UroLift Visual Obturator, 20Ch-The visual obturator is used with the UroLift¿ system. Instrument for insertion of a cystoscope shaft into the bladder. Model: UL-VO

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI-DI: 04250480106617 Lot Number: 141980
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Scholly Fiberoptic Gmbh
Reason for Recall:
There is an improperly performed weld between the cone of the sheath lock and tube of the obturator; mechanical load capacity is impaired, and fracture of the welded joint may occur. Contamination can occur in this capillary gap, which is not covered by the reprocessing validations performed.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

UroLift Visual Obturator, 20Ch-The visual obturator is used with the UroLift¿ system. Instrument for insertion of a cystoscope shaft into the bladder. Model: UL-VO

Product Codes/Lot Numbers:

UDI-DI: 04250480106617 Lot Number: 141980

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2364-2023