Recall Details
- Company:
- Stryker GmbH
- Reason for Recall:
- The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-conforming cannulation of the nails may result in reduced component strength and potentially premature nail breakage.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Ankle Arthrodesis Nail, right T2 Ankle ¿11x300mm, Product Number 18191130S
Product Codes/Lot Numbers:
GTIN: 04546540673428; Lot code: K081639
Distribution:
Distributed in: US, AL, FL, KY, NC, TN, UT, WA
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2408-2018
Related Recalls
Investigation determined that one lot of the T2 Greater Trochanter Nail (GTN) was labeled as left version but contained a right nail and another lot was labeled as right version but contained a left nail.
Long Nail, RIGHT Gamma4 10x320mm x 125, component of the Gamma4 System. Catalog Number: 84250320S.
Stryker GmbH
Investigation confirmed that the GAMMA4 Right Long Nail was manufactured with the geometry of a left nail.
Investigation determined that one lot of the T2 Greater Trochanter Nail (GTN) was labeled as left version but contained a right nail and another lot was labeled as right version but contained a left nail.