Medtronic EnTrust Dual Chamber Implantable Cardioverter Defibrillator with Atrial and Ventricular Therapies and RapidRead Telemetry, D154ATG Product Usage: The device indicated for use in ICD patients with atrial tachyarrhythmias, or who are at significant risk of developing atrial tachyarrhythmias. The implantable cardioverter defibrillator is intended to provide ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of life threatening ventricular arrhythmias.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Product UPN/GTIN: 00613994170927 Serial Numbers: PNR606917S PNR606918S PNR606923S Product UPN/GTIN: 00885074299938 Serial Numbers: PNR414990H PNR417905H PNR600110S PNR600121S PNR600123S PNR600136S PNR600141S PNR600147S PNR600148S PNR600157S PNR600163S PNR600166S PNR600176S PNR600177S PNR600183S PNR600189S PNR600190S PNR600192S PNR600194S PNR600195S PNR600198S PNR600203S PNR600290S PNR600306S PNR600381S PNR600382S PNR600383S PNR600385S PNR600387S PNR600388S PNR600389S PNR600390S PNR600392S PNR600393S PNR600396S PNR600397S PNR600398S PNR600401S PNR600402S PNR600404S PNR600405S PNR600406S PNR600407S PNR600409S PNR600410S PNR600412S PNR600415S PNR600416S PNR600417S PNR600418S PNR600419S PNR600422S PNR600423S PNR600424S PNR600425S PNR600426S PNR600427S PNR600428S PNR600429S PNR600432S PNR600433S PNR600434S PNR600435S PNR600436S PNR600437S PNR600438S PNR600439S PNR600441S PNR600442S PNR600443S PNR600448S PNR600450S PNR600452S PNR600453S PNR600454S PNR600455S PNR600458S PNR600459S PNR600461S PNR600462S PNR600463S PNR600464S PNR600503S PNR600504S PNR600505S PNR600506S PNR600510S PNR600511S PNR600545S PNR600548S PNR600549S PNR600551S PNR600552S PNR600553S PNR600554S PNR600555S PNR600557S PNR600561S PNR600562S PNR600563S PNR600564S PNR600565S PNR600566S PNR600570S PNR600572S PNR600576S PNR600577S PNR600578S PNR600581S PNR600583S PNR600585S PNR600599S PNR600600S PNR600601S PNR600603S PNR600604S PNR600609S PNR600610S PNR600614S PNR600615S PNR600616S PNR600619S PNR600620S PNR600621S PNR600623S PNR600624S PNR600626S PNR600629S PNR600631S PNR600731S PNR600733S PNR600770S PNR600773S PNR600775S PNR600781S PNR600784S PNR600786S PNR600789S PNR600790S PNR600792S PNR600798S PNR600801S PNR600802S PNR600804S PNR600811S PNR600813S PNR600816S PNR600818S PNR600819S PNR600821S PNR600824S PNR600849S PNR600886S PNR600887S PNR600891S PNR600892S PNR600904S PNR600933S PNR600948S PNR600949S PNR600950S PNR600953S PNR600956S PNR600957S PNR600958S PNR600964S PNR600978S PNR600979S PNR600987S PNR600990S PNR600996S PNR600997S PNR600998S PNR601017S PNR601029S PNR601031S PNR601036S PNR601044S PNR601045S PNR601046S PNR601054S PNR601056S PNR601058S PNR601060S PNR601061S PNR601063S PNR601070S PNR601075S PNR601076S PNR601079S PNR601080S PNR601082S PNR601083S PNR601089S PNR601090S PNR601097S PNR601102S PNR601109S PNR601113S PNR601115S PNR601116S PNR601117S PNR601118S PNR601120S PNR601121S PNR601124S PNR601125S PNR601126S PNR601128S PNR601129S PNR601131S PNR601133S PNR601134S PNR601136S PNR601137S PNR601140S PNR601149S PNR601150S PNR601151S PNR601154S PNR601155S PNR601164S PNR601170S PNR601174S PNR601178S PNR601180S PNR601195S PNR601197S PNR601199S PNR601201S PNR601203S PNR601206S PNR601210S PNR601212S PNR601216S PNR601225S PNR601235S PNR601248S PNR601250S PNR601329S PNR601351S PNR601354S PNR601364S PNR601372S PNR601373S PNR601375S PNR601391S PNR601393S PNR601395S PNR601406S PNR601416S PNR601419S PNR601426S PNR601431S PNR601441S PNR601444S PNR601446S PNR601447S PNR601451S PNR601452S PNR601453S PNR601472S PNR601484S PNR601487S PNR601524S PNR601528S PNR601546S PNR601548S PNR601550S PNR601555S PNR601559S PNR601564S PNR601567S PNR601570S PNR601572S PNR601573S PNR601574S PNR601582S PNR601584S PNR601590S PNR601591S PNR601594S PNR601603S PNR601609S PNR601617S PNR601620S PNR601624S PNR601628S PNR601645S PNR601646S PNR601647S PNR601651S PNR601652S PNR601653S PNR601661S PNR601662S PNR601674S PNR601676S PNR601684S PNR601685S PNR601690S PNR601693S PNR601709S PNR601710S PNR601712S PNR601714S PNR601716S PNR601742S PNR601748S PNR601750S PNR601763S PNR601765S PNR601821S PNR601824S PNR601825S PNR601834S PNR601836S PNR601852S PNR601854S PNR601860S PNR601863S PNR601871S PNR601875S PNR601880S PNR601884S PNR601888S PNR601939S PNR601940S PNR601948S PNR601952S PNR601955S PNR601956S PNR601957S PNR601958S PNR601959S PNR601962S PNR601963S PNR601964S PNR601967S PNR601968S PNR601972S PNR601974S PNR601975S PNR601978S PNR601981S PNR601982S PNR601983S PNR601985S PNR601986S PNR601987S PNR601988S PNR601991S PNR601993S PNR601994S PNR601995S PNR601996S PNR601997S PNR601998S PNR601999S PNR602001S PNR602004S PNR602005S PNR602008S PNR602049S PNR602089S PNR602091S PNR602094S PNR602098S PNR602099S PNR602101S PNR602102S PNR602106S PNR602107S PNR602108S PNR602110S PNR602112S PNR602113S PNR602114S PNR602115S PNR602120S PNR602126S PNR602138S PNR602139S PNR602152S PNR602166S PNR602174S PNR602177S PNR602178S PNR602192S PNR602193S PNR602194S PNR602196S PNR602202S PNR602205S PNR602242S PNR602243S PNR602246S PNR602248S PNR602262S PNR602265S PNR602269S PNR602275S PNR602278S PNR602279S PNR602282S PNR602284S PNR602287S PNR602290S PNR602296S PNR602300S PNR602303S PNR602318S PNR602320S PNR602330S PNR602344S PNR602355S PNR602372S PNR602386S PNR602390S PNR602393S PNR602395S PNR602396S PNR602397S PNR602398S PNR602399S PNR602400S PNR602402S PNR602404S PNR602406S PNR602407S PNR602409S PNR602412S PNR602413S PNR602416S PNR602419S PNR602422S PNR602426S PNR602427S PNR602432S PNR602438S PNR602452S PNR602511S PNR602526S PNR602538S PNR602539S PNR602541S PNR602543S PNR602567S PNR602574S PNR602577S PNR602580S PNR602585S PNR602586S PNR602587S PNR602588S PNR602674S PNR602686S PNR602701S PNR602716S PNR602719S PNR602721S PNR602726S PNR602732S PNR602734S PNR602735S PNR602739S PNR602740S PNR602741S PNR602743S PNR602744S PNR602748S PNR602749S PNR602750S PNR602753S PNR602755S PNR602756S PNR602759S PNR602760S PNR602761S PNR602764S PNR602766S PNR602768S PNR602769S PNR602770S PNR602771S PNR602772S PNR602773S PNR602776S PNR602779S PNR602781S PNR602782S PNR602783S PNR602786S PNR602787S PNR602788S PNR602791S PNR602792S PNR602794S PNR602795S PNR602802S PNR602804S PNR602807S PNR602808S PNR602809S PNR602811S PNR602812S PNR602814S PNR602815S PNR602816S PNR602818S PNR602819S PNR602820S PNR602821S PNR602822S PNR602823S PNR602824S PNR602825S PNR602827S PNR602828S PNR602829S PNR602830S PNR602831S PNR602833S PNR602834S PNR602835S PNR602836S PNR602837S PNR602838S PNR602841S PNR602842S PNR602843S PNR602849S PNR602852S PNR602853S PNR602854S PNR602855S PNR602856S PNR602874S PNR602885S PNR602895S PNR602910S PNR603573S PNR604518S PNR604520S PNR604521S PNR604522S PNR604523S PNR604835S PNR604840S PNR604888S PNR604889S PNR604891S PNR604894S PNR604895S PNR604896S PNR604898S PNR604900S PNR604903S PNR604905S PNR604928S PNR604933S PNR604936S PNR604937S PNR604939S PNR604940S PNR604941S PNR604942S PNR604943S PNR604945S PNR604946S PNR604948S PNR604950S PNR604952S PNR604955S PNR604956S PNR604959S PNR604963S PNR604964S PNR604965S PNR604966S PNR604967S PNR604969S PNR604970S PNR604971S PNR604974S PNR604975S PNR604976S PNR604979S PNR604980S PNR604981S PNR604983S PNR604984S PNR604985S PNR604988S PNR604989S PNR604990S PNR605001S PNR605003S PNR605004S PNR605005S PNR605006S PNR605007S PNR605008S PNR605010S PNR605011S PNR605013S PNR605014S PNR605015S PNR605017S PNR605018S PNR605019S PNR605020S PNR605022S PNR605023S PNR605031S PNR605032S PNR605033S PNR605034S PNR605036S PNR605057S PNR605058S PNR605059S PNR605060S PNR605065S PNR605070S PNR605075S PNR605077S PNR605079S PNR605080S PNR605082S PNR605083S PNR605084S PNR605085S PNR605086S PNR605087S PNR605088S PNR605089S PNR605093S PNR605096S PNR605098S PNR605103S PNR605105S PNR605107S PNR605108S PNR605109S PNR605113S PNR605115S PNR605116S PNR605118S PNR605119S PNR605120S PNR605122S PNR605125S PNR605126S PNR605202S PNR605227S PNR605407S PNR605408S PNR605421S PNR605465S PNR605468S PNR605470S PNR605554S PNR605555S PNR605556S PNR605557S PNR605558S PNR605559S PNR605560S PNR605561S PNR605562S PNR605563S PNR605565S PNR605567S PNR605569S PNR605570S PNR605572S PNR605573S PNR605575S PNR605576S PNR605577S PNR605578S PNR605579S PNR605581S PNR605582S PNR605583S PNR605584S PNR605585S PNR605588S PNR605589S PNR605591S PNR605592S PNR605593S PNR605594S PNR605596S PNR605597S PNR605599S PNR605601S PNR605602S PNR605603S PNR605604S PNR605605S PNR605606S PNR605607S PNR605608S PNR605610S PNR605709S PNR605710S PNR605711S PNR605712S PNR605713S PNR605717S PNR605718S PNR605719S PNR605722S PNR605723S PNR605725S PNR605726S PNR605786S PNR605830S PNR605831S PNR605832S PNR605833S PNR605834S PNR605835S PNR605836S PNR605837S PNR605838S PNR605839S PNR605841S PNR605842S PNR605843S PNR605844S PNR605845S PNR605846S PNR605872S PNR605881S PNR605902S PNR605909S PNR605916S PNR605917S PNR605918S PNR605919S PNR605920S PNR605921S PNR605922S PNR605923S PNR605924S PNR605925S PNR605926S PNR605927S PNR605928S PNR605930S PNR605931S PNR605932S PNR605933S PNR605965S PNR605966S PNR605967S PNR605970S PNR605971S PNR605972S PNR605973S PNR605974S PNR605975S PNR605980S PNR605981S PNR605983S PNR605984S PNR605985S PNR605986S PNR605987S PNR606057S PNR606078S PNR606080S PNR606083S PNR606084S PNR606085S PNR606090S PNR606091S PNR606126S PNR606127S PNR606128S PNR606139S PNR606149S PNR606151S PNR606192S PNR606193S PNR606194S PNR606195S PNR606196S PNR606197S PNR606198S PNR606201S PNR606203S PNR606206S PNR606207S PNR606208S PNR606209S PNR606210S PNR606212S PNR606283S PNR606345S PNR606347S PNR606348S PNR606349S PNR606350S PNR606351S PNR606352S PNR606355S PNR606356S PNR606358S PNR606360S PNR606362S PNR606365S PNR606367S PNR606370S PNR606371S PNR606372S PNR606373S PNR606375S PNR606376S PNR606377S PNR606378S PNR606379S PNR606380S PNR606457S PNR606458S PNR606459S PNR606460S PNR606461S PNR606465S PNR606466S PNR606497S PNR606498S PNR606499S PNR606500S PNR606501S PNR606503S PNR606504S PNR606505S PNR606508S PNR606509S PNR606510S PNR606512S PNR606513S PNR606515S PNR606518S PNR606520S PNR606521S PNR606610S PNR606624S PNR606625S PNR606626S PNR606627S PNR606630S PNR606632S PNR606633S PNR606634S PNR606635S PNR606636S PNR606638S PNR606641S PNR606642S PNR606643S PNR606644S PNR606645S PNR606646S PNR606648S PNR606649S PNR606650S PNR606671S PNR606672S PNR606673S PNR606674S PNR606676S PNR606677S PNR606732S PNR606734S PNR606737S PNR606742S PNR606744S PNR606756S PNR606758S PNR606759S PNR606763S PNR606768S PNR606775S PNR606778S PNR606779S PNR606781S PNR606783S PNR606788S PNR606789S PNR606790S PNR606796S PNR606797S PNR606806S PNR606809S PNR606813S PNR606815S PNR606816S PNR606820S PNR606821S PNR606834S PNR606841S PNR606842S PNR606843S PNR606844S PNR606862S PNR606901S PNR606905S PNR606908S PNR606909S PNR606913S PNR606919S PNR606920S PNR606928S PNR606929S PNR606932S PNR606966S PNR607036S PNR607077S PNR607175S PNR607176S PNR607177S PNR607178S PNR607179S PNR607180S PNR607182S PNR607183S PNR607185S PNR607186S PNR607264S PNR607266S PNR607304S PNR607323S PNR607324S PNR607326S PNR607328S PNR607379S PNR607380S PNR607381S PNR607383S PNR607384S PNR607385S PNR607386S PNR607387S PNR607413S PNR607415S PNR607419S PNR607425S PNR607435S PNR607493S PNR607500S PNR607501S PNR607505S PNR607507S PNR607596S PNR607597S PNR607598S PNR607599S PNR607607S PNR607654S PNR607669S PNR607696S PNR607697S PNR607698S PNR607761S PNR607762S PNR607774S PNR607775S PNR607776S PNR607777S PNR607787S PNR607788S PNR607789S PNR607790S PNR607791S PNR607792S PNR607793S PNR607794S PNR607795S PNR607796S PNR607802S PNR607823S PNR607834S PNR607851S PNR607866S PNR607871S PNR607876S PNR607927S PNR608029S PNR608030S PNR608031S PNR608038S PNR608044S PNR608141S PNR608148S PNR608162S PNR608176S PNR608177S PNR608178S PNR608179S PNR608181S PNR608184S PNR608186S PNR608188S PNR608197S PNR608202S PNR608257S PNR608271S PNR608311S PNR608432S PNR608485S PNR608539S PNR608547S PNR608581S PNR608582S PNR608589S PNR608596S PNR608597S PNR608605S PNR608606S PNR608607S PNR608610S PNR608628S PNR608641S PNR608643S PNR608650S PNR608651S PNR608652S PNR608657S PNR608659S PNR608672S PNR608673S PNR608680S PNR608685S PNR609011S PNR609031S PNR609033S PNR609331S PNR609332S PNR609339S PNR609642S PNR609948S PNR609949S PNR610198S PNR610199S PNR610200S PNR610201S PNR610202S PNR610203S PNR610204S PNR610207S PNR610210S PNR610212S PNR610214S PNR610218S PNR610220S PNR610222S PNR610224S PNR610229S PNR610231S PNR610232S PNR610235S PNR610236S PNR610237S PNR610241S PNR610247S PNR610249S PNR610253S PNR610255S PNR610257S PNR610258S PNR610262S PNR610264S PNR610265S PNR611045S PNR611053S PNR611117S PNR611121S PNR611414S PNR611470S PNR611471S PNR611476S PNR611478S PNR611482S PNR611562S PNR611569S PNR611599S PNR611614S PNR611788S PNR611840S PNR611853S PNR611969S PNR612439S PNR612440S PNR612442S PNR612447S PNR612453S PNR612674S PNR612686S PNR612822S PNR612920S PNR612937S PNR613031S PNR613146S PNR613973S PNR614009S PVV617668S Product UPN/GTIN: 00885074300191 Serial Numbers: PNR600077S PNR600080S PNR600081S PNR600085S PNR600086S PNR600087S PNR600089S PNR600090S PNR600091S PNR600093S PNR600094S PNR600096S PNR600098S PNR600101S PNR600103S PNR600104S PNR600106S PNR600107S PNR600108S PNR600109S PNR600112S PNR600113S PNR600114S PNR600115S PNR600116S PNR600117S PNR600118S PNR600122S PNR600124S PNR600126S PNR600127S PNR600128S PNR600130S PNR600132S PNR600133S PNR600134S PNR600135S PNR600137S PNR600138S PNR600139S PNR600140S PNR600142S PNR600146S PNR600149S PNR600150S PNR600151S PNR600152S PNR600153S PNR600155S PNR600164S PNR600168S PNR600171S PNR600172S PNR600173S PNR600175S PNR600178S PNR600181S PNR600188S PNR600201S PNR600202S PNR600205S PNR600206S PNR600207S PNR600208S PNR600210S PNR600211S PNR600212S PNR600213S PNR600216S PNR600217S PNR600219S PNR600220S PNR600221S PNR600222S PNR600223S PNR600224S PNR600225S PNR600226S PNR600228S PNR600229S PNR600230S PNR600231S PNR600232S PNR600233S PNR600234S PNR600235S PNR600236S PNR600237S PNR600238S PNR600239S PNR600240S PNR600241S PNR600242S PNR600243S PNR600244S PNR600245S PNR600246S PNR600247S PNR600249S PNR600250S PNR600252S PNR600253S PNR600255S PNR600256S PNR600257S PNR600258S PNR600259S PNR600260S PNR600261S PNR600262S PNR600263S PNR600264S PNR600265S PNR600266S PNR600267S PNR600268S PNR600269S PNR600270S PNR600271S PNR600272S PNR600273S PNR600274S PNR600275S PNR600276S PNR600277S PNR600278S PNR600279S PNR600280S PNR600281S PNR600282S PNR600283S PNR600284S PNR600285S PNR600286S PNR600287S PNR600289S PNR600291S PNR600292S PNR600293S PNR600295S PNR600296S PNR600297S PNR600300S PNR600301S PNR600302S PNR600303S PNR600304S PNR600305S PNR600307S PNR600308S PNR600309S PNR600310S PNR600311S PNR600313S PNR600314S PNR600315S PNR600316S PNR600317S PNR600318S PNR600319S PNR600320S PNR600321S PNR600322S PNR600378S PNR600395S PNR600430S PNR600582S PNR600598S PNR600605S PNR600612S PNR600618S PNR600622S PNR600625S PNR600627S PNR600628S PNR600632S PNR600633S PNR600729S PNR600730S PNR600732S PNR600734S PNR600735S PNR600736S PNR600737S PNR600739S PNR600740S PNR600741S PNR600742S PNR600743S PNR600744S PNR600745S PNR600746S PNR600747S PNR600748S PNR600749S PNR600750S PNR600751S PNR600752S PNR600753S PNR600754S PNR600755S PNR600756S PNR600757S PNR600758S PNR600759S PNR600760S PNR600761S PNR600762S PNR600764S PNR600765S PNR600766S PNR600767S PNR600768S PNR600769S PNR600772S PNR600774S PNR600776S PNR600777S PNR600779S PNR600780S PNR600782S PNR600785S PNR600787S PNR600788S PNR600791S PNR600794S PNR600796S PNR600799S PNR600803S PNR600805S PNR600806S PNR600808S PNR600812S PNR600814S PNR600817S PNR600833S PNR600834S PNR600835S PNR600837S PNR600838S PNR600839S PNR600840S PNR600841S PNR600842S PNR600846S PNR600847S PNR600848S PNR600850S PNR600851S PNR600852S PNR600853S PNR600854S PNR600855S PNR600856S PNR600857S PNR600859S PNR600860S PNR600861S PNR600862S PNR600863S PNR600864S PNR600867S PNR600868S PNR600869S PNR600870S PNR600872S PNR600873S PNR600874S PNR600875S PNR600876S PNR600877S PNR600879S PNR600881S PNR600882S PNR600884S PNR600885S PNR600888S PNR600889S PNR600893S PNR600895S PNR600896S PNR600899S PNR600900S PNR600901S PNR600902S PNR600906S PNR600907S PNR600908S PNR600911S PNR600912S PNR600913S PNR600914S PNR600915S PNR600917S PNR600918S PNR600920S PNR600921S PNR600923S PNR600924S PNR600925S PNR600926S PNR600928S PNR600929S PNR600930S PNR600931S PNR600932S PNR600935S PNR600936S PNR600937S PNR600941S PNR600942S PNR600960S PNR600970S PNR600977S PNR600981S PNR600986S PNR600991S PNR600994S PNR601016S PNR601024S PNR601028S PNR601033S PNR601035S PNR601041S PNR601042S PNR601047S PNR601049S PNR601053S PNR601057S PNR601064S PNR601065S PNR601066S PNR601069S PNR601071S PNR601072S PNR601073S PNR601081S PNR601088S PNR601094S PNR601096S PNR601104S PNR601111S PNR601148S PNR601160S PNR601161S PNR601163S PNR601167S PNR601169S PNR601172S PNR601173S PNR601176S PNR601179S PNR601181S PNR601183S PNR601184S PNR601185S PNR601186S PNR601187S PNR601188S PNR601190S PNR601192S PNR601193S PNR601196S PNR601198S PNR601200S PNR601204S PNR601205S PNR601207S PNR601208S PNR601211S PNR601214S PNR601218S PNR601219S PNR601220S PNR601221S PNR601222S PNR601223S PNR601224S PNR601227S PNR601228S PNR601229S PNR601230S PNR601231S PNR601232S PNR601233S PNR601234S PNR601236S PNR601237S PNR601238S PNR601239S PNR601245S PNR601247S PNR601251S PNR601254S PNR601255S PNR601256S PNR601257S PNR601258S PNR601259S PNR601260S PNR601261S PNR601262S PNR601263S PNR601264S PNR601265S PNR601266S PNR601267S PNR601268S PNR601269S PNR601270S PNR601271S PNR601273S PNR601274S PNR601275S PNR601277S PNR601278S PNR601279S PNR601281S PNR601282S PNR601283S PNR601284S PNR601285S PNR601286S PNR601287S PNR601288S PNR601289S PNR601290S PNR601291S PNR601292S PNR601293S PNR601294S PNR601295S PNR601296S PNR601297S PNR601298S PNR601299S PNR601300S PNR601301S PNR601303S PNR601304S PNR601305S PNR601306S PNR601308S PNR601310S PNR601311S PNR601312S PNR601314S PNR601315S PNR601316S PNR601317S PNR601318S PNR601319S PNR601320S PNR601321S PNR601322S PNR601323S PNR601324S PNR601326S PNR601327S PNR601328S PNR601330S PNR601331S PNR601332S PNR601333S PNR601334S PNR601335S PNR601336S PNR601337S PNR601338S PNR601339S PNR601340S PNR601341S PNR601342S PNR601343S PNR601344S PNR601345S PNR601346S PNR601347S PNR601348S PNR601349S PNR601352S PNR601353S PNR601355S PNR601356S PNR601357S PNR601358S PNR601359S PNR601360S PNR601365S PNR601367S PNR601369S PNR601370S PNR601371S PNR601374S PNR601376S PNR601377S PNR601378S PNR601379S PNR601380S PNR601382S PNR601383S PNR601384S PNR601385S PNR601386S PNR601387S PNR601389S PNR601392S PNR601394S PNR601396S PNR601397S PNR601399S PNR601400S PNR601401S PNR601402S PNR601403S PNR601404S PNR601405S PNR601407S PNR601408S PNR601409S PNR601410S PNR601411S PNR601412S PNR601413S PNR601415S PNR601417S PNR601418S PNR601423S PNR601424S PNR601427S PNR601428S PNR601429S PNR601430S PNR601432S PNR601434S PNR601435S PNR601436S PNR601437S PNR601438S PNR601439S PNR601445S PNR601449S PNR601454S PNR601455S PNR601457S PNR601458S PNR601459S PNR601461S PNR601463S PNR601464S PNR601465S PNR601467S PNR601468S PNR601469S PNR601471S PNR601473S PNR601474S PNR601475S PNR601476S PNR601477S PNR601478S PNR601479S PNR601480S PNR601481S PNR601482S PNR601483S PNR601485S PNR601486S PNR601488S PNR601489S PNR601490S PNR601491S PNR601492S PNR601493S PNR601494S PNR601495S PNR601496S PNR601497S PNR601498S PNR601499S PNR601500S PNR601501S PNR601502S PNR601503S PNR601504S PNR601505S PNR601506S PNR601508S PNR601509S PNR601510S PNR601511S PNR601512S PNR601513S PNR601514S PNR601515S PNR601517S PNR601518S PNR601519S PNR601520S PNR601521S PNR601522S PNR601523S PNR601525S PNR601526S PNR601527S PNR601529S PNR601530S PNR601531S PNR601532S PNR601533S PNR601534S PNR601536S PNR601537S PNR601538S PNR601539S PNR601540S PNR601543S PNR601544S PNR601545S PNR601547S PNR601549S PNR601551S PNR601552S PNR601553S PNR601554S PNR601556S PNR601557S PNR601560S PNR601561S PNR601562S PNR601563S PNR601575S PNR601577S PNR601580S PNR601585S PNR601586S PNR601587S PNR601588S PNR601589S PNR601595S PNR601596S PNR601597S PNR601598S PNR601599S PNR601600S PNR601601S PNR601602S PNR601605S PNR601606S PNR601607S PNR601608S PNR601610S PNR601612S PNR601613S PNR601614S PNR601615S PNR601616S PNR601619S PNR601621S PNR601623S PNR601625S PNR601627S PNR601629S PNR601630S PNR601631S PNR601632S PNR601633S PNR601635S PNR601636S PNR601637S PNR601639S PNR601642S PNR601643S PNR601644S PNR601654S PNR601655S PNR601656S PNR601658S PNR601660S PNR601663S PNR601666S PNR601671S PNR601673S PNR601675S PNR601677S PNR601678S PNR601679S PNR601680S PNR601681S PNR601682S PNR601683S PNR601686S PNR601687S PNR601688S PNR601689S PNR601691S PNR601694S PNR601695S PNR601696S PNR601697S PNR601698S PNR601699S PNR601702S PNR601703S PNR601704S PNR601705S PNR601707S PNR601708S PNR601711S PNR601715S PNR601717S PNR601718S PNR601719S PNR601721S PNR601724S PNR601725S PNR601726S PNR601727S PNR601728S PNR601730S PNR601731S PNR601732S PNR601734S PNR601735S PNR601736S PNR601737S PNR601738S PNR601739S PNR601740S PNR601741S PNR601743S PNR601744S PNR601746S PNR601747S PNR601749S PNR601751S PNR601752S PNR601753S PNR601754S PNR601755S PNR601756S PNR601757S PNR601758S PNR601760S PNR601762S PNR601764S PNR601766S PNR601768S PNR601769S PNR601770S PNR601772S PNR601773S PNR601774S PNR601775S PNR601777S PNR601778S PNR601779S PNR601780S PNR601782S PNR601783S PNR601785S PNR601786S PNR601787S PNR601788S PNR601789S PNR601790S PNR601791S PNR601792S PNR601793S PNR601794S PNR601795S PNR601796S PNR601797S PNR601798S PNR601799S PNR601800S PNR601802S PNR601803S PNR601804S PNR601806S PNR601807S PNR601808S PNR601810S PNR601811S PNR601812S PNR601813S PNR601814S PNR601815S PNR601816S PNR601817S PNR601819S PNR601820S PNR601823S PNR601826S PNR601827S PNR601828S PNR601830S PNR601832S PNR601833S PNR601835S PNR601837S PNR601842S PNR601843S PNR601844S PNR601845S PNR601846S PNR601847S PNR601848S PNR601849S PNR601850S PNR601853S PNR601862S PNR601864S PNR601865S PNR601866S PNR601867S PNR601868S PNR601869S PNR601870S PNR601873S PNR601876S PNR601877S PNR601879S PNR601881S PNR601882S PNR601885S PNR601886S PNR601887S PNR601889S PNR601890S PNR601891S PNR601893S PNR601894S PNR601895S PNR601896S PNR601897S PNR601898S PNR601899S PNR601900S PNR601901S PNR601902S PNR601903S PNR601904S PNR601905S PNR601906S PNR601907S PNR601908S PNR601909S PNR601910S PNR601911S PNR601912S PNR601914S PNR601915S PNR601916S PNR601917S PNR601918S PNR601920S PNR601921S PNR601922S PNR601923S PNR601925S PNR601926S PNR601927S PNR601928S PNR601931S PNR601933S PNR601934S PNR601935S PNR601936S PNR601937S PNR601938S PNR601942S PNR601945S PNR601946S PNR601949S PNR601950S PNR601960S PNR601971S PNR602000S PNR602006S PNR602009S PNR602010S PNR602011S PNR602012S PNR602014S PNR602017S PNR602018S PNR602019S PNR602020S PNR602022S PNR602023S PNR602024S PNR602026S PNR602027S PNR602028S PNR602029S PNR602030S PNR602031S PNR602032S PNR602033S PNR602034S PNR602035S PNR602036S PNR602038S PNR602039S PNR602042S PNR602043S PNR602044S PNR602045S PNR602046S PNR602047S PNR602048S PNR602051S PNR602053S PNR602054S PNR602055S PNR602056S PNR602057S PNR602058S PNR602059S PNR602060S PNR602061S PNR602062S PNR602063S PNR602064S PNR602065S PNR602066S PNR602067S PNR602069S PNR602070S PNR602071S PNR602072S PNR602073S PNR602074S PNR602075S PNR602076S PNR602077S PNR602078S PNR602079S PNR602081S PNR602082S PNR602083S PNR602085S PNR602086S PNR602087S PNR602090S PNR602092S PNR602093S PNR602095S PNR602096S PNR602097S PNR602100S PNR602104S PNR602111S PNR602121S PNR602122S PNR602123S PNR602127S PNR602128S PNR602129S PNR602130S PNR602131S PNR602133S PNR602134S PNR602135S PNR602137S PNR602140S PNR602142S PNR602143S PNR602144S PNR602145S PNR602146S PNR602147S PNR602148S PNR602149S PNR602151S PNR602153S PNR602154S PNR602155S PNR602156S PNR602157S PNR602158S PNR602161S PNR602162S PNR602163S PNR602164S PNR602167S PNR602168S PNR602169S PNR602170S PNR602171S PNR602172S PNR602173S PNR602175S PNR602176S PNR602180S PNR602181S PNR602182S PNR602183S PNR602184S PNR602185S PNR602186S PNR602187S PNR602188S PNR602190S PNR602191S PNR602195S PNR602197S PNR602198S PNR602199S PNR602203S PNR602206S PNR602207S PNR602208S PNR602209S PNR602210S PNR602211S PNR602212S PNR602214S PNR602215S PNR602216S PNR602217S PNR602218S PNR602219S PNR602220S PNR602221S PNR602222S PNR602223S PNR602224S PNR602226S PNR602227S PNR602228S PNR602229S PNR602230S PNR602231S PNR602234S PNR602235S PNR602237S PNR602238S PNR602239S PNR602240S PNR602241S PNR602244S PNR602245S PNR602249S PNR602250S PNR602251S PNR602252S PNR602253S PNR602254S PNR602256S PNR602257S PNR602258S PNR602259S PNR602260S PNR602261S PNR602263S PNR602264S PNR602266S PNR602267S PNR602268S PNR602270S PNR602271S PNR602272S PNR602273S PNR602274S PNR602276S PNR602280S PNR602281S PNR602283S PNR602285S PNR602286S PNR602288S PNR602291S PNR602293S PNR602294S PNR602295S PNR602297S PNR602299S PNR602301S PNR602302S PNR602304S PNR602306S PNR602307S PNR602310S PNR602311S PNR602312S PNR602314S PNR602315S PNR602316S PNR602317S PNR602319S PNR602325S PNR602326S PNR602327S PNR602328S PNR602329S PNR602331S PNR602332S PNR602334S PNR602336S PNR602337S PNR602338S PNR602339S PNR602340S PNR602341S PNR602342S PNR602345S PNR602346S PNR602348S PNR602349S PNR602352S PNR602354S PNR602356S PNR602357S PNR602358S PNR602359S PNR602360S PNR602362S PNR602363S PNR602364S PNR602365S PNR602366S PNR602367S PNR602368S PNR602369S PNR602370S PNR602371S PNR602373S PNR602374S PNR602375S PNR602376S PNR602377S PNR602378S PNR602379S PNR602380S PNR602381S PNR602389S PNR602391S PNR602392S PNR602408S PNR602415S PNR602418S PNR602423S PNR602424S PNR602429S PNR602430S PNR602431S PNR602433S PNR602434S PNR602436S PNR602437S PNR602439S PNR602441S PNR602442S PNR602443S PNR602445S PNR602446S PNR602447S PNR602448S PNR602449S PNR602451S PNR602453S PNR602455S PNR602456S PNR602457S PNR602458S PNR602460S PNR602461S PNR602462S PNR602463S PNR602464S PNR602465S PNR602466S PNR602467S PNR602468S PNR602469S PNR602470S PNR602471S PNR602472S PNR602476S PNR602477S PNR602478S PNR602480S PNR602481S PNR602482S PNR602483S PNR602487S PNR602488S PNR602489S PNR602490S PNR602491S PNR602492S PNR602493S PNR602494S PNR602495S PNR602496S PNR602497S PNR602498S PNR602499S PNR602500S PNR602501S PNR602502S PNR602504S PNR602507S PNR602508S PNR602509S PNR602510S PNR602513S PNR602514S PNR602515S PNR602516S PNR602517S PNR602518S PNR602521S PNR602524S PNR602525S PNR602527S PNR602528S PNR602529S PNR602531S PNR602532S PNR602533S PNR602534S PNR602542S PNR602545S PNR602547S PNR602548S PNR602551S PNR602552S PNR602553S PNR602554S PNR602556S PNR602557S PNR602559S PNR602560S PNR602561S PNR602562S PNR602564S PNR602566S PNR602568S PNR602569S PNR602571S PNR602573S PNR602575S PNR602576S PNR602578S PNR602579S PNR602581S PNR602583S PNR602584S PNR602589S PNR602591S PNR602593S PNR602594S PNR602595S PNR602596S PNR602597S PNR602598S PNR602600S PNR602601S PNR602602S PNR602605S PNR602607S PNR602608S PNR602609S PNR602611S PNR602612S PNR602613S PNR602615S PNR602616S PNR602618S PNR602619S PNR602620S PNR602623S PNR602625S PNR602627S PNR602628S PNR602629S PNR602630S PNR602631S PNR602632S PNR602633S PNR602634S PNR602635S PNR602636S PNR602637S PNR602638S PNR602639S PNR602640S PNR602641S PNR602642S PNR602643S PNR602645S PNR602646S PNR602647S PNR602648S PNR602649S PNR602650S PNR602651S PNR602652S PNR602653S PNR602654S PNR602655S PNR602656S PNR602657S PNR602658S PNR602659S PNR602660S PNR602661S PNR602662S PNR602663S PNR602664S PNR602665S PNR602667S PNR602668S PNR602669S PNR602670S PNR602671S PNR602673S PNR602675S PNR602676S PNR602679S PNR602681S PNR602682S PNR602683S PNR602684S PNR602685S PNR602688S PNR602689S PNR602690S PNR602692S PNR602695S PNR602696S PNR602697S PNR602698S PNR602699S PNR602700S PNR602702S PNR602703S PNR602704S PNR602705S PNR602706S PNR602707S PNR602709S PNR602710S PNR602715S PNR602722S PNR602725S PNR602796S PNR602797S PNR602798S PNR602799S PNR602869S PNR602894S PNR602899S PNR602907S PNR602942S PNR602964S PNR602974S PNR602976S PNR602978S PNR602979S PNR602980S PNR602982S PNR602985S PNR602986S PNR602991S PNR602992S PNR602994S PNR602995S PNR602997S PNR602999S PNR603001S PNR603005S PNR603007S PNR603013S PNR603014S PNR603018S PNR603019S PNR603023S PNR603024S PNR603027S PNR603028S PNR603033S PNR603034S PNR603036S PNR603037S PNR603040S PNR603041S PNR603045S PNR603046S PNR603050S PNR603051S PNR603052S PNR603063S PNR603066S PNR603071S PNR603072S PNR603074S PNR603081S PNR603083S PNR603085S PNR603089S PNR603091S PNR603123S PNR603265S PNR603280S PNR603291S PNR603293S PNR603311S PNR603313S PNR603315S PNR603316S PNR603321S PNR603323S PNR603325S PNR603329S PNR603330S PNR603333S PNR603334S PNR603336S PNR603338S PNR603341S PNR603345S PNR603347S PNR603349S PNR603352S PNR603355S PNR603356S PNR603358S PNR603359S PNR603366S PNR603367S PNR603369S PNR603370S PNR603374S PNR603376S PNR603379S PNR603385S PNR603386S PNR603389S PNR603390S PNR603391S PNR603392S PNR603393S PNR603397S PNR603398S PNR603403S PNR603404S PNR603406S PNR603408S PNR603410S PNR603411S PNR603416S PNR603420S PNR603428S PNR603430S PNR603433S PNR603440S PNR603441S PNR603444S PNR603445S PNR603447S PNR603454S PNR603455S PNR603459S PNR603462S PNR603463S PNR603464S PNR603470S PNR603475S PNR603482S PNR603486S PNR603488S PNR603489S PNR603490S PNR603491S PNR603494S PNR603495S PNR603496S PNR603503S PNR603511S PNR603514S PNR603515S PNR603516S PNR603519S PNR603520S PNR603521S PNR603529S PNR603564S PNR603646S PNR603670S PNR603982S PNR603995S PNR604158S PNR604160S PNR604161S PNR604223S PNR604411S PNR604418S PNR604425S PNR604454S PNR604463S PNR604465S PNR604469S PNR604477S PNR604484S PNR604489S PNR604495S PNR604496S PNR604497S PNR604498S PNR604499S PNR604500S PNR604501S PNR604502S PNR604503S PNR604504S PNR604505S PNR604506S PNR604508S PNR604509S PNR604510S PNR604511S PNR604512S PNR604513S PNR604515S PNR604519S PNR604524S PNR604525S PNR604526S PNR604527S PNR604528S PNR604529S PNR604530S PNR604531S PNR604532S PNR604533S PNR604534S PNR604535S PNR604536S PNR604537S PNR604538S PNR604539S PNR604540S PNR604541S PNR604550S PNR604551S PNR604552S PNR604553S PNR604554S PNR604556S PNR604557S PNR604558S PNR604559S PNR604560S PNR604561S PNR604562S PNR604563S PNR604564S PNR604565S PNR604566S PNR604567S PNR604568S PNR604569S PNR604570S PNR604571S PNR604572S PNR604573S PNR604574S PNR604575S PNR604577S PNR604578S PNR604579S PNR604580S PNR604581S PNR604582S PNR604583S PNR604584S PNR604585S PNR604586S PNR604588S PNR604589S PNR604590S PNR604591S PNR604592S PNR604593S PNR604594S PN
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
- Reason for Recall:
- Medtronic has become aware of a potential for loss of high voltage and anti-tachycardia pacing therapy in EnTrust and Escudo implantable cardioverter defibrillators (ICDs) as they near elective replacement indicator (ERI) voltage. Under certain circumstances, the device may display an immediate End of Life (EOL) Observation with no prior ERI alert. Though no ERI alert is triggered, there may not be enough remaining battery capacity to charge the high voltage circuits, resulting in an excessive charge time EOL Observation, leading to a loss of high voltage and anti-tachycardia pacing therapy.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Medtronic EnTrust Dual Chamber Implantable Cardioverter Defibrillator with Atrial and Ventricular Therapies and RapidRead Telemetry, D154ATG Product Usage: The device indicated for use in ICD patients with atrial tachyarrhythmias, or who are at significant risk of developing atrial tachyarrhythmias. The implantable cardioverter defibrillator is intended to provide ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of life threatening ventricular arrhythmias.
Product Codes/Lot Numbers:
Product UPN/GTIN: 00613994170927 Serial Numbers: PNR606917S PNR606918S PNR606923S Product UPN/GTIN: 00885074299938 Serial Numbers: PNR414990H PNR417905H PNR600110S PNR600121S PNR600123S PNR600136S PNR600141S PNR600147S PNR600148S PNR600157S PNR600163S PNR600166S PNR600176S PNR600177S PNR600183S PNR600189S PNR600190S PNR600192S PNR600194S PNR600195S PNR600198S PNR600203S PNR600290S PNR600306S PNR600381S PNR600382S PNR600383S PNR600385S PNR600387S PNR600388S PNR600389S PNR600390S PNR600392S PNR600393S PNR600396S PNR600397S PNR600398S PNR600401S PNR600402S PNR600404S PNR600405S PNR600406S PNR600407S PNR600409S PNR600410S PNR600412S PNR600415S PNR600416S PNR600417S PNR600418S PNR600419S PNR600422S PNR600423S PNR600424S PNR600425S PNR600426S PNR600427S PNR600428S PNR600429S PNR600432S PNR600433S PNR600434S PNR600435S PNR600436S PNR600437S PNR600438S PNR600439S PNR600441S PNR600442S PNR600443S PNR600448S PNR600450S PNR600452S PNR600453S PNR600454S PNR600455S PNR600458S PNR600459S PNR600461S PNR600462S PNR600463S PNR600464S PNR600503S PNR600504S PNR600505S PNR600506S PNR600510S PNR600511S PNR600545S PNR600548S PNR600549S PNR600551S PNR600552S PNR600553S PNR600554S PNR600555S PNR600557S PNR600561S PNR600562S PNR600563S PNR600564S PNR600565S PNR600566S PNR600570S PNR600572S PNR600576S PNR600577S PNR600578S PNR600581S PNR600583S PNR600585S PNR600599S PNR600600S PNR600601S PNR600603S PNR600604S PNR600609S PNR600610S PNR600614S PNR600615S PNR600616S PNR600619S PNR600620S PNR600621S PNR600623S PNR600624S PNR600626S PNR600629S PNR600631S PNR600731S PNR600733S PNR600770S PNR600773S PNR600775S PNR600781S PNR600784S PNR600786S PNR600789S PNR600790S PNR600792S PNR600798S PNR600801S PNR600802S PNR600804S PNR600811S PNR600813S PNR600816S PNR600818S PNR600819S PNR600821S PNR600824S PNR600849S PNR600886S PNR600887S PNR600891S PNR600892S PNR600904S PNR600933S PNR600948S PNR600949S PNR600950S PNR600953S PNR600956S PNR600957S PNR600958S PNR600964S PNR600978S PNR600979S PNR600987S PNR600990S PNR600996S PNR600997S PNR600998S PNR601017S PNR601029S PNR601031S PNR601036S PNR601044S PNR601045S PNR601046S PNR601054S PNR601056S PNR601058S PNR601060S PNR601061S PNR601063S PNR601070S PNR601075S PNR601076S PNR601079S PNR601080S PNR601082S PNR601083S PNR601089S PNR601090S PNR601097S PNR601102S PNR601109S PNR601113S PNR601115S PNR601116S PNR601117S PNR601118S PNR601120S PNR601121S PNR601124S PNR601125S PNR601126S PNR601128S PNR601129S PNR601131S PNR601133S PNR601134S PNR601136S PNR601137S PNR601140S PNR601149S PNR601150S PNR601151S PNR601154S PNR601155S PNR601164S PNR601170S PNR601174S PNR601178S PNR601180S PNR601195S PNR601197S PNR601199S PNR601201S PNR601203S PNR601206S PNR601210S PNR601212S PNR601216S PNR601225S PNR601235S PNR601248S PNR601250S PNR601329S PNR601351S PNR601354S PNR601364S PNR601372S PNR601373S PNR601375S PNR601391S PNR601393S PNR601395S PNR601406S PNR601416S PNR601419S PNR601426S PNR601431S PNR601441S PNR601444S PNR601446S PNR601447S PNR601451S PNR601452S PNR601453S PNR601472S PNR601484S PNR601487S PNR601524S PNR601528S PNR601546S PNR601548S PNR601550S PNR601555S PNR601559S PNR601564S PNR601567S PNR601570S PNR601572S PNR601573S PNR601574S PNR601582S PNR601584S PNR601590S PNR601591S PNR601594S PNR601603S PNR601609S PNR601617S PNR601620S PNR601624S PNR601628S PNR601645S PNR601646S PNR601647S PNR601651S PNR601652S PNR601653S PNR601661S PNR601662S PNR601674S PNR601676S PNR601684S PNR601685S PNR601690S PNR601693S PNR601709S PNR601710S PNR601712S PNR601714S PNR601716S PNR601742S PNR601748S PNR601750S PNR601763S PNR601765S PNR601821S PNR601824S PNR601825S PNR601834S PNR601836S PNR601852S PNR601854S PNR601860S PNR601863S PNR601871S PNR601875S PNR601880S PNR601884S PNR601888S PNR601939S PNR601940S PNR601948S PNR601952S PNR601955S PNR601956S PNR601957S PNR601958S PNR601959S PNR601962S PNR601963S PNR601964S PNR601967S PNR601968S PNR601972S PNR601974S PNR601975S PNR601978S PNR601981S PNR601982S PNR601983S PNR601985S PNR601986S PNR601987S PNR601988S PNR601991S PNR601993S PNR601994S PNR601995S PNR601996S PNR601997S PNR601998S PNR601999S PNR602001S PNR602004S PNR602005S PNR602008S PNR602049S PNR602089S PNR602091S PNR602094S PNR602098S PNR602099S PNR602101S PNR602102S PNR602106S PNR602107S PNR602108S PNR602110S PNR602112S PNR602113S PNR602114S PNR602115S PNR602120S PNR602126S PNR602138S PNR602139S PNR602152S PNR602166S PNR602174S PNR602177S PNR602178S PNR602192S PNR602193S PNR602194S PNR602196S PNR602202S PNR602205S PNR602242S PNR602243S PNR602246S PNR602248S PNR602262S PNR602265S PNR602269S PNR602275S PNR602278S PNR602279S PNR602282S PNR602284S PNR602287S PNR602290S PNR602296S PNR602300S PNR602303S PNR602318S PNR602320S PNR602330S PNR602344S PNR602355S PNR602372S PNR602386S PNR602390S PNR602393S PNR602395S PNR602396S PNR602397S PNR602398S PNR602399S PNR602400S PNR602402S PNR602404S PNR602406S PNR602407S PNR602409S PNR602412S PNR602413S PNR602416S PNR602419S PNR602422S PNR602426S PNR602427S PNR602432S PNR602438S PNR602452S PNR602511S PNR602526S PNR602538S PNR602539S PNR602541S PNR602543S PNR602567S PNR602574S PNR602577S PNR602580S PNR602585S PNR602586S PNR602587S PNR602588S PNR602674S PNR602686S PNR602701S PNR602716S PNR602719S PNR602721S PNR602726S PNR602732S PNR602734S PNR602735S PNR602739S PNR602740S PNR602741S PNR602743S PNR602744S PNR602748S PNR602749S PNR602750S PNR602753S PNR602755S PNR602756S PNR602759S PNR602760S PNR602761S PNR602764S PNR602766S PNR602768S PNR602769S PNR602770S PNR602771S PNR602772S PNR602773S PNR602776S PNR602779S PNR602781S PNR602782S PNR602783S PNR602786S PNR602787S PNR602788S PNR602791S PNR602792S PNR602794S PNR602795S PNR602802S PNR602804S PNR602807S PNR602808S PNR602809S PNR602811S PNR602812S PNR602814S PNR602815S PNR602816S PNR602818S PNR602819S PNR602820S PNR602821S PNR602822S PNR602823S PNR602824S PNR602825S PNR602827S PNR602828S PNR602829S PNR602830S PNR602831S PNR602833S PNR602834S PNR602835S PNR602836S PNR602837S PNR602838S PNR602841S PNR602842S PNR602843S PNR602849S PNR602852S PNR602853S PNR602854S PNR602855S PNR602856S PNR602874S PNR602885S PNR602895S PNR602910S PNR603573S PNR604518S PNR604520S PNR604521S PNR604522S PNR604523S PNR604835S PNR604840S PNR604888S PNR604889S PNR604891S PNR604894S PNR604895S PNR604896S PNR604898S PNR604900S PNR604903S PNR604905S PNR604928S PNR604933S PNR604936S PNR604937S PNR604939S PNR604940S PNR604941S PNR604942S PNR604943S PNR604945S PNR604946S PNR604948S PNR604950S PNR604952S PNR604955S PNR604956S PNR604959S PNR604963S PNR604964S PNR604965S PNR604966S PNR604967S PNR604969S PNR604970S PNR604971S PNR604974S PNR604975S PNR604976S PNR604979S PNR604980S PNR604981S PNR604983S PNR604984S PNR604985S PNR604988S PNR604989S PNR604990S PNR605001S PNR605003S PNR605004S PNR605005S PNR605006S PNR605007S PNR605008S PNR605010S PNR605011S PNR605013S PNR605014S PNR605015S PNR605017S PNR605018S PNR605019S PNR605020S PNR605022S PNR605023S PNR605031S PNR605032S PNR605033S PNR605034S PNR605036S PNR605057S PNR605058S PNR605059S PNR605060S PNR605065S PNR605070S PNR605075S PNR605077S PNR605079S PNR605080S PNR605082S PNR605083S PNR605084S PNR605085S PNR605086S PNR605087S PNR605088S PNR605089S PNR605093S PNR605096S PNR605098S PNR605103S PNR605105S PNR605107S PNR605108S PNR605109S PNR605113S PNR605115S PNR605116S PNR605118S PNR605119S PNR605120S PNR605122S PNR605125S PNR605126S PNR605202S PNR605227S PNR605407S PNR605408S PNR605421S PNR605465S PNR605468S PNR605470S PNR605554S PNR605555S PNR605556S PNR605557S PNR605558S PNR605559S PNR605560S PNR605561S PNR605562S PNR605563S PNR605565S PNR605567S PNR605569S PNR605570S PNR605572S PNR605573S PNR605575S PNR605576S PNR605577S PNR605578S PNR605579S PNR605581S PNR605582S PNR605583S PNR605584S PNR605585S PNR605588S PNR605589S PNR605591S PNR605592S PNR605593S PNR605594S PNR605596S PNR605597S PNR605599S PNR605601S PNR605602S PNR605603S PNR605604S PNR605605S PNR605606S PNR605607S PNR605608S PNR605610S PNR605709S PNR605710S PNR605711S PNR605712S PNR605713S PNR605717S PNR605718S PNR605719S PNR605722S PNR605723S PNR605725S PNR605726S PNR605786S PNR605830S PNR605831S PNR605832S PNR605833S PNR605834S PNR605835S PNR605836S PNR605837S PNR605838S PNR605839S PNR605841S PNR605842S PNR605843S PNR605844S PNR605845S PNR605846S PNR605872S PNR605881S PNR605902S PNR605909S PNR605916S PNR605917S PNR605918S PNR605919S PNR605920S PNR605921S PNR605922S PNR605923S PNR605924S PNR605925S PNR605926S PNR605927S PNR605928S PNR605930S PNR605931S PNR605932S PNR605933S PNR605965S PNR605966S PNR605967S PNR605970S PNR605971S PNR605972S PNR605973S PNR605974S PNR605975S PNR605980S PNR605981S PNR605983S PNR605984S PNR605985S PNR605986S PNR605987S PNR606057S PNR606078S PNR606080S PNR606083S PNR606084S PNR606085S PNR606090S PNR606091S PNR606126S PNR606127S PNR606128S PNR606139S PNR606149S PNR606151S PNR606192S PNR606193S PNR606194S PNR606195S PNR606196S PNR606197S PNR606198S PNR606201S PNR606203S PNR606206S PNR606207S PNR606208S PNR606209S PNR606210S PNR606212S PNR606283S PNR606345S PNR606347S PNR606348S PNR606349S PNR606350S PNR606351S PNR606352S PNR606355S PNR606356S PNR606358S PNR606360S PNR606362S PNR606365S PNR606367S PNR606370S PNR606371S PNR606372S PNR606373S PNR606375S PNR606376S PNR606377S PNR606378S PNR606379S PNR606380S PNR606457S PNR606458S PNR606459S PNR606460S PNR606461S PNR606465S PNR606466S PNR606497S PNR606498S PNR606499S PNR606500S PNR606501S PNR606503S PNR606504S PNR606505S PNR606508S PNR606509S PNR606510S PNR606512S PNR606513S PNR606515S PNR606518S PNR606520S PNR606521S PNR606610S PNR606624S PNR606625S PNR606626S PNR606627S PNR606630S PNR606632S PNR606633S PNR606634S PNR606635S PNR606636S PNR606638S PNR606641S PNR606642S PNR606643S PNR606644S PNR606645S PNR606646S PNR606648S PNR606649S PNR606650S PNR606671S PNR606672S PNR606673S PNR606674S PNR606676S PNR606677S PNR606732S PNR606734S PNR606737S PNR606742S PNR606744S PNR606756S PNR606758S PNR606759S PNR606763S PNR606768S PNR606775S PNR606778S PNR606779S PNR606781S PNR606783S PNR606788S PNR606789S PNR606790S PNR606796S PNR606797S PNR606806S PNR606809S PNR606813S PNR606815S PNR606816S PNR606820S PNR606821S PNR606834S PNR606841S PNR606842S PNR606843S PNR606844S PNR606862S PNR606901S PNR606905S PNR606908S PNR606909S PNR606913S PNR606919S PNR606920S PNR606928S PNR606929S PNR606932S PNR606966S PNR607036S PNR607077S PNR607175S PNR607176S PNR607177S PNR607178S PNR607179S PNR607180S PNR607182S PNR607183S PNR607185S PNR607186S PNR607264S PNR607266S PNR607304S PNR607323S PNR607324S PNR607326S PNR607328S PNR607379S PNR607380S PNR607381S PNR607383S PNR607384S PNR607385S PNR607386S PNR607387S PNR607413S PNR607415S PNR607419S PNR607425S PNR607435S PNR607493S PNR607500S PNR607501S PNR607505S PNR607507S PNR607596S PNR607597S PNR607598S PNR607599S PNR607607S PNR607654S PNR607669S PNR607696S PNR607697S PNR607698S PNR607761S PNR607762S PNR607774S PNR607775S PNR607776S PNR607777S PNR607787S PNR607788S PNR607789S PNR607790S PNR607791S PNR607792S PNR607793S PNR607794S PNR607795S PNR607796S PNR607802S PNR607823S PNR607834S PNR607851S PNR607866S PNR607871S PNR607876S PNR607927S PNR608029S PNR608030S PNR608031S PNR608038S PNR608044S PNR608141S PNR608148S PNR608162S PNR608176S PNR608177S PNR608178S PNR608179S PNR608181S PNR608184S PNR608186S PNR608188S PNR608197S PNR608202S PNR608257S PNR608271S PNR608311S PNR608432S PNR608485S PNR608539S PNR608547S PNR608581S PNR608582S PNR608589S PNR608596S PNR608597S PNR608605S PNR608606S PNR608607S PNR608610S PNR608628S PNR608641S PNR608643S PNR608650S PNR608651S PNR608652S PNR608657S PNR608659S PNR608672S PNR608673S PNR608680S PNR608685S PNR609011S PNR609031S PNR609033S PNR609331S PNR609332S PNR609339S PNR609642S PNR609948S PNR609949S PNR610198S PNR610199S PNR610200S PNR610201S PNR610202S PNR610203S PNR610204S PNR610207S PNR610210S PNR610212S PNR610214S PNR610218S PNR610220S PNR610222S PNR610224S PNR610229S PNR610231S PNR610232S PNR610235S PNR610236S PNR610237S PNR610241S PNR610247S PNR610249S PNR610253S PNR610255S PNR610257S PNR610258S PNR610262S PNR610264S PNR610265S PNR611045S PNR611053S PNR611117S PNR611121S PNR611414S PNR611470S PNR611471S PNR611476S PNR611478S PNR611482S PNR611562S PNR611569S PNR611599S PNR611614S PNR611788S PNR611840S PNR611853S PNR611969S PNR612439S PNR612440S PNR612442S PNR612447S PNR612453S PNR612674S PNR612686S PNR612822S PNR612920S PNR612937S PNR613031S PNR613146S PNR613973S PNR614009S PVV617668S Product UPN/GTIN: 00885074300191 Serial Numbers: PNR600077S PNR600080S PNR600081S PNR600085S PNR600086S PNR600087S PNR600089S PNR600090S PNR600091S PNR600093S PNR600094S PNR600096S PNR600098S PNR600101S PNR600103S PNR600104S PNR600106S PNR600107S PNR600108S PNR600109S PNR600112S PNR600113S PNR600114S PNR600115S PNR600116S PNR600117S PNR600118S PNR600122S PNR600124S PNR600126S PNR600127S PNR600128S PNR600130S PNR600132S PNR600133S PNR600134S PNR600135S PNR600137S PNR600138S PNR600139S PNR600140S PNR600142S PNR600146S PNR600149S PNR600150S PNR600151S PNR600152S PNR600153S PNR600155S PNR600164S PNR600168S PNR600171S PNR600172S PNR600173S PNR600175S PNR600178S PNR600181S PNR600188S PNR600201S PNR600202S PNR600205S PNR600206S PNR600207S PNR600208S PNR600210S PNR600211S PNR600212S PNR600213S PNR600216S PNR600217S PNR600219S PNR600220S PNR600221S PNR600222S PNR600223S PNR600224S PNR600225S PNR600226S PNR600228S PNR600229S PNR600230S PNR600231S PNR600232S PNR600233S PNR600234S PNR600235S PNR600236S PNR600237S PNR600238S PNR600239S PNR600240S PNR600241S PNR600242S PNR600243S PNR600244S PNR600245S PNR600246S PNR600247S PNR600249S PNR600250S PNR600252S PNR600253S PNR600255S PNR600256S PNR600257S PNR600258S PNR600259S PNR600260S PNR600261S PNR600262S PNR600263S PNR600264S PNR600265S PNR600266S PNR600267S PNR600268S PNR600269S PNR600270S PNR600271S PNR600272S PNR600273S PNR600274S PNR600275S PNR600276S PNR600277S PNR600278S PNR600279S PNR600280S PNR600281S PNR600282S PNR600283S PNR600284S PNR600285S PNR600286S PNR600287S PNR600289S PNR600291S PNR600292S PNR600293S PNR600295S PNR600296S PNR600297S PNR600300S PNR600301S PNR600302S PNR600303S PNR600304S PNR600305S PNR600307S PNR600308S PNR600309S PNR600310S PNR600311S PNR600313S PNR600314S PNR600315S PNR600316S PNR600317S PNR600318S PNR600319S PNR600320S PNR600321S PNR600322S PNR600378S PNR600395S PNR600430S PNR600582S PNR600598S PNR600605S PNR600612S PNR600618S PNR600622S PNR600625S PNR600627S PNR600628S PNR600632S PNR600633S PNR600729S PNR600730S PNR600732S PNR600734S PNR600735S PNR600736S PNR600737S PNR600739S PNR600740S PNR600741S PNR600742S PNR600743S PNR600744S PNR600745S PNR600746S PNR600747S PNR600748S PNR600749S PNR600750S PNR600751S PNR600752S PNR600753S PNR600754S PNR600755S PNR600756S PNR600757S PNR600758S PNR600759S PNR600760S PNR600761S PNR600762S PNR600764S PNR600765S PNR600766S PNR600767S PNR600768S PNR600769S PNR600772S PNR600774S PNR600776S PNR600777S PNR600779S PNR600780S PNR600782S PNR600785S PNR600787S PNR600788S PNR600791S PNR600794S PNR600796S PNR600799S PNR600803S PNR600805S PNR600806S PNR600808S PNR600812S PNR600814S PNR600817S PNR600833S PNR600834S PNR600835S PNR600837S PNR600838S PNR600839S PNR600840S PNR600841S PNR600842S PNR600846S PNR600847S PNR600848S PNR600850S PNR600851S PNR600852S PNR600853S PNR600854S PNR600855S PNR600856S PNR600857S PNR600859S PNR600860S PNR600861S PNR600862S PNR600863S PNR600864S PNR600867S PNR600868S PNR600869S PNR600870S PNR600872S PNR600873S PNR600874S PNR600875S PNR600876S PNR600877S PNR600879S PNR600881S PNR600882S PNR600884S PNR600885S PNR600888S PNR600889S PNR600893S PNR600895S PNR600896S PNR600899S PNR600900S PNR600901S PNR600902S PNR600906S PNR600907S PNR600908S PNR600911S PNR600912S PNR600913S PNR600914S PNR600915S PNR600917S PNR600918S PNR600920S PNR600921S PNR600923S PNR600924S PNR600925S PNR600926S PNR600928S PNR600929S PNR600930S PNR600931S PNR600932S PNR600935S PNR600936S PNR600937S PNR600941S PNR600942S PNR600960S PNR600970S PNR600977S PNR600981S PNR600986S PNR600991S PNR600994S PNR601016S PNR601024S PNR601028S PNR601033S PNR601035S PNR601041S PNR601042S PNR601047S PNR601049S PNR601053S PNR601057S PNR601064S PNR601065S PNR601066S PNR601069S PNR601071S PNR601072S PNR601073S PNR601081S PNR601088S PNR601094S PNR601096S PNR601104S PNR601111S PNR601148S PNR601160S PNR601161S PNR601163S PNR601167S PNR601169S PNR601172S PNR601173S PNR601176S PNR601179S PNR601181S PNR601183S PNR601184S PNR601185S PNR601186S PNR601187S PNR601188S PNR601190S PNR601192S PNR601193S PNR601196S PNR601198S PNR601200S PNR601204S PNR601205S PNR601207S PNR601208S PNR601211S PNR601214S PNR601218S PNR601219S PNR601220S PNR601221S PNR601222S PNR601223S PNR601224S PNR601227S PNR601228S PNR601229S PNR601230S PNR601231S PNR601232S PNR601233S PNR601234S PNR601236S PNR601237S PNR601238S PNR601239S PNR601245S PNR601247S PNR601251S PNR601254S PNR601255S PNR601256S PNR601257S PNR601258S PNR601259S PNR601260S PNR601261S PNR601262S PNR601263S PNR601264S PNR601265S PNR601266S PNR601267S PNR601268S PNR601269S PNR601270S PNR601271S PNR601273S PNR601274S PNR601275S PNR601277S PNR601278S PNR601279S PNR601281S PNR601282S PNR601283S PNR601284S PNR601285S PNR601286S PNR601287S PNR601288S PNR601289S PNR601290S PNR601291S PNR601292S PNR601293S PNR601294S PNR601295S PNR601296S PNR601297S PNR601298S PNR601299S PNR601300S PNR601301S PNR601303S PNR601304S PNR601305S PNR601306S PNR601308S PNR601310S PNR601311S PNR601312S PNR601314S PNR601315S PNR601316S PNR601317S PNR601318S PNR601319S PNR601320S PNR601321S PNR601322S PNR601323S PNR601324S PNR601326S PNR601327S PNR601328S PNR601330S PNR601331S PNR601332S PNR601333S PNR601334S PNR601335S PNR601336S PNR601337S PNR601338S PNR601339S PNR601340S PNR601341S PNR601342S PNR601343S PNR601344S PNR601345S PNR601346S PNR601347S PNR601348S PNR601349S PNR601352S PNR601353S PNR601355S PNR601356S PNR601357S PNR601358S PNR601359S PNR601360S PNR601365S PNR601367S PNR601369S PNR601370S PNR601371S PNR601374S PNR601376S PNR601377S PNR601378S PNR601379S PNR601380S PNR601382S PNR601383S PNR601384S PNR601385S PNR601386S PNR601387S PNR601389S PNR601392S PNR601394S PNR601396S PNR601397S PNR601399S PNR601400S PNR601401S PNR601402S PNR601403S PNR601404S PNR601405S PNR601407S PNR601408S PNR601409S PNR601410S PNR601411S PNR601412S PNR601413S PNR601415S PNR601417S PNR601418S PNR601423S PNR601424S PNR601427S PNR601428S PNR601429S PNR601430S PNR601432S PNR601434S PNR601435S PNR601436S PNR601437S PNR601438S PNR601439S PNR601445S PNR601449S PNR601454S PNR601455S PNR601457S PNR601458S PNR601459S PNR601461S PNR601463S PNR601464S PNR601465S PNR601467S PNR601468S PNR601469S PNR601471S PNR601473S PNR601474S PNR601475S PNR601476S PNR601477S PNR601478S PNR601479S PNR601480S PNR601481S PNR601482S PNR601483S PNR601485S PNR601486S PNR601488S PNR601489S PNR601490S PNR601491S PNR601492S PNR601493S PNR601494S PNR601495S PNR601496S PNR601497S PNR601498S PNR601499S PNR601500S PNR601501S PNR601502S PNR601503S PNR601504S PNR601505S PNR601506S PNR601508S PNR601509S PNR601510S PNR601511S PNR601512S PNR601513S PNR601514S PNR601515S PNR601517S PNR601518S PNR601519S PNR601520S PNR601521S PNR601522S PNR601523S PNR601525S PNR601526S PNR601527S PNR601529S PNR601530S PNR601531S PNR601532S PNR601533S PNR601534S PNR601536S PNR601537S PNR601538S PNR601539S PNR601540S PNR601543S PNR601544S PNR601545S PNR601547S PNR601549S PNR601551S PNR601552S PNR601553S PNR601554S PNR601556S PNR601557S PNR601560S PNR601561S PNR601562S PNR601563S PNR601575S PNR601577S PNR601580S PNR601585S PNR601586S PNR601587S PNR601588S PNR601589S PNR601595S PNR601596S PNR601597S PNR601598S PNR601599S PNR601600S PNR601601S PNR601602S PNR601605S PNR601606S PNR601607S PNR601608S PNR601610S PNR601612S PNR601613S PNR601614S PNR601615S PNR601616S PNR601619S PNR601621S PNR601623S PNR601625S PNR601627S PNR601629S PNR601630S PNR601631S PNR601632S PNR601633S PNR601635S PNR601636S PNR601637S PNR601639S PNR601642S PNR601643S PNR601644S PNR601654S PNR601655S PNR601656S PNR601658S PNR601660S PNR601663S PNR601666S PNR601671S PNR601673S PNR601675S PNR601677S PNR601678S PNR601679S PNR601680S PNR601681S PNR601682S PNR601683S PNR601686S PNR601687S PNR601688S PNR601689S PNR601691S PNR601694S PNR601695S PNR601696S PNR601697S PNR601698S PNR601699S PNR601702S PNR601703S PNR601704S PNR601705S PNR601707S PNR601708S PNR601711S PNR601715S PNR601717S PNR601718S PNR601719S PNR601721S PNR601724S PNR601725S PNR601726S PNR601727S PNR601728S PNR601730S PNR601731S PNR601732S PNR601734S PNR601735S PNR601736S PNR601737S PNR601738S PNR601739S PNR601740S PNR601741S PNR601743S PNR601744S PNR601746S PNR601747S PNR601749S PNR601751S PNR601752S PNR601753S PNR601754S PNR601755S PNR601756S PNR601757S PNR601758S PNR601760S PNR601762S PNR601764S PNR601766S PNR601768S PNR601769S PNR601770S PNR601772S PNR601773S PNR601774S PNR601775S PNR601777S PNR601778S PNR601779S PNR601780S PNR601782S PNR601783S PNR601785S PNR601786S PNR601787S PNR601788S PNR601789S PNR601790S PNR601791S PNR601792S PNR601793S PNR601794S PNR601795S PNR601796S PNR601797S PNR601798S PNR601799S PNR601800S PNR601802S PNR601803S PNR601804S PNR601806S PNR601807S PNR601808S PNR601810S PNR601811S PNR601812S PNR601813S PNR601814S PNR601815S PNR601816S PNR601817S PNR601819S PNR601820S PNR601823S PNR601826S PNR601827S PNR601828S PNR601830S PNR601832S PNR601833S PNR601835S PNR601837S PNR601842S PNR601843S PNR601844S PNR601845S PNR601846S PNR601847S PNR601848S PNR601849S PNR601850S PNR601853S PNR601862S PNR601864S PNR601865S PNR601866S PNR601867S PNR601868S PNR601869S PNR601870S PNR601873S PNR601876S PNR601877S PNR601879S PNR601881S PNR601882S PNR601885S PNR601886S PNR601887S PNR601889S PNR601890S PNR601891S PNR601893S PNR601894S PNR601895S PNR601896S PNR601897S PNR601898S PNR601899S PNR601900S PNR601901S PNR601902S PNR601903S PNR601904S PNR601905S PNR601906S PNR601907S PNR601908S PNR601909S PNR601910S PNR601911S PNR601912S PNR601914S PNR601915S PNR601916S PNR601917S PNR601918S PNR601920S PNR601921S PNR601922S PNR601923S PNR601925S PNR601926S PNR601927S PNR601928S PNR601931S PNR601933S PNR601934S PNR601935S PNR601936S PNR601937S PNR601938S PNR601942S PNR601945S PNR601946S PNR601949S PNR601950S PNR601960S PNR601971S PNR602000S PNR602006S PNR602009S PNR602010S PNR602011S PNR602012S PNR602014S PNR602017S PNR602018S PNR602019S PNR602020S PNR602022S PNR602023S PNR602024S PNR602026S PNR602027S PNR602028S PNR602029S PNR602030S PNR602031S PNR602032S PNR602033S PNR602034S PNR602035S PNR602036S PNR602038S PNR602039S PNR602042S PNR602043S PNR602044S PNR602045S PNR602046S PNR602047S PNR602048S PNR602051S PNR602053S PNR602054S PNR602055S PNR602056S PNR602057S PNR602058S PNR602059S PNR602060S PNR602061S PNR602062S PNR602063S PNR602064S PNR602065S PNR602066S PNR602067S PNR602069S PNR602070S PNR602071S PNR602072S PNR602073S PNR602074S PNR602075S PNR602076S PNR602077S PNR602078S PNR602079S PNR602081S PNR602082S PNR602083S PNR602085S PNR602086S PNR602087S PNR602090S PNR602092S PNR602093S PNR602095S PNR602096S PNR602097S PNR602100S PNR602104S PNR602111S PNR602121S PNR602122S PNR602123S PNR602127S PNR602128S PNR602129S PNR602130S PNR602131S PNR602133S PNR602134S PNR602135S PNR602137S PNR602140S PNR602142S PNR602143S PNR602144S PNR602145S PNR602146S PNR602147S PNR602148S PNR602149S PNR602151S PNR602153S PNR602154S PNR602155S PNR602156S PNR602157S PNR602158S PNR602161S PNR602162S PNR602163S PNR602164S PNR602167S PNR602168S PNR602169S PNR602170S PNR602171S PNR602172S PNR602173S PNR602175S PNR602176S PNR602180S PNR602181S PNR602182S PNR602183S PNR602184S PNR602185S PNR602186S PNR602187S PNR602188S PNR602190S PNR602191S PNR602195S PNR602197S PNR602198S PNR602199S PNR602203S PNR602206S PNR602207S PNR602208S PNR602209S PNR602210S PNR602211S PNR602212S PNR602214S PNR602215S PNR602216S PNR602217S PNR602218S PNR602219S PNR602220S PNR602221S PNR602222S PNR602223S PNR602224S PNR602226S PNR602227S PNR602228S PNR602229S PNR602230S PNR602231S PNR602234S PNR602235S PNR602237S PNR602238S PNR602239S PNR602240S PNR602241S PNR602244S PNR602245S PNR602249S PNR602250S PNR602251S PNR602252S PNR602253S PNR602254S PNR602256S PNR602257S PNR602258S PNR602259S PNR602260S PNR602261S PNR602263S PNR602264S PNR602266S PNR602267S PNR602268S PNR602270S PNR602271S PNR602272S PNR602273S PNR602274S PNR602276S PNR602280S PNR602281S PNR602283S PNR602285S PNR602286S PNR602288S PNR602291S PNR602293S PNR602294S PNR602295S PNR602297S PNR602299S PNR602301S PNR602302S PNR602304S PNR602306S PNR602307S PNR602310S PNR602311S PNR602312S PNR602314S PNR602315S PNR602316S PNR602317S PNR602319S PNR602325S PNR602326S PNR602327S PNR602328S PNR602329S PNR602331S PNR602332S PNR602334S PNR602336S PNR602337S PNR602338S PNR602339S PNR602340S PNR602341S PNR602342S PNR602345S PNR602346S PNR602348S PNR602349S PNR602352S PNR602354S PNR602356S PNR602357S PNR602358S PNR602359S PNR602360S PNR602362S PNR602363S PNR602364S PNR602365S PNR602366S PNR602367S PNR602368S PNR602369S PNR602370S PNR602371S PNR602373S PNR602374S PNR602375S PNR602376S PNR602377S PNR602378S PNR602379S PNR602380S PNR602381S PNR602389S PNR602391S PNR602392S PNR602408S PNR602415S PNR602418S PNR602423S PNR602424S PNR602429S PNR602430S PNR602431S PNR602433S PNR602434S PNR602436S PNR602437S PNR602439S PNR602441S PNR602442S PNR602443S PNR602445S PNR602446S PNR602447S PNR602448S PNR602449S PNR602451S PNR602453S PNR602455S PNR602456S PNR602457S PNR602458S PNR602460S PNR602461S PNR602462S PNR602463S PNR602464S PNR602465S PNR602466S PNR602467S PNR602468S PNR602469S PNR602470S PNR602471S PNR602472S PNR602476S PNR602477S PNR602478S PNR602480S PNR602481S PNR602482S PNR602483S PNR602487S PNR602488S PNR602489S PNR602490S PNR602491S PNR602492S PNR602493S PNR602494S PNR602495S PNR602496S PNR602497S PNR602498S PNR602499S PNR602500S PNR602501S PNR602502S PNR602504S PNR602507S PNR602508S PNR602509S PNR602510S PNR602513S PNR602514S PNR602515S PNR602516S PNR602517S PNR602518S PNR602521S PNR602524S PNR602525S PNR602527S PNR602528S PNR602529S PNR602531S PNR602532S PNR602533S PNR602534S PNR602542S PNR602545S PNR602547S PNR602548S PNR602551S PNR602552S PNR602553S PNR602554S PNR602556S PNR602557S PNR602559S PNR602560S PNR602561S PNR602562S PNR602564S PNR602566S PNR602568S PNR602569S PNR602571S PNR602573S PNR602575S PNR602576S PNR602578S PNR602579S PNR602581S PNR602583S PNR602584S PNR602589S PNR602591S PNR602593S PNR602594S PNR602595S PNR602596S PNR602597S PNR602598S PNR602600S PNR602601S PNR602602S PNR602605S PNR602607S PNR602608S PNR602609S PNR602611S PNR602612S PNR602613S PNR602615S PNR602616S PNR602618S PNR602619S PNR602620S PNR602623S PNR602625S PNR602627S PNR602628S PNR602629S PNR602630S PNR602631S PNR602632S PNR602633S PNR602634S PNR602635S PNR602636S PNR602637S PNR602638S PNR602639S PNR602640S PNR602641S PNR602642S PNR602643S PNR602645S PNR602646S PNR602647S PNR602648S PNR602649S PNR602650S PNR602651S PNR602652S PNR602653S PNR602654S PNR602655S PNR602656S PNR602657S PNR602658S PNR602659S PNR602660S PNR602661S PNR602662S PNR602663S PNR602664S PNR602665S PNR602667S PNR602668S PNR602669S PNR602670S PNR602671S PNR602673S PNR602675S PNR602676S PNR602679S PNR602681S PNR602682S PNR602683S PNR602684S PNR602685S PNR602688S PNR602689S PNR602690S PNR602692S PNR602695S PNR602696S PNR602697S PNR602698S PNR602699S PNR602700S PNR602702S PNR602703S PNR602704S PNR602705S PNR602706S PNR602707S PNR602709S PNR602710S PNR602715S PNR602722S PNR602725S PNR602796S PNR602797S PNR602798S PNR602799S PNR602869S PNR602894S PNR602899S PNR602907S PNR602942S PNR602964S PNR602974S PNR602976S PNR602978S PNR602979S PNR602980S PNR602982S PNR602985S PNR602986S PNR602991S PNR602992S PNR602994S PNR602995S PNR602997S PNR602999S PNR603001S PNR603005S PNR603007S PNR603013S PNR603014S PNR603018S PNR603019S PNR603023S PNR603024S PNR603027S PNR603028S PNR603033S PNR603034S PNR603036S PNR603037S PNR603040S PNR603041S PNR603045S PNR603046S PNR603050S PNR603051S PNR603052S PNR603063S PNR603066S PNR603071S PNR603072S PNR603074S PNR603081S PNR603083S PNR603085S PNR603089S PNR603091S PNR603123S PNR603265S PNR603280S PNR603291S PNR603293S PNR603311S PNR603313S PNR603315S PNR603316S PNR603321S PNR603323S PNR603325S PNR603329S PNR603330S PNR603333S PNR603334S PNR603336S PNR603338S PNR603341S PNR603345S PNR603347S PNR603349S PNR603352S PNR603355S PNR603356S PNR603358S PNR603359S PNR603366S PNR603367S PNR603369S PNR603370S PNR603374S PNR603376S PNR603379S PNR603385S PNR603386S PNR603389S PNR603390S PNR603391S PNR603392S PNR603393S PNR603397S PNR603398S PNR603403S PNR603404S PNR603406S PNR603408S PNR603410S PNR603411S PNR603416S PNR603420S PNR603428S PNR603430S PNR603433S PNR603440S PNR603441S PNR603444S PNR603445S PNR603447S PNR603454S PNR603455S PNR603459S PNR603462S PNR603463S PNR603464S PNR603470S PNR603475S PNR603482S PNR603486S PNR603488S PNR603489S PNR603490S PNR603491S PNR603494S PNR603495S PNR603496S PNR603503S PNR603511S PNR603514S PNR603515S PNR603516S PNR603519S PNR603520S PNR603521S PNR603529S PNR603564S PNR603646S PNR603670S PNR603982S PNR603995S PNR604158S PNR604160S PNR604161S PNR604223S PNR604411S PNR604418S PNR604425S PNR604454S PNR604463S PNR604465S PNR604469S PNR604477S PNR604484S PNR604489S PNR604495S PNR604496S PNR604497S PNR604498S PNR604499S PNR604500S PNR604501S PNR604502S PNR604503S PNR604504S PNR604505S PNR604506S PNR604508S PNR604509S PNR604510S PNR604511S PNR604512S PNR604513S PNR604515S PNR604519S PNR604524S PNR604525S PNR604526S PNR604527S PNR604528S PNR604529S PNR604530S PNR604531S PNR604532S PNR604533S PNR604534S PNR604535S PNR604536S PNR604537S PNR604538S PNR604539S PNR604540S PNR604541S PNR604550S PNR604551S PNR604552S PNR604553S PNR604554S PNR604556S PNR604557S PNR604558S PNR604559S PNR604560S PNR604561S PNR604562S PNR604563S PNR604564S PNR604565S PNR604566S PNR604567S PNR604568S PNR604569S PNR604570S PNR604571S PNR604572S PNR604573S PNR604574S PNR604575S PNR604577S PNR604578S PNR604579S PNR604580S PNR604581S PNR604582S PNR604583S PNR604584S PNR604585S PNR604586S PNR604588S PNR604589S PNR604590S PNR604591S PNR604592S PNR604593S PNR604594S PN
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2417-2018
Related Recalls
ICD-DR DDBB2D1 EVERA XT IS1/DF1 INTL, Model Number DDBB2D1; Implantable Cardioverter Defibrillators
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
ICD-DR DDMC3D1 EVERA MRI S OUS/US DF1, Model Number DDMC3D1; Implantable Cardioverter Defibrillators
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
ICD DDPA2D1G COBALT XT DR MRI DF1 GOLD, Model Number DDPA2D1G; Implantable Cardioverter Defibrillators
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.