RTL190395: 2008K@HOME MACHINE, SHORT CAB,OLC/DP,HP - Blood Pump Rotor spare part, Model Number: F40015481 Rev A

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Model Number: RTL190395 (Rental Units); UDI/DI: 00840861100958; Serial Numbers: 4K0S-R138, 5K0S-Z652, 5K0S-462B, 5K0S-Z669, 6K0S-505F, 6K0S-513F, 6K0S-850G, 6K0S-480J, 6K0S-467K, 6K0S-995P, 7K0S103784, 6K0S-009M, 7K0S110463, 7K0S110427, 7K0S110771, 6K0S-725R, 6K0S-324L,
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Fresenius Medical Care Holdings, Inc.
Reason for Recall:
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

RTL190395: 2008K@HOME MACHINE, SHORT CAB,OLC/DP,HP - Blood Pump Rotor spare part, Model Number: F40015481 Rev A

Product Codes/Lot Numbers:

Model Number: RTL190395 (Rental Units); UDI/DI: 00840861100958; Serial Numbers: 4K0S-R138, 5K0S-Z652, 5K0S-462B, 5K0S-Z669, 6K0S-505F, 6K0S-513F, 6K0S-850G, 6K0S-480J, 6K0S-467K, 6K0S-995P, 7K0S103784, 6K0S-009M, 7K0S110463, 7K0S110427, 7K0S110771, 6K0S-725R, 6K0S-324L,

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2344-2024

Related Recalls

Bicarby Dialysate; Model number: RFP-400-G;

Fresenius Medical Care Holdings

Class I - Dangerous

There is an error in the labeling of the GTIN on the product. When scanned; however, the case labeling, lot number, and printed labeling on the product correctly identifies the product code and the solution contained in the bag. In the event the incorrect GTIN labeling is used as the singular source for product identification, unrecognized use of a potassium solution with a concentration differing from the prescription may result.

May 7, 2026 Other Medical Devices View Details →

Given the increased number of leaks reported by users, it was determined to be the appropriate course of action to notify users of the potential for the occurrence of Bicarby luer-lock leaks during manipulation of the frangible and tubing when preparing the product for use for treatment. This issue presents the potential of a slip and fall hazard associated with fluid on the floor.

May 12, 2026 Other Medical Devices Nationwide View Details →