microTargeting" STar" Drive System (Manual) Distributed by Medtronic, Product Number FC8001
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI: 00873263001486, 00873263006351
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- FHC, Inc.
- Reason for Recall:
- The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
microTargeting" STar" Drive System (Manual) Distributed by Medtronic, Product Number FC8001
Product Codes/Lot Numbers:
UDI: 00873263001486, 00873263006351
Distribution:
Nationwide distribution
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2357-2018
Related Recalls
A complaint was received (MDR 3002250546-2024-00003) prompting an investigation. The investigation revealed a schematic for a spacer tube which is too long relative to the size of other components was included in the build order for Lot 244517 of P/N 66-IT-06P. Use of the product with a spacer tube which is too long could result in intercranial hemorrhaging and brain damage.
An electrical sub-circuit was not properly connected. Its purpose is to eliminate the possibility of interference with proper performance of the Guideline 5 Main Processing Unit, if an ESD event (greater than 6kV) were to occur.
The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.