Accu-Chek Performa Test Strip, Blood Glucose Monitoring System, Model number 07299702001
Class I - DangerousWhat Should You Do?
- Check if you have this product: Since this is a labeling correction, this recall is not lot specific and is applicable to all non-expired lots.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Roche Diabetes Care, Inc.
- Reason for Recall:
- Test strip vials may open while inside sealed cartons during shipment. An open vial might expose the test strips to humidity, which might damage the strips and result in inaccurate results, potentially leading to inappropriate therapy decisions and adverse health consequences.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Accu-Chek Performa Test Strip, Blood Glucose Monitoring System, Model number 07299702001
Product Codes/Lot Numbers:
Since this is a labeling correction, this recall is not lot specific and is applicable to all non-expired lots.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2352-2021
Related Recalls
The reason for recall is the meters may show an incorrect measurement unit (mmol/L rather than mg/dL) and may result in the wrong unit of measure. Which may subsequently cause an error in interpreting a result as approximately 18 times lower that the actual blood glucose. Potential immediate health consequences of the glucose meter providing a reading in the wrong units of measurement (UOM) may result in inappropriate rescue therapy for presumed hypoglycemia including urgent and repeat administration of carbohydrates. This in turn could lead to frequent episodes of hyperglycemia which is unrecognized and undertreated.Long range consequences of continued use of the affected meters could include mismanagement of the patient s diabetes for longer periods of time potentially resulting in a major health event (in general persistent and/or severe hyperglycemia) that may lead them to seek medical attention, and depending on the duration of hyperglycemia, could contribute to the development of microvascular (i.e., retinopathy, neuropathy, nephropathy) and/or macrovascular (i.e., myocardial infarction, ischemic stroke) complications of diabetes and could lead to fetal damage (abnormal fetal growth and altered organ development and maturation, fetal hypoxemia, spontaneous abortion, and congenital anomalies) in pregnant women. This issue was identified by three initial consumer complaints, with reports that the meters were presenting the incorrect unit of measure.
RocheDiabetes Care Platform Software version 2.5.2 material number 08888353001. This is a web application software hosted on a server. There is no physical product.
Roche Diabetes Care
Potential for patient data mismatch when using browser "back" button to navigate between patients when using the diabetes management software.
Accu-Chek Aviva Plus Test Strip, Blood Glucose Monitoring System, Model numbers: 06908152001, 06908217001, 06908268001, 06908349001, 06908373001
Roche Diabetes Care
Test strip vials may open while inside sealed cartons during shipment. An open vial might expose the test strips to humidity, which might damage the strips and result in inaccurate results, potentially leading to inappropriate therapy decisions and adverse health consequences.