Recall Details

Company:
Biomerieux Inc
Reason for Recall:
The integrity of the product container may be compromised thereby exposing product to moisture which could lead to antibiotic degradation (loss of potency).
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

VITEK¿ 2 Gram Positive Susceptibility card (AST-P606), REF 22330, 20 cards per carton.

Product Codes/Lot Numbers:

4860165103 10/May/18 4860228203 12/Jul/18

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2365-2017

Related Recalls

Potential for false resistant results for Colistin (cs02n) with multi-drug-resistant Pseudomonas aeruginosa and Acinetobacter baumannii complex used in specific cards

Jun 26, 2025 Diagnostic Equipment View Details →

Ceftriaxone (cro02n) concentration errors associated with specific recently manufactured test kits which would have the potential for for false susceptible results. The risk is present for isolates that have a result of MIC=0.5, 1, 2.

Mar 14, 2024 Diagnostic Equipment View Details →

For users with MYLA V4.8.X / V4.9 that use VITEK MS to identify organisms as part of their workflow, AST filter rules that have been activated are not always being applied to AST results when sending the results to the clinician.

Feb 6, 2023 Surgical Instruments Nationwide View Details →