Battery Operated infusion pumps. t:slim X2 G5, Replacement, Refurbished, REF: 1000911; t:slim X2 GS Classic, Replacement, Refurbished, REF: 1003808; t: slim X2, Basal-IQ, mg/dl, REF: 1004219; Pump, t:slim X2, Clinical Use Only, REF: 1004484; Pump, t:slim X2, Control-IQ, REF: 1005611; Pump, t:slim X2, Control-IQ, mmoUL, Refurbished Replacement, REF: 1005613; Pump, t:slim X2, Control-IQ, mg/dl, REF: 1005615; Pump, t:slim X2, Control-IQ, mg/dl, Refurbished Replacement, REF: 1005617; Pump, t:slim X2, Basal-IQ 6.4, Refurbished Replacement, REF: 1006379; Pump, t:slim X2, Control-IQ 7.4, REF: 1006402; Pump, t:slim X2, Control-IQ 7.4, Replacement, REF: 1006406; Pump, t:slim X2, Control-IQ 7.4, REF: 1006408; Pump, t:slim X2, Basal-IQ, , REF: 1006419; Pump, t:slim X2, Basal-IQ, mg/dL, REF: 1006420; Pump, t:slim X2, Basal-IQ, mmoUL, Refurbished Replacement, REF: 1006535; Pump, t:slim X2, Basal-IQ, mg/dL, Refurbished Replacement, REF: 1006537; Pump, t:slim X2, Control-IQ, mg/dl , Replacement, REF: 1008255; Pump, t:slim X2, Control-IQ, mmoUL, Replacement, REF: 1008256; Pump, t:slim X2, Basal-IQ, mmoUL, Replacement, REF: 1008257; Pump, t:slim X2, Basal-IQ, mg/dl ,Replacement, REF: 1008258; Pump, t:slim X2, Basal-IQ, REF: 1010004; Pump, t:slim X2, Basal-IQ, Replacement, REF: 1010005; Pump, t:slim X2, Basal-IQ, Medicare, REF: 1010006; Pump, t:slim X2, Basal-IQ, Medicare, REF: 1010007; Pump, t:slim X2, Control-IQ, REF: 1010009; Pump, t:slim X2, Control-IQ Replacement, REF: 1010010; Pump, t:slim X2, Control-IQ, Medicare, REF: 1010011; Pump, t:slim X2, Control-IQ, Refurbished Replacement, REF: 1010012; Pump, t:slim X2, Control-IQ, REF: 1012867; Pump, t:slim X2, Control-IQ, Medicare, REF: 1012868; Pump, t:slim X2, Control-IQ, Replacement, REF: 1012869; Pump, t:slim X2, Control-IQ 7.8, REF: 1013195; Pump, t:slim X2, Control-IQ 7.8, Medicare, REF: 1013196; Pump, t :slim X2, Control-IQ 7.8, Replacement, REF: 1013197; Pump, t:slim X2, Control-IQ 7.8, Refurbished Replacement, REF: 1013198
Class I - DangerousWhat Should You Do?
- Check if you have this product: REF/UDI-DI: 1000911/t:slim X2 G5, Replacement, Refurbished/00853052007318; 1003808/t:slim X2 GS Classic, Replacement, Refurbished/ 00853052007981; 1004219/t: slim X2, Basal-IQ, mg/dl/ 00853052007998; 1004484/Pump, t:slim X2, Clinical Use Only; 1005611/Pump, t:slim X2, Control-IQ, mmoUL/ 00850006613304/; 1005613/Pump, t:slim X2, Control-IQ, mmoUL, Refurbished Replacement/ 00850006613311/; 1005615/Pump, t:slim X2, Control-IQ, mg/dl/ 00850006613328/; 1005617/Pump, t:slim X2, Control-IQ, mg/dl, Refurbished Replacement/ 00850006613335/; 1006379/Pump, t:slim X2, Basal-IQ 6.4, Refurbished Replacement/ 00850006613731/; 1006402/Pump, t:slim X2, Control-IQ 7.4/ 00850006613762/; 1006406/Pump, t:slim X2, Control-IQ 7.4, Replacement/ 00850006613786/; 1006408/Pump, t:slim X2, Control-IQ 7.4/ 00850006613793/; 1006419/Pump, t:slim X2, Basal-IQ, mmoUL/ 00850006613809/; 1006420/Pump, t:slim X2, Basal-IQ, mg/dL 00850006613816; 1006535/Pump, t:slim X2, Basal-IQ, mmoUL, Refurbished Replacement/ 00850006613823/; 1006537/Pump, t:slim X2, Basal-IQ, mg/dL, Refurbished Replacement/ 00850006613830/; 1008255/Pump, t:slim X2, Control-IQ, mg/dl , Replacement/00850018992671; 1008256/Pump, t:slim X2, Control-IQ, mmoUL, Replacement/00850018992664; 1008257/Pump, t:slim X2, Basal-IQ, mmoUL, Replacement/00850018992657; 1008258/Pump, t:slim X2, Basal-IQ, mg/dl ,Replacement/00850018992640; 1010004/Pump, t:slim X2, Basal-IQ/00389152000015; 1010005/Pump, t:slim X2, Basal-IQ, Replacement/00389152000022; 1010006/Pump, t:slim X2, Basal-IQ, Medicare/00389152000039; 1010007/Pump, t:slim X2, Basal-IQ, Medicare/00389152000046; 1010009/Pump, t:slim X2, Control-IQ/00389152000077; 1010010/Pump, t:slim X2, Control-IQ Replacement/00389152000084; 1010011/Pump, t:slim X2, Control-IQ, Medicare/00389152000091; 1010012/Pump, t:slim X2, Control-IQ, Refurbished Replacement/00389152000107; 1012867/Pump, t:slim X2, Control-IQ/00389152286716; 1012868/Pump, t:slim X2, Control-IQ, Medicare/00389152286815; 1012869/Pump, t:slim X2, Control-IQ, Replacement/00389152286914; 1013195/Pump, t:slim X2, Control-IQ 7.8/00389152319513; 1013196/Pump, t:slim X2, Control-IQ 7.8, Medicare/00389152319612; 1013197/Pump, t :slim X2, Control-IQ 7.8, Replacement/00389152319711; 1013198/Pump, t:slim X2, Control-IQ 7.8, Refurbished Replacement/00389152319810 Serial Numbers: 951824 1009214 1009818 1010205 1010216 1010280 1010282 1124339 1124359 1120792 1284185 1096196 1011443 1127741 1017739 1088141 1023688 1084425 1028167 1028228 1033168 1109355 1035333 1044731 1037937 1038725 1040458 1040729 1044512 1041052 1041217 1041397 1041457 1156947 1043829 1044400 1044800 1284661 1049585 1049643 1053348 1186711 1191406 1121799 1132861 1124364 1123627 1063736 1064271 1064593 1106444 1067539 1067828 1067887 1172701 1174318 1098764 1110835 1073710 1128723 1124356 1121152 1127745 1076858 1085036 1101544 1088173 1113296 1109496 1132398 1099207 1134096 1109488 1188241 1188233 1187866 1174100 1177299 1186285 1085203 1086053 1169060 1098778 1099063 1098525 1132410 1134897 1139347 1149413 1149109 1109118 1147667 1149987 1193922 1095865 1148061 1120208 1125365 1127634 1127974 1139150 1138109 1098244 1149017 1099022 1115201 1115510 1101725 1195419 1160890 1195424 1190520 1195271 1190631 1193090 1188686 1151553 1196393 1157533 1011994 1011686 1010370 1010330 1010149 1010339 1009580 1010186 1018441 1018522 1020329 1020438 1020404 1021017 1020215 1018419 1018458 1020226 1018560 1016948 1020578 1020553 1020572 1020531 1020263 1020113 1020100 1020054 1020712 1020104 1020560 1020655 1020487 1023365 1022648 1021118 1020917 1020373 1024186 1023401 1023533 1024261 1023478 1022834 1022835 1022828 1024374 1024315 1024201 1024386 1010323 1008806 1008828 1009705 1009241 1009735 1009666 1009722 1008939 1009686 1009704 1009284 1009191 1009123 1010025 1010059 1010067 1010279 1009063 1008575 1010022 1010277 1010338 1010221 1010107 1010315 1010211 1008487 1008531 1008486 1008514 1008467 1008820 1008760 1008615 1008904 1008694 1021317 1011702 1019671 1021237 1021192 1021279 1021324 1021318 1010379 1021299 1021301 1010249 1019683 1011761 1021325 1021261 1019418 1021284 1010359 1010356 1064999 1044969 1044176 1191892 1191939 1152446 1195279 1185661 1134013 1072356 1125364 1103406 1128033 1127345 1098206 1098758 1132314 1138833 1138876 1137809 1104141 1164437 1148205 1148307 1148565 1148163 1148709 1149419 1158667 1180647 1138036 1139540 1139510 1037782 1098256 1041793 1041139 1119177 1117738 1190568 1044931 1044351 1040764 1160757 1147747 1097466 1159784 1148530 1151030 1174104 1110283 1101561 1152960 1136798 1186356 1186639 1188142 1186579 1186590 1099572 1102128 1059066 1101740 1134163 1186587 1049144 1085902 1086054 1151146 1089176 1080141 1095314 1109999 1042173 1031387 1038019 1038543 1190762 1160289 1160568 1167268 1196846 1151805 1164867 1077168 1172909 1065134 1177890 1139118 1124251 1182202 1190766 1195370 1139765 1139675 1078571 1169487 1154526 1131902 1192577 1174388 1177382 1179618 1186544 1178243 1185555 1186681 1193040 1127529 1124667 1048221 1043088 1062012 1064036 1072664 1072131 1069163 1085921 1085990 1077115 1098578 1100659 1028390 1133023 1139699 1134164 1139733 1142649 1141467 1027083 1027674 1098353 1115931 1040780 1133914 1133659 1148066 1147771 1149703 1151165 1085987 1103320 1120657 1125266 1120217 1120109 1133124 1125657 1199271 1089240 1044786 1050174 1127450 1051372 1085255 1084995 1098309 1063996 1098494 1060892 1041467 1040982 1132281 1019268 1136846 1138865 1040584 1041466 1034158 1167994 1030767 1030872 1028097 1088069 1148309 1094754 1094258 1095557 1109170 1079138 1154688 1129307 1117370 1121728 1112236 1129842 1038028 1139532 1167107 1132327 1132871 1046471 1049938 1044325 1046565 1046564 1050219 1127590 1055444 1161899 1171637 1061294 1189986 1190593 1064898 1059461 1109196 1190998 1113847 1152496 1109208 1041407 1041427 1023013 1033115 1032409 1174373 1184801 1186628 1128584 1135284 1136631 1139728 1139139 1191337 1193055 1191584 1190739 1031121 1028166 1169532 1177589 1174320 1132865 1133050 1133066 1128536 1177786 1178743 1098255 1120100 1184807 1085097 1101946 1133946 1138817 1139310 1122071 1185606 1085624 1086793 1088003 1052447 1038695 1106503 1041455 1038079 1037491 1148174 1028206 1043709 1040735 1151381 1043039 1027142 1072103 1097921 1097858 1159873 1117464 1100615 1104217 1187916 1077458 1050400 1102546 1141218 1128104 1113002 1085984 1095329 1051040 1115605 1052544 1098977 1074138 1148055 1132301 1136965 1138765 1041090 1188435 1188531 1051922 1190303 1177860 1029585 1195324 1182299 1187987 1188086 1149405 1115558 1151447 1044354 1152129 1057291 1041275 1038086 1060564 1133413 1060325 1190571 1077524 1189551 1098054 1056414 1103886 1154504 1133046 1094886 1132349 1134204 1139218 1113760 1141155 1148273 1164002 1063494 1085954 1106589 1089272 1190549 1111378 1199209 1106029 1152398 1125187 1122241 1095339 1121226 1174387 1159322 1034564 1186414 1186434 1186695 1085212 1085996 1086062 1139585 1143409 1117530 1150664 1046461 1095525 1050221 1113299 1056972 1044808 1038618 1037893 1040139 1040321 1128621 1128738 1160499 1124362 1125406 1125412 1168090 1064612 1161981 1161850 1159817 1190014 1043955 1190554 1191489 1163817 1038869 1156941 1190596 1194938 1124421 1034507 1139312 1190574 1046076 1177468 1028402 1068857 1099456 1098683 1186496 1044838 1100561 1187964 1186187 1169622 1132079 1137698 1065972 1149210 1146554 1069931 1067479 1024742 1031087 1072323 1119171 1077619 1037833 1103990 1100662 1100466 1102989 1101036 1134244 1041149 1139603 1046299 1142858 1078771 1174301 1174374 1177452 1146633 1199251 1085985 1085226 1085894 1046818 1161628 1094147 1068465 1162288 1038471 1072180 1188393 1072527 1185512 1190016 1019613 1191375 1067950 1073201 1041501 1132992 1076960 1124746 1132847 1044421 1103924 1102352 1124716 1097939 1167884 1134466 1134190 1138711 1102578 1034728 1109924 1149188 1086035 1149548 1151394 1188087 1186350 1029532 1072247 1031238 1086465 1176903 1184435 1099369 1133054 1037832 1030675 1109724 1068650 1044439 1044436 1044727 1044611 1128117 1127691 1123578 1153155 1161425 1038767 1093990 1119254 1109199 1188010 1113467 1117360 1197840 1188498 1027732 1041409 1040093 1044609 1044703 1068991 1030818 1161773 1189870 1159552 1161365 1162154 1077060 1095841 1102502 1104200 1102135 1150870 1033805 1134529 1034421 1133771 1148250 1149268 1188008 1153098 1047582 1188273 1150811 1095950 1195399 1188051 1125557 1127686 1177271 1182206 1186523 1044241 1017686 1028044 1028146 1109156 1064368 1064601 1041162 1128371 1124376 1124270 1122292 1127475 1189958 1189827 1086789 1065528 1099196 1095887 1084648 1107418 1094874 1094364 1109212 1113472 1040593 1040679 1037646 1037836 1041200 1160540 1160866 1025253 1160404 1148604 1148243 1098437 1161376 1130682 1150125 1012001 1188432 1188319 1028110 1044951 1115175 1059516 1191080 1189967 1189988 1188528 1167280 1190546 1115976 1153216 1128877 1164126 1185483 1070036 1186746 1128150 1124116 1088171 1094522 1125304 1124645 1171630 1128058 1192760 1199256 1041436 1028194 1046514 1051446 1186366 1068685 1069098 1190306 1134047 1134171 1139177 1178806 1077251 1186704 1187882 1153484 1192833 1097968 1174185 1177302 1078805 1062474 1188444 1068442 1017620 1133966 1134168 1133923 1045006 1049986 1027826 1027863 1028082 1115043 1188065 1195410 1132412 1190810 1060916 1064324 1127690 1189618 1192014 1050166 1148219 1149181 1088207 1087517 1190690 1055413 1049148 1194945 1113841 1097944 1098373 1094776 1100542 1098741 1078809 1104411 1102879 1153115 1134330 1119451 1177308 1139370 1137433 1134230 1138902 1127530 1115352 1126191 1088668 1070852 1071114 1150805 1028093 1142869 1115573 1060251 1119165 1148103 1148252 1155202 1037868 1044505 1041767 1037657 1044876 1040916 1040005 1064578 1068543 1190656 1191920 1190333 1189995 1107638 1132985 1127796 1161655 1119431 1164693 1157998 1127399 1090061 1034401 1199272 1169560 1064463 1090307 1085098 1119291 1117988 1038235 1042748 1034171 1145625 1125416 1090125 1085931 1085838 1060204 1189925 1161359 1169009 1094360 1186439 1186538 1147450 1098342 1104538 1095317 1105508 1149079 1041092 1041533 1041486 1044501 1040775 1160502 1195540 1192977 1067804 1151347 1072222 1158592 1133254 1121023 1088225 1040013 1099204 1130727 1089982 1188481 1050238 1109317 1117711 1113350 1044344 1040749 1135367 1063760 1121177 1059012 1059471 1182547 1184806 1079229 1115470 1030849 1186423 1116132 1008590 1098717 1083132 1192291 1114635 1177305 1124112 1134200 1138825 1056228 1195432 1125559 1125290 1097926 1125556 1062527 1072412 1103856 1086051 1085025 1077553 1139569 1098152 1096490 1099195 1131060 1138074 1136777 1139306 1118953 1148637 1148729 1064590 1064916 1127185 1044690 1044467 1151356 1171599 1044350 1160551 1116465 1166998 1150231 1189954 1061548 1061146 1149390 1161522 1157905 1063919 1064079 1060630 1059475 1114013 1071463 1153225 1114841 1071453 1044962 1049380 1041727 1177290 1060246 1060236 1186683 1190579 1128697 1032549 1068036 1085123 1028225 1072280 1063276 1115063 1095572 1076631 1099030 1098557 1098359 1109614 1099144 1100557 1188098 1080325 1139542 1125073 1094724 1149452 1150910 1095484 1109277 1101908 1142612 1060508 1118046 1117332 1149130 1034520 1037796 1037816 1069732 1041036 1032923 1034497 1044568 1017342 1101382 1149024 1158788 1163737 1063648 1059982 1164633 1132988 1189985 1120180 1049739 1122583 1049998 1111329 1086125 1177253 1064104 1186604 1187938 1095354 1194727 1098260 1106462 1152243 1177278 1177269 1182503 1068489 1041353 1031439 1098934 1097815 1099273 1098726 1121132 1087706 1102777 1127904 1127338 1085090 1161934 1134157 1089465 1115429 1037815 1037703 1109129 1173392 1044385 1186747 1169017 1161868 1068618 1072186 1167134 1187974 1033174 1095247 1192106 1048761 1063254 1033455 1027508 1030136 1187792 1064295 1076982 1186368 1098294 1099094 1186818 1098112 1098055 1100645 1124382 1067912 1125410 1125402 1049172 1049696 1127688 1138789 1179783 1135115 1137041 1139734 1133952 1077354 1098575 1128630 1128712 1102876 1099561 1049620 1109410 1134093 1148218 1149416 1153194 1134174 1134050 1150943 1044807 1044620 1086018 1085926 1152840 1088231 1115356 1141072 1060615 1060076 1027896 1106415 1097458 1070480 1070448 1153086 1148505 1116750 1061558 1040397 1038622 1050354 1144475 1051422 1191951 1191891 1190328 1155508 1190008 1190196 1190006 1190663 1191937 1189918 1188582 1190606 1120405 1096622 1167414 1188466 1046140 1045901 1186484 1185484 1063512 1063670 1191503 1073798 1174994 1068863 1049185 1030948 1124342 1117368 1156851 1060174 1060320 1076605 1076730 1098845 1097977 1084407 1134161 1146604 1175041 1051500 1133486 1153930 1127500 1069019 1127800 1128047 1124827 1190098 1124608 1064304 1124104 1191885 1191956 1190334 1190602 1128086 1128087 1124641 1085288 1032553 1094482 1031286 1024594 1031354 1049158 1118011 1106402 1076788 1085941 1078871 1097851 1044552 1094453 1117753 1040804 1042866 1038065 1035276 1041435 1041233 1037866 1081704 1136199 1139219 1108628 1167284 1157343 1177824 1027249 1169023 1188330 1064109 1163696 1186489 1179527 1190082 1189824 1182619 1182353 1027983 1185485 1166888 1191377 1194930 1195409 1035068 1154480 1195421 1195391 1125226 1034689 1086749 1067438 1072670 1094485 1077003 1186119 1044726 1164400 1161903 1113310 1101764 1124316 1120628 1063545 1149162 1181954 1134940 1061232 1185574 1187930 1186727 1085854 1191970 1184504 1156595 1178859 1056262 1049215 1195398 1027820 1094358 1091771 1100661 1027555 1043721 1072515 1031192 1044528 1050484 1040144 1098005 1098451 1141295 1134268 1103980 1055105 1135362 1139125 1065336 1101609 1120688 1049985 1142605 1141356 1042095 1149201 1115670 1064088 1053714 1124213 1113948 1134218 1120213 1101526 1084932 1028088 1142290 1109210 1144138 1124343 1063582 1085982 1040829 1085241 1085217 1085940 1022971 1188912 1034297 1095489 1188452 1109289 1161298 1041449 1194877 1179645 1044631 1040788 1044693 1186190 1160894 1167064 1069959 1190893 1161855 1159491 1125566 1061308 1064912 1060317 1127455 1191926 1111133 1086081 1088304 1153370 1101555 1095603 1030776 1049191 1118065 1040970 1038897 1149148 1151668 1161441 1164152 1069480 1099647 1167447 1134901 1132167 1147483 1068437 1186490 1150967 1132713 1098911 1150834 1164318 1022267 1153490 1134918 1132407 1139115 1085234 1022830 1089263 1090073 1133616 1094826 1120771 1119981 1046591 1049955 1051324 1095522 1087543 1037840 1064747 1051356 1044831 1095536 1019923 1061433 1064819 1044644 1044585 1085636 1109184 1059113 1134380 1086017 1101524 1085929 1089515 1146530 1189583 1098241 1188385 1109319 1066520 1148548 1146575 1149674 1149465 1177512 1041438 1037917 1186372 1037757 1187975 1041422 1182350 1145622 1122917 1192233 1065217 1161380 1044761 1059846 1110262 1191959 1189723 1093248 1099354 1153509 1023740 1191465 1034542 1139318 1034382 1125363 1125782 1074984 1115567 1085419 1089169 1127460 1086078 1088216 1086919 1044531 1104618 1110101 1113332 1119086 1113437 1040561 1063441 1064738 1064193 1061352 1059436 1060881 1085867 1088259 1188492 1177901 1033381 1186437 1186041 1188014 1186269 1099358 1102654 1102509 1138772 1133105 1148791 1038616 1085238 1105372 1161841 1098381 1119173 1113506 1041468 1160219 1159828 1160345 1168104 1039160 1284307 1186389 1284662 1106216 1186723 1174415 1182650 1179555 1185595 1186713 1188085 1186520 1072097 1026867 1098426 1076525 1098013 1103902 1104410 1099419 1134034 1138858 1136697 1138692 1111577 1152515 1007779 1141921 1009582 1123198 1088387 1195226 1098015 1040166 1042165 1077356 1048610 1049869 1051085 1128152 1088122 1086164 1064217 1066240 1072151 1157022 1157339 1197529 1085100 1186085 1072331 1197010 1077425 1085194 1123773 1087036 1157719 1157891 1157446 1169018 1167428 1101189 1099989 1085721 1086145 1184593 1141708 1153644 1097700 1134175 1151946 1158577 1191734 1191385 1137892 1134141 1191368 1193510 1140415 1140409 1140696 1018364 1023269 1023507 1023453 1023626 1023583 1023634 1023586 1023597 1023744 1023621 1023759 1023580 1157758 1153629 1069554 1072068 1195156 1088435 1140441 1108005 1142019 1144697 1085904 1098245 1037850 1038657 1044508 1062594 1055901 1162305 1072182 1149935 1171814 1061482 1051176 1151678 1141709 1072147 1127048 1122960 1124499 1087850 1087424 1065924 1061231 1022609 1086730 1103525 1168984 1167397 1136740 1137724 1151923 1154614 1149775 1077692 1072304 1035914 1134008 1134179 1149830 1092170 1148431 1123594 1128175 1124502 1072141 1074018 1170290 1049382 1053499 1096514 1133960 1148293 1142030 1150328 1169284 1040103 1039372 1161010 1059024 1157770 1107419 1104476 1197136 1197026 1075249 1153722 1153707 1176672 1161816 1187152 1178988 1035898 1124864 1098802 1085270 1135847 1088409 1194004 1088125 1089257 1140787 1072150 1042677 1041520 1149888 1054032 1137913 1039404 1104432 1121702 1066239 1062064 1070633 1072436 1140379 1103680 1104023 1104104 1149622 1137674 1137675 1193437 1193123 1177918 1055454 1148355 1188352 1050441 1049370 1049407 1053343 1070690 1039443 1085969 1085395 1085132 1101455 1086355 1189921 1122991 1088385 1128397 1193967 1140328 1088174 1095399 1184632 1037401 1088366 1088617 1085702 1063561 1140531 1037682 1128248 1053439 1161564 1153939 1157138 1157105 1195601 1165259 1162624 1163328 1161194 1153306 1094165 1182343 1176669 1176666 1056725 1035918 1167109 1186112 1175310 1141705 1149621 1062323 1128537 1038653 1134082 1053372 1055960 1056080 1123256 1151350 1197434 1086157 1168670 1053748 1072148 1197047 1122482 1101824 1153511 1141702 1148444 1123558 1124533 1070386 1088130 1086325 1086044 1187254 1085884 1049399 1097295 1082190 1189990 1053323 1191478 1062536 1128875 1085204 1068591 1040751 1123796 1103638 1089249 1091200 1088344 1165685 1184601 1106374 1176683 1184540 1187238 1161095 1187197 1055013 1058045 1157098 1162142 1165828 1197009 1152367 1149828 1195693 1150983 1195113 1127662 1157294 1162364 1127828 1052705 1065950 1088128 1187708 1066156 1128281 1064301 1103068 1085723 1088276 1157688 1088603 1035279 1107433 1196954 1197353 1190523 1193579 1140343 1162175 1157762 1147166 1176679 1193902 1193532 1051859 1100333 1140782 1056229 1197102 1159168 1149832 1169051 1169168 1167317 1042103 1123679 1057021 1119626 1197688 1121664 1137849 1042134 1096247 1086148 1056462 1109181 1066162 1068583 1187703 1108919 1080065 1136775 1157728 1196975 1182892 1099266 1053244 1150289 1054088 1124424 1123737 1142056 1135093 1121652 1149600 1049978 1053790 1053360 1088343 1152325 1101803 1154621 1120742 1123077 1123628 1035904 1140437 1140377 1140778 1151883 1053300 1153558 1071170 1156233 1157760 1120740 1157423 1119602 1160178 1121731 1059446 1088439 1035891 1097404 1122453 1165404 1091153 1196706 1182824 1165956 1152534 1178012 1184834 1160636 1035843 1041496 1086895 1161548 1169158 1123086 1120857 1194119 1037671 1122509 1121109 1091146 1108613 1048640 1096165 1049346 1051421 1050774 1059335 1064450 1134283 1184656 1195486 1065807 1165304 1086725 1106355 1101792 1104398 1104077 1104012 1103657 1066102 1070574 1128187 1103567 1085227 1074459 1085224 1134183 1194009 1148063 1193566 1062887 1149789 1085727 1157725 1168747 1169292 1151927 1035832 1100160 1159224 1121284 1062369 1049393 1156881 1086784 1103956 1104331 1195622 1103027 1069797 1036592 1098068 1032385 1157428 1087086 1086505 1081062 1107461 1059021 1086883 1060335 1031879 1135590 1124466 1157356 1067305 1148543 1042098 1095312 1148348 1148531 1085257 1041055 1070837 1161656 1112077 1072287 1078359 1108113 1085031 1128239 1024029 1103984 1104507 1133970 1133924 1133922 1137577 1133783 1193471 1192455 1195166 1135507 1142064 1108509 1140619 1149577 1149618 1150958 1151925 1152102 1154639 1187104 1156086 1157067 1097724 1152428 1158803 1151841 1128263 1113846 1168732 1039406 1081194 1161079 1053443 1157443 1161145 1050858 1052283 1056000 1055910 1167424 1168673 1196963 1197050 1191057 1176667 1054726 1186161 1131808 1061493 1082743 1065904 1140370 1124643 1128120 1124601 1195605 1153023 1123126 1070562 1070686 1091088 1088759 1088369 1035247 1098345 1085225 1133576 1131793 1040825 1040041 1038643 1152007 1105009 1134192 1172715 1161183 1164810 1139596 1187612 1176658 1175212 1184668 1187156 1197391 1035032 1089550 1081995 1072209 1081930 1085584 1081697 1133614 1133746 1141759 1094034 1053371 1120819 1142071 1040897 1161107 1187260 1148339 1178019 1039382 1107394 1062584 1187802 1084984 1022880 1061670 1139313 1074072 1103545 1104024 1161245 1120888 1120775 1158892 1164914 1023370 1149568 1131814 1088411 1193175 1140232 1162600 1157102 1195629 1197140 1084302 1108154 1034744 1176401 1095458 1086245 1157942 1023656 1185184 1113590 1157768 1157645 1164815 1198162 1149910 1039408 1086631 1103963 1101761 1081618 1074286 1184708 1128252 1142079 1141701 1061552 1083666 1178038 1184831 1087101 1035929 1035911 1066139 1123149 1121498 1069569 1107422 1091140 1128322 1023605 1139828 1063474 1140380 1168693 1169203 1108474 1088055 1086000 1193546 1088206 1089254 1100385 1108694 1147219 1176700 1169259 1165004 1053412 1152378 1163234 1053234 1059624 1101431 1151825 1081606 1128156 1120912 1120825 1134987 1131927 1135799 1086345 1090743 1174116 1088404 1107426 1040999 1042178 1096179 1088060 1157685 1157687 1086255 1150006 1041489 1050494 1157448 1157177 1197037 1190567 1152326 1107603 1085252 1039785 1140671 1161551 1112108 1119572 1115128 1176811 1058137 1066114 1284927 1066022 1101765 1134069 1137779 1136781 1027772 1150680 1167603 1007780 1007796 1007800 1007805 1017422 1017590 1017945 1018079 1285329 1075408 1020023 1020044 1092504 1069414 1094853 1094383 1074811 1112217 1126643 1023211 1023213 1023858 1023878 1023979 1025182 1284880 1027144 1027280 1027313 1028555 1028672 1030008 1030331 1030603 1031057 1031069 1031130 1106654 1106584 1107245 91647667 1031746 1031753 1031830 1032009 1032047 1033303 1033310 1033617 1033663 1050939 1052195 1052261 1052262 1111191 1111623 1111637 1112270 1112538 1112638 1112679 1116610 1118532 1119635 1119774 1035385 1035511 1040142 1284987 1041034 1047424 1046598 1046685 1047881 1041440 1041576 1041890 1042393 1041754 1043484 1042350 1044092 1043532 1043069 1041272 1167640 1167278 1167460 1168296 1168853 1168872 1169226 1168768 1168272 1168854 1165654 1166406 1062231 1131407 1166948 1166294 1041484 1041526 1041537 1170235 1174854 1041699 1172037 1172947 1173241 1174351 1041783 1041877 1041879 1041894 1041973 1042180 1042391 1042436 1042482 1042487 1042534 1042537 1042580 1042729 1043024 1043094 1043112 1043194 1043228 1043248 1043250 1043272 1043278 1043289 1125030 1125313 1082806 1119700 1158314 1158622 1158684 1159587 1159140 1159149 1168191 1168537 1168615 1168589 1168665 1170774 1170498 1170456 1170492 1170553 1170547 1171943 1169581 1170457 1170140 1169779 1170173 1167811 1173080 1172065 1174797 1175077 1171573 1174817 1043447 1124371 1124865 1129746 1130143 1130965 1131151 1131173 1163102 1139868 1157748 1167844 1168288 1168298 1175344 1175787 1162832 1170674 1170937 1173383 1171760 1043510 1043899 1043953 1112274 1045786 1045803 1045921 1046219 1046372 1046381 1046424 1177537 1177876 1178002 1046880 1177088 1181559 1182814 1181290 1190759 1046967 1047062 1047135 1047136 1047353 1047463 1047494 1191204 1191213 1191227 1047699 1047727 1047771 1047824 1047953 1048064 1048136 1048174 1048338 1048481 1174653 1173840 1186005 1051066 1051713 1052275 1052286 1053385 1054023 1054049 1054315 1054518 1054673 1054714 1054740 1054761 1054887 1054891 1055196 1055297 1055536 1055552 1057363 1057373 1057406 1057522 1057635 1057674 1057689 1057772 1057966 1058004 1058315 1058382 1058432 1058442 1058800 1059047 1059119 1059221 1059261 1059465 1059637 1075662 1075695 1061799 1061999 1173238 1174271 1176427 1177489 1178937 1179361 1179366 1179373 1179434 1173578 1177709 1092650 1074739 1096973 1177717 1177935 1177942 1178167 1178179 1170361 1064520 1094389 1106606 1107131 1107211 1107368 90958420 1179037 1179793 1174189 1176166 1177416 1178854 1178996 1174978 1178051 1178083 1170300 1177146 1178127 1178987 1179059 1179816 1072640 1074234 90651138 90658045 1091605 1095377 1111685 1112418 1112634 1112650 1118501 1119758 90800205 90631885 1166510 1168134 1167563 1168814 1169425 1169400 1166517 1169654 1169721 1075607 1124969 1166682 1169344 1169172 1168127 1168689 1168838 1168458 1169032 1165660 1166304 1166094 1166483 1167054 1166530 1072864 1074027 1174788 1169555 1172738 1174912 1176656 1071163 1122295 1123897 1067809 1124894 1125016 1119569 1144857 1144868 1144620 1072122 1157659 1158086 1158375 1158452 1159065 1159332 1159618 1168264 1169528 1168568 1168599 1168412 1168540 1168126 1169289 1170573 1170488 1170577 1170570 1170144 1170045 1170186 1169838 1172060 1171269 1171724 1171580 1174866 1174862 1174871 1125061 1125158 1131099 1131324 1163332 1163821 1180438 1151785 1162092 1167854 1176110 1162770 1171015 1170745 1170687 1170684 1170739 1171221 1171784 1117550 1069212 1168393 1168741 1181469 1181600 1173126 1173233 1173231 1173333 1070486 1176617 1172074 1177085 1176917 1178452 1177168 1177909 1177960 1178442 1072046 1072892 1073720 1074593 1074930 1075006 1075103 1075134 1075143 1075319 1075488 1075598 1075660 1138349 1138202 1138452 1137836 1076651 1134555 1134531 1134922 1092612 1092583 90938369 1093768 1094443 1094372 1107286 90880144 90763354 901028644 1095290 1108612 1112542 1112818 1116785 1116849 1119605 1122571 1120643 1122916 1121996 1121844 1099504 90953164 1167907 1167557 1167593 1169024 1168808 1169412 1169674 1169792 1168784 1168913 1168850 1167796 1168236 1168341 1165741 1165782 1166567 1165851 1166497 1169587 1172962 1174957 1172533 1174411 1178504 1171638 1174309 1124993 1125045 1125373 1100608 1119553 1125628 1125981 1144678 1157566 1157650 1157678 1158138 1158191 1158474 1158495 1158602 1159036 1159315 1159477 1159834 1168656 1169447 1169479 1169239 1170863 1170526 1184112 1185556 1180588 1183407 1169452 1170586 1171199 1170539 1170517 1171279 1169875 1170310 1172563 1173100 1171523 1174954 1174814 1130984 1131354 1163730 1163741 1180429 1145347 1145727 1162133 1168269 1175693 1175865 1158669 1162783 1171342 1171150 1170637 1171106 1170709 1171280 1171086 1171831 1171848 1171786 1162459 1171941 1171217 1172394 1118621 1167309 1168199 1177175 1177439 1181571 1190755 1191288 1191290 1174276 1171290 1176608 1177643 1176971 1177883 1179551 1176992 1112583 1112729 1174989 1167900 1169760 1177011 1177161 1177402 1181605 1180803 1190786 1172447 1139652 1139783 1118985 1138469 1145629 1139190 1173484 1173335 1174661 1173457 1176502 1176709 1176978 1179416 1177927 90992607 1106933 1119577 1173588 1173639 1177813 1177820 1177836 1169745 1170005 1173407 1173489 1145174 1144524 1144597 1146173 1144987 1145062 1133977 1145801 1112910 1124042 1127991 1112871 1125211 1129901 1130305 1130962 1131036 1090890 1111235 1111646 1111687 1111710 1111809 1111868 1111875 1112054 1112085 1112419 1112513 1112577 1166957 1125814 1125229 1125310 1163285 1177613 1092580 1092867 1094191 1129193 1129587 1113732 1129876 1162800 1163155 1123956 1123226 1126574 1131418 1130005 90758814 90806848 1163707 1173243 1173591 1175027 1160930 1169872 1173554 1139273 1145030 1145403 1145520 1144511 1144791 1107153 1107103 91763592 90689615 1127909 1129535 1129545 1126130 1130961 1130974 1101681 1167761 1167294 1167093 1168844 1169334 1169228 1169214 1168196 1169302 1168567 1169176 1168522 1166310 1165587 1166175 1165916 1129761 1166367 1165898 1172872 1172925 1172346 1178192 1176603 1176628 1164802 1111527 1111841 1112520 1113055 1131647 1121666 1125153 1125326 1125772 1125799 1125908 1103095 1119884 1150599 1144666 1145313 1158078 1158238 1158989 1159786 1159745 1171348 1170137 1169389 1171666 1172048 1172009 1174733 1174929 1129976 1127790 1131007 1131127 1131621 1163085 1163226 1131504 1163631 1163690 1151784 1145695 1157694 1158151 1175783 1176142 1171266 1170607 1170645 1170635 1171020 1166069 1170562 1171950 1171817 1179797 1173829 1124795 1131365 1162683 1175818 1176633 1178502 1178721 1178948 1179002 1177816 1177839 1178042 1178148 1178193 1112666 1155162 1174333 1174337 1178593 1178748 1178865 1178995 1179200 1177596 1170423 1178182 1179001 1179465 1179594 1179665 1179795 90698171 90760146 1111719 1111760 1112527 1112535 1112900 90840031 90808773 1118381 1173656 1177615 1178183 1161489 1170098 1170124 1173543 1145797 1144681 1144401 1138485 1131453 1140028 1123243 1123895 1126005 1129210 1129552 1129644 1143914 1130007 1131128 1131477 1162666 90543977 90562298 90597199 90620745 90664669 90699018 90708711 90723865 90808788 90819142 90830825 90833035 90837726 90839916 90855089 90859110 90862557 90864150 91876260 90955714 90953502 91760192 91669459 91616579 90946426 90965344 90933315 90936603 91626832 91541221 90880230 90916905 91594373 91881844 90951697 93660099 90954696 90882897 90938016 90667319 90617516 92700407 90624921 90604411 90873935 90744321 90820433 90751723 90922328 90800108 90882701 91543326 90852569 90649230 90617050 90620742 90807691 90855872 91537196 90929038 90658919 91546630 90876769 90618376 90531690 90640856 90606296 90659037 90719582 90757788 91862128 90619481 90919084 90876474 90939264 90943803 91839337 91627862 90840728 90775904 90594214 90623835 90823029 91695410 90542596 91899869 90853266 90900585 90870794 90825290 93673741 90842451 1007799 1017591 1017584 1017601 1017459 1017277 1017629 1017470 1017445 1017480 1017465 1017289 1017485 1017474 1017193 1017340 1017604 1017242 1017449 1017488 1017615 1017592 1017585 1017454 1017273 1017204 1017254 1017596 1017175 1017172 1017248 1017205 1017053 1017161 1016981 1017209 1016995 1017253 1017347 1017262 1017619 1017290 1017212 1017224 1017633 1017110 1017626 1017221 1017167 1017223 1017029 1017586 1016943 1025115 1016951 1025228 1025346 1025210 1025220 1017234 1017162 1017228 1017270 1025327 1017001 1017278 1016992 1017025 1017206 1017000 1017249 1017219 1017177 1017061 1025257 1017317 1017211 1017266 1025251 1025239 90877586 90736708 91648180 90821387 90693602 90852535 90800941 90677285 91847970 90938881 90844939 90601666 90623950 91720421 91636613 90958874 90701896 91612399 90739079 90926890 90708797 90946720 90853010 90598229 90834152 90865953 92669671 91594979 90730206 91620284 91656086 90824504 90824330 91676805 90586965 90541248 90675673 90673080 90739452 1043694 1059690 1057575 1057551 1058067 1058296 1059282 1048305 1071567 1072905 1052589 1047734 1046968 1030048 1026607 1048112 1033379 1023665 1019379 1020022 1017727 1024671 1023103 1177008 1178423 1179242 1179399 1173580 1174983 1177680 1178298 1177139 1178354 1179850 1137843 1138329 1137898 1137884 1135348 1138011 1059144 1092883 1073734 1092606 1106483 1118548 1118734 1134990 1134229 1135134 1174693 1173948 1173342 1173471 1173553 1177451 1177821 1170303 1173497 1173413 1176912 1176908 1176742 1157232 1177339 1176933 1144789 1144629 1144330 1143257 1074800 1145007 1144856 1144589 1145048 1074889 1094703 90987533 1128011 1127499 1127784 1128003 1129444 1113563 1130012 91670734 90831004 1106558 1118696 1106570 1106631 1111004 1111332 1131424 1131439 1173687 1177977 1178130 1170394 1175652 1052284 1052946 1062862 1111557 1112386 1112531 1112536 1112572 1112950 1118456 1122645 1123623 1123552 90728473 90643332 1123045 1123709 1122024 1121873 1123121 1055039 1096978 1107140 1106597 1035310 1095271 1092601 1112306 1112675 1047110 1047515 1047954 1047957 1032215 1033384 1040666 1041391 1041686 1042538 1042575 1043464 1130924 1131122 1131143 1131184 1131230 1131400 1131534 1131941 1163293 1041511 1046375 1046624 1047995 1032090 1023207 1041390 1042703 1042342 1042493 1042368 1043231 1043185 1042856 1042769 90846779 1171566 1173130 1174765 1175349 1176002 1179019 1173620 1178074 1178082 1169931 1179095 1179281 1179424 1070190 1143215 1139110 1155566 1138636 1144505 1145733 1167581 1168019 1169005 1167367 1168301 1169006 1167864 1167516 1169715 1169733 1169618 1168852 1168941 1168629 1168318 1165475 1102873 1165784 1166492 1166647 1166516 1167465 1168978 1167123 1167886 1169129 1169089 1168243 1169710 1169724 1075581 1168911 1169067 1168324 1168518 1168869 1165613 1165931 1165890 1166088 1175074 1172927 1176438 1173174 1176644 1172461 1171561 1175172 1175170 1176037 1113084 1124228 1124301 1124803 1125399 1109035 1119422 1125720 1144679 1158240 1158878 1158949 1059425 1072032 1074669 1092935 1171257 1171253 1171036 1169243 1092886 1047785 90915138 1170183 1170050 1172721 1172570 1172331 1170975 1171546 1174898 1175067 1174918 1121495 1130982 1131156 1163110 1059075 1124987 1125003 1125118 1119763 1126634 1157438 1157669 1157716 1158224 1158421 1158470 1158506 1158960 1159080 1158448 1159069 1159046 1165855 1168597 1169619 1169237 1168837 1170719 1170761 1171188 1170803 1171246 1171919 1170117 1170453 1170383 1170338 1169429 1171025 1174726 1171550 1171522 1150631
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Tandem Diabetes Care, Inc.
- Reason for Recall:
- Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result in a hyperglycemic event, which may lead to injury and/or hospitalization.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Battery Operated infusion pumps. t:slim X2 G5, Replacement, Refurbished, REF: 1000911; t:slim X2 GS Classic, Replacement, Refurbished, REF: 1003808; t: slim X2, Basal-IQ, mg/dl, REF: 1004219; Pump, t:slim X2, Clinical Use Only, REF: 1004484; Pump, t:slim X2, Control-IQ, REF: 1005611; Pump, t:slim X2, Control-IQ, mmoUL, Refurbished Replacement, REF: 1005613; Pump, t:slim X2, Control-IQ, mg/dl, REF: 1005615; Pump, t:slim X2, Control-IQ, mg/dl, Refurbished Replacement, REF: 1005617; Pump, t:slim X2, Basal-IQ 6.4, Refurbished Replacement, REF: 1006379; Pump, t:slim X2, Control-IQ 7.4, REF: 1006402; Pump, t:slim X2, Control-IQ 7.4, Replacement, REF: 1006406; Pump, t:slim X2, Control-IQ 7.4, REF: 1006408; Pump, t:slim X2, Basal-IQ, , REF: 1006419; Pump, t:slim X2, Basal-IQ, mg/dL, REF: 1006420; Pump, t:slim X2, Basal-IQ, mmoUL, Refurbished Replacement, REF: 1006535; Pump, t:slim X2, Basal-IQ, mg/dL, Refurbished Replacement, REF: 1006537; Pump, t:slim X2, Control-IQ, mg/dl , Replacement, REF: 1008255; Pump, t:slim X2, Control-IQ, mmoUL, Replacement, REF: 1008256; Pump, t:slim X2, Basal-IQ, mmoUL, Replacement, REF: 1008257; Pump, t:slim X2, Basal-IQ, mg/dl ,Replacement, REF: 1008258; Pump, t:slim X2, Basal-IQ, REF: 1010004; Pump, t:slim X2, Basal-IQ, Replacement, REF: 1010005; Pump, t:slim X2, Basal-IQ, Medicare, REF: 1010006; Pump, t:slim X2, Basal-IQ, Medicare, REF: 1010007; Pump, t:slim X2, Control-IQ, REF: 1010009; Pump, t:slim X2, Control-IQ Replacement, REF: 1010010; Pump, t:slim X2, Control-IQ, Medicare, REF: 1010011; Pump, t:slim X2, Control-IQ, Refurbished Replacement, REF: 1010012; Pump, t:slim X2, Control-IQ, REF: 1012867; Pump, t:slim X2, Control-IQ, Medicare, REF: 1012868; Pump, t:slim X2, Control-IQ, Replacement, REF: 1012869; Pump, t:slim X2, Control-IQ 7.8, REF: 1013195; Pump, t:slim X2, Control-IQ 7.8, Medicare, REF: 1013196; Pump, t :slim X2, Control-IQ 7.8, Replacement, REF: 1013197; Pump, t:slim X2, Control-IQ 7.8, Refurbished Replacement, REF: 1013198
Product Codes/Lot Numbers:
REF/UDI-DI: 1000911/t:slim X2 G5, Replacement, Refurbished/00853052007318; 1003808/t:slim X2 GS Classic, Replacement, Refurbished/ 00853052007981; 1004219/t: slim X2, Basal-IQ, mg/dl/ 00853052007998; 1004484/Pump, t:slim X2, Clinical Use Only; 1005611/Pump, t:slim X2, Control-IQ, mmoUL/ 00850006613304/; 1005613/Pump, t:slim X2, Control-IQ, mmoUL, Refurbished Replacement/ 00850006613311/; 1005615/Pump, t:slim X2, Control-IQ, mg/dl/ 00850006613328/; 1005617/Pump, t:slim X2, Control-IQ, mg/dl, Refurbished Replacement/ 00850006613335/; 1006379/Pump, t:slim X2, Basal-IQ 6.4, Refurbished Replacement/ 00850006613731/; 1006402/Pump, t:slim X2, Control-IQ 7.4/ 00850006613762/; 1006406/Pump, t:slim X2, Control-IQ 7.4, Replacement/ 00850006613786/; 1006408/Pump, t:slim X2, Control-IQ 7.4/ 00850006613793/; 1006419/Pump, t:slim X2, Basal-IQ, mmoUL/ 00850006613809/; 1006420/Pump, t:slim X2, Basal-IQ, mg/dL 00850006613816; 1006535/Pump, t:slim X2, Basal-IQ, mmoUL, Refurbished Replacement/ 00850006613823/; 1006537/Pump, t:slim X2, Basal-IQ, mg/dL, Refurbished Replacement/ 00850006613830/; 1008255/Pump, t:slim X2, Control-IQ, mg/dl , Replacement/00850018992671; 1008256/Pump, t:slim X2, Control-IQ, mmoUL, Replacement/00850018992664; 1008257/Pump, t:slim X2, Basal-IQ, mmoUL, Replacement/00850018992657; 1008258/Pump, t:slim X2, Basal-IQ, mg/dl ,Replacement/00850018992640; 1010004/Pump, t:slim X2, Basal-IQ/00389152000015; 1010005/Pump, t:slim X2, Basal-IQ, Replacement/00389152000022; 1010006/Pump, t:slim X2, Basal-IQ, Medicare/00389152000039; 1010007/Pump, t:slim X2, Basal-IQ, Medicare/00389152000046; 1010009/Pump, t:slim X2, Control-IQ/00389152000077; 1010010/Pump, t:slim X2, Control-IQ Replacement/00389152000084; 1010011/Pump, t:slim X2, Control-IQ, Medicare/00389152000091; 1010012/Pump, t:slim X2, Control-IQ, Refurbished Replacement/00389152000107; 1012867/Pump, t:slim X2, Control-IQ/00389152286716; 1012868/Pump, t:slim X2, Control-IQ, Medicare/00389152286815; 1012869/Pump, t:slim X2, Control-IQ, Replacement/00389152286914; 1013195/Pump, t:slim X2, Control-IQ 7.8/00389152319513; 1013196/Pump, t:slim X2, Control-IQ 7.8, Medicare/00389152319612; 1013197/Pump, t :slim X2, Control-IQ 7.8, Replacement/00389152319711; 1013198/Pump, t:slim X2, Control-IQ 7.8, Refurbished Replacement/00389152319810 Serial Numbers: 951824 1009214 1009818 1010205 1010216 1010280 1010282 1124339 1124359 1120792 1284185 1096196 1011443 1127741 1017739 1088141 1023688 1084425 1028167 1028228 1033168 1109355 1035333 1044731 1037937 1038725 1040458 1040729 1044512 1041052 1041217 1041397 1041457 1156947 1043829 1044400 1044800 1284661 1049585 1049643 1053348 1186711 1191406 1121799 1132861 1124364 1123627 1063736 1064271 1064593 1106444 1067539 1067828 1067887 1172701 1174318 1098764 1110835 1073710 1128723 1124356 1121152 1127745 1076858 1085036 1101544 1088173 1113296 1109496 1132398 1099207 1134096 1109488 1188241 1188233 1187866 1174100 1177299 1186285 1085203 1086053 1169060 1098778 1099063 1098525 1132410 1134897 1139347 1149413 1149109 1109118 1147667 1149987 1193922 1095865 1148061 1120208 1125365 1127634 1127974 1139150 1138109 1098244 1149017 1099022 1115201 1115510 1101725 1195419 1160890 1195424 1190520 1195271 1190631 1193090 1188686 1151553 1196393 1157533 1011994 1011686 1010370 1010330 1010149 1010339 1009580 1010186 1018441 1018522 1020329 1020438 1020404 1021017 1020215 1018419 1018458 1020226 1018560 1016948 1020578 1020553 1020572 1020531 1020263 1020113 1020100 1020054 1020712 1020104 1020560 1020655 1020487 1023365 1022648 1021118 1020917 1020373 1024186 1023401 1023533 1024261 1023478 1022834 1022835 1022828 1024374 1024315 1024201 1024386 1010323 1008806 1008828 1009705 1009241 1009735 1009666 1009722 1008939 1009686 1009704 1009284 1009191 1009123 1010025 1010059 1010067 1010279 1009063 1008575 1010022 1010277 1010338 1010221 1010107 1010315 1010211 1008487 1008531 1008486 1008514 1008467 1008820 1008760 1008615 1008904 1008694 1021317 1011702 1019671 1021237 1021192 1021279 1021324 1021318 1010379 1021299 1021301 1010249 1019683 1011761 1021325 1021261 1019418 1021284 1010359 1010356 1064999 1044969 1044176 1191892 1191939 1152446 1195279 1185661 1134013 1072356 1125364 1103406 1128033 1127345 1098206 1098758 1132314 1138833 1138876 1137809 1104141 1164437 1148205 1148307 1148565 1148163 1148709 1149419 1158667 1180647 1138036 1139540 1139510 1037782 1098256 1041793 1041139 1119177 1117738 1190568 1044931 1044351 1040764 1160757 1147747 1097466 1159784 1148530 1151030 1174104 1110283 1101561 1152960 1136798 1186356 1186639 1188142 1186579 1186590 1099572 1102128 1059066 1101740 1134163 1186587 1049144 1085902 1086054 1151146 1089176 1080141 1095314 1109999 1042173 1031387 1038019 1038543 1190762 1160289 1160568 1167268 1196846 1151805 1164867 1077168 1172909 1065134 1177890 1139118 1124251 1182202 1190766 1195370 1139765 1139675 1078571 1169487 1154526 1131902 1192577 1174388 1177382 1179618 1186544 1178243 1185555 1186681 1193040 1127529 1124667 1048221 1043088 1062012 1064036 1072664 1072131 1069163 1085921 1085990 1077115 1098578 1100659 1028390 1133023 1139699 1134164 1139733 1142649 1141467 1027083 1027674 1098353 1115931 1040780 1133914 1133659 1148066 1147771 1149703 1151165 1085987 1103320 1120657 1125266 1120217 1120109 1133124 1125657 1199271 1089240 1044786 1050174 1127450 1051372 1085255 1084995 1098309 1063996 1098494 1060892 1041467 1040982 1132281 1019268 1136846 1138865 1040584 1041466 1034158 1167994 1030767 1030872 1028097 1088069 1148309 1094754 1094258 1095557 1109170 1079138 1154688 1129307 1117370 1121728 1112236 1129842 1038028 1139532 1167107 1132327 1132871 1046471 1049938 1044325 1046565 1046564 1050219 1127590 1055444 1161899 1171637 1061294 1189986 1190593 1064898 1059461 1109196 1190998 1113847 1152496 1109208 1041407 1041427 1023013 1033115 1032409 1174373 1184801 1186628 1128584 1135284 1136631 1139728 1139139 1191337 1193055 1191584 1190739 1031121 1028166 1169532 1177589 1174320 1132865 1133050 1133066 1128536 1177786 1178743 1098255 1120100 1184807 1085097 1101946 1133946 1138817 1139310 1122071 1185606 1085624 1086793 1088003 1052447 1038695 1106503 1041455 1038079 1037491 1148174 1028206 1043709 1040735 1151381 1043039 1027142 1072103 1097921 1097858 1159873 1117464 1100615 1104217 1187916 1077458 1050400 1102546 1141218 1128104 1113002 1085984 1095329 1051040 1115605 1052544 1098977 1074138 1148055 1132301 1136965 1138765 1041090 1188435 1188531 1051922 1190303 1177860 1029585 1195324 1182299 1187987 1188086 1149405 1115558 1151447 1044354 1152129 1057291 1041275 1038086 1060564 1133413 1060325 1190571 1077524 1189551 1098054 1056414 1103886 1154504 1133046 1094886 1132349 1134204 1139218 1113760 1141155 1148273 1164002 1063494 1085954 1106589 1089272 1190549 1111378 1199209 1106029 1152398 1125187 1122241 1095339 1121226 1174387 1159322 1034564 1186414 1186434 1186695 1085212 1085996 1086062 1139585 1143409 1117530 1150664 1046461 1095525 1050221 1113299 1056972 1044808 1038618 1037893 1040139 1040321 1128621 1128738 1160499 1124362 1125406 1125412 1168090 1064612 1161981 1161850 1159817 1190014 1043955 1190554 1191489 1163817 1038869 1156941 1190596 1194938 1124421 1034507 1139312 1190574 1046076 1177468 1028402 1068857 1099456 1098683 1186496 1044838 1100561 1187964 1186187 1169622 1132079 1137698 1065972 1149210 1146554 1069931 1067479 1024742 1031087 1072323 1119171 1077619 1037833 1103990 1100662 1100466 1102989 1101036 1134244 1041149 1139603 1046299 1142858 1078771 1174301 1174374 1177452 1146633 1199251 1085985 1085226 1085894 1046818 1161628 1094147 1068465 1162288 1038471 1072180 1188393 1072527 1185512 1190016 1019613 1191375 1067950 1073201 1041501 1132992 1076960 1124746 1132847 1044421 1103924 1102352 1124716 1097939 1167884 1134466 1134190 1138711 1102578 1034728 1109924 1149188 1086035 1149548 1151394 1188087 1186350 1029532 1072247 1031238 1086465 1176903 1184435 1099369 1133054 1037832 1030675 1109724 1068650 1044439 1044436 1044727 1044611 1128117 1127691 1123578 1153155 1161425 1038767 1093990 1119254 1109199 1188010 1113467 1117360 1197840 1188498 1027732 1041409 1040093 1044609 1044703 1068991 1030818 1161773 1189870 1159552 1161365 1162154 1077060 1095841 1102502 1104200 1102135 1150870 1033805 1134529 1034421 1133771 1148250 1149268 1188008 1153098 1047582 1188273 1150811 1095950 1195399 1188051 1125557 1127686 1177271 1182206 1186523 1044241 1017686 1028044 1028146 1109156 1064368 1064601 1041162 1128371 1124376 1124270 1122292 1127475 1189958 1189827 1086789 1065528 1099196 1095887 1084648 1107418 1094874 1094364 1109212 1113472 1040593 1040679 1037646 1037836 1041200 1160540 1160866 1025253 1160404 1148604 1148243 1098437 1161376 1130682 1150125 1012001 1188432 1188319 1028110 1044951 1115175 1059516 1191080 1189967 1189988 1188528 1167280 1190546 1115976 1153216 1128877 1164126 1185483 1070036 1186746 1128150 1124116 1088171 1094522 1125304 1124645 1171630 1128058 1192760 1199256 1041436 1028194 1046514 1051446 1186366 1068685 1069098 1190306 1134047 1134171 1139177 1178806 1077251 1186704 1187882 1153484 1192833 1097968 1174185 1177302 1078805 1062474 1188444 1068442 1017620 1133966 1134168 1133923 1045006 1049986 1027826 1027863 1028082 1115043 1188065 1195410 1132412 1190810 1060916 1064324 1127690 1189618 1192014 1050166 1148219 1149181 1088207 1087517 1190690 1055413 1049148 1194945 1113841 1097944 1098373 1094776 1100542 1098741 1078809 1104411 1102879 1153115 1134330 1119451 1177308 1139370 1137433 1134230 1138902 1127530 1115352 1126191 1088668 1070852 1071114 1150805 1028093 1142869 1115573 1060251 1119165 1148103 1148252 1155202 1037868 1044505 1041767 1037657 1044876 1040916 1040005 1064578 1068543 1190656 1191920 1190333 1189995 1107638 1132985 1127796 1161655 1119431 1164693 1157998 1127399 1090061 1034401 1199272 1169560 1064463 1090307 1085098 1119291 1117988 1038235 1042748 1034171 1145625 1125416 1090125 1085931 1085838 1060204 1189925 1161359 1169009 1094360 1186439 1186538 1147450 1098342 1104538 1095317 1105508 1149079 1041092 1041533 1041486 1044501 1040775 1160502 1195540 1192977 1067804 1151347 1072222 1158592 1133254 1121023 1088225 1040013 1099204 1130727 1089982 1188481 1050238 1109317 1117711 1113350 1044344 1040749 1135367 1063760 1121177 1059012 1059471 1182547 1184806 1079229 1115470 1030849 1186423 1116132 1008590 1098717 1083132 1192291 1114635 1177305 1124112 1134200 1138825 1056228 1195432 1125559 1125290 1097926 1125556 1062527 1072412 1103856 1086051 1085025 1077553 1139569 1098152 1096490 1099195 1131060 1138074 1136777 1139306 1118953 1148637 1148729 1064590 1064916 1127185 1044690 1044467 1151356 1171599 1044350 1160551 1116465 1166998 1150231 1189954 1061548 1061146 1149390 1161522 1157905 1063919 1064079 1060630 1059475 1114013 1071463 1153225 1114841 1071453 1044962 1049380 1041727 1177290 1060246 1060236 1186683 1190579 1128697 1032549 1068036 1085123 1028225 1072280 1063276 1115063 1095572 1076631 1099030 1098557 1098359 1109614 1099144 1100557 1188098 1080325 1139542 1125073 1094724 1149452 1150910 1095484 1109277 1101908 1142612 1060508 1118046 1117332 1149130 1034520 1037796 1037816 1069732 1041036 1032923 1034497 1044568 1017342 1101382 1149024 1158788 1163737 1063648 1059982 1164633 1132988 1189985 1120180 1049739 1122583 1049998 1111329 1086125 1177253 1064104 1186604 1187938 1095354 1194727 1098260 1106462 1152243 1177278 1177269 1182503 1068489 1041353 1031439 1098934 1097815 1099273 1098726 1121132 1087706 1102777 1127904 1127338 1085090 1161934 1134157 1089465 1115429 1037815 1037703 1109129 1173392 1044385 1186747 1169017 1161868 1068618 1072186 1167134 1187974 1033174 1095247 1192106 1048761 1063254 1033455 1027508 1030136 1187792 1064295 1076982 1186368 1098294 1099094 1186818 1098112 1098055 1100645 1124382 1067912 1125410 1125402 1049172 1049696 1127688 1138789 1179783 1135115 1137041 1139734 1133952 1077354 1098575 1128630 1128712 1102876 1099561 1049620 1109410 1134093 1148218 1149416 1153194 1134174 1134050 1150943 1044807 1044620 1086018 1085926 1152840 1088231 1115356 1141072 1060615 1060076 1027896 1106415 1097458 1070480 1070448 1153086 1148505 1116750 1061558 1040397 1038622 1050354 1144475 1051422 1191951 1191891 1190328 1155508 1190008 1190196 1190006 1190663 1191937 1189918 1188582 1190606 1120405 1096622 1167414 1188466 1046140 1045901 1186484 1185484 1063512 1063670 1191503 1073798 1174994 1068863 1049185 1030948 1124342 1117368 1156851 1060174 1060320 1076605 1076730 1098845 1097977 1084407 1134161 1146604 1175041 1051500 1133486 1153930 1127500 1069019 1127800 1128047 1124827 1190098 1124608 1064304 1124104 1191885 1191956 1190334 1190602 1128086 1128087 1124641 1085288 1032553 1094482 1031286 1024594 1031354 1049158 1118011 1106402 1076788 1085941 1078871 1097851 1044552 1094453 1117753 1040804 1042866 1038065 1035276 1041435 1041233 1037866 1081704 1136199 1139219 1108628 1167284 1157343 1177824 1027249 1169023 1188330 1064109 1163696 1186489 1179527 1190082 1189824 1182619 1182353 1027983 1185485 1166888 1191377 1194930 1195409 1035068 1154480 1195421 1195391 1125226 1034689 1086749 1067438 1072670 1094485 1077003 1186119 1044726 1164400 1161903 1113310 1101764 1124316 1120628 1063545 1149162 1181954 1134940 1061232 1185574 1187930 1186727 1085854 1191970 1184504 1156595 1178859 1056262 1049215 1195398 1027820 1094358 1091771 1100661 1027555 1043721 1072515 1031192 1044528 1050484 1040144 1098005 1098451 1141295 1134268 1103980 1055105 1135362 1139125 1065336 1101609 1120688 1049985 1142605 1141356 1042095 1149201 1115670 1064088 1053714 1124213 1113948 1134218 1120213 1101526 1084932 1028088 1142290 1109210 1144138 1124343 1063582 1085982 1040829 1085241 1085217 1085940 1022971 1188912 1034297 1095489 1188452 1109289 1161298 1041449 1194877 1179645 1044631 1040788 1044693 1186190 1160894 1167064 1069959 1190893 1161855 1159491 1125566 1061308 1064912 1060317 1127455 1191926 1111133 1086081 1088304 1153370 1101555 1095603 1030776 1049191 1118065 1040970 1038897 1149148 1151668 1161441 1164152 1069480 1099647 1167447 1134901 1132167 1147483 1068437 1186490 1150967 1132713 1098911 1150834 1164318 1022267 1153490 1134918 1132407 1139115 1085234 1022830 1089263 1090073 1133616 1094826 1120771 1119981 1046591 1049955 1051324 1095522 1087543 1037840 1064747 1051356 1044831 1095536 1019923 1061433 1064819 1044644 1044585 1085636 1109184 1059113 1134380 1086017 1101524 1085929 1089515 1146530 1189583 1098241 1188385 1109319 1066520 1148548 1146575 1149674 1149465 1177512 1041438 1037917 1186372 1037757 1187975 1041422 1182350 1145622 1122917 1192233 1065217 1161380 1044761 1059846 1110262 1191959 1189723 1093248 1099354 1153509 1023740 1191465 1034542 1139318 1034382 1125363 1125782 1074984 1115567 1085419 1089169 1127460 1086078 1088216 1086919 1044531 1104618 1110101 1113332 1119086 1113437 1040561 1063441 1064738 1064193 1061352 1059436 1060881 1085867 1088259 1188492 1177901 1033381 1186437 1186041 1188014 1186269 1099358 1102654 1102509 1138772 1133105 1148791 1038616 1085238 1105372 1161841 1098381 1119173 1113506 1041468 1160219 1159828 1160345 1168104 1039160 1284307 1186389 1284662 1106216 1186723 1174415 1182650 1179555 1185595 1186713 1188085 1186520 1072097 1026867 1098426 1076525 1098013 1103902 1104410 1099419 1134034 1138858 1136697 1138692 1111577 1152515 1007779 1141921 1009582 1123198 1088387 1195226 1098015 1040166 1042165 1077356 1048610 1049869 1051085 1128152 1088122 1086164 1064217 1066240 1072151 1157022 1157339 1197529 1085100 1186085 1072331 1197010 1077425 1085194 1123773 1087036 1157719 1157891 1157446 1169018 1167428 1101189 1099989 1085721 1086145 1184593 1141708 1153644 1097700 1134175 1151946 1158577 1191734 1191385 1137892 1134141 1191368 1193510 1140415 1140409 1140696 1018364 1023269 1023507 1023453 1023626 1023583 1023634 1023586 1023597 1023744 1023621 1023759 1023580 1157758 1153629 1069554 1072068 1195156 1088435 1140441 1108005 1142019 1144697 1085904 1098245 1037850 1038657 1044508 1062594 1055901 1162305 1072182 1149935 1171814 1061482 1051176 1151678 1141709 1072147 1127048 1122960 1124499 1087850 1087424 1065924 1061231 1022609 1086730 1103525 1168984 1167397 1136740 1137724 1151923 1154614 1149775 1077692 1072304 1035914 1134008 1134179 1149830 1092170 1148431 1123594 1128175 1124502 1072141 1074018 1170290 1049382 1053499 1096514 1133960 1148293 1142030 1150328 1169284 1040103 1039372 1161010 1059024 1157770 1107419 1104476 1197136 1197026 1075249 1153722 1153707 1176672 1161816 1187152 1178988 1035898 1124864 1098802 1085270 1135847 1088409 1194004 1088125 1089257 1140787 1072150 1042677 1041520 1149888 1054032 1137913 1039404 1104432 1121702 1066239 1062064 1070633 1072436 1140379 1103680 1104023 1104104 1149622 1137674 1137675 1193437 1193123 1177918 1055454 1148355 1188352 1050441 1049370 1049407 1053343 1070690 1039443 1085969 1085395 1085132 1101455 1086355 1189921 1122991 1088385 1128397 1193967 1140328 1088174 1095399 1184632 1037401 1088366 1088617 1085702 1063561 1140531 1037682 1128248 1053439 1161564 1153939 1157138 1157105 1195601 1165259 1162624 1163328 1161194 1153306 1094165 1182343 1176669 1176666 1056725 1035918 1167109 1186112 1175310 1141705 1149621 1062323 1128537 1038653 1134082 1053372 1055960 1056080 1123256 1151350 1197434 1086157 1168670 1053748 1072148 1197047 1122482 1101824 1153511 1141702 1148444 1123558 1124533 1070386 1088130 1086325 1086044 1187254 1085884 1049399 1097295 1082190 1189990 1053323 1191478 1062536 1128875 1085204 1068591 1040751 1123796 1103638 1089249 1091200 1088344 1165685 1184601 1106374 1176683 1184540 1187238 1161095 1187197 1055013 1058045 1157098 1162142 1165828 1197009 1152367 1149828 1195693 1150983 1195113 1127662 1157294 1162364 1127828 1052705 1065950 1088128 1187708 1066156 1128281 1064301 1103068 1085723 1088276 1157688 1088603 1035279 1107433 1196954 1197353 1190523 1193579 1140343 1162175 1157762 1147166 1176679 1193902 1193532 1051859 1100333 1140782 1056229 1197102 1159168 1149832 1169051 1169168 1167317 1042103 1123679 1057021 1119626 1197688 1121664 1137849 1042134 1096247 1086148 1056462 1109181 1066162 1068583 1187703 1108919 1080065 1136775 1157728 1196975 1182892 1099266 1053244 1150289 1054088 1124424 1123737 1142056 1135093 1121652 1149600 1049978 1053790 1053360 1088343 1152325 1101803 1154621 1120742 1123077 1123628 1035904 1140437 1140377 1140778 1151883 1053300 1153558 1071170 1156233 1157760 1120740 1157423 1119602 1160178 1121731 1059446 1088439 1035891 1097404 1122453 1165404 1091153 1196706 1182824 1165956 1152534 1178012 1184834 1160636 1035843 1041496 1086895 1161548 1169158 1123086 1120857 1194119 1037671 1122509 1121109 1091146 1108613 1048640 1096165 1049346 1051421 1050774 1059335 1064450 1134283 1184656 1195486 1065807 1165304 1086725 1106355 1101792 1104398 1104077 1104012 1103657 1066102 1070574 1128187 1103567 1085227 1074459 1085224 1134183 1194009 1148063 1193566 1062887 1149789 1085727 1157725 1168747 1169292 1151927 1035832 1100160 1159224 1121284 1062369 1049393 1156881 1086784 1103956 1104331 1195622 1103027 1069797 1036592 1098068 1032385 1157428 1087086 1086505 1081062 1107461 1059021 1086883 1060335 1031879 1135590 1124466 1157356 1067305 1148543 1042098 1095312 1148348 1148531 1085257 1041055 1070837 1161656 1112077 1072287 1078359 1108113 1085031 1128239 1024029 1103984 1104507 1133970 1133924 1133922 1137577 1133783 1193471 1192455 1195166 1135507 1142064 1108509 1140619 1149577 1149618 1150958 1151925 1152102 1154639 1187104 1156086 1157067 1097724 1152428 1158803 1151841 1128263 1113846 1168732 1039406 1081194 1161079 1053443 1157443 1161145 1050858 1052283 1056000 1055910 1167424 1168673 1196963 1197050 1191057 1176667 1054726 1186161 1131808 1061493 1082743 1065904 1140370 1124643 1128120 1124601 1195605 1153023 1123126 1070562 1070686 1091088 1088759 1088369 1035247 1098345 1085225 1133576 1131793 1040825 1040041 1038643 1152007 1105009 1134192 1172715 1161183 1164810 1139596 1187612 1176658 1175212 1184668 1187156 1197391 1035032 1089550 1081995 1072209 1081930 1085584 1081697 1133614 1133746 1141759 1094034 1053371 1120819 1142071 1040897 1161107 1187260 1148339 1178019 1039382 1107394 1062584 1187802 1084984 1022880 1061670 1139313 1074072 1103545 1104024 1161245 1120888 1120775 1158892 1164914 1023370 1149568 1131814 1088411 1193175 1140232 1162600 1157102 1195629 1197140 1084302 1108154 1034744 1176401 1095458 1086245 1157942 1023656 1185184 1113590 1157768 1157645 1164815 1198162 1149910 1039408 1086631 1103963 1101761 1081618 1074286 1184708 1128252 1142079 1141701 1061552 1083666 1178038 1184831 1087101 1035929 1035911 1066139 1123149 1121498 1069569 1107422 1091140 1128322 1023605 1139828 1063474 1140380 1168693 1169203 1108474 1088055 1086000 1193546 1088206 1089254 1100385 1108694 1147219 1176700 1169259 1165004 1053412 1152378 1163234 1053234 1059624 1101431 1151825 1081606 1128156 1120912 1120825 1134987 1131927 1135799 1086345 1090743 1174116 1088404 1107426 1040999 1042178 1096179 1088060 1157685 1157687 1086255 1150006 1041489 1050494 1157448 1157177 1197037 1190567 1152326 1107603 1085252 1039785 1140671 1161551 1112108 1119572 1115128 1176811 1058137 1066114 1284927 1066022 1101765 1134069 1137779 1136781 1027772 1150680 1167603 1007780 1007796 1007800 1007805 1017422 1017590 1017945 1018079 1285329 1075408 1020023 1020044 1092504 1069414 1094853 1094383 1074811 1112217 1126643 1023211 1023213 1023858 1023878 1023979 1025182 1284880 1027144 1027280 1027313 1028555 1028672 1030008 1030331 1030603 1031057 1031069 1031130 1106654 1106584 1107245 91647667 1031746 1031753 1031830 1032009 1032047 1033303 1033310 1033617 1033663 1050939 1052195 1052261 1052262 1111191 1111623 1111637 1112270 1112538 1112638 1112679 1116610 1118532 1119635 1119774 1035385 1035511 1040142 1284987 1041034 1047424 1046598 1046685 1047881 1041440 1041576 1041890 1042393 1041754 1043484 1042350 1044092 1043532 1043069 1041272 1167640 1167278 1167460 1168296 1168853 1168872 1169226 1168768 1168272 1168854 1165654 1166406 1062231 1131407 1166948 1166294 1041484 1041526 1041537 1170235 1174854 1041699 1172037 1172947 1173241 1174351 1041783 1041877 1041879 1041894 1041973 1042180 1042391 1042436 1042482 1042487 1042534 1042537 1042580 1042729 1043024 1043094 1043112 1043194 1043228 1043248 1043250 1043272 1043278 1043289 1125030 1125313 1082806 1119700 1158314 1158622 1158684 1159587 1159140 1159149 1168191 1168537 1168615 1168589 1168665 1170774 1170498 1170456 1170492 1170553 1170547 1171943 1169581 1170457 1170140 1169779 1170173 1167811 1173080 1172065 1174797 1175077 1171573 1174817 1043447 1124371 1124865 1129746 1130143 1130965 1131151 1131173 1163102 1139868 1157748 1167844 1168288 1168298 1175344 1175787 1162832 1170674 1170937 1173383 1171760 1043510 1043899 1043953 1112274 1045786 1045803 1045921 1046219 1046372 1046381 1046424 1177537 1177876 1178002 1046880 1177088 1181559 1182814 1181290 1190759 1046967 1047062 1047135 1047136 1047353 1047463 1047494 1191204 1191213 1191227 1047699 1047727 1047771 1047824 1047953 1048064 1048136 1048174 1048338 1048481 1174653 1173840 1186005 1051066 1051713 1052275 1052286 1053385 1054023 1054049 1054315 1054518 1054673 1054714 1054740 1054761 1054887 1054891 1055196 1055297 1055536 1055552 1057363 1057373 1057406 1057522 1057635 1057674 1057689 1057772 1057966 1058004 1058315 1058382 1058432 1058442 1058800 1059047 1059119 1059221 1059261 1059465 1059637 1075662 1075695 1061799 1061999 1173238 1174271 1176427 1177489 1178937 1179361 1179366 1179373 1179434 1173578 1177709 1092650 1074739 1096973 1177717 1177935 1177942 1178167 1178179 1170361 1064520 1094389 1106606 1107131 1107211 1107368 90958420 1179037 1179793 1174189 1176166 1177416 1178854 1178996 1174978 1178051 1178083 1170300 1177146 1178127 1178987 1179059 1179816 1072640 1074234 90651138 90658045 1091605 1095377 1111685 1112418 1112634 1112650 1118501 1119758 90800205 90631885 1166510 1168134 1167563 1168814 1169425 1169400 1166517 1169654 1169721 1075607 1124969 1166682 1169344 1169172 1168127 1168689 1168838 1168458 1169032 1165660 1166304 1166094 1166483 1167054 1166530 1072864 1074027 1174788 1169555 1172738 1174912 1176656 1071163 1122295 1123897 1067809 1124894 1125016 1119569 1144857 1144868 1144620 1072122 1157659 1158086 1158375 1158452 1159065 1159332 1159618 1168264 1169528 1168568 1168599 1168412 1168540 1168126 1169289 1170573 1170488 1170577 1170570 1170144 1170045 1170186 1169838 1172060 1171269 1171724 1171580 1174866 1174862 1174871 1125061 1125158 1131099 1131324 1163332 1163821 1180438 1151785 1162092 1167854 1176110 1162770 1171015 1170745 1170687 1170684 1170739 1171221 1171784 1117550 1069212 1168393 1168741 1181469 1181600 1173126 1173233 1173231 1173333 1070486 1176617 1172074 1177085 1176917 1178452 1177168 1177909 1177960 1178442 1072046 1072892 1073720 1074593 1074930 1075006 1075103 1075134 1075143 1075319 1075488 1075598 1075660 1138349 1138202 1138452 1137836 1076651 1134555 1134531 1134922 1092612 1092583 90938369 1093768 1094443 1094372 1107286 90880144 90763354 901028644 1095290 1108612 1112542 1112818 1116785 1116849 1119605 1122571 1120643 1122916 1121996 1121844 1099504 90953164 1167907 1167557 1167593 1169024 1168808 1169412 1169674 1169792 1168784 1168913 1168850 1167796 1168236 1168341 1165741 1165782 1166567 1165851 1166497 1169587 1172962 1174957 1172533 1174411 1178504 1171638 1174309 1124993 1125045 1125373 1100608 1119553 1125628 1125981 1144678 1157566 1157650 1157678 1158138 1158191 1158474 1158495 1158602 1159036 1159315 1159477 1159834 1168656 1169447 1169479 1169239 1170863 1170526 1184112 1185556 1180588 1183407 1169452 1170586 1171199 1170539 1170517 1171279 1169875 1170310 1172563 1173100 1171523 1174954 1174814 1130984 1131354 1163730 1163741 1180429 1145347 1145727 1162133 1168269 1175693 1175865 1158669 1162783 1171342 1171150 1170637 1171106 1170709 1171280 1171086 1171831 1171848 1171786 1162459 1171941 1171217 1172394 1118621 1167309 1168199 1177175 1177439 1181571 1190755 1191288 1191290 1174276 1171290 1176608 1177643 1176971 1177883 1179551 1176992 1112583 1112729 1174989 1167900 1169760 1177011 1177161 1177402 1181605 1180803 1190786 1172447 1139652 1139783 1118985 1138469 1145629 1139190 1173484 1173335 1174661 1173457 1176502 1176709 1176978 1179416 1177927 90992607 1106933 1119577 1173588 1173639 1177813 1177820 1177836 1169745 1170005 1173407 1173489 1145174 1144524 1144597 1146173 1144987 1145062 1133977 1145801 1112910 1124042 1127991 1112871 1125211 1129901 1130305 1130962 1131036 1090890 1111235 1111646 1111687 1111710 1111809 1111868 1111875 1112054 1112085 1112419 1112513 1112577 1166957 1125814 1125229 1125310 1163285 1177613 1092580 1092867 1094191 1129193 1129587 1113732 1129876 1162800 1163155 1123956 1123226 1126574 1131418 1130005 90758814 90806848 1163707 1173243 1173591 1175027 1160930 1169872 1173554 1139273 1145030 1145403 1145520 1144511 1144791 1107153 1107103 91763592 90689615 1127909 1129535 1129545 1126130 1130961 1130974 1101681 1167761 1167294 1167093 1168844 1169334 1169228 1169214 1168196 1169302 1168567 1169176 1168522 1166310 1165587 1166175 1165916 1129761 1166367 1165898 1172872 1172925 1172346 1178192 1176603 1176628 1164802 1111527 1111841 1112520 1113055 1131647 1121666 1125153 1125326 1125772 1125799 1125908 1103095 1119884 1150599 1144666 1145313 1158078 1158238 1158989 1159786 1159745 1171348 1170137 1169389 1171666 1172048 1172009 1174733 1174929 1129976 1127790 1131007 1131127 1131621 1163085 1163226 1131504 1163631 1163690 1151784 1145695 1157694 1158151 1175783 1176142 1171266 1170607 1170645 1170635 1171020 1166069 1170562 1171950 1171817 1179797 1173829 1124795 1131365 1162683 1175818 1176633 1178502 1178721 1178948 1179002 1177816 1177839 1178042 1178148 1178193 1112666 1155162 1174333 1174337 1178593 1178748 1178865 1178995 1179200 1177596 1170423 1178182 1179001 1179465 1179594 1179665 1179795 90698171 90760146 1111719 1111760 1112527 1112535 1112900 90840031 90808773 1118381 1173656 1177615 1178183 1161489 1170098 1170124 1173543 1145797 1144681 1144401 1138485 1131453 1140028 1123243 1123895 1126005 1129210 1129552 1129644 1143914 1130007 1131128 1131477 1162666 90543977 90562298 90597199 90620745 90664669 90699018 90708711 90723865 90808788 90819142 90830825 90833035 90837726 90839916 90855089 90859110 90862557 90864150 91876260 90955714 90953502 91760192 91669459 91616579 90946426 90965344 90933315 90936603 91626832 91541221 90880230 90916905 91594373 91881844 90951697 93660099 90954696 90882897 90938016 90667319 90617516 92700407 90624921 90604411 90873935 90744321 90820433 90751723 90922328 90800108 90882701 91543326 90852569 90649230 90617050 90620742 90807691 90855872 91537196 90929038 90658919 91546630 90876769 90618376 90531690 90640856 90606296 90659037 90719582 90757788 91862128 90619481 90919084 90876474 90939264 90943803 91839337 91627862 90840728 90775904 90594214 90623835 90823029 91695410 90542596 91899869 90853266 90900585 90870794 90825290 93673741 90842451 1007799 1017591 1017584 1017601 1017459 1017277 1017629 1017470 1017445 1017480 1017465 1017289 1017485 1017474 1017193 1017340 1017604 1017242 1017449 1017488 1017615 1017592 1017585 1017454 1017273 1017204 1017254 1017596 1017175 1017172 1017248 1017205 1017053 1017161 1016981 1017209 1016995 1017253 1017347 1017262 1017619 1017290 1017212 1017224 1017633 1017110 1017626 1017221 1017167 1017223 1017029 1017586 1016943 1025115 1016951 1025228 1025346 1025210 1025220 1017234 1017162 1017228 1017270 1025327 1017001 1017278 1016992 1017025 1017206 1017000 1017249 1017219 1017177 1017061 1025257 1017317 1017211 1017266 1025251 1025239 90877586 90736708 91648180 90821387 90693602 90852535 90800941 90677285 91847970 90938881 90844939 90601666 90623950 91720421 91636613 90958874 90701896 91612399 90739079 90926890 90708797 90946720 90853010 90598229 90834152 90865953 92669671 91594979 90730206 91620284 91656086 90824504 90824330 91676805 90586965 90541248 90675673 90673080 90739452 1043694 1059690 1057575 1057551 1058067 1058296 1059282 1048305 1071567 1072905 1052589 1047734 1046968 1030048 1026607 1048112 1033379 1023665 1019379 1020022 1017727 1024671 1023103 1177008 1178423 1179242 1179399 1173580 1174983 1177680 1178298 1177139 1178354 1179850 1137843 1138329 1137898 1137884 1135348 1138011 1059144 1092883 1073734 1092606 1106483 1118548 1118734 1134990 1134229 1135134 1174693 1173948 1173342 1173471 1173553 1177451 1177821 1170303 1173497 1173413 1176912 1176908 1176742 1157232 1177339 1176933 1144789 1144629 1144330 1143257 1074800 1145007 1144856 1144589 1145048 1074889 1094703 90987533 1128011 1127499 1127784 1128003 1129444 1113563 1130012 91670734 90831004 1106558 1118696 1106570 1106631 1111004 1111332 1131424 1131439 1173687 1177977 1178130 1170394 1175652 1052284 1052946 1062862 1111557 1112386 1112531 1112536 1112572 1112950 1118456 1122645 1123623 1123552 90728473 90643332 1123045 1123709 1122024 1121873 1123121 1055039 1096978 1107140 1106597 1035310 1095271 1092601 1112306 1112675 1047110 1047515 1047954 1047957 1032215 1033384 1040666 1041391 1041686 1042538 1042575 1043464 1130924 1131122 1131143 1131184 1131230 1131400 1131534 1131941 1163293 1041511 1046375 1046624 1047995 1032090 1023207 1041390 1042703 1042342 1042493 1042368 1043231 1043185 1042856 1042769 90846779 1171566 1173130 1174765 1175349 1176002 1179019 1173620 1178074 1178082 1169931 1179095 1179281 1179424 1070190 1143215 1139110 1155566 1138636 1144505 1145733 1167581 1168019 1169005 1167367 1168301 1169006 1167864 1167516 1169715 1169733 1169618 1168852 1168941 1168629 1168318 1165475 1102873 1165784 1166492 1166647 1166516 1167465 1168978 1167123 1167886 1169129 1169089 1168243 1169710 1169724 1075581 1168911 1169067 1168324 1168518 1168869 1165613 1165931 1165890 1166088 1175074 1172927 1176438 1173174 1176644 1172461 1171561 1175172 1175170 1176037 1113084 1124228 1124301 1124803 1125399 1109035 1119422 1125720 1144679 1158240 1158878 1158949 1059425 1072032 1074669 1092935 1171257 1171253 1171036 1169243 1092886 1047785 90915138 1170183 1170050 1172721 1172570 1172331 1170975 1171546 1174898 1175067 1174918 1121495 1130982 1131156 1163110 1059075 1124987 1125003 1125118 1119763 1126634 1157438 1157669 1157716 1158224 1158421 1158470 1158506 1158960 1159080 1158448 1159069 1159046 1165855 1168597 1169619 1169237 1168837 1170719 1170761 1171188 1170803 1171246 1171919 1170117 1170453 1170383 1170338 1169429 1171025 1174726 1171550 1171522 1150631
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2343-2025
Related Recalls
Insulin pump includes a vibration motor that gives tactile feedback for any alerts, alarms, or malfunctions. Pump may exhibit false vibration motor failure due to a software issue causing Malfunction 12: "Pump cannot operate, the mobile app can no longer receive data from the pump. Insulin delivery and any active CGM Sessions have been stopped", which could result in hyperglycemia.
A software defect in versions 7.9.0.1 and 7.10.1 of the pump software for Tandem t:slim X2 and version 7.9.0.1 of the Tandem Mobi pump with Control-IQ+ technology that are paired with a G7 sensor may result in an unexpected automatic insulin correction bolus (autobolus ) which could result in hypoglycemia.
t:slim X2 Insulin Pump with Interoperable Technology (with Tandem t:slim mobile app)
Tandem Diabetes Care
an app defect that occurs when the phone is set to a right-to-left language, which causes app-pump pairing issues and graphical defects, which can potentially lead to incorrect therapy decisions and subsequent hypoglycemia or hyperglycemia.