microTargeting" STar" Drive System (Manual), Product Number ST-DS-MA

Class I - Dangerous
🏥 Medical Devices Recalled: May 10, 2018 FHC Infusion Pumps Nationwide

What Should You Do?

  1. Check if you have this product:
    UDI: 00873263004418
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
FHC, Inc.
Reason for Recall:
The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

microTargeting" STar" Drive System (Manual), Product Number ST-DS-MA

Product Codes/Lot Numbers:

UDI: 00873263004418

Distribution:

Nationwide distribution

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2349-2018

Related Recalls

A complaint was received (MDR 3002250546-2024-00003) prompting an investigation. The investigation revealed a schematic for a spacer tube which is too long relative to the size of other components was included in the build order for Lot 244517 of P/N 66-IT-06P. Use of the product with a spacer tube which is too long could result in intercranial hemorrhaging and brain damage.

Oct 4, 2024 Surgical Instruments View Details →

An electrical sub-circuit was not properly connected. Its purpose is to eliminate the possibility of interference with proper performance of the Guideline 5 Main Processing Unit, if an ESD event (greater than 6kV) were to occur.

Jun 17, 2020 Surgical Instruments Nationwide View Details →

The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

May 10, 2018 Infusion Pumps Nationwide View Details →