Alinity ci -series System Control¿Module, Part Number 03R70 01 Product Usage: The Alinity ci series System Control Module has a scalable design to provide full integration of clinical chemistry and immunoassay analysis. The Alinity ci series System Control Module includes a clinical chemistry module and an immunoassay module, each performing all sample processing activities, and a system control module to provide a single user friendly interface. Each system can be customized by combining one or more sample processing modules, in multiple configurations, with a system control module to form a single workstation. The Alinity i Processing Module is a fully automated immunoassay analyzer allowing random and continuous access, as well as priority and automated retest processing using chemiluminescent microparticle immunoassay (CMIA) technology. CMIA technology is used to determine the presence of antigens, antibodies, and analytes in samples.
Class I - DangerousWhat Should You Do?
- Check if you have this product: All serial numbers
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Abbott Gmbh & Co. KG
- Reason for Recall:
- Abbott has identified an issue with all on market versions (v2.6.0 and v2.6.1) of Alinity ci series Software where incorrect results may occur after a system Stop due to the Alinity i re use of reaction vessels (RVs). This issue only occurs if the system is transitioned from Processing to Stopped to Idle.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Alinity ci -series System Control¿Module, Part Number 03R70 01 Product Usage: The Alinity ci series System Control Module has a scalable design to provide full integration of clinical chemistry and immunoassay analysis. The Alinity ci series System Control Module includes a clinical chemistry module and an immunoassay module, each performing all sample processing activities, and a system control module to provide a single user friendly interface. Each system can be customized by combining one or more sample processing modules, in multiple configurations, with a system control module to form a single workstation. The Alinity i Processing Module is a fully automated immunoassay analyzer allowing random and continuous access, as well as priority and automated retest processing using chemiluminescent microparticle immunoassay (CMIA) technology. CMIA technology is used to determine the presence of antigens, antibodies, and analytes in samples.
Product Codes/Lot Numbers:
All serial numbers
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2363-2019
Related Recalls
Alinity ci series System Control Module (SCM); LN 3R70-01 Alinity ci series software versions, 3.1.1 and below
Abbott Gmbh & Co. KG
Quality Control results were not properly evaluated for QC failures.
Alinity C, Processing Module. Chemistry analyzer for in-vitro diagnostics.
Abbott Gmbh & Co. KG
There is a potential to generate incorrect patient results for the following assays: Alkaline Phosphatase, Amylase, Creatine Kinase, and Gamma Glutamyl Transferase.
Abbott identified that samples tested for ARCHITECT STAT High Sensitive Troponin I and STAT High Sensitivity Troponin I assays (LN 3P25 and LN 2R98) may show interaction when processed directly after the ARCHITECT iGentamicin (LN 1P31) assay and patient results might be impacted.