MassHunter Quantitative Analysis Software, Version B.07.01, used with Triple Quadrupole Mass Spectrometers, models K6460 and K6420 Product Usage: A mass spectrometer for clinical use is a device intended to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens by ionizing the compound under investigation and separating the resulting ions by means of an electrical and magnetic field according to their mass.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Software versions B.07 or below
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Agilent Technologies, Inc.
Reason for Recall:
When utilizing specific software versions of an LC/MS device, and Batch at a Glance in Compound Table view, a defect occurs when a new sample(s) is inserted. This defect creates a mismatch between the sample name and the column header whereby the sample header will be offset by one sample. The recalling firm requests that you discontinue using this workflow to generate reports because the Quant batch table will not display fully analyzed results or save any changes made.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

MassHunter Quantitative Analysis Software, Version B.07.01, used with Triple Quadrupole Mass Spectrometers, models K6460 and K6420 Product Usage: A mass spectrometer for clinical use is a device intended to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens by ionizing the compound under investigation and separating the resulting ions by means of an electrical and magnetic field according to their mass.

Product Codes/Lot Numbers:

Software versions B.07 or below

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2359-2019

Related Recalls

DNA Labeling Kit for use in sample labeling to be hybridized to oligonucleotide microarrays has a component, Human Reference DNA Female aliquot, affected by a mosaic aberration that may result in Chromosome 12 baseline shift, which may lead to false positive/negative results, resulting in diagnosis delay, and may lead to delayed treatment of a treatable genetic disorder.