Femoral Nail, A/R T2 Femur ¿10x340 mm, Product Number 18251034S

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    GTIN 04546540199508; Lot code: K0957B1
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Stryker GmbH
Reason for Recall:
The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-conforming cannulation of the nails may result in reduced component strength and potentially premature nail breakage.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Femoral Nail, A/R T2 Femur ¿10x340 mm, Product Number 18251034S

Product Codes/Lot Numbers:

GTIN 04546540199508; Lot code: K0957B1

Distribution:

Distributed in: US, AL, FL, KY, NC, TN, UT, WA

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2395-2018

Related Recalls