RenLane Renal Denervation Catheter; Cat No.: D135601. Used in adults patients (>18 years) with drug resistant hypertension to denervate the renal arteries to reduce blood pressure.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Model Number: D135601 Lots #'s 15958389LA, 15959088LA, 15959090LA, 15968753LA, 15968806LA, 15968810LA, and 15968813LA, with Exp Date 8/31/2014 Lots #'s 15976735L, 15976820L, 15976823L, 15978174L, 15978179L, 159784180L, and 15978181L with Exp Date 9/30/2014
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Cordis Corporation
Reason for Recall:
Potential Damage may occur to the helical tip of the RENLANE Renal Denervation Catheter, if the hemostasis valve of the arterial access device is not fully open before insertion or withdrawal of the catheter.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

RenLane Renal Denervation Catheter; Cat No.: D135601. Used in adults patients (>18 years) with drug resistant hypertension to denervate the renal arteries to reduce blood pressure.

Product Codes/Lot Numbers:

Model Number: D135601 Lots #'s 15958389LA, 15959088LA, 15959090LA, 15968753LA, 15968806LA, 15968810LA, and 15968813LA, with Exp Date 8/31/2014 Lots #'s 15976735L, 15976820L, 15976823L, 15978174L, 15978179L, 159784180L, and 15978181L with Exp Date 9/30/2014

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2405-2015