SEDASYS Computer-Assisted Personalized Sedation System, SEDPRU01. An integrated physiological monitoring and drug delivery system intended to provide physician-led teams with a means of administering 1% propofol (10 mg/mL) injectable emulsion to sedate patients > 18 (ASA I-II) undergoing esophagogastroduodenoscopy (EGD) and colonoscopy procedures. Intended to provide physician-led teams with a means of administering 1% propofol (10 mg/mL) injectable emulsion to sedate patients > 18 (ASA I-II) undergoing esophagogastroduodenoscopy (EGD) and colonoscopy procedures.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Model #SEDPRU01
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Ethicon Endo-Surgery Inc
Reason for Recall:
Ethicon has found that disinfecting methods not specified in the Instructions For Use (IFU) have led to the degradation of the plastic material of the SEDASYS System Control Unit.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

SEDASYS Computer-Assisted Personalized Sedation System, SEDPRU01. An integrated physiological monitoring and drug delivery system intended to provide physician-led teams with a means of administering 1% propofol (10 mg/mL) injectable emulsion to sedate patients > 18 (ASA I-II) undergoing esophagogastroduodenoscopy (EGD) and colonoscopy procedures. Intended to provide physician-led teams with a means of administering 1% propofol (10 mg/mL) injectable emulsion to sedate patients > 18 (ASA I-II) undergoing esophagogastroduodenoscopy (EGD) and colonoscopy procedures.

Product Codes/Lot Numbers:

Model #SEDPRU01

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2374-2015

Related Recalls

Firm has received an increase in reports regarding inadvertent instrument lockout during surgical procedures involving the device. If lockout occurs, the device will momentarily activate but will not cut or staple tissue, and additional steps will be required to open it and remove it from tissue. If user is unable to remove the device from the tissue, potential harms to patient include surgical delay, bleeding/hemorrhage, life threatening hemorrhage/hemorrhagic shock, and conversion to open therapy. Two adverse events received, including one patient death.

Apr 22, 2025 Other Medical Devices Nationwide View Details →