 1024497, 1024567, 1025085, 1025086, 1025089, 1025093, 1025094, 1025095, 1025096, 1025099, 1025100, 1025117, 1025770, 1025776, 1025781, 1025792, 1025798, 1025802, 1025819, 1025834, 1025839, 1026270, 1026357, 1026687, 1017622, 1026851, 1026868, 1026873, 1026891, 1026914, 1026917, 1026920, 1026927, 1026931, 1026934, 1026971, 1026979, 1026982, 1026997, 1026999, 1027005, 1027006, 1027884, 1027900, 1027901, 1027906, 1027907, 1027908, 1027910, 1027913, 1027919, 1027920, 1027921, 1027922, 1027923, 1027925, 1027927, 1027952, 1027981, 1027983, 1027984, 1027985, 1028419, 1028549, 1028554, 1028596, 1028667, 1028668, 1028670, 1028673, 1028675, 1028678, 1028683, 1028685, 1028686, 1028687, 1028689, 1028691, 1028696, 1028699, 1028972, 1029018, 1029019, 1029023, 1029027, 1029051, 1029055, 1029492, 1029508, 1029515, 1029518, 1029549, 1029562, 1029567, 1029570, 1029571, 1029701, 1029703, 1029766, 1029770, 1029782, 1029799, 1029803, 1030407, 1030479, 1030483, 1030621, 1030626, 1030645, 1030648, 1031533, 1039351, 1031625, 1031644, 1014448, 1014452, 1032265, 1032403, 1032416, 1032428, 1032454, 1032456, 1032459, 1032514, 1032518, 1032529, 1032532, 1032537, 1032546, 1032551, 1032552, 1032655, 1032698, 1032719, 1032723, 1032724, 1032725, 1032727, 1032731, 1032745, 1032775, 1032776, 1032784, 1032795, 1032816, 1032846, 1032847, 1032850, 1032853, 1032875, 1032891, 1032895, 1032942, 1032947, 1032948, 1032949, 1032959, 1033109, 1023205, 1035367, 1035992, 1036003, 1036018, 1036082, 1036089, 1036092, 1036095, 1036096, 1036099, 1036105, 1036106, 1036107, 1036114, 1036116, 1036118, 1036129, 1036132, 1036134, 1036146, 1036148, 1036149, 1036157, 1036160, 1036166, 1036169, 1036173, 1036295, 1036305, 1036393, 1036415, 1036465, 1036472, 1036473, 1036493, 1036714, 1036716, 1036839, 1036956, 1036961, 1037405, 1037411, 1037413, 1037422, 1037423, 1037457, 1037474, 1038197, 1038213, 1038215, 1038224, 1038225, 1038238, 1038303, 1038309, 1038316, 1038323, 1038339, 1038340, 1038342, 1038343, 1038345, 1038347, 1038356, 1038362, 1038369, 1038415, 1038427, 1038476, 1038487, 1038561, 1038611, 1038784, 1039042, 1039047, 1039061, 1039084, 1039103, 1039115, 1039130, 1039324, 1039344, 1039352, 1039366, 1039369, 1039370, 1039379, 1039382, 1039385, 1039389, 1039391, 1039392, 1039393, 1039394, 1039396, 1039397, 1039400, 1039401, 1039407, 1039409, 1039412, 1039413, 1039415, 1039416, 1039417, 1039671, 1039673, 1039674, 1039676, 1039679, 1039682, 1039685, 1039686, 1039691, 1039697, 1039704, 1039708, 1039709, 1039713, 1039717, 1039720, 1039722, 1039723, 1039724, 1039725, 1039726, 1039728, 1039729, 1039731, 1040253, 1040263, 1040305, 1040513, 1040528, 1040555, 1040768, 1041265, 1041595, 1041597, 1041602, 1041972, 1042000, 1044123, 1044135, 1044137, 1044140, 1044148, 1044166, 1044167, 1044169, 1044204, 1044231, 1044329, 1044390, 1044710, 1044720, 1044727, 1044730, 1044741, 1044745, 1044762, 1044767, 1044786, 1044789, 1014874, 1044795, 1014875, 1047279, 1047358, 1017344, 1029477, 1029565, 1026631, 1026639, 1014480, 1028136, 1028139, 1039190, 1039191, 1024398, 1024399, 1024401, 1024406, 1022713, 1012639, 1015834, 1016051, 1022374, 1014677, 1014691, 1020369, 1068313, 1013445, 1036958, 1032283, 1032310, 1025074, 1027934, 1014813, 1039323, 1015791, 1023366, 1020683, 1020688, 1020704, 1020723, 1020729, 1033415, 1014833, 1014866, 1028304, 1013435, 1035554, 1047264, 1148607, 1016830, 1016831, 1016834, 1016839, 1016845, 1016851, 1016854, 1016855, 1042943, 1098608, 1020787, 1020025, 1020027, 1020227, 1020302, 1020303, 1012992, 1012597, 1030643, 1030644, 1014963, 1014969, 1014975, 1018757, 1018791, 1035931, 1035939, 1037398, 1041610, 1035377, 1024574, 1028705, 1015324, 1014422, 1014430, 1030566, 1030580, 1030582; REF List Number 21-2111-0200-02, UDI/DI 30610586039397, Serial Numbers: 1012544, 1024990, 1024992, 1013501; REF List Number 21-2111-0200-03, UDI/DI 10610586039416, Serial Numbers: 1015056, 1015057, 1015058, 1015059, 1015060, 1015061, 1017665, 1017675, 1017692, 1017693, 1017695, 1017696, 1017704, 1017705, 1017707, 1021927, 1021928, 1021933, 1021936, 1021937, 1022496, 1022507, 1022521, 1022952, 1022954, 1022967, 1022978, 1022985, 1022988, 1022999, 1023000, 1023003, 1022970, 1016119, 1014632, 1016437, 1016447, 1016470, 1017741, 1017749, 1015964, 1022005, 1022213, 1022428, 1022497, 1022685, 1022705, 1023147, 1023318, 1023703, 1023704, 1023705, 1023706, 1023713, 1023714, 1023718, 1023721, 1023730, 1023764, 1023767, 1023776, 1023778, 1023780, 1023782, 1023801, 1023815, 1024199, 1024200, 1024201, 1024202, 1024218, 1024219, 1024223, 1024224, 1024225, 1024230, 1024232, 1024233, 1024234, 1016471, 1024690, 1024691, 1025322, 1025344, 1025346, 1029435, 1015489, 1022032, 1013091, 1024665, 1024666, 1024704, 1024705, 1024706, 1024707, 1024708, 1024709, 1024710, 1024712, 1024714, 1024715, 1024716, 1024718, 1024719, 1024720, 1024721, 1024722, 1024723, 1024724, 1024725, 1016343, 1025324, 1021442, 1021443, 1021444, 1021448, 1021511, 1114780, 1022175, 1017759, 1015065; REF List Number 21-2111-0200-06, UDI/DI 10610586039430, Serial Numbers: 1021051, 1021052, 1021053, 1021054, 1021055, 1021056, 1021057, 1021058, 1021512, 1021513; REF List Number 21-2111-0300-00, UDI/DI 10610586038792, Serial Numbers: 1047706, 1078090, 1078327, 1078337; REF List Number 21-2111-0300-01, UDI/DI 10610586038778, Serial Numbers: 1003302, 1003614, 1005310, 1005321, 1011210, 1012744, 1014286, 1063228, 1014808, 1014877, 1016779, 1017035, 1017038, 1017156, 1017169, 1017178, 1017182, 1017183, 1017184, 1017185, 1017192, 1017194, 1017204, 1017239, 1017245, 1017247, 1017250, 1017251, 1017257, 1017279, 1017280, 1017287, 1017311, 1017314, 1017318, 1017320, 1017324, 1017328, 1017329, 1017338, 1017345, 1017351, 1017353, 1017355, 1017428, 1017439, 1017440, 1017446, 1017451, 1017454, 1017458, 1017459, 1017464, 1017469, 1017470, 1017479, 1017480, 1017482, 1017483, 1017489, 1017494, 1017497, 1017499, 1017500, 1017502, 1017519, 1017524, 1017531, 1052696, 1052700, 1097947, 1018234, 1084235, 1085310, 1079869, 1067819, 1124242, 1124338, 1022275, 1022437, 1085055, 1024111, 1098821, 1058329, 1028578, 1064986, 1029524, 1099542, 1032412, 1099587, 1032922, 1032983, 1099637, 1099645, 1099659, 1084423, 1075642, 1072386, 1053632, 1053633, 1051244, 1052021, 1053407, 1053532, 1053535, 1053546, 1053547, 1053592, 1053601, 1053620, 1053626, 1053627, 1053656, 1053662, 1053663, 1053736, 1053737, 1053743, 1053745, 1053749, 1053750, 1036147, 1058053, 1153346, 1038227, 1038897, 1038908, 1076129, 1065016, 1065043, 1040035, 1075430, 1075448, 1075838, 1044129, 1044132, 1044141, 1044144, 1044145, 1044146, 1044151, 1044156, 1044157, 1044165, 1044170, 1067147, 1067148, 1044895, 1069460, 1046788, 1046794, 1046803, 1046812, 1046814, 1059831, 1059850, 1048771, 1047821, 1017468, 1017515, 1048248, 1048251, 1048259, 1118271, 1049473, 1049517, 1050421, 1050436, 1050451, 1044937, 1050776, 1050809, 1050972, 1044979, 1051126, 1045019, 1051249, 1051250, 1051252, 1051253, 1051254, 1051255, 1051256, 1051259, 1051260, 1051261, 1051262, 1051263, 1051264, 1051266, 1051267, 1051268, 1051269, 1051270, 1051271, 1051272, 1051273, 1051274, 1051275, 1051277, 1051278, 1051279, 1051280, 1051281, 1051283, 1051284, 1051285, 1051287, 1051288, 1051290, 1051291, 1051293, 1051295, 1051297, 1051304, 1051305, 1051306, 1051307, 1051425, 1051427, 1051428, 1051429, 1051431, 1051432, 1051433, 1051434, 1051435, 1051436, 1051437, 1051438, 1051439, 1051440, 1051442, 1051443, 1051445, 1051446, 1051447, 1051448, 1051451, 1051452, 1051453, 1051454, 1051455, 1051926, 1052073, 1052235, 1052683, 1052746, 1052748, 1052757, 1052816, 1052853, 1052860, 1052861, 1052875, 1052883, 1052886, 1052901, 1052902, 1052908, 1052910, 1052912, 1052915, 1052922, 1052925, 1053091, 1053093, 1053099, 1053100, 1053107, 1097346, 1053279, 1053321, 1053324, 1053325, 1053326, 1053328, 1053329, 1053333, 1053334, 1053335, 1053336, 1053338, 1053339, 1053340, 1053341, 1053343, 1053344, 1053345, 1053398, 1053402, 1053436, 1053506, 1053522, 1053527, 1053541, 1053579, 1053585, 1053590, 1053597, 1053600, 1053602, 1053617, 1053621, 1053623, 1053649, 1053741, 1053742, 1053747, 1054688, 1056254, 1056297, 1056304, 1056312, 1056558, 1056633, 1056714, 1098405, 1098407, 1058045, 1058120, 1058178, 1087970, 1058209, 1087995, 1088004, 1093056, 1058255, 1058266, 1058282, 1058296, 1058310, 1058344, 1097533, 1097547, 1097549, 1097550, 1097562, 1097681, 1059241, 1059245, 1048031, 1060639, 1060951, 1060961, 1060962, 1060968, 1060974, 1060978, 1060983, 1060988, 1060989, 1045060, 1057816, 1057837, 1058336, 1058356, 1058364, 1062359, 1062121, 1062122, 1062124, 1062126, 1062230, 1062241, 1062243, 1062251, 1069560, 1069563, 1069564, 1069566, 1069570, 1053455, 1053463, 1053470, 1056327, 1048353, 1048355, 1048361, 1048365, 1088029, 1064771, 1064854, 1064891, 1064893, 1064905, 1064929, 1064989, 1065101, 1065107, 1065211, 1065212, 1119182, 1065719, 1066323, 1066324, 1066326, 1066327, 1066328, 1066330, 1066331, 1066332, 1066333, 1067046, 1067095, 1067096, 1067097, 1097918, 1097920, 1079183, 1079251, 1079379, 1079524, 1079568, 1079623, 1079916, 1080447, 1080449, 1080450, 1080455, 1080465, 1080515, 1080524, 1080525, 1080527, 1080528, 1080530, 1080531, 1080532, 1080574, 1080577, 1080580, 1080581, 1080583, 1080588, 1080591, 1080592, 1080611, 1080612, 1080621, 1080622, 1080624, 1080625, 1080629, 1080632, 1080634, 1080635, 1080638, 1080640, 1080643, 1080651, 1080652, 1068739, 1068753, 1068781, 1068800, 1097035, 1069154, 1069157, 1069160, 1069164, 1098645, 1098848, 1098857, 1069234, 1098858, 1098859, 1069241, 1069242, 1069243, 1098861, 1069250, 1069251, 1098879, 1098892, 1098907, 1099828, 1099836, 1069509, 1069510, 1069512, 1069514, 1069515, 1069516, 1069520, 1069532, 1069534, 1099099, 1065924, 1087862, 1099037, 1099042, 1099059, 1099063, 1099065, 1099066, 1099071, 1099073, 1099076, 1166599, 1070071, 1070325, 1097913, 1070868, 1070869, 1070984, 1071072, 1071097, 1071098, 1071321, 1071411, 1071442, 1071460, 1071480, 1071612, 1071622, 1071783, 1071830, 1071853, 1071877, 1072020, 1072031, 1072062, 1072063, 1072064, 1072074, 1072098, 1072124, 1072228, 1072300, 1072365, 1072401, 1093374, 1072483, 1072493, 1072528, 1072674, 1072722, 1072729, 1072730, 1072751, 1072755, 1072772, 1073048, 1073073, 1073097, 1073105, 1073118, 1073141, 1073149, 1073399, 1073413, 1073430, 1073568, 1073570, 1015795, 1074432, 1065086, 1074808, 1074809, 1074810, 1074988, 1075271, 1075368, 1075385, 1075529, 1075605, 1075606, 1075607, 1075609, 1075610, 1075611, 1075612, 1075613, 1075626, 1075753, 1075799, 1075949, 1075952, 1075956, 1075958, 1075959, 1075961, 1076011, 1076019, 1076022, 1099414, 1099416, 1099417, 1072996, 1080526, 1080620, 1080678, 1080856, 1080878, 1080896, 1080930, 1080932, 1080934, 1080938, 1080939, 1080940, 1080943, 1080946, 1080960, 1080966, 1080967, 1081017, 1081019, 1081020, 1081021, 1081023, 1081024, 1081025, 1081026, 1081027, 1081028, 1081057, 1081083, 1081089, 1081093, 1081094, 1081573, 1081583, 1081649, 1081840, 1081876, 1088380, 1088381, 1088382, 1088383, 1074825, 1065494, 1064435, 1097314, 1083622, 1083637, 1083643, 1134952, 1084194, 1084204, 1084218, 1084223, 1084367, 1084417, 1084487, 1084503, 1084508, 1084519, 1084525, 1084599, 1084608, 1084609, 1084625, 1084630, 1084637, 1097527, 1084859, 1084863, 1051289, 1051294, 1051430, 1057941, 1057942, 1086551, 1124156, 1075404, 1088398, 1087643, 1087645, 1087646, 1087648, 1087685, 1087692, 1088000, 1088081, 1088090, 1088478, 1088491, 1092992, 1095249, 1096819, 1097067, 1097076, 1097088, 1097093, 1097096, 1097110, 1097115, 1097130, 1097135, 1097206, 1097351, 1097600, 1097603, 1097607, 1097613, 1097684, 1097711, 1097715, 1097721, 1097743, 1097746, 1097788, 1097854, 1097862, 1097863, 1097865, 1097866, 1097869, 1097884, 1097888, 1097892, 1097894, 1097905, 1097910, 1097928, 1097930, 1097944, 1097946, 1097948, 1097950, 1097953, 1097957, 1097960, 1097983, 1098194, 1098196, 1098205, 1098272, 1098278, 1098284, 1098290, 1098307, 1098308, 1098310, 1098315, 1098317, 1098322, 1098323, 1098327, 1098336, 1098340, 1098342, 1098348, 1098352, 1098362, 1098365, 1098368, 1098379, 1098381, 1098382, 1098384, 1098480, 1098569, 1098570, 1098572, 1098575, 1098577, 1098581, 1098584, 1098586, 1098587, 1098590, 1098591, 1098603, 1098630, 1098674, 1098832, 1098839, 1098841, 1098845, 1048468, 1076093, 1076105, 1076118, 1072955, 1072960, 1099466, 1099467, 1099481, 1099482, 1099483, 1099484, 1099485, 1099486, 1099488, 1099490, 1099876, 1099877, 1099879, 1099882, 1099886, 1099889, 1099895, 1099901, 1099903, 1099904, 1100014, 1100024, 1100192, 1100206, 1100209, 1100212, 1100214, 1100221, 1100229, 1100232, 1100235, 1100242, 1100247, 1100250, 1100256, 1100261, 1100397, 1101971, 1101972, 1101976, 1101977, 1101981, 1101985, 1101987, 1101989, 1101992, 1101994, 1102001, 1102002, 1102004, 1102005, 1102006, 1102008, 1088578, 1080137, 1080140, 1080542, 1080553, 1080566, 1119270, 1088725, 1088750, 1088763, 1044685, 1014976, 1097468, 1097479, 1075923, 1115859, 1115882, 1115883, 1115887, 1115896, 1117327, 1117371, 1117374, 1117392, 1117930, 1117937, 1117941, 1117945, 1117955, 1117958, 1117976, 1117981, 1117982, 1117989, 1117990, 1117996, 1117999, 1118001, 1118004, 1118008, 1118010, 1118012, 1118013, 1118018, 1118026, 1118029, 1118033, 1118035, 1118036, 1118037, 1118038, 1118039, 1118040, 1118047, 1118100, 1118101, 1118102, 1118111, 1118112, 1118118, 1118126, 1118127, 1118129, 1118131, 1118134, 1118136, 1118137, 1118140, 1118156, 1118170, 1118183, 1118184, 1118196, 1118208, 1118213, 1118214, 1118216, 1118228, 1118229, 1118230, 1118235, 1118239, 1118242, 1118243, 1118244, 1118245, 1118250, 1118253, 1118255, 1118281, 1118282, 1118286, 1118292, 1118298, 1118300, 1118309, 1118310, 1118315, 1118321, 1118323, 1118327, 1118344, 1118347, 1118348, 1118357, 1118358, 1118367, 1118372, 1118374, 1118379, 1118381, 1118383, 1118386, 1118388, 1118389, 1118393, 1118395, 1118402, 1118403, 1118409, 1118410, 1118411, 1118413, 1118416, 1118419, 1118421, 1118423, 1118424, 1118428, 1118431, 1118438, 1118439, 1118441, 1118443, 1118826, 1118908, 1119227, 1119309, 1119323, 1119332, 1119336, 1119337, 1119339, 1119402, 1119407, 1069523, 1119539, 1119552, 1119562, 1119720, 1084420, 1119834, 1119836, 1119862, 1119870, 1119915, 1119940, 1119952, 1119957, 1119985, 1084962, 1084964, 1084965, 1120322, 1120328, 1120329, 1120833, 1120834, 1120844, 1120847, 1120848, 1120849, 1120850, 1120852, 1120854, 1120855, 1120905, 1120915, 1120919, 1120923, 1120926, 1120928, 1120932, 1120938, 1120958, 1120960, 1120962, 1120964, 1120965, 1120967, 1120969, 1120970, 1120987, 1121013, 1121184, 1121199, 1121216, 1121219, 1121272, 1121282, 1121292, 1121336, 1121340, 1121343, 1121347, 1121350, 1121353, 1121356, 1121360, 1121361, 1121373, 1121381, 1121386, 1121387, 1121389, 1121394, 1121611, 1121621, 1123858, 1072917, 1124104, 1124105, 1124177, 1124186, 1124196, 1124334, 1124335, 1062991, 1063034, 1124913, 1063220, 1063221, 1064467, 1040747, 1048664, 1052680, 1052742, 1052773, 1052881, 1052884, 1052904, 1052943, 1052958, 1133649, 1133685, 1134965, 1135183, 1053005, 1053040, 1053110, 1053143, 1143564, 1145868, 1145876, 1099777, 1153404, 1153333, 1153409, 1153410, 1153412, 1153419, 1153421, 1153423, 1153424, 1153426, 1058423, 1058425, 1058426, 1058431, 1051257, 1058466, 1058483, 1069819, 1170008, 1170116, 1170731, 1170923, 1030577, 1121364, 1049006, 1049014, 1049015, 1049018, 1049010, 1053674, 1054137, 1054206, 1054210, 1054224, 1054942, 1054946, 1054953, 1097254, 1097289, 1098000; REF List Number 21-2111-0300-02, UDI/DI 10610586038891, Serial Numbers: 1047659, 1079778, 1079880, 1079927, 1079935, 1080039, 1080296, 1080404, 1080405, 1080705, 1080706, 1080707, 1080709, 1080774, 1080789, 1080793, 1080794, 1041790, 1041791, 1041792, 1078019, 1047649, 1077673, 1094561, 1049499, 1041785, 1041786, 1041787, 1041788, 1041789, 1089591, 1089597, 1089599, 1092736, 1114614, 1114626, 1114628, 1050236, 1050239, 1050240, 1081319, 1081320, 1081321, 1081322, 1081323, 1081325, 1081326, 1081327, 1024974, 1033849, 1033692, 1033694, 1033889, 1036629; REF List Number 21-2111-0300-03, UDI/DI 10610586038938, Serial Numbers: 1025549, 1025550, 1028173, 1028175, 1034021, 1035138, 1035139, 1082157, 1102976, 1113445, 1118881, 1118882, 1119022, 1119023, 1119024, 1119025, 1119026, 1119027, 1119028, 1119030, 1119031, 1119564, 1119565, 1119566, 1119567, 1119568, 1119569, 1119570, 1119572, 1117486, 1080749, 1045084, 1025533, 1032098, 1055691, 1025458, 1025459, 1123409, 1123411, 1123412, 1123413, 1029649, 1029912, 1030262, 1030731, 1031035, 1031357, 1031363, 1031366, 1031367, 1031368, 1031376, 1032099, 1033665, 1033740, 1033741, 1034000, 1035135, 1035143, 1035145, 1035146, 1035150, 1035155, 1035173, 1119016, 1119651, 1037891, 1082255, 1113431, 1045226, 1111513, 1115756, 1115760, 1115794, 1091538, 1091541, 1091546, 1117701, 1129158, 1024341, 1030248, 1030265, 1070159, 1070160, 1070161, 1070162, 1100852, 1100853, 1100863, 1047507, 1048346, 1048837, 1049242, 1050292, 1054652, 1054674, 1054675, 1054676, 1055274, 1031375, 1032094, 1137853, 1108890, 1055267, 1055266, 1087323, 1025450, 1032105, 1080693, 1080739, 1066557, 1047509, 1048056, 1048057, 1048058, 1048059, 1048060, 1048061, 1048063, 1048064, 1048065, 1048066, 1048067, 1131196, 1080741, 1080742, 1080743, 1080745, 1080747, 1080748, 1080750, 1080751, 1080752, 1080831, 1080832, 1080833, 1080834, 1080835, 1080836, 1080838, 1080839, 1080840, 1081270, 1081271, 1081294, 1081300, 1082350, 1082346, 1082347, 1033613, 1031636, 1047506, 1089718, 1093465, 1100859, 1102972, 1103051, 1103054, 1103059, 1108050, 1111633, 1112409, 1092340, 1095118, 1115246, 1115260, 1117473, 1117483, 1117718, 1117724, 1117739, 1117744, 1100856, 1102453, 1102459, 1102665, 1102668, 1102669, 1119013, 1119599, 1119647, 1119648, 1119660, 1041859, 1123380, 1123388, 1126634, 1129179, 1131195, 1131197, 1131323, 1135493, 1137683, 1144640, 1149550, 1150592, 1155075, 1024977, 1035237, 1035238, 1034018, 1092337, 1052027, 1133743, 1133748, 1045154, 1045096, 1037544, 1033609, 1114853, 1033614; REF List Number 21-2111-0300-06, UDI/DI 10610586038969, Serial Numbers: 1006546, 1032061, 1036631, 1007059, 1074596, 1074597, 1074598, 1074599, 1074600, 1074601, 1074602, 1074603, 1074604, 1074605, 1115720, 1115721, 1115722, 1115723, 1115724, 1016516; REF List Number 21-2111-0300-08, UDI/DI 10610586039010, Serial Numbers: 1046527, 1110009, 1110020, 1155180, 1157317, 1163923; REF List Number 21-2111-0300-14, UDI/DI 10610586039065, Serial Numbers: 1115208; REF List Number 21-2111-0300-50, UDI/DI 10610586038839, Serial Numbers: 1037514, 1066478, 1066785, 1066791, 1066815, 1027152, 1027157, 1028030; REF List Number 21-2111-0400-51, UDI/DI 10610586041037, Serial Numbers: 1118744, 1108192; REF List Number 21-2111-0401-02L, UDI/DI 35019517135323, Serial Numbers: 1001723, 1004057, 1006543, 1159374, 1177517, 1159408, 1159409, 1152299, 1152300, 1152301, 1152302, 1152303, 1152304, 1152305, 1152306, 1152307, 1152308, 1152309, 1152310, 1152311, 1152312, 1152313, 1152314, 1152315, 1152316, 1152317, 1152318, 1152319, 1152320, 1152321, 1152322, 1152323, 1152324, 1152325, 1152326, 1152327, 1152328, 1152329, 1152330, 1152331, 1152332, 1152333, 1152334, 1152335, 1152336, 1152337, 1152338, 1152339, 1152340, 1152341, 1152342, 1152343, 1152344, 1152345, 1152346, 1152435, 1152436, 1153523, 1153525, 1153526, 1153527, 1153528, 1153529, 1153530, 1153531, 1153532, 1153533, 1153534, 1153535, 1153536, 1153537, 1153538, 1153539, 1153540, 1153541, 1153543, 1153544, 1153545, 1153546, 1153547, 1153548, 1153549, 1153550, 1153551, 1153552, 1153553, 1153554, 1153555, 1153556, 1153557, 1153558, 1153559, 1153560, 1153561, 1153562, 1153563, 1153564, 1153565, 1153566, 1153567, 1153568, 1153569, 1153570, 1153571, 1153572, 1153573, 1153574, 1153575, 1153576, 1153577, 1153578, 1153579, 1153580, 1153581, 1153582, 1153583, 1153584, 1153585, 1153586, 1153587, 1153588, 1153589, 1153590, 1153591, 1153592, 1153593, 1153594, 1153595, 1153596, 1153597, 1153598, 1153599, 1153600, 1153601, 1153602, 1155290, 1155291, 1155292, 1155293, 1155294, 1155295, 1155296, 1155297, 1155298, 1155299, 1155300, 1155301, 1155302, 1155303, 1155304, 1155305, 1155306, 1155307, 1155309, 1155310, 1155311, 1155312, 1155313, 1155314, 1155315, 1155316, 1155317, 1155319, 1155320, 1155321, 1155322, 1155323, 1155324, 1155325, 1155326, 1155327, 1155328, 1155329, 1155330, 1155331, 1155332, 1155333, 1155334, 1155335, 1155336, 1155338, 1155339, 1155340, 1155341, 1155343, 1155344, 1155345, 1155346, 1155347, 1155348, 1155349, 1155350, 1155351, 1155352, 1155353, 1155355, 1155356, 1155357, 1155358, 1155359, 1155360, 1155361, 1155362, 1155363, 1155364, 1155365, 1155366, 1155367, 1155368, 1155369, 1155370, 1159327, 1159328, 1159329, 1159330, 1159331, 1159332, 1159333, 1159334, 1159335, 1159336, 1159337, 1159338, 1159340, 1159341, 1159342, 1159343, 1159344, 1159345, 1159346, 1159347, 1159348, 1159349, 1159350, 1159351, 1159352, 1159353, 1159354, 1159355, 1159357, 1159358, 1159359, 1159360, 1159361, 1159362, 1159363, 1159364, 1159365, 1159366, 1159368, 1159369, 1159370, 1159371, 1159372, 1159373, 1159375, 1159376, 1159377, 1159378, 1159379, 1159380, 1159383, 1159384, 1159385, 1159386, 1159387, 1159388, 1159389, 1159390, 1159391, 1159392, 1159393, 1159395, 1159396, 1159397, 1159398, 1159399, 1159400, 1159401, 1159402, 1159403, 1159404, 1159405, 1159406, 1159407, 1159410, 1159411, 1159412, 1159413, 1159414, 1159415, 1159416, 1159417, 1159418, 1159419, 1159420, 1159421, 1159422, 1159423, 1159424, 1159425, 1159426, 1159427, 1159428, 1159429, 1159430, 1159431, 1159432, 1159433, 1159434, 1159435, 1159436, 1159437, 1159438, 1159439, 1159440, 1159441, 1159775, 1159779, 1159780, 1159781, 1159782, 1159784, 1159785, 1159786, 1159787, 1159788, 1159789, 1159790, 1159791, 1159792, 1159793, 1159794, 1159795, 1159796, 1159797, 1159798, 1159799, 1159800, 1159802, 1159803, 1159804, 1159805, 1159806, 1159807, 1159808, 1159809, 1159810, 1159811, 1159812, 1159813, 1159814, 1159815, 1159816, 1159817, 1159818, 1159819, 1159820, 1159821, 1159822, 1159823, 1159824, 1159825, 1159827, 1159828, 1159829, 1159831, 1159832, 1159833, 1159834, 1159835, 1159836, 1159837, 1159838, 1159839, 1159840, 1159842, 1159843, 1159844, 1159845, 1159848, 1159849, 1159850, 1159852, 1159853, 1159854, 1159855, 1159856, 1159857, 1159858, 1159859, 1087064, 1160254, 1160259, 1160260, 1160261, 1160262, 1160263, 1160264, 1160265, 1160266, 1160267, 1160268, 1160269, 1160270, 1160271, 1160272, 1160273, 1160274, 1160275, 1160276, 1160277, 1160278, 1160279, 1160280, 1160281, 1160283, 1164110, 1164125, 1164126, 1164127, 1164128, 1164129, 1164130, 1164131, 1164132, 1164133, 1164134, 1164135, 1164136, 1164137, 1164138, 1164139, 1164140, 1164141, 1164142, 1164143, 1164144, 1164145, 1164146, 1164147, 1164148, 1164149, 1164150, 1164151, 1164152, 1164153, 1164154, 1164155, 1164156, 1164157, 1164158, 1164159, 1164160, 1164161, 1164162, 1164163, 1164164, 1164165, 1164166, 1164167, 1164168, 1164169, 1164170, 1164171, 1164172, 1164173, 1164174, 1164175, 1164176, 1164177, 1164178, 1164179, 1164180, 1164181, 1164182, 1164183, 1164184, 1164185, 1164186, 1164187, 1164188, 1164189, 1164190, 1164191, 1164192, 1164193, 1164194, 1165081, 1167531, 1167532, 1167533, 1167534, 1167535, 1167536, 1167537, 1167538, 1167539, 1167540, 1167541, 1167542, 1167543, 1167544, 1167545, 1167546, 1167557, 1167558, 1167559, 1167560, 1167561, 1167562, 1167563, 1167564, 1167565, 1167566, 1167567, 1167568, 1167569, 1167570, 1170552, 1170553, 1170554, 1170555, 1170556, 1170557, 1170558, 1170559, 1170560, 1170561, 1170562, 1170563, 1170564, 1170566, 1170567, 1170571, 1170572, 1170573, 1170574, 1170575, 1170576, 1170578, 1170580, 1170581, 1170582, 1170583, 1170584, 1170585, 1170586, 1170587, 1170589, 1170802, 1170803, 1170805, 1170806, 1170807, 1170808, 1170809, 1170810, 1170811, 1170812, 1170813, 1170815, 1170816, 1170817, 1170818, 1170819, 1170820, 1170821, 1170822, 1170823, 1170824, 1170826, 1170827, 1170830, 1170831, 1170832, 1170833, 1170834, 1170835, 1170836, 1170837, 1170838, 1170839, 1170840, 1170841, 1170843, 1170844, 1170846, 1170847, 1170848, 1170849, 1170850, 1170853, 1170854, 1170859, 1170860, 1170862, 1170865, 1170904, 1176830, 1176831, 1176873, 1176875, 1176876, 1176877, 1176879, 1176880, 1176881, 1176882, 1176883, 1176884, 1176885, 1176886, 1176887, 1176897, 1176898, 1176899, 1176900, 1176901, 1176902, 1176903, 1176904, 1176905, 1176906, 1176907, 1176908, 1176909, 1176910, 1176911, 1176912, 1176913, 1176914, 1176915, 1176916, 1176917, 1176918, 1176919, 1176920, 1176921, 1176922, 1176923, 1176924, 1176925, 1176926, 1176927, 1176928, 1176929, 1177230, 1177237, 1177239, 1177241, 1177243, 1177245, 1177246, 1177247, 1177248, 1177249, 1177250, 1177251, 1177253, 1177254, 1177255, 1177257, 1177258, 1177259, 1177260, 1177261, 1177262, 1177312, 1177360, 1177361, 1177362, 1177367, 1177368, 1177369, 1177370, 1177371, 1177372, 1177373, 1177374, 1177375, 1177376, 1177377, 1177378, 1177379, 1177380, 1177381, 1177382, 1177383, 1177384, 1177385, 1177386, 1177387, 1177388, 1177389, 1177390, 1177391, 1177392, 1177393, 1177394, 1177395, 1177464, 1177475, 1177476, 1177477, 1177478, 1177479, 1177480, 1177481, 1177482, 1177483, 1177484, 1177485, 1177486, 1177487, 1177488, 1177489, 1177490, 1177491, 1177492, 1177493, 1177494, 1177495, 1177496, 1177497, 1177498, 1177499, 1177500, 1177501, 1177502, 1177503, 1177504, 1177505, 1177506, 1177507, 1177508, 1177509, 1177510, 1177511, 1177512, 1177513, 1177514, 1177515, 1177516, 1177518, 1177519, 1177520, 1177521, 1177522, 1177523, 1177524, 1177525, 1177526, 1177527, 1177528, 1177529, 1177530, 1177553, 1181160, 1181161, 1181162, 1181163, 1181164, 1181420; REF List Number 21-2111-0401-03L, UDI/DI 35019517128677, Serial Numbers: 1013538, 1015946, 1014903, 1014901, 1158620, 1029080, 1029428, 1158595, 1158697, 1181143, 1165410, 1144636, 1150613, 1152280, 1152347, 1152348, 1152349, 1152350, 1152351, 1152352, 1152353, 1152354, 1152355, 1152356, 1152357, 1152358, 1152359, 1152360, 1152361, 1152362, 1152363, 1152364, 1152365, 1152366, 1152367, 1152368, 1152369, 1152370, 1152371, 1152372, 1152373, 1152374, 1152375, 1152376, 1152377, 1152378, 1152379, 1152380, 1152381, 1152382, 1152383, 1152384, 1152385, 1152386, 1152387, 1152388, 1152389, 1152390, 1152391, 1152392, 1152393, 1152394, 1152395, 1152396, 1152397, 1153603, 1153604, 1153605, 1153606, 1153607, 1153608, 1153609, 1153610, 1153611, 1153612, 1153613, 1153614, 1153615, 1153616, 1153617, 1153618, 1153619, 1153620, 1153621, 1153622, 1153623, 1153624, 1153625, 1153626, 1153627, 1153628, 1153629, 1153630, 1153631, 1153632, 1154727, 1154728, 1154729, 1154730, 1154731, 1154732, 1154733, 1154734, 1154735, 1154736, 1154737, 1154738, 1154739, 1155040, 1155041, 1155042, 1155043, 1155044, 1155045, 1155046, 1155047, 1155048, 1155049, 1155050, 1155051, 1155052, 1155053, 1155187, 1155188, 1155189, 1155190, 1155191, 1155192, 1155193, 1155194, 1155195, 1155196, 1155197, 1155198, 1155199, 1155200, 1155201, 1155202, 1155203, 1155204, 1155205, 1155206, 1155207, 1155208, 1155209, 1155210, 1155211, 1155212, 1155213, 1155214, 1155215, 1155216, 1155217, 1155218, 1155219, 1155220, 1155221, 1155222, 1155223, 1155224, 1155225, 1155226, 1155227, 1155228, 1155229, 1155230, 1155231, 1155232, 1155233, 1155234, 1155235, 1155236, 1155237, 1155238, 1155239, 1155240, 1155241, 1155242, 1155243, 1155244, 1155245, 1155246, 1155247, 1155248, 1155249, 1155250, 1155251, 1155252, 1155253, 1155254, 1155255, 1155256, 1155257, 1155258, 1155259, 1155260, 1155261, 1155262, 1155263, 1155264, 1155265, 1155266, 1155505, 1155506, 1155507, 1155508, 1155509, 1155510, 1155511, 1155512, 1155513, 1155514, 1155515, 1155516, 1155517, 1155518, 1155519, 1155520, 1155521, 1155522, 1155523, 1155524, 1155525, 1155526, 1155527, 1155528, 1155529, 1155530, 1155531, 1155532, 1155533, 1155534, 1155535, 1155536, 1155537, 1155539, 1155540, 1155541, 1155542, 1155543, 1155544, 1155545, 1155546, 1155547, 1155548, 1155549, 1155550, 1155551, 1155552, 1155553, 1155554, 1158489, 1158490, 1158491, 1158492, 1158493, 1158494, 1158495, 1158496, 1158497, 1158498, 1158499, 1158500, 1158501, 1158502, 1158503, 1158504, 1158505, 1158506, 1158507, 1158508, 1158509, 1158510, 1158511, 1158512, 1158513, 1158514, 1158515, 1158516, 1158517, 1158518, 1158519, 1158520, 1158521, 1158522, 1158523, 1158524, 1158525, 1158526, 1158527, 1158528, 1158529, 1158530, 1158531, 1158532, 1158533, 1158534, 1158535, 1158536, 1158537, 1158538, 1158539, 1158540, 1158541, 1158542, 1158543, 1158544, 1158545, 1158546, 1158547, 1158548, 1158551, 1158552, 1158553, 1158583, 1158584, 1158585, 1158586, 1158587, 1158588, 1158589, 1158590, 1158591, 1158592, 1158593, 1158594, 1158596, 1158597, 1158598, 1158599, 1158600, 1158601, 1158602, 1158603, 1158604, 1158605, 1158606, 1158607, 1158608, 1158609, 1158610, 1158611, 1158612, 1158613, 1158614, 1158615, 1158616, 1158617, 1158618, 1158619, 1158621, 1158622, 1158623, 1158624, 1158625, 1158626, 1158627, 1158628, 1158629, 1158630, 1158631, 1158632, 1158633, 1158634, 1158635, 1158636, 1158637, 1158638, 1158639, 1158640, 1158641, 1158642, 1158643, 1158644, 1158645, 1158648, 1158649, 1158652, 1158653, 1158656, 1158657, 1158660, 1158661, 1158662, 1158663, 1158672, 1158673, 1158680, 1158681, 1158682, 1158683, 1158684, 1158685, 1158686, 1158687, 1158688, 1158689, 1158690, 1158691, 1158692, 1158693, 1158694, 1158695, 1158696, 1158698, 1158699, 1165380, 1165381, 1165382, 1165383, 1165384, 1165385, 1165386, 1165387, 1165388, 1165389, 1165390, 1165391, 1165392, 1165393, 1165394, 1165395, 1165396, 1165397, 1165398, 1165399, 1165400, 1165401, 1165402, 1165403, 1165404, 1165405, 1165406, 1165407, 1165408, 1165409, 1165411, 1165412, 1165413, 1165414, 1165415, 1165416, 1165417, 1165418, 1165419, 1165420, 1165421, 1165422, 1165423, 1165424, 1165425, 1165426, 1165427, 1165428, 1165429, 1167519, 1167520, 1167521, 1167522, 1167523, 1167524, 1167525, 1167526, 1167527, 1167528, 1167529, 1167530, 1170320, 1170321, 1170323, 1170324, 1170325, 1170326, 1170327, 1170328, 1170329, 1170330, 1170331, 1170332, 1170333, 1170335, 1170336, 1170337, 1170338, 1170339, 1170341, 1170342, 1170356, 1170357, 1170358, 11
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Smiths Medical ASD Inc.
Reason for Recall:
The CADD Solis VIP Ambulatory Infusion Pump is indicated for the following uses: " For intravenous, intraarterial, subcutaneous, intraperitoneal, perineural, surgical site, epidural space, or subarachnoid space infusion.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

CADD-Solis VIP ambulatory Infusion Pumps, Model 2110. Indicated for the following uses: Intra-arterial, subcutaneous, intraperitoneal, in close proximity to nerves, into an intraoperative site, epidural space or subarachnoid space. Therapies that require a continuous rate of infusion, intermittent bolus, and/or with patient-controlled demand doses.

Product Codes/Lot Numbers:

REF List Number 21-2111-0100-00, UDI/DI 10610586039195, Serial Numbers: 10610586039195; REF List Number 21-2111-0100-51, UDI/DI 30610586039205, Serial Numbers: 1008871, 1011314, 1011695, 1012140, 1012157, 1012387, 1012393, 1012430, 1012607, 1012611, 1012749, 1012751, 1012752, 1012758, 1012765, 1012767, 1012769, 1012773, 1012815, 1012988, 1013182, 1013185, 1013186, 1013187, 1013219, 1013221, 1013223, 1013230, 1013235, 1013237, 1013239, 1013251, 1013256, 1013264, 1013265, 1013278, 1013282, 1013291, 1013295, 1013301, 1013562, 1013817, 1013824, 1013853, 1013866, 1013953, 1013979, 1014000, 1014064, 1014103, 1014117, 1014279, 1014282, 1014288, 1014302, 1014308, 1014447, 1014463, 1016862, 1016876, 1016881, 1014804, 1014806, 1014811, 1014831, 1014844, 1014861, 1014880, 1014895, 1015034, 1015375, 1015382, 1016292, 1016522, 1016709, 1038360, 1016879, 1016927, 1016961, 1023905, 1016989, 1016990, 1017003, 1017005, 1017006, 1017049, 1017051, 1017075, 1017096, 1017143, 1017146, 1017159, 1017161, 1017163, 1017172, 1017173, 1017174, 1017175, 1017180, 1017189, 1017200, 1017202, 1017221, 1017230, 1017237, 1017241, 1017244, 1017252, 1017261, 1017264, 1017267, 1017276, 1017284, 1017286, 1017290, 1017293, 1017294, 1017295, 1017302, 1017325, 1017327, 1017332, 1017333, 1017339, 1017347, 1017359, 1017407, 1017431, 1017433, 1017435, 1017436, 1017444, 1017447, 1017448, 1017455, 1017463, 1017471, 1017486, 1017488, 1017491, 1017493, 1017508, 1017509, 1017511, 1017512, 1017514, 1017516, 1017518, 1017522, 1017525, 1017532, 1017788, 1017970, 1018001, 1018005, 1018018, 1018048, 1018057, 1018062, 1018095, 1018115, 1018222, 1018286, 1018303, 1018339, 1018376, 1018387, 1018393, 1018415, 1018434, 1018528, 1018538, 1018561, 1018620, 1019029, 1019103, 1019208, 1019262, 1019051, 1020066, 1020067, 1020108, 1020109, 1020110, 1020111, 1020125, 1020128, 1020181, 1020185, 1020197, 1020205, 1020210, 1020221, 1026697, 1026698, 1026702, 1026709, 1026828, 1020836, 1020866, 1021919, 1021943, 1021978, 1022324, 1022346, 1022354, 1022368, 1022396, 1022414, 1022449, 1022467, 1022483, 1022502, 1022511, 1022532, 1022541, 1022751, 1022852, 1022864, 1022971, 1020421, 1023190, 1023192, 1023203, 1023206, 1023213, 1023364, 1023409, 1023449, 1023458, 1023460, 1023476, 1023630, 1023633, 1023673, 1023712, 1023763, 1022468, 1022471, 1023830, 1023839, 1023884, 1023912, 1024043, 1024050, 1024062, 1024070, 1024092, 1024099, 1024113, 1024120, 1024126, 1024146, 1024163, 1024460, 1024497, 1024567, 1025085, 1025086, 1025089, 1025093, 1025094, 1025095, 1025096, 1025099, 1025100, 1025117, 1025770, 1025776, 1025781, 1025792, 1025798, 1025802, 1025819, 1025834, 1025839, 1026270, 1026357, 1026687, 1017622, 1026851, 1026868, 1026873, 1026891, 1026914, 1026917, 1026920, 1026927, 1026931, 1026934, 1026971, 1026979, 1026982, 1026997, 1026999, 1027005, 1027006, 1027884, 1027900, 1027901, 1027906, 1027907, 1027908, 1027910, 1027913, 1027919, 1027920, 1027921, 1027922, 1027923, 1027925, 1027927, 1027952, 1027981, 1027983, 1027984, 1027985, 1028419, 1028549, 1028554, 1028596, 1028667, 1028668, 1028670, 1028673, 1028675, 1028678, 1028683, 1028685, 1028686, 1028687, 1028689, 1028691, 1028696, 1028699, 1028972, 1029018, 1029019, 1029023, 1029027, 1029051, 1029055, 1029492, 1029508, 1029515, 1029518, 1029549, 1029562, 1029567, 1029570, 1029571, 1029701, 1029703, 1029766, 1029770, 1029782, 1029799, 1029803, 1030407, 1030479, 1030483, 1030621, 1030626, 1030645, 1030648, 1031533, 1039351, 1031625, 1031644, 1014448, 1014452, 1032265, 1032403, 1032416, 1032428, 1032454, 1032456, 1032459, 1032514, 1032518, 1032529, 1032532, 1032537, 1032546, 1032551, 1032552, 1032655, 1032698, 1032719, 1032723, 1032724, 1032725, 1032727, 1032731, 1032745, 1032775, 1032776, 1032784, 1032795, 1032816, 1032846, 1032847, 1032850, 1032853, 1032875, 1032891, 1032895, 1032942, 1032947, 1032948, 1032949, 1032959, 1033109, 1023205, 1035367, 1035992, 1036003, 1036018, 1036082, 1036089, 1036092, 1036095, 1036096, 1036099, 1036105, 1036106, 1036107, 1036114, 1036116, 1036118, 1036129, 1036132, 1036134, 1036146, 1036148, 1036149, 1036157, 1036160, 1036166, 1036169, 1036173, 1036295, 1036305, 1036393, 1036415, 1036465, 1036472, 1036473, 1036493, 1036714, 1036716, 1036839, 1036956, 1036961, 1037405, 1037411, 1037413, 1037422, 1037423, 1037457, 1037474, 1038197, 1038213, 1038215, 1038224, 1038225, 1038238, 1038303, 1038309, 1038316, 1038323, 1038339, 1038340, 1038342, 1038343, 1038345, 1038347, 1038356, 1038362, 1038369, 1038415, 1038427, 1038476, 1038487, 1038561, 1038611, 1038784, 1039042, 1039047, 1039061, 1039084, 1039103, 1039115, 1039130, 1039324, 1039344, 1039352, 1039366, 1039369, 1039370, 1039379, 1039382, 1039385, 1039389, 1039391, 1039392, 1039393, 1039394, 1039396, 1039397, 1039400, 1039401, 1039407, 1039409, 1039412, 1039413, 1039415, 1039416, 1039417, 1039671, 1039673, 1039674, 1039676, 1039679, 1039682, 1039685, 1039686, 1039691, 1039697, 1039704, 1039708, 1039709, 1039713, 1039717, 1039720, 1039722, 1039723, 1039724, 1039725, 1039726, 1039728, 1039729, 1039731, 1040253, 1040263, 1040305, 1040513, 1040528, 1040555, 1040768, 1041265, 1041595, 1041597, 1041602, 1041972, 1042000, 1044123, 1044135, 1044137, 1044140, 1044148, 1044166, 1044167, 1044169, 1044204, 1044231, 1044329, 1044390, 1044710, 1044720, 1044727, 1044730, 1044741, 1044745, 1044762, 1044767, 1044786, 1044789, 1014874, 1044795, 1014875, 1047279, 1047358, 1017344, 1029477, 1029565, 1026631, 1026639, 1014480, 1028136, 1028139, 1039190, 1039191, 1024398, 1024399, 1024401, 1024406, 1022713, 1012639, 1015834, 1016051, 1022374, 1014677, 1014691, 1020369, 1068313, 1013445, 1036958, 1032283, 1032310, 1025074, 1027934, 1014813, 1039323, 1015791, 1023366, 1020683, 1020688, 1020704, 1020723, 1020729, 1033415, 1014833, 1014866, 1028304, 1013435, 1035554, 1047264, 1148607, 1016830, 1016831, 1016834, 1016839, 1016845, 1016851, 1016854, 1016855, 1042943, 1098608, 1020787, 1020025, 1020027, 1020227, 1020302, 1020303, 1012992, 1012597, 1030643, 1030644, 1014963, 1014969, 1014975, 1018757, 1018791, 1035931, 1035939, 1037398, 1041610, 1035377, 1024574, 1028705, 1015324, 1014422, 1014430, 1030566, 1030580, 1030582; REF List Number 21-2111-0200-02, UDI/DI 30610586039397, Serial Numbers: 1012544, 1024990, 1024992, 1013501; REF List Number 21-2111-0200-03, UDI/DI 10610586039416, Serial Numbers: 1015056, 1015057, 1015058, 1015059, 1015060, 1015061, 1017665, 1017675, 1017692, 1017693, 1017695, 1017696, 1017704, 1017705, 1017707, 1021927, 1021928, 1021933, 1021936, 1021937, 1022496, 1022507, 1022521, 1022952, 1022954, 1022967, 1022978, 1022985, 1022988, 1022999, 1023000, 1023003, 1022970, 1016119, 1014632, 1016437, 1016447, 1016470, 1017741, 1017749, 1015964, 1022005, 1022213, 1022428, 1022497, 1022685, 1022705, 1023147, 1023318, 1023703, 1023704, 1023705, 1023706, 1023713, 1023714, 1023718, 1023721, 1023730, 1023764, 1023767, 1023776, 1023778, 1023780, 1023782, 1023801, 1023815, 1024199, 1024200, 1024201, 1024202, 1024218, 1024219, 1024223, 1024224, 1024225, 1024230, 1024232, 1024233, 1024234, 1016471, 1024690, 1024691, 1025322, 1025344, 1025346, 1029435, 1015489, 1022032, 1013091, 1024665, 1024666, 1024704, 1024705, 1024706, 1024707, 1024708, 1024709, 1024710, 1024712, 1024714, 1024715, 1024716, 1024718, 1024719, 1024720, 1024721, 1024722, 1024723, 1024724, 1024725, 1016343, 1025324, 1021442, 1021443, 1021444, 1021448, 1021511, 1114780, 1022175, 1017759, 1015065; REF List Number 21-2111-0200-06, UDI/DI 10610586039430, Serial Numbers: 1021051, 1021052, 1021053, 1021054, 1021055, 1021056, 1021057, 1021058, 1021512, 1021513; REF List Number 21-2111-0300-00, UDI/DI 10610586038792, Serial Numbers: 1047706, 1078090, 1078327, 1078337; REF List Number 21-2111-0300-01, UDI/DI 10610586038778, Serial Numbers: 1003302, 1003614, 1005310, 1005321, 1011210, 1012744, 1014286, 1063228, 1014808, 1014877, 1016779, 1017035, 1017038, 1017156, 1017169, 1017178, 1017182, 1017183, 1017184, 1017185, 1017192, 1017194, 1017204, 1017239, 1017245, 1017247, 1017250, 1017251, 1017257, 1017279, 1017280, 1017287, 1017311, 1017314, 1017318, 1017320, 1017324, 1017328, 1017329, 1017338, 1017345, 1017351, 1017353, 1017355, 1017428, 1017439, 1017440, 1017446, 1017451, 1017454, 1017458, 1017459, 1017464, 1017469, 1017470, 1017479, 1017480, 1017482, 1017483, 1017489, 1017494, 1017497, 1017499, 1017500, 1017502, 1017519, 1017524, 1017531, 1052696, 1052700, 1097947, 1018234, 1084235, 1085310, 1079869, 1067819, 1124242, 1124338, 1022275, 1022437, 1085055, 1024111, 1098821, 1058329, 1028578, 1064986, 1029524, 1099542, 1032412, 1099587, 1032922, 1032983, 1099637, 1099645, 1099659, 1084423, 1075642, 1072386, 1053632, 1053633, 1051244, 1052021, 1053407, 1053532, 1053535, 1053546, 1053547, 1053592, 1053601, 1053620, 1053626, 1053627, 1053656, 1053662, 1053663, 1053736, 1053737, 1053743, 1053745, 1053749, 1053750, 1036147, 1058053, 1153346, 1038227, 1038897, 1038908, 1076129, 1065016, 1065043, 1040035, 1075430, 1075448, 1075838, 1044129, 1044132, 1044141, 1044144, 1044145, 1044146, 1044151, 1044156, 1044157, 1044165, 1044170, 1067147, 1067148, 1044895, 1069460, 1046788, 1046794, 1046803, 1046812, 1046814, 1059831, 1059850, 1048771, 1047821, 1017468, 1017515, 1048248, 1048251, 1048259, 1118271, 1049473, 1049517, 1050421, 1050436, 1050451, 1044937, 1050776, 1050809, 1050972, 1044979, 1051126, 1045019, 1051249, 1051250, 1051252, 1051253, 1051254, 1051255, 1051256, 1051259, 1051260, 1051261, 1051262, 1051263, 1051264, 1051266, 1051267, 1051268, 1051269, 1051270, 1051271, 1051272, 1051273, 1051274, 1051275, 1051277, 1051278, 1051279, 1051280, 1051281, 1051283, 1051284, 1051285, 1051287, 1051288, 1051290, 1051291, 1051293, 1051295, 1051297, 1051304, 1051305, 1051306, 1051307, 1051425, 1051427, 1051428, 1051429, 1051431, 1051432, 1051433, 1051434, 1051435, 1051436, 1051437, 1051438, 1051439, 1051440, 1051442, 1051443, 1051445, 1051446, 1051447, 1051448, 1051451, 1051452, 1051453, 1051454, 1051455, 1051926, 1052073, 1052235, 1052683, 1052746, 1052748, 1052757, 1052816, 1052853, 1052860, 1052861, 1052875, 1052883, 1052886, 1052901, 1052902, 1052908, 1052910, 1052912, 1052915, 1052922, 1052925, 1053091, 1053093, 1053099, 1053100, 1053107, 1097346, 1053279, 1053321, 1053324, 1053325, 1053326, 1053328, 1053329, 1053333, 1053334, 1053335, 1053336, 1053338, 1053339, 1053340, 1053341, 1053343, 1053344, 1053345, 1053398, 1053402, 1053436, 1053506, 1053522, 1053527, 1053541, 1053579, 1053585, 1053590, 1053597, 1053600, 1053602, 1053617, 1053621, 1053623, 1053649, 1053741, 1053742, 1053747, 1054688, 1056254, 1056297, 1056304, 1056312, 1056558, 1056633, 1056714, 1098405, 1098407, 1058045, 1058120, 1058178, 1087970, 1058209, 1087995, 1088004, 1093056, 1058255, 1058266, 1058282, 1058296, 1058310, 1058344, 1097533, 1097547, 1097549, 1097550, 1097562, 1097681, 1059241, 1059245, 1048031, 1060639, 1060951, 1060961, 1060962, 1060968, 1060974, 1060978, 1060983, 1060988, 1060989, 1045060, 1057816, 1057837, 1058336, 1058356, 1058364, 1062359, 1062121, 1062122, 1062124, 1062126, 1062230, 1062241, 1062243, 1062251, 1069560, 1069563, 1069564, 1069566, 1069570, 1053455, 1053463, 1053470, 1056327, 1048353, 1048355, 1048361, 1048365, 1088029, 1064771, 1064854, 1064891, 1064893, 1064905, 1064929, 1064989, 1065101, 1065107, 1065211, 1065212, 1119182, 1065719, 1066323, 1066324, 1066326, 1066327, 1066328, 1066330, 1066331, 1066332, 1066333, 1067046, 1067095, 1067096, 1067097, 1097918, 1097920, 1079183, 1079251, 1079379, 1079524, 1079568, 1079623, 1079916, 1080447, 1080449, 1080450, 1080455, 1080465, 1080515, 1080524, 1080525, 1080527, 1080528, 1080530, 1080531, 1080532, 1080574, 1080577, 1080580, 1080581, 1080583, 1080588, 1080591, 1080592, 1080611, 1080612, 1080621, 1080622, 1080624, 1080625, 1080629, 1080632, 1080634, 1080635, 1080638, 1080640, 1080643, 1080651, 1080652, 1068739, 1068753, 1068781, 1068800, 1097035, 1069154, 1069157, 1069160, 1069164, 1098645, 1098848, 1098857, 1069234, 1098858, 1098859, 1069241, 1069242, 1069243, 1098861, 1069250, 1069251, 1098879, 1098892, 1098907, 1099828, 1099836, 1069509, 1069510, 1069512, 1069514, 1069515, 1069516, 1069520, 1069532, 1069534, 1099099, 1065924, 1087862, 1099037, 1099042, 1099059, 1099063, 1099065, 1099066, 1099071, 1099073, 1099076, 1166599, 1070071, 1070325, 1097913, 1070868, 1070869, 1070984, 1071072, 1071097, 1071098, 1071321, 1071411, 1071442, 1071460, 1071480, 1071612, 1071622, 1071783, 1071830, 1071853, 1071877, 1072020, 1072031, 1072062, 1072063, 1072064, 1072074, 1072098, 1072124, 1072228, 1072300, 1072365, 1072401, 1093374, 1072483, 1072493, 1072528, 1072674, 1072722, 1072729, 1072730, 1072751, 1072755, 1072772, 1073048, 1073073, 1073097, 1073105, 1073118, 1073141, 1073149, 1073399, 1073413, 1073430, 1073568, 1073570, 1015795, 1074432, 1065086, 1074808, 1074809, 1074810, 1074988, 1075271, 1075368, 1075385, 1075529, 1075605, 1075606, 1075607, 1075609, 1075610, 1075611, 1075612, 1075613, 1075626, 1075753, 1075799, 1075949, 1075952, 1075956, 1075958, 1075959, 1075961, 1076011, 1076019, 1076022, 1099414, 1099416, 1099417, 1072996, 1080526, 1080620, 1080678, 1080856, 1080878, 1080896, 1080930, 1080932, 1080934, 1080938, 1080939, 1080940, 1080943, 1080946, 1080960, 1080966, 1080967, 1081017, 1081019, 1081020, 1081021, 1081023, 1081024, 1081025, 1081026, 1081027, 1081028, 1081057, 1081083, 1081089, 1081093, 1081094, 1081573, 1081583, 1081649, 1081840, 1081876, 1088380, 1088381, 1088382, 1088383, 1074825, 1065494, 1064435, 1097314, 1083622, 1083637, 1083643, 1134952, 1084194, 1084204, 1084218, 1084223, 1084367, 1084417, 1084487, 1084503, 1084508, 1084519, 1084525, 1084599, 1084608, 1084609, 1084625, 1084630, 1084637, 1097527, 1084859, 1084863, 1051289, 1051294, 1051430, 1057941, 1057942, 1086551, 1124156, 1075404, 1088398, 1087643, 1087645, 1087646, 1087648, 1087685, 1087692, 1088000, 1088081, 1088090, 1088478, 1088491, 1092992, 1095249, 1096819, 1097067, 1097076, 1097088, 1097093, 1097096, 1097110, 1097115, 1097130, 1097135, 1097206, 1097351, 1097600, 1097603, 1097607, 1097613, 1097684, 1097711, 1097715, 1097721, 1097743, 1097746, 1097788, 1097854, 1097862, 1097863, 1097865, 1097866, 1097869, 1097884, 1097888, 1097892, 1097894, 1097905, 1097910, 1097928, 1097930, 1097944, 1097946, 1097948, 1097950, 1097953, 1097957, 1097960, 1097983, 1098194, 1098196, 1098205, 1098272, 1098278, 1098284, 1098290, 1098307, 1098308, 1098310, 1098315, 1098317, 1098322, 1098323, 1098327, 1098336, 1098340, 1098342, 1098348, 1098352, 1098362, 1098365, 1098368, 1098379, 1098381, 1098382, 1098384, 1098480, 1098569, 1098570, 1098572, 1098575, 1098577, 1098581, 1098584, 1098586, 1098587, 1098590, 1098591, 1098603, 1098630, 1098674, 1098832, 1098839, 1098841, 1098845, 1048468, 1076093, 1076105, 1076118, 1072955, 1072960, 1099466, 1099467, 1099481, 1099482, 1099483, 1099484, 1099485, 1099486, 1099488, 1099490, 1099876, 1099877, 1099879, 1099882, 1099886, 1099889, 1099895, 1099901, 1099903, 1099904, 1100014, 1100024, 1100192, 1100206, 1100209, 1100212, 1100214, 1100221, 1100229, 1100232, 1100235, 1100242, 1100247, 1100250, 1100256, 1100261, 1100397, 1101971, 1101972, 1101976, 1101977, 1101981, 1101985, 1101987, 1101989, 1101992, 1101994, 1102001, 1102002, 1102004, 1102005, 1102006, 1102008, 1088578, 1080137, 1080140, 1080542, 1080553, 1080566, 1119270, 1088725, 1088750, 1088763, 1044685, 1014976, 1097468, 1097479, 1075923, 1115859, 1115882, 1115883, 1115887, 1115896, 1117327, 1117371, 1117374, 1117392, 1117930, 1117937, 1117941, 1117945, 1117955, 1117958, 1117976, 1117981, 1117982, 1117989, 1117990, 1117996, 1117999, 1118001, 1118004, 1118008, 1118010, 1118012, 1118013, 1118018, 1118026, 1118029, 1118033, 1118035, 1118036, 1118037, 1118038, 1118039, 1118040, 1118047, 1118100, 1118101, 1118102, 1118111, 1118112, 1118118, 1118126, 1118127, 1118129, 1118131, 1118134, 1118136, 1118137, 1118140, 1118156, 1118170, 1118183, 1118184, 1118196, 1118208, 1118213, 1118214, 1118216, 1118228, 1118229, 1118230, 1118235, 1118239, 1118242, 1118243, 1118244, 1118245, 1118250, 1118253, 1118255, 1118281, 1118282, 1118286, 1118292, 1118298, 1118300, 1118309, 1118310, 1118315, 1118321, 1118323, 1118327, 1118344, 1118347, 1118348, 1118357, 1118358, 1118367, 1118372, 1118374, 1118379, 1118381, 1118383, 1118386, 1118388, 1118389, 1118393, 1118395, 1118402, 1118403, 1118409, 1118410, 1118411, 1118413, 1118416, 1118419, 1118421, 1118423, 1118424, 1118428, 1118431, 1118438, 1118439, 1118441, 1118443, 1118826, 1118908, 1119227, 1119309, 1119323, 1119332, 1119336, 1119337, 1119339, 1119402, 1119407, 1069523, 1119539, 1119552, 1119562, 1119720, 1084420, 1119834, 1119836, 1119862, 1119870, 1119915, 1119940, 1119952, 1119957, 1119985, 1084962, 1084964, 1084965, 1120322, 1120328, 1120329, 1120833, 1120834, 1120844, 1120847, 1120848, 1120849, 1120850, 1120852, 1120854, 1120855, 1120905, 1120915, 1120919, 1120923, 1120926, 1120928, 1120932, 1120938, 1120958, 1120960, 1120962, 1120964, 1120965, 1120967, 1120969, 1120970, 1120987, 1121013, 1121184, 1121199, 1121216, 1121219, 1121272, 1121282, 1121292, 1121336, 1121340, 1121343, 1121347, 1121350, 1121353, 1121356, 1121360, 1121361, 1121373, 1121381, 1121386, 1121387, 1121389, 1121394, 1121611, 1121621, 1123858, 1072917, 1124104, 1124105, 1124177, 1124186, 1124196, 1124334, 1124335, 1062991, 1063034, 1124913, 1063220, 1063221, 1064467, 1040747, 1048664, 1052680, 1052742, 1052773, 1052881, 1052884, 1052904, 1052943, 1052958, 1133649, 1133685, 1134965, 1135183, 1053005, 1053040, 1053110, 1053143, 1143564, 1145868, 1145876, 1099777, 1153404, 1153333, 1153409, 1153410, 1153412, 1153419, 1153421, 1153423, 1153424, 1153426, 1058423, 1058425, 1058426, 1058431, 1051257, 1058466, 1058483, 1069819, 1170008, 1170116, 1170731, 1170923, 1030577, 1121364, 1049006, 1049014, 1049015, 1049018, 1049010, 1053674, 1054137, 1054206, 1054210, 1054224, 1054942, 1054946, 1054953, 1097254, 1097289, 1098000; REF List Number 21-2111-0300-02, UDI/DI 10610586038891, Serial Numbers: 1047659, 1079778, 1079880, 1079927, 1079935, 1080039, 1080296, 1080404, 1080405, 1080705, 1080706, 1080707, 1080709, 1080774, 1080789, 1080793, 1080794, 1041790, 1041791, 1041792, 1078019, 1047649, 1077673, 1094561, 1049499, 1041785, 1041786, 1041787, 1041788, 1041789, 1089591, 1089597, 1089599, 1092736, 1114614, 1114626, 1114628, 1050236, 1050239, 1050240, 1081319, 1081320, 1081321, 1081322, 1081323, 1081325, 1081326, 1081327, 1024974, 1033849, 1033692, 1033694, 1033889, 1036629; REF List Number 21-2111-0300-03, UDI/DI 10610586038938, Serial Numbers: 1025549, 1025550, 1028173, 1028175, 1034021, 1035138, 1035139, 1082157, 1102976, 1113445, 1118881, 1118882, 1119022, 1119023, 1119024, 1119025, 1119026, 1119027, 1119028, 1119030, 1119031, 1119564, 1119565, 1119566, 1119567, 1119568, 1119569, 1119570, 1119572, 1117486, 1080749, 1045084, 1025533, 1032098, 1055691, 1025458, 1025459, 1123409, 1123411, 1123412, 1123413, 1029649, 1029912, 1030262, 1030731, 1031035, 1031357, 1031363, 1031366, 1031367, 1031368, 1031376, 1032099, 1033665, 1033740, 1033741, 1034000, 1035135, 1035143, 1035145, 1035146, 1035150, 1035155, 1035173, 1119016, 1119651, 1037891, 1082255, 1113431, 1045226, 1111513, 1115756, 1115760, 1115794, 1091538, 1091541, 1091546, 1117701, 1129158, 1024341, 1030248, 1030265, 1070159, 1070160, 1070161, 1070162, 1100852, 1100853, 1100863, 1047507, 1048346, 1048837, 1049242, 1050292, 1054652, 1054674, 1054675, 1054676, 1055274, 1031375, 1032094, 1137853, 1108890, 1055267, 1055266, 1087323, 1025450, 1032105, 1080693, 1080739, 1066557, 1047509, 1048056, 1048057, 1048058, 1048059, 1048060, 1048061, 1048063, 1048064, 1048065, 1048066, 1048067, 1131196, 1080741, 1080742, 1080743, 1080745, 1080747, 1080748, 1080750, 1080751, 1080752, 1080831, 1080832, 1080833, 1080834, 1080835, 1080836, 1080838, 1080839, 1080840, 1081270, 1081271, 1081294, 1081300, 1082350, 1082346, 1082347, 1033613, 1031636, 1047506, 1089718, 1093465, 1100859, 1102972, 1103051, 1103054, 1103059, 1108050, 1111633, 1112409, 1092340, 1095118, 1115246, 1115260, 1117473, 1117483, 1117718, 1117724, 1117739, 1117744, 1100856, 1102453, 1102459, 1102665, 1102668, 1102669, 1119013, 1119599, 1119647, 1119648, 1119660, 1041859, 1123380, 1123388, 1126634, 1129179, 1131195, 1131197, 1131323, 1135493, 1137683, 1144640, 1149550, 1150592, 1155075, 1024977, 1035237, 1035238, 1034018, 1092337, 1052027, 1133743, 1133748, 1045154, 1045096, 1037544, 1033609, 1114853, 1033614; REF List Number 21-2111-0300-06, UDI/DI 10610586038969, Serial Numbers: 1006546, 1032061, 1036631, 1007059, 1074596, 1074597, 1074598, 1074599, 1074600, 1074601, 1074602, 1074603, 1074604, 1074605, 1115720, 1115721, 1115722, 1115723, 1115724, 1016516; REF List Number 21-2111-0300-08, UDI/DI 10610586039010, Serial Numbers: 1046527, 1110009, 1110020, 1155180, 1157317, 1163923; REF List Number 21-2111-0300-14, UDI/DI 10610586039065, Serial Numbers: 1115208; REF List Number 21-2111-0300-50, UDI/DI 10610586038839, Serial Numbers: 1037514, 1066478, 1066785, 1066791, 1066815, 1027152, 1027157, 1028030; REF List Number 21-2111-0400-51, UDI/DI 10610586041037, Serial Numbers: 1118744, 1108192; REF List Number 21-2111-0401-02L, UDI/DI 35019517135323, Serial Numbers: 1001723, 1004057, 1006543, 1159374, 1177517, 1159408, 1159409, 1152299, 1152300, 1152301, 1152302, 1152303, 1152304, 1152305, 1152306, 1152307, 1152308, 1152309, 1152310, 1152311, 1152312, 1152313, 1152314, 1152315, 1152316, 1152317, 1152318, 1152319, 1152320, 1152321, 1152322, 1152323, 1152324, 1152325, 1152326, 1152327, 1152328, 1152329, 1152330, 1152331, 1152332, 1152333, 1152334, 1152335, 1152336, 1152337, 1152338, 1152339, 1152340, 1152341, 1152342, 1152343, 1152344, 1152345, 1152346, 1152435, 1152436, 1153523, 1153525, 1153526, 1153527, 1153528, 1153529, 1153530, 1153531, 1153532, 1153533, 1153534, 1153535, 1153536, 1153537, 1153538, 1153539, 1153540, 1153541, 1153543, 1153544, 1153545, 1153546, 1153547, 1153548, 1153549, 1153550, 1153551, 1153552, 1153553, 1153554, 1153555, 1153556, 1153557, 1153558, 1153559, 1153560, 1153561, 1153562, 1153563, 1153564, 1153565, 1153566, 1153567, 1153568, 1153569, 1153570, 1153571, 1153572, 1153573, 1153574, 1153575, 1153576, 1153577, 1153578, 1153579, 1153580, 1153581, 1153582, 1153583, 1153584, 1153585, 1153586, 1153587, 1153588, 1153589, 1153590, 1153591, 1153592, 1153593, 1153594, 1153595, 1153596, 1153597, 1153598, 1153599, 1153600, 1153601, 1153602, 1155290, 1155291, 1155292, 1155293, 1155294, 1155295, 1155296, 1155297, 1155298, 1155299, 1155300, 1155301, 1155302, 1155303, 1155304, 1155305, 1155306, 1155307, 1155309, 1155310, 1155311, 1155312, 1155313, 1155314, 1155315, 1155316, 1155317, 1155319, 1155320, 1155321, 1155322, 1155323, 1155324, 1155325, 1155326, 1155327, 1155328, 1155329, 1155330, 1155331, 1155332, 1155333, 1155334, 1155335, 1155336, 1155338, 1155339, 1155340, 1155341, 1155343, 1155344, 1155345, 1155346, 1155347, 1155348, 1155349, 1155350, 1155351, 1155352, 1155353, 1155355, 1155356, 1155357, 1155358, 1155359, 1155360, 1155361, 1155362, 1155363, 1155364, 1155365, 1155366, 1155367, 1155368, 1155369, 1155370, 1159327, 1159328, 1159329, 1159330, 1159331, 1159332, 1159333, 1159334, 1159335, 1159336, 1159337, 1159338, 1159340, 1159341, 1159342, 1159343, 1159344, 1159345, 1159346, 1159347, 1159348, 1159349, 1159350, 1159351, 1159352, 1159353, 1159354, 1159355, 1159357, 1159358, 1159359, 1159360, 1159361, 1159362, 1159363, 1159364, 1159365, 1159366, 1159368, 1159369, 1159370, 1159371, 1159372, 1159373, 1159375, 1159376, 1159377, 1159378, 1159379, 1159380, 1159383, 1159384, 1159385, 1159386, 1159387, 1159388, 1159389, 1159390, 1159391, 1159392, 1159393, 1159395, 1159396, 1159397, 1159398, 1159399, 1159400, 1159401, 1159402, 1159403, 1159404, 1159405, 1159406, 1159407, 1159410, 1159411, 1159412, 1159413, 1159414, 1159415, 1159416, 1159417, 1159418, 1159419, 1159420, 1159421, 1159422, 1159423, 1159424, 1159425, 1159426, 1159427, 1159428, 1159429, 1159430, 1159431, 1159432, 1159433, 1159434, 1159435, 1159436, 1159437, 1159438, 1159439, 1159440, 1159441, 1159775, 1159779, 1159780, 1159781, 1159782, 1159784, 1159785, 1159786, 1159787, 1159788, 1159789, 1159790, 1159791, 1159792, 1159793, 1159794, 1159795, 1159796, 1159797, 1159798, 1159799, 1159800, 1159802, 1159803, 1159804, 1159805, 1159806, 1159807, 1159808, 1159809, 1159810, 1159811, 1159812, 1159813, 1159814, 1159815, 1159816, 1159817, 1159818, 1159819, 1159820, 1159821, 1159822, 1159823, 1159824, 1159825, 1159827, 1159828, 1159829, 1159831, 1159832, 1159833, 1159834, 1159835, 1159836, 1159837, 1159838, 1159839, 1159840, 1159842, 1159843, 1159844, 1159845, 1159848, 1159849, 1159850, 1159852, 1159853, 1159854, 1159855, 1159856, 1159857, 1159858, 1159859, 1087064, 1160254, 1160259, 1160260, 1160261, 1160262, 1160263, 1160264, 1160265, 1160266, 1160267, 1160268, 1160269, 1160270, 1160271, 1160272, 1160273, 1160274, 1160275, 1160276, 1160277, 1160278, 1160279, 1160280, 1160281, 1160283, 1164110, 1164125, 1164126, 1164127, 1164128, 1164129, 1164130, 1164131, 1164132, 1164133, 1164134, 1164135, 1164136, 1164137, 1164138, 1164139, 1164140, 1164141, 1164142, 1164143, 1164144, 1164145, 1164146, 1164147, 1164148, 1164149, 1164150, 1164151, 1164152, 1164153, 1164154, 1164155, 1164156, 1164157, 1164158, 1164159, 1164160, 1164161, 1164162, 1164163, 1164164, 1164165, 1164166, 1164167, 1164168, 1164169, 1164170, 1164171, 1164172, 1164173, 1164174, 1164175, 1164176, 1164177, 1164178, 1164179, 1164180, 1164181, 1164182, 1164183, 1164184, 1164185, 1164186, 1164187, 1164188, 1164189, 1164190, 1164191, 1164192, 1164193, 1164194, 1165081, 1167531, 1167532, 1167533, 1167534, 1167535, 1167536, 1167537, 1167538, 1167539, 1167540, 1167541, 1167542, 1167543, 1167544, 1167545, 1167546, 1167557, 1167558, 1167559, 1167560, 1167561, 1167562, 1167563, 1167564, 1167565, 1167566, 1167567, 1167568, 1167569, 1167570, 1170552, 1170553, 1170554, 1170555, 1170556, 1170557, 1170558, 1170559, 1170560, 1170561, 1170562, 1170563, 1170564, 1170566, 1170567, 1170571, 1170572, 1170573, 1170574, 1170575, 1170576, 1170578, 1170580, 1170581, 1170582, 1170583, 1170584, 1170585, 1170586, 1170587, 1170589, 1170802, 1170803, 1170805, 1170806, 1170807, 1170808, 1170809, 1170810, 1170811, 1170812, 1170813, 1170815, 1170816, 1170817, 1170818, 1170819, 1170820, 1170821, 1170822, 1170823, 1170824, 1170826, 1170827, 1170830, 1170831, 1170832, 1170833, 1170834, 1170835, 1170836, 1170837, 1170838, 1170839, 1170840, 1170841, 1170843, 1170844, 1170846, 1170847, 1170848, 1170849, 1170850, 1170853, 1170854, 1170859, 1170860, 1170862, 1170865, 1170904, 1176830, 1176831, 1176873, 1176875, 1176876, 1176877, 1176879, 1176880, 1176881, 1176882, 1176883, 1176884, 1176885, 1176886, 1176887, 1176897, 1176898, 1176899, 1176900, 1176901, 1176902, 1176903, 1176904, 1176905, 1176906, 1176907, 1176908, 1176909, 1176910, 1176911, 1176912, 1176913, 1176914, 1176915, 1176916, 1176917, 1176918, 1176919, 1176920, 1176921, 1176922, 1176923, 1176924, 1176925, 1176926, 1176927, 1176928, 1176929, 1177230, 1177237, 1177239, 1177241, 1177243, 1177245, 1177246, 1177247, 1177248, 1177249, 1177250, 1177251, 1177253, 1177254, 1177255, 1177257, 1177258, 1177259, 1177260, 1177261, 1177262, 1177312, 1177360, 1177361, 1177362, 1177367, 1177368, 1177369, 1177370, 1177371, 1177372, 1177373, 1177374, 1177375, 1177376, 1177377, 1177378, 1177379, 1177380, 1177381, 1177382, 1177383, 1177384, 1177385, 1177386, 1177387, 1177388, 1177389, 1177390, 1177391, 1177392, 1177393, 1177394, 1177395, 1177464, 1177475, 1177476, 1177477, 1177478, 1177479, 1177480, 1177481, 1177482, 1177483, 1177484, 1177485, 1177486, 1177487, 1177488, 1177489, 1177490, 1177491, 1177492, 1177493, 1177494, 1177495, 1177496, 1177497, 1177498, 1177499, 1177500, 1177501, 1177502, 1177503, 1177504, 1177505, 1177506, 1177507, 1177508, 1177509, 1177510, 1177511, 1177512, 1177513, 1177514, 1177515, 1177516, 1177518, 1177519, 1177520, 1177521, 1177522, 1177523, 1177524, 1177525, 1177526, 1177527, 1177528, 1177529, 1177530, 1177553, 1181160, 1181161, 1181162, 1181163, 1181164, 1181420; REF List Number 21-2111-0401-03L, UDI/DI 35019517128677, Serial Numbers: 1013538, 1015946, 1014903, 1014901, 1158620, 1029080, 1029428, 1158595, 1158697, 1181143, 1165410, 1144636, 1150613, 1152280, 1152347, 1152348, 1152349, 1152350, 1152351, 1152352, 1152353, 1152354, 1152355, 1152356, 1152357, 1152358, 1152359, 1152360, 1152361, 1152362, 1152363, 1152364, 1152365, 1152366, 1152367, 1152368, 1152369, 1152370, 1152371, 1152372, 1152373, 1152374, 1152375, 1152376, 1152377, 1152378, 1152379, 1152380, 1152381, 1152382, 1152383, 1152384, 1152385, 1152386, 1152387, 1152388, 1152389, 1152390, 1152391, 1152392, 1152393, 1152394, 1152395, 1152396, 1152397, 1153603, 1153604, 1153605, 1153606, 1153607, 1153608, 1153609, 1153610, 1153611, 1153612, 1153613, 1153614, 1153615, 1153616, 1153617, 1153618, 1153619, 1153620, 1153621, 1153622, 1153623, 1153624, 1153625, 1153626, 1153627, 1153628, 1153629, 1153630, 1153631, 1153632, 1154727, 1154728, 1154729, 1154730, 1154731, 1154732, 1154733, 1154734, 1154735, 1154736, 1154737, 1154738, 1154739, 1155040, 1155041, 1155042, 1155043, 1155044, 1155045, 1155046, 1155047, 1155048, 1155049, 1155050, 1155051, 1155052, 1155053, 1155187, 1155188, 1155189, 1155190, 1155191, 1155192, 1155193, 1155194, 1155195, 1155196, 1155197, 1155198, 1155199, 1155200, 1155201, 1155202, 1155203, 1155204, 1155205, 1155206, 1155207, 1155208, 1155209, 1155210, 1155211, 1155212, 1155213, 1155214, 1155215, 1155216, 1155217, 1155218, 1155219, 1155220, 1155221, 1155222, 1155223, 1155224, 1155225, 1155226, 1155227, 1155228, 1155229, 1155230, 1155231, 1155232, 1155233, 1155234, 1155235, 1155236, 1155237, 1155238, 1155239, 1155240, 1155241, 1155242, 1155243, 1155244, 1155245, 1155246, 1155247, 1155248, 1155249, 1155250, 1155251, 1155252, 1155253, 1155254, 1155255, 1155256, 1155257, 1155258, 1155259, 1155260, 1155261, 1155262, 1155263, 1155264, 1155265, 1155266, 1155505, 1155506, 1155507, 1155508, 1155509, 1155510, 1155511, 1155512, 1155513, 1155514, 1155515, 1155516, 1155517, 1155518, 1155519, 1155520, 1155521, 1155522, 1155523, 1155524, 1155525, 1155526, 1155527, 1155528, 1155529, 1155530, 1155531, 1155532, 1155533, 1155534, 1155535, 1155536, 1155537, 1155539, 1155540, 1155541, 1155542, 1155543, 1155544, 1155545, 1155546, 1155547, 1155548, 1155549, 1155550, 1155551, 1155552, 1155553, 1155554, 1158489, 1158490, 1158491, 1158492, 1158493, 1158494, 1158495, 1158496, 1158497, 1158498, 1158499, 1158500, 1158501, 1158502, 1158503, 1158504, 1158505, 1158506, 1158507, 1158508, 1158509, 1158510, 1158511, 1158512, 1158513, 1158514, 1158515, 1158516, 1158517, 1158518, 1158519, 1158520, 1158521, 1158522, 1158523, 1158524, 1158525, 1158526, 1158527, 1158528, 1158529, 1158530, 1158531, 1158532, 1158533, 1158534, 1158535, 1158536, 1158537, 1158538, 1158539, 1158540, 1158541, 1158542, 1158543, 1158544, 1158545, 1158546, 1158547, 1158548, 1158551, 1158552, 1158553, 1158583, 1158584, 1158585, 1158586, 1158587, 1158588, 1158589, 1158590, 1158591, 1158592, 1158593, 1158594, 1158596, 1158597, 1158598, 1158599, 1158600, 1158601, 1158602, 1158603, 1158604, 1158605, 1158606, 1158607, 1158608, 1158609, 1158610, 1158611, 1158612, 1158613, 1158614, 1158615, 1158616, 1158617, 1158618, 1158619, 1158621, 1158622, 1158623, 1158624, 1158625, 1158626, 1158627, 1158628, 1158629, 1158630, 1158631, 1158632, 1158633, 1158634, 1158635, 1158636, 1158637, 1158638, 1158639, 1158640, 1158641, 1158642, 1158643, 1158644, 1158645, 1158648, 1158649, 1158652, 1158653, 1158656, 1158657, 1158660, 1158661, 1158662, 1158663, 1158672, 1158673, 1158680, 1158681, 1158682, 1158683, 1158684, 1158685, 1158686, 1158687, 1158688, 1158689, 1158690, 1158691, 1158692, 1158693, 1158694, 1158695, 1158696, 1158698, 1158699, 1165380, 1165381, 1165382, 1165383, 1165384, 1165385, 1165386, 1165387, 1165388, 1165389, 1165390, 1165391, 1165392, 1165393, 1165394, 1165395, 1165396, 1165397, 1165398, 1165399, 1165400, 1165401, 1165402, 1165403, 1165404, 1165405, 1165406, 1165407, 1165408, 1165409, 1165411, 1165412, 1165413, 1165414, 1165415, 1165416, 1165417, 1165418, 1165419, 1165420, 1165421, 1165422, 1165423, 1165424, 1165425, 1165426, 1165427, 1165428, 1165429, 1167519, 1167520, 1167521, 1167522, 1167523, 1167524, 1167525, 1167526, 1167527, 1167528, 1167529, 1167530, 1170320, 1170321, 1170323, 1170324, 1170325, 1170326, 1170327, 1170328, 1170329, 1170330, 1170331, 1170332, 1170333, 1170335, 1170336, 1170337, 1170338, 1170339, 1170341, 1170342, 1170356, 1170357, 1170358, 11

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2420-2024

Related Recalls

CADD-Solis HPCA Ambulatory Infusion Pump; Model/Number: 21-2111-0402-09-KYO 21-2111-0402-09-KYO CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP 21-2111-0300-09_LOAN CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR CMSA 21-2111-0300-09_CMSA CADD Solis PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR FUKUOKA 21-2111-0300-09_FUK CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR KYOTO 21-2111-0300-09_KYO CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR OKAYAMA 21-2111-0300-09_OKA CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR OOMIYA 21-2111-0300-09_OMI CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR OSAKA 21-2111-0300-09_OSA CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR SAPPORO 21-2111-0300-09_SAP CADD SOLIS PIB PUMP FOR CLINICAL EVALUATION/DEMO PUMP FOR TOKYO 21-2111-0300-09_TOK CADD_SOLIS_MODEL 2110_ENGLISH_1/EA 21-2111-0200-50JP CADD-Solis VIP CADD-SOLIS V4.2 LOAN FUK 21-2111-0402-09-FUK CADD-SOLIS V4.2 LOAN NAG 21-2111-0402-09-NAG CADD-SOLIS V4.2 LOAN OKA 21-2111-0402-09-OKA CADD-SOLIS V4.2 LOAN OSA 21-2111-0402-09-OSA CADD-SOLIS V4.2 LOAN SAI 21-2111-0402-09-SAI CADD-SOLIS V4.2 LOAN SAP 21-2111-0402-09-SAP CADD-SOLIS V4.2 LOAN TOKE 21-2111-0402-09-TOKE CADD-SOLIS V4.2 LOAN TOKW 21-2111-0402-09-TOKW CADD-SOLIS V4.2 LOAN YELLOW KYO 21-2112-0402-09-KYO CADD-SOLIS V4.2 LOAN YELLOW NAG 21-2112-0402-09-NAG CADD-SOLIS V4.2 LOAN YELLOW YOK ¿ 21-2112-0402-09-YOK CADD-SOLIS V4.2 LOAN YOK 21-2111-0402-09-YOK Number:21-2111-0403-02Part Type:List Number/Finished GoodLifecycle Phase:03-ReleasedDescription:PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, FRENCH, 1/EA 21-2111-0403-02 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, ENGLISH, 1/EA 21-2111-0403-51 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, TRIAL, LOANER, ENGLISH, 1/EA 21-2111-0403-78 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, TRIAL, LOANER, YELLOW, ENGLISH, 1/EA 21-2112-0403-78 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, YELLOW, ENGLISH, 1/EA 21-2112-0403-51 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, YELLOW, FRENCH, 1/EA 21-2112-0403-02 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110 TRIAL/LOANER, NA ENGLISH 1/EA 21-2111-0300-00 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, DANISH 1/EA ¿ 21-2111-0300-12 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, DUTCH 1/EA ¿ 21-2111-0300-06 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, FRENCH 1/EA ¿ 21-2111-0300-02 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, GERMAN 1/EA ¿ 21-2111-0300-03 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, ITALIAN 1/EA ¿ 21-2111-0300-07 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, JAPANESE 21-2111-0300-09JP PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NA ENGLISH 1/EA ¿ 21-2111-0300-01 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NFHU, FRENCH 1/EA ¿ 21-2111-0300-232 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NFHU, NA ENGLISH 1/EA ¿ 21-2111-0300-231 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NFHU, SWEDISH 1/EA 21-2111-0300-244 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, OUS ENGLISH 1/EA ¿ 21-2111-0300-50 Not PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, PORTUGUESE 1/EA ¿ 21-2111-0300-17 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, SPANISH 1/EA ¿ 21-2111-0300-08 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, SWEDISH 1/EA ¿ 21-2111-0300-14 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW KEYPAD, MODEL 2110, NA ENGLISH 1/EA ¿ 21-2112-0300-01 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, DUTCH 1/EA 21-2112-0300-06 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, FRENCH 1/EA ¿ 21-2112-0300-02 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, GERMAN 1/EA 21-2112-0300-03 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, OUS ENGLISH 1/EA 21-2112-0300-50 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110,TRIAL/LOANER, NA ENGLISH 1/EA 21-2112-0300-00 PUMP KIT, CADD-SOLIS HPCA, MDL 2100, DANISH 1/EA ¿ 21-2101-0200-12 PUMP KIT, CADD-SOLIS HPCA, MDL 2100, FRENCH 1/EA ¿ 21-2101-0200-02 PUMP KIT, CADD-SOLIS HPCA, MDL 2100, GERMAN 1/EA ¿ 21-2101-0200-03 PUMP KIT, CADD-SOLIS HPCA, MDL 2100, ITALIAN 1/EA ¿ 21-2101-02

Smiths Medical ASD

Class I - Dangerous

Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.

Apr 10, 2025 Infusion Pumps Nationwide View Details →

Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.

Apr 10, 2025 Infusion Pumps Nationwide View Details →

CADD Solis VIP Ambulatory Infusion Pump Models: 1) PUMP KIT, CADD-SOLIS VIP, MDL 2120, NA ENGLISH, MANUAL MODE, 1/EA, Model Number: 21-2127-0105-01; 2) PUMP KIT, CADD-SOLIS VIP, MDL 2120, NA ENGLISH, PHARMGUARD ENABLED, 1/EA, Model Number: 21-2120-0105-01; 3) PUMP KIT, CADD-SOLIS VIP, MDL 2120, NA ENGLISH 1/EA, Model Number: 21-2120-0103-01; 4) PUMP, CADD-SOLIS VIP, MDL 2120, ENGLISH 1/EA, Model Number: 21-2120-0102-51; 5) PUMP KIT, CADD-SOLIS VIP, MDL 2120, NA ENGLISH, PHARMGUARD ENABLED 1/EA, Model Number: 21-2120-0104-01; 6) PUMP KIT, CADD-SOLIS VIP, MDL 2120, CE GLOBAL ENGLISH 1/EA, Model Number: 21-2120-0103-51; 7) PUMP KIT, CADD-SOLIS VIP, MDL 2120, NA ENGLISH, MANUAL MODE 1/EA, Model Number: 21-2127-0104-01; 8) PUMP KIT, CADD-SOLIS VIP, MDL 2120, TRIAL/LOANER, NA ENGLISH, PHARMGUARD, 1/EA, Model Number: 21-2120-0105-00; 9) PUMP, CADD-SOLIS VIP, MDL 2120, TRIAL/LOANER CE GLOBAL ENGLISH 1/EA, Model Number: 21-2120-0103-78; 10) PUMP, CADD-SOLIS VIP, MDL 2120, ENGLISH, TRIAL/LOANER 1/EA, Model Number: 21-2120-0102-78; 11) PUMP, CADD-SOLIS VIP, MDL 2120, NFHU, NA ENGLISH, PHARMGUARD ENABLED 1/EA, Model Number: 21-2120-0104-231; 12) PUMP, CADD-SOLIS VIP, MDL 2120, ENGLISH, NFHU 1/EA, Model Number: 21-2120-0102-249; 13) PUMP, CADD-SOLIS VIP, MDL 2120, NFHU CE GLOBAL ENGLISH 1/EA, Model Number: 21-2120-0103-249; 14) PUMP KIT, CADD-SOLIS VIP, MDL 2120, NA ENGLISH, FACTORY LIBRARY 1/EA, Model Number: 21-2125-0104-01; 15) PUMP, CADD-SOLIS VIP, MDL 2120, ENGLISH, NFHU 1/EA, Model Number: 21-2120-0102-249.

Smiths Medical ASD

Class I - Dangerous

There is a potential for thermal damage in CADD-Solis and CADD- Solis VIP infusion pumps.

Apr 10, 2025 Infusion Pumps Nationwide View Details →