GE Healthcare CARESCAPE Monitor B650
Class I - DangerousWhat Should You Do?
- Check if you have this product: Serial Numbers: SEW13497373HA SEW13497374HA SEW13497375HA SEW13497376HA SEW13508068HA SEW12494591HA SEW14280294HA SEW14280298HA SEW14289769HA SEW14289777HA SK416341018HA SK416341019HA SK416341026HA SEW12494422HA SEW12494425HA SEW12494435HA SEW12494584HA SEW12494585HA SEW12494592HA SEW12494593HA SEW12494596HA SEW12494598HA SEW12494619HA SEW12494621HA SEW12494622HA SEW12494625HA SEW12494626HA SEW12504918HA SEW12504923HA SEW12504940HA SEW12504942HA SEW12504944HA SEW12504954HA SEW14278996HA SEW14279002HA SEW14279003HA SEW14279005HA SK416061047HA SK416061048HA SK416061088HA SK416081102HA SK416461063HA SK416461092HA SEW13109786HA SEW12059577HA SEW12059583HA SEW12059587HA SEW12059589HA SEW12059592HA SEW12059593HA SEW12059594HA SEW12059596HA SEW12059597HA SEW12059598HA SEW12059599HA SEW12059601HA SEW13275956HA SEW13276028HA SEW13276029HA SEW13275967HA SEW13276066HA SEW13276073HA SEW13286164HA SEW13286175HA SEW13286183HA SK415491246HA SK415491249HA SK415491253HA SK415491257HA SK415491259HA SK415491264HA SK415491265HA SQC17490035HA SQC17490038HA SQC17490039HA SQC17490040HA SQC17490042HA SEW15134264HA SEW15134269HA SEW15134270HA SK415211108HA SK415211110HA SK415211116HA SK415221002HA SK415221006HA SK415361118HA SK415361119HA SK415361120HA SK415361121HA SK415361122HA SK415361123HA SK416041059HA SK416051085HA SK416051119HA SK416041053HA SK416041054HA SK416041062HA SK416041063HA SK416041065HA SK415491254HA SK415521029HA SK415531048HA SK416011088HA SK416011113HA SK416501112HA SK416501119HA SK416501122HA SK417051035HA SK417051036HA SK417051038HA SK417051099HA SK417051100HA SK417051118HA SK417051121HA SK417051122HA SK417051125HA SK417051126HA SK417051130HA SK417051131HA SK417051139HA SK417051144HA SK417051145HA SK417051149HA SK417051150HA SK417051151HA SK417051152HA SK417051153HA SK417051154HA SK417051157HA SK417061002HA SK417061023HA SK417061024HA SK417061025HA SEW13508579HA SEW13369186HA SEW13369234HA SEW13369245HA SEW13369246HA SEW13369249HA SEW13369251HA SEW13369252HA SEW13369254HA SEW13369255HA SEW13369256HA SEW14381680HA SEW15166221HA SEW15166222HA SEW15166223HA SEW15166224HA SEW15166226HA SEW15166230HA SEW15166231HA SEW15166235HA SEW15166236HA SEW15166237HA SEW14063178HA SEW14062864HA SEW14062865HA SEW14062927HA SEW14063174HA SEW14063175HA SEW14063176HA SEW14063179HA SEW14062866HA SEW14235956HA SEW14235957HA SEW14235958HA SEW14235993HA SEW14235994HA SEW14235995HA SEW14246223HA SK416051106HA SK416051113HA SK416051114HA SK416051115HA SEW11202291HA SK416121077HA SK416121083HA SEW12245068HA SEW12245071HA SEW12245072HA SEW12245074HA SEW12245075HA SEW12245076HA SEW12482932HA SQC18150040HA SQC18150042HA SK416321071HA SK416321073HA SK416321075HA SK416321076HA SEW13317664HA SQC17500211HA SQC17500212HA SQC17500213HA SQC17500214HA SQC17500215HA SQC17500216HA SQC17500217HA SQC17500218HA SQC17500219HA SQC17500220HA SQC17500221HA SQC17500222HA SQC17500224HA SQC17500225HA SQC17500226HA SQC17500228HA SQC17500230HA SQC17500234HA SQC17500235HA SEW14063243HA SQC18010001HA SK417081005HA SEW13369260HA SEW13369277HA SEW13379768HA SEW13379778HA SEW13379779HA SEW13379780HA SEW13379782HA SEW13379784HA SEW13379790HA SEW13379791HA SEW14031040HA SEW14031041HA SEW14031042HA SEW14031044HA SEW14031045HA SEW14031046HA SEW14031047HA SEW14031060HA SEW14031061HA SEW14031062HA SEW14031066HA SEW14041261HA SEW14041262HA SEW14041263HA SEW14041264HA SEW14041267HA SEW14041268HA SEW14041269HA SEW14041271HA SEW14041282HA SEW14041283HA SEW14041284HA SEW14041286HA SEW14041287HA SEW14041289HA SEW14041290HA SEW14041291HA SEW14063210HA SEW14105484HA SEW14105968HA SEW14105969HA SEW14116795HA SEW14127301HA SEW14127307HA SEW14127308HA SEW14127312HA SEW14127313HA SEW14127314HA SEW14127315HA SEW14127316HA SEW14127317HA SEW14127318HA SEW14127319HA SEW14127320HA SEW14127321HA SEW14127322HA SEW14127323HA SEW14127324HA SEW14127325HA SEW14127326HA SEW14127327HA SEW14127328HA SEW14127329HA SEW14127330HA SEW14127331HA SEW14127332HA SEW14127333HA SEW14127334HA SEW14127335HA SEW14127631HA SEW14127632HA SEW14127633HA SEW14127634HA SEW14127635HA SEW14127636HA SEW14127637HA SEW14127638HA SEW14127639HA SEW14127863HA SEW14127864HA SEW14127865HA SEW14127866HA SEW14127867HA SEW13379789HA SEW13379793HA SQC17450282HA SQC17450283HA SQC17450284HA SQC17450285HA SQC17450287HA SQC17450289HA SQC17450292HA SQC17450293HA SEW12388960HA SEW12388971HA SEW12389025HA SEW12389026HA SEW12494289HA SEW12494291HA SEW12494292HA SEW12494293HA SEW12494294HA SEW12494296HA SEW12494316HA SEW12494319HA SEW12494320HA SEW12494321HA SEW12494323HA SEW12494324HA SEW12494325HA SEW12494328HA SEW12494559HA SEW12494561HA SEW12494562HA SEW12494568HA SEW12494603HA SEW12494605HA SEW12494606HA SEW12494607HA SEW12494608HA SEW12494610HA SEW12494611HA SEW12494612HA SEW12494613HA SEW12494615HA SEW13265672HA SEW13265673HA SEW13265674HA SEW13275965HA SK416071199HA SK416071202HA SK416071205HA SK416071206HA SK416071207HA SK416071208HA SK416071209HA SK416071210HA SK416071211HA SK416071212HA SK416071213HA SK416071214HA SK416071225HA SK416071227HA SK416071228HA SK416071229HA SK416071230HA SK416071231HA SK416071232HA SK416071233HA SEW11243341HA SEW11243343HA SEW11243344HA SEW11243345HA SEW11243350HA SEW11446229HA SEW11446230HA SEW11446231HA SEW11446236HA SEW14127659HA SEW14127663HA SEW14127671HA SEW14127672HA SEW14127673HA SEW15155904HA SEW15166217HA SEW15166225HA SEW15166227HA SK415361034HA SEW12245031HA SK417011071HA SK417011072HA SK417011073HA SK417011077HA SK417011079HA SQC18010097HA SQC18010098HA SQC18010109HA SEW14496326HA SEW14496327HA SEW14496333HA SEW14496334HA SEW14496353HA SK417111097HA SK417111092HA SK417111098HA SK417111100HA SK417111101HA SK417111103HA SK417111104HA SK417111105HA SK417111106HA SK417111110HA SK417111112HA SK417111113HA SEW12131766HA SEW12244579HA SEW12244580HA SEW12357582HA SEW14459881HA SEW15054763HA SEW15054764HA SEW15054765HA SEW15054805HA SK416421111HA SK416421119HA SK416421120HA SK417021019HA SK417021025HA SK417021027HA SQC17520016HA SQC17520017HA SEW12399425HA SEW12399426HA SEW12399427HA SEW12399428HA SEW12399429HA SEW12399430HA SEW12399432HA SEW12399434HA SEW12399435HA SEW12399436HA SEW12399437HA SK415311122HA SK415311131HA SK415311132HA SK416371226HA SK416371230HA SEW14029842HR SQC18110196HA SQC18110198HA SQC18120082HA SQC18120083HA SQC18120084HA SEW11202299HA SK415361010HA SK415361124HA SK415361126HA SK415361127HA SK415361128HA SK415361129HA SK415361130HA SK415361197HA SK415361203HA SK415361206HA SK415361209HA SK415361213HA SK415361216HA SK415361220HA SK415371002HA SK415371006HA SEW13475994HA SEW13476000HA SEW13476001HA SEW13475998HA SEW13476003HA SK416071011HA SK416071020HA SK416071021HA SK416071022HA SK416071023HA SK416071024HA SK416071025HA SK416071026HA SK416071027HA SK416071029HA SEW11243443HA SEW11243450HA SEW11243451HA SEW11243452HA SEW11243456HA SEW11243457HA SEW12059455HA SEW12059456HA SEW12059461HA SEW12059462HA SK415390013HA SK416151168HA SK416151169HA SK416151170HA SK417341201HA SK417021061HA SEW13286262HA SEW13296800HA SEW13296801HA SEW13296803HA SEW13296804HA SEW13296805HA SEW13296806HA SEW13296821HA SEW13296826HA SEW13296827HA SEW13296828HA SEW13296831HA SEW13296843HA SEW13296844HA SEW13296848HA SEW13296849HA SEW13296850HA SK416461126HA SK416461133HA SK416461134HA SK416461136HA SK416461137HA SK416461138HA SK416461139HA SK416461140HA SK416461141HA SK416461142HA SK416461144HA SK416461149HA SEW12018076HA SEW12018094HA SEW13203376HA SQC18120171HA SK417261197HA SK417281074HA SK417281106HA SK417291202HA SK417301011HA SK417301029HA SK417301030HA SK417301031HA SK417301038HA SK417301227HA SK417301230HA SK417301231HA SK417301232HA SK417301233HA SK417301234HA SK417301238HA SK417301239HA SK417311017HA SK417311018HA SK417311019HA SK417311020HA SK417311024HA SK417311025HA SK417311036HA SK417311037HA SK417311038HA SK417311039HA SK417311040HA SK417311041HA SK417311042HA SK417311043HA SK417311044HA SK417311045HA SK417311046HA SK417311053HA SK417311055HA SK417311056HA SK417311057HA SK417311058HA SK417311059HA SK417311060HA SK417311061HA SK417311062HA SK417311063HA SK417311064HA SK417311065HA SK417311066HA SK417311067HA SK417311089HA SK417311090HA SK417311113HA SK417311115HA SK417311116HA SK417311118HA SK417311138HA SK417311139HA SK417311140HA SK417311141HA SK417311142HA SK417311143HA SK417311144HA SK417311145HA SK417311146HA SK417311147HA SK417311148HA SK417311149HA SK417311150HA SK417311151HA SK417311152HA SK417311153HA SK417311154HA SK417311155HA SK417311156HA SK417311157HA SK417311158HA SK417311159HA SK417311161HA SK417311198HA SK417321070HA SK417321114HA SK417321115HA SK417321116HA SK417321117HA SK417321119HA SK417321120HA SK417321121HA SK417321125HA SK417321147HA SK417331003HA SK417331004HA SK417331005HA SK417331008HA SK417331028HA SK417331041HA SK417331042HA SK417331043HA SK417331044HA SK417331157HA SQC17430011HA SEW11202775HA SEW11202778HA SEW11223011HA SEW11223022HA SEW11223123HA SEW11263888HA SEW12234397HA SEW14302254HA SEW14302255HA SEW14302257HA SEW14302263HA SEW14347390HA SEW12028675HA SEW12028676HA SEW12028677HA SEW12028678HA SEW12028680HA SEW12028681HA SEW12028682HA SEW12182716HA SEW12245080HA SEW12245082HA SQC17420031HA SQC17420053HA SQC18100142HA SQC18100143HA SQC18100144HA SQC18100153HA SQC18100154HA SQC18040085HA SEW14473939HA SEW14484145HA SK417111024HA SK417111026HA SK417111027HA SK417111028HA SEW15101852HA SK417341202HA SK417341205HA SK417341206HA SK417341207HA SK417341210HA SK417341212HA SK417341213HA SK417341214HA SK417341229HA SK417381247HA SK417381250HA SQC17420215HA SQC17430002HA SQC17430003HA SQC17430012HA SEW10354778HA SEW12285885HA SK417061148HA SEW13131087HA SEW13131088HA SEW14280584HA SEW14280598HA SEW15271035HA SEW15271038HA SEW15381015HA SEW15381017HA SEW11079419HA SEW11079425HA SEW11079436HA SEW11079441HA SEW11079465HA SEW11079479HA SEW11079496HA SEW11059204HA SEW11079373HA SEW11079378HA SEW11079463HA SEW11079466HA SEW11079469HA SEW11079493HA SEW11079507HA SEW11079509HA SEW11079511HA SEW11079512HA SEW11079534HA SEW13172414HA SEW13182787HA SEW13519440HA SEW13519441HA SEW13519443HA SEW13519444HA SEW14160610HA SEW14160613HA SEW14170856HA SEW14170868HA SEW14171070HA SEW14171071HA SK417081003HA SK417081130HA SK417101089HA SK417121090HA SK417121095HA SK417121096HA SK417121097HA SK417121098HA SK417121099HA SK417121100HA SK417121101HA SK417121102HA SK417121103HA SK417121104HA SK417121105HA SK417121106HA SK417121107HA SK417121108HA SK417121110HA SK417121111HA SK417221043HA SK416331162HA SK416331185HA SK416331186HA SK416341048HA SK416341060HA SK416341065HA SK416341066HA SK416341067HA SK416341068HA SK416341070HA SK416341072HA SK416341073HA SK416341074HA SK416341077HA SK416341078HA SK416341079HA SK416341080HA SK416341081HA SK416341082HA SK416341083HA SK416341084HA SK416341085HA SK416341086HA SK416341087HA SK416341088HA SK416341089HA SK416341090HA SK416341091HA SK416341092HA SK416341093HA SK416341094HA SK416341095HA SK416341096HA SK416341097HA SK416341098HA SK416341099HA SK416341100HA SK416341116HA SEW14392029HA SEW14392037HA SEW14392038HA SEW14392042HA SEW14392044HA SK416371112HA SQC17400053HA SQC17400054HA SQC17400055HA SQC17400056HA SQC17400057HA SQC17400058HA SQC17440152HA SQC17440180HA SQC17440182HA SQC17440198HA SQC17440199HA SQC17450102HA SEW13454551HR SEW15065986HA SEW15065987HA SK415390009HR SK415390012HR SK415390014HR SEW13411759HA SEW13411760HA SEW13411761HA SEW13411762HA SEW13411764HA SEW13411765HA SEW13411766HA SEW13411767HA SEW13411925HA SEW13411928HA SEW13432945HA SEW13432947HA SEW13432948HA SEW13432952HA SEW13432953HA SEW13432954HA SEW13432955HA SEW13432958HA SEW13432959HA SEW13432962HA SEW13464760HA SEW13464764HA SEW13464765HA SEW13464768HA SEW13464769HA SEW13464770HA SEW13464772HA SEW13508138HA SEW13518959HA SEW13518960HA SEW13518962HA SEW14246763HA SEW14246773HA SEW14246774HA SEW14246775HA SEW14247266HA SEW14247287HA SEW14224894HA SK415511077HA SK415511090HA SK415511092HA SK415511094HA SK415511095HA SK415511096HA SK415511098HA SK415511100HA SK415511101HA SK415511102HA SK415511103HA SK415511104HA SK415511123HA SK415511124HA SK415511126HA SK415511127HA SK415511128HA SK415511130HA SK416191169HA SK415511165HA SK415511177HA SK416301151HA SK416311002HA SK416311003HA SK416311004HA SK416311005HA SK416311009HA SK416311050HA SK416311052HA SK416311053HA SK416311054HA SK416311055HA SK416311056HA SK416311057HA SK416311063HA SK416311064HA SK416311065HA SK416311066HA SK416341055HA SK416341056HA SK416341057HA SK416341059HA SK416341062HA SK416341063HA SK416341064HA SK416221117HA SK416221118HA SK416221119HA SK416221120HA SK416221121HA SK416221122HA SK416221123HA SK416221200HA SK416221201HA SK416221205HA SK416221206HA SK416221207HA SK416221209HA SK416231032HA SK416341058HA SK416401057HA SK416401058HA SK416401059HA SK416401062HA SK416401063HA SK416401064HA SK416401066HA SK416401068HA SK416401071HA SK417211179HA SK417221231HA SK417221232HA SK417221234HA SK417221235HA SK417221236HA SK416231214HA SEW12142057HA SEW12152088HA SEW12193058HA SEW12203363HA SEW12203364HA SEW12203365HA SEW12203366HA SEW12203367HA SEW12203368HA SEW12265428HA SEW12265429HA SEW12265430HA SEW12265431HA SEW12265432HA SEW12265592HA SEW12265593HA SEW12265596HA SEW12265597HA SEW12265598HA SEW12265599HA SEW12265600HA SEW12275631HA SEW12275632HA SEW12275633HA SEW12275656HA SEW12275678HA SEW12275679HA SEW12275685HA SEW12275686HA SEW12275687HA SEW12275688HA SEW12275689HA SEW12275690HA SEW12275691HA SEW12275692HA SEW12275694HA SEW12275695HA SEW12275696HA SEW12275698HA SEW12275699HA SEW12275704HA SEW12285973HA SEW12357257HA SEW12368068HA SEW13286647HA SEW13296671HA SEW13296674HA SEW13296681HA SEW13296682HA SEW13296683HA SEW13296685HA SEW13296687HA SEW13296688HA SEW13296689HA SEW13296690HA SEW13296692HA SEW13296697HA SEW13296699HA SEW13296701HA SEW13296702HA SEW13296706HA SEW13296708HA SEW13296710HA SEW13296712HA SK417071135HA SK417071137HA SK417071138HA SK417071140HA SK417231322HA SK417231323HA SK417231324HA SK417231326HA SK417231327HA SK417231330HA SK417231331HA SK417241001HA SK417381100HA SK417381101HA SK417381102HA SK417381103HA SEW12131856HA SEW12131859HA SEW12131860HA SEW12131877HA SEW12131879HA SEW12131886HA SEW12131887HA SEW12142061HA SEW12142062HA SEW12142063HA SEW12244544HA SEW12244545HA SEW12244547HA SEW12244552HA SEW12244554HA SEW12244556HA SEW12244717HA SEW12244718HA SK416431024HA SK416431026HA SK416431027HA SK416431037HA SK417211175HA SK417231206HA SK417231210HA SK417231222HA SK417241042HA SK417241043HA SK417241044HA SK417241045HA SK417241046HA SK417241047HA SK417241048HA SK417241050HA SK417251051HA SK417251052HA SK417251053HA SK417251054HA SK417251056HA SK417251058HA SK417251060HA SK416031042HA SK416031053HA SK416031054HA SEW12131778HA SEW12131779HA SEW12131791HA SEW12131794HA SEW12131797HA SK415260001HA SK416041098HA SEW12296316HA SEW12296317HA SEW12296318HA SEW12306478HA SEW12306479HA SEW12306484HA SEW11477028HA SEW11517992HA SEW11517997HA SK416231013HA SK416231017HA SK416231023HA SK416231025HA SK416231026HA SK416231027HA SK416231028HA SK416231029HA SEW11202381HA SEW14019781HA SK416181031HA SK416181032HA SK416181033HA SK416181034HA SK416181036HA SK416181037HA SK416181038HA SK416181040HA SK416181041HA SK416181042HA SK416181043HA SK416181044HA SK416181045HA SK416181046HA SK416181050HA SK416181051HA SK417141069HA SK417141070HA SEW12494263HA SEW12494269HA SEW12494271HA SEW12494272HA SEW12494275HA SEW12494587HA SEW12494628HA SEW11059217HA SEW11202769HA SEW12245129HA SEW15190043HA SEW15190046HA SEW15190047HA SEW15190048HA SEW15190049HA SEW15200001HA SEW15200002HA SEW15200003HA SEW15200004HA SEW15200005HA SEW15200006HA SEW15200007HA SEW15200008HA SEW15200009HA SEW15200010HA SEW15220001HA SEW15220002HA SEW15220004HA SEW15220005HA SEW15220006HA SEW15220007HA SEW15220008HA SEW15220009HA SEW15220010HA SEW15220011HA SEW15220012HA SEW15220018HA SEW15220020HA SEW15241048HA SEW15241068HA SEW15241097HA SEW15241098HA SEW15241099HA SEW15241100HA SEW15241101HA SEW15241102HA SEW15241103HA SEW15241122HA SEW15241123HA SEW15241124HA SEW15241125HA SEW15241126HA SEW15251001HA SEW15251002HA SEW15251003HA SEW15251004HA SEW15251005HA SEW15251006HA SEW15251007HA SEW15251008HA SEW15251009HA SEW15251011HA SEW15251012HA SEW15251013HA SEW15251014HA SEW15251016HA SEW15251017HA SEW15251022HA SK415141076HA SEW14381659HA SEW14381661HA SEW14381665HA SEW14381666HA SEW14381668HA SEW14381669HA SEW14381670HA SEW14381673HA SEW14381675HA SEW14381676HA SEW14381677HA SEW14381682HA SEW14381683HA SEW14415291HA SEW14415292HA SEW14415298HA SEW14415301HA SEW14415302HA SEW14426253HA SEW14426257HA SEW14426263HA SEW14437098HA SEW14437118HA SEW14437120HA SEW14437125HA SEW14437126HA SEW14437127HA SEW14437129HA SEW14437139HA SEW14437140HA SEW14437141HA SEW14437142HA SEW14437143HA SEW14437149HA SEW14437156HA SEW14437157HA SEW14437158HA SEW14437159HA SEW14148891HA SEW14148892HA SEW14148899HA SK417311199HA SK417311200HA SK417311201HA SK417311204HA SK417311205HA SK417321002HA SK417321004HA SK417321013HA SK415511164HR SK415511167HR SK416021098HR SK416061019HR SQC17470152HA SQC17470153HA SQC17470154HA SQC17470155HA SQC17470161HA SQC17470162HA SQC17470163HA SQC17470164HA SEW13475573HA SEW13475574HA SEW13475576HA SEW14160609HA SEW14170855HA SEW14170864HA SEW14508174HA SEW11263783HA SEW13275963HA SEW13275977HA SEW12378484HA SEW12378485HA SEW12378487HA SEW12378488HA SEW12378489HA SEW12378490HA SEW12378491HA SEW11405568HA SEW11477093HA SEW11487240HA SEW11487251HA SEW11487279HA SEW11487322HA SEW11497477HA SEW12049008HA SEW12244833HA SEW12245017HA SEW12245018HA SEW12245019HA SEW12245029HA SEW12245030HA SEW13464740HA SEW13464742HA SEW13464743HA SK417231204HA SK417231290HA SK417241084HA SK417241086HA SK417241104HA SK417241105HA SK417241106HA SK417241107HA SK417241108HA SK417241109HA SK417241110HA SK417241111HA SK417241112HA SK417241113HA SK417241114HA SK417241115HA SK417241116HA SK417241117HA SK417241118HA SK417241119HA SK417241120HA SK417241121HA SK417241122HA SK417241123HA SK417241124HA SEW14313345HA SEW14313356HA SEW15055425HA SEW15065967HA SEW15065984HA SEW12505570HA SEW13026793HA SEW13026844HA SEW13026861HA SEW13026862HA SEW13026863HA SEW13026865HA SEW13026870HA SEW13026875HA SEW13026936HA SEW13026939HA SEW13026954HA SEW13026955HA SEW12409606HA SEW12483210HA SEW12483214HA SEW13234375HA SEW13234439HA SEW13234440HA SEW13234441HA SEW13234442HA SEW13234444HA SEW13234450HA SK417271092HA SK417271095HA SQC17510134HA SEW12368046HA SEW12389045HA SEW13475604HA SK416461035HA SK417021013HA SK417021022HA SEW11202317HA SEW13519436HA SEW14160612HA SEW14160632HA SEW14170854HA SK415341074HA SK415341075HA SK415341076HA SK415341082HA SK415341091HA SK415341092HA SK415341093HA SK415341108HA SK415341111HA SK415341113HA SK416051018HA SK417291054HA SK417291056HA SK417291082HA SK417291083HA SK417291090HA SK417291091HA SK417291092HA SK417291093HA SK417291097HA SK417291098HA SK417291099HA SK417291100HA SK417291102HA SK417291135HA SK417291136HA SK417291137HA SQC18010091HA SQC18010123HA SQC18020097HA SQC18030119HA SQC18040023HA SQC18040112HA SQC18060132HA SQC18070048HA SQC18070049HA SQC18070055HA SQC18070056HA SQC18070057HA SQC18070058HA SQC18070059HA SQC18070060HA SQC18070061HA SQC18070062HA SQC18070063HA SQC18070064HA SQC18070065HA SQC18070066HA SQC18070077HA SQC18070078HA SQC18070079HA SQC18070080HA SQC18070081HA SQC18070082HA SQC18070084HA SQC18070085HA SQC18070086HA SK416251008HA SK416251011HA SK416511038HA SK416511148HA SK416511149HA SK417031013HA SK416221096HA SK416221097HA SK416221100HA SK416221105HA SK416221173HA SK416221175HA SK416221177HA SK416221192HA SK416221193HA SK416221194HA SK416251004HA SK416251006HA SK416251007HA SK416251012HA SK416251013HA SK416251017HA SK416251018HA SK416251020HA SEW15020439HA SEW15020440HA SEW15020441HA SEW15020442HA SEW15020443HA SEW15020444HA SEW15020445HA SEW15020447HA SEW15020448HA SEW15020449HA SEW14268779HA SEW14268781HA SEW14278924HA SEW14278929HA SEW14278932HA SEW14278936HA SEW14278937HA SEW14278938HA SEW14278942HA SEW14278957HA SEW15020457HA SEW15020458HA SEW15020459HA SEW15020460HA SEW15020462HA SEW15020463HA SEW15020464HA SEW15020465HA SEW15020466HA SEW15020467HA SEW15020469HA SEW15020470HA SEW15020474HA SEW15020478HA SEW15020479HA SEW15020480HA SEW15020481HA SEW15020491HA SEW15020496HA SEW15124003HA SEW12121391HA SEW12121431HA SEW12121432HA SEW12121434HA SEW12121435HA SEW12131743HA SEW12141994HA SEW12142006HA SEW12142010HA SEW12142020HA SEW12142021HA SEW12142022HA SEW12142023HA SEW12049295HA SEW12049297HA SEW14448191HA SEW14448193HA SEW14448196HA SEW15066817HA SEW11436062HA SEW11436070HA SEW11436073HA SEW11436099HA SEW12245028HA SEW14127293HA SK417241067HA SK417251024HA SK417251028HA SK417251029HA SK417251030HA SK417251031HA SK417251033HA SK417251034HA SK417251035HA SK417251036HA SK417251037HA SK417251039HA SK417251041HA SK417251042HA SEW14127870HA SEW14127871HA SEW14127878HA SEW14127879HA SEW14127880HA SEW14127881HA SEW14127882HA SEW14127883HA SEW14127884HA SEW14127885HA SEW14127886HA SEW14127888HA SEW14127889HA SEW14127890HA SEW14127891HA SEW14127892HA SEW14127893HA SEW14127894HA SEW14127895HA SEW14127896HA SEW14127897HA SEW14127898HA SEW14127899HA SEW14127900HA SEW14127907HA SQC18090129HA SQC18090132HA SQC18090133HA SQC18090134HA SQC18090135HA SQC18090136HA SK416221111HA SK416221112HA SEW13327772HA SEW13327802HA SEW13327803HA SEW13327804HA SEW14508178HA SEW12505585HA SEW12505620HA SEW12505772HA SEW12505783HA SEW12505786HA SEW12505789HA SEW12505800HA SEW12505804HA SEW12505805HA SEW12505806HA SEW12505827HA SEW12515638HA SEW12515639HA SK415411123HA SK415411124HA SK415491336HA SK415501020HA SK415501022HA SK415501023HA SK417271132HA SK416211122HA SK416211128HA SK416211174HA SK416221071HA SK416221142HA SK416221143HA SK416221145HA SK416221185HA SK416221186HA SEW14313327HA SEW14313653HA SEW14313663HA SEW14313664HA SEW14313679HA SEW14313680HA SEW14313681HA SEW14313682HA SQC17420061HA SQC17420067HA SQC17420069HA SQC17420075HA SQC18030184HA SQC18040075HA SEW11079353HA SEW11079394HA SEW11079396HA SEW11079526HA SEW11079533HA SEW11109962HA SEW11109977HA SEW11109978HA SEW11202309HA SEW11202322HA SEW14448266HA SK416011112HA SK416011114HA SEW14073459HA SEW14073460HA SEW14149136HA SK415521036HA SK415521039HA SK415521040HA SK415521042HA SK415521043HA SK415521045HA SK415521046HA SK415521048HA SK415521050HA SK417141062HA SK417141063HA SK417141064HA SK417141065HA SK415391019HA SK416051081HA SK416051082HA SK417291094HA SK417291095HA SK417291096HA SK415401027HA SK416011035HA SEW13026364HA SEW13026368HA SEW13026369HA SEW13026377HA SK416261088HA SK416261089HA SK416261090HA SK416261091HA SK416261093HA SK416261095HA SK416261096HA SK416261098HA SK416261099HA SK416261100HA SK416261101HA SK416261104HA SK416261105HA SK416261115HA SK416271171HA SK416271174HA SK416281009HA SK416281017HA SK417381233HA SK416341022HA SEW11079370HA SEW12142002HA SEW13255296HA SEW13327793HA SK415421058HA SK415421063HA SEW14392092HA SK417251040HA SK417251149HA SK417251150HA SK417251151HA SK417251152HA SK417251153HA SK417251154HA SK417251156HA SK417261102HA SK417271146HA SK417271147HA SK417271148HA SK417271153HA SK417291244HA SK417291273HA SK417301006HA SK417361071HA SK417361072HA SK417361073HA SK417121003HA SK417121004HA SK417121005HA SK417121006HA SK417121030HA SK417121031HA SK417121032HA SK417121058HA SK417131011HA SK417131012HA SK417131013HA SK417131014HA SK417131015HA SK417131016HA SK417131017HA SK417131018HA SK417131020HA SK417131021HA SK417131023HA SK417131024HA SK417131025HA SK417131026HA SK417131027HA SK417131028HA SK417131029HA SK417131030HA SK417131031HA SK417131032HA SK417131033HA SK417131034HA SK417131035HA SK417131037HA SK417131038HA SK417131039HA SK417131040HA SK417131041HA SK417131042HA SK417131043HA SK417131044HA SK417131045HA SK417131046HA SK417131047HA SK417131048HA SK417131006HA SK417141072HA SK417141108HA SK417141109HA SK417141110HA SK417141111HA SK417141112HA SK417141113HA SK417141114HA SK417141115HA SK417141116HA SK417141117HA SK417141118HA SK417141119HA SK417141120HA SK417141121HA SK417141122HA SEW13234433HA SEW13244743HA SEW13244760HA SEW13244770HA SEW13244973HA SEW13244974HA SEW13244975HA SEW13244982HA SEW13244983HA SEW13244987HA SEW13244988HA SEW13244989HA SEW13244990HA SEW13244992HA SEW13244993HA SEW13244994HA SEW13244995HA SEW13245019HA SEW13245020HA SEW13245021HA SEW14279006HA SEW14279014HA SEW14279016HA SEW14279017HA SEW14279019HA SEW14279448HA SEW14279451HA SEW14279452HA SEW14279453HA SEW14279454HA SEW14279455HA SEW14279456HA SEW14279457HA SEW14279458HA SEW14279459HA SEW14279460HA SEW14279465HA SEW14291041HA SEW14325257HA SEW14325430HA SEW14325431HA SEW14325432HA SEW14325452HA SEW14325569HA SEW14325579HA SEW14325580HA SEW14325581HA SEW14325582HA SEW14325589HA SK415241033HA SK415251001HA SK415261011HA SK415261012HA SK415261013HA SK415261014HA SK415261021HA SK415261024HA SK415281018HA SK416101070HA SK417201063HA SK417201064HA SK417201065HA SEW11181943HA SEW14127710HA SEW13348789HA SEW13348816HA SEW13348817HA SEW15090492HA SEW15090494HA SEW15090495HA SEW15090496HA SEW15090497HA SEW15101833HA SEW15101834HA SEW15101857HA SK415511060HA SEW14149122HA SEW14138489HA SEW14149117HA SEW14019784HA SK415491115HA SK415491116HA SK415491119HA SK415491172HA SK415491177HA SK415491178HA SEW15134271HA SEW15134272HA SEW15134276HA SEW15134277HA SEW15134278HA SEW15134279HA SEW15134280HA SEW15134281HA SEW15134282HA SEW15134283HA SEW15134284HA SEW15134285HA SEW15134286HA SEW15134287HA SEW15134288HA SEW15134289HA SEW15134290HA SEW15134291HA SEW15134292HA SEW15134293HA SEW15134294HA SEW15134295HA SEW15134296HA SEW15134297HA SEW15134298HA SEW15134299HA SEW15134300HA SEW15134301HA SEW15134302HA SEW15134303HA SEW15134304HA SEW15134305HA SEW15134306HA SEW15134307HA SEW15134308HA SEW15134309HA SEW15134310HA SEW15134311HA SEW15134312HA SEW15134313HA SEW15134314HA SEW15134315HA SEW15134316HA SEW15134317HA SEW15134318HA SEW15134319HA SEW15134320HA SEW15134321HA SK417251104HA SEW13047471HA SEW13099688HA SEW12378574HA SK415280001HA SK415280003HA SK415280004HA SK415280005HA SK415280007HA SEW11365069HA SEW11375254HA SEW11375257HA SEW11405460HA SEW11405461HA SEW11405462HA SEW11405463HA SEW11405464HA SEW11405465HA SEW11405467HA SEW11405472HA SEW11405478HA SEW11405507HA SEW11405509HA SEW11425922HA SEW12090754HA SEW12152084HA SEW12152085HA SEW12152086HA SEW12152093HA SEW12152094HA SEW12152098HA SEW12152099HA SEW12152100HA SEW12152101HA SEW12152102HA SEW12182628HA SEW12182630HA SEW12182702HA SEW12193064HA SEW12193074HA SEW12193076HA SEW12193077HA SEW12193078HA SEW12193079HA SEW12193080HA SEW12193100HA SEW12203354HA SEW12203356HA SEW12203359HA SEW12203360HA SEW12265457HA SEW12275712HA SEW12285848HA SEW12285865HA SEW12285866HA SEW12285867HA SEW12285869HA SEW12285911HA SEW12285915HA SEW12296279HA SEW12296313HA SEW12367718HA SEW12368047HA SEW12368049HA SEW12368051HA SEW12368052HA SEW12368058HA SEW12368060HA SEW12368062HA SEW12368063HA SEW12378573HA SEW12378575HA SEW12378576HA SEW12378577HA SEW12378578HA SEW12378579HA SEW12378580HA SEW12378581HA SEW12378582HA SEW12378585HA SEW12378586HA SEW12378590HA SEW12378591HA SEW12378592HA SEW12378640HA SEW12378643HA SEW12378644HA SEW12378645HA SEW12378646HA SEW12378648HA SEW12378651HA SEW12378652HA SEW12431357HA SEW12472724HA SEW12472725HA SEW12472726HA SEW12482895HA SEW12482946HA SEW12482947HA SEW12483206HA SEW12493488HA SEW12493511HA SEW12493514HA SEW12493595HA SEW12494349HA SEW13026828HA SEW13026829HA SEW13026830HA SEW13026874HA SEW13026880HA SEW13026881HA SEW13026882HA SEW13026916HA SEW13026918HA SEW13026919HA SEW13026920HA SEW13026921HA SEW13026922HA SEW13026923HA SEW13026924HA SEW13026925HA SEW13047425HA SEW13047427HA SEW13047428HA SEW13047431HA SEW13047432HA SEW13047433HA SEW13047434HA SEW13047435HA SEW13047436HA SEW13047437HA SEW13047438HA SEW13047441HA SEW13047443HA SEW13047719HA SEW13057780HA SEW13057786HA SEW13057787HA SEW13057788HA SEW13057790HA SEW13057794HA SEW13057795HA SEW13057796HA SEW13057800HA SEW13057824HA SEW13057827HA SEW13057828HA SEW13057829HA SEW13068514HA SEW13068581HA SEW13068582HA SEW13068587HA SEW13068588HA SEW13068591HA SEW13068607HA SEW13068625HA SEW13068626HA SEW13099402HA SEW13100129HA SEW13100130HA SEW13100132HA SEW13100133HA SEW13100135HA SEW13100136HA SEW13100137HA SEW13100138HA SEW13100140HA SEW13100164HA SEW13100168HA SEW13100171HA SEW13109755HA SEW13109803HA SEW13109804HA SEW13110201HA SEW13110202HA SEW12367703HA SEW12483266HA SEW12483268HA SEW12483270HA SEW12483271HA SEW12483276HA SEW12483277HA SEW12483278HA SEW12483279HA SEW12483280HA SEW12493419HA SEW12493457HA SEW12493458HA SEW12493479HA SEW12493483HA SEW12493487HA SEW12493497HA SEW12493500HA SEW12493510HA SEW12493512HA SEW12493513HA SEW12493515HA SEW12493589HA SEW12493594HA SEW12493596HA SEW12493597HA SEW12493598HA SEW12493611HA SEW12493612HA SEW12493613HA SEW12493614HA SEW12494353HA SEW14461674HA SEW14461677HA SEW14462161HA SEW14462184HA SK417391055HA SK417391147HA SK415501025HA SK415501250HA SK416011022HA SK416021072HA SEW11365014HA SEW11365051HA SEW11365058HA SEW11365065HA SEW13131133HA SEW14073741HA SEW14073742HA SEW13131126HA SEW11364928HA SEW11364973HA SEW11364988HA SEW11364996HA SEW11365005HA SEW11365041HA SEW11365043HA SEW11365045HA SEW11365046HA SEW11365047HA SEW11365048HA SEW11365049HA SEW11365050HA SEW11365052HA SEW11365054HA SEW11365055HA SEW11365056HA SEW11365057HA SEW11365059HA SEW12121539HA SEW12275718HA SEW12285907HA SEW12285908HA SEW12285921HA SEW12285923HA SEW13234418HA SEW12285900HA SEW12234241HA SEW12234484HA SEW14105547HA SEW14106040HA SEW14106041HA SEW14116728HA SEW14116730HA SEW14116731HA SEW14116732HA SEW14116733HA SEW14116740HA SEW14116741HA SEW14116747HA SEW14138452HA SEW14138462HA SEW14346898HA SEW14346899HA SEW14357697HA SEW14357701HA SEW14357704HA SEW14357705HA SEW14357710HA SEW14357712HA SEW14357716HA SEW14357717HA SEW14357734HA SEW14357736HA SEW14357737HA SEW14496309HA SEW15221063HA SEW15221064HA SEW15221065HA SEW15221066HA SEW15221067HA SEW15221069HA SK415261019HA SK415261032HA SK415371001HA SK415371004HA SK415371005HA SK415371007HA SK415371008HA SK415371009HA SK415371010HA SK415371011HA SK415371012HA SK415371013HA SK415371014HA SK415371015HA SK415371017HA SK415371018HA SK415371019HA SK415371022HA SK415371024HA SK416051182HA SK416061003HA SK416061004HA SK416061005HA SK416061007HA SK416061008HA SK416231249HA SK416231284HA SK416241076HA SK416241088HA SK416251176HA SK416251198HA SK416251199HA SK416251200HA SK416251201HA SK416251203HA SK416261034HA SK416261036HA SK417371097HA SK417371098HA SK417371099HA SK417371100HA SK417371101HA SK417371102HA SK417371103HA SK417371104HA SK417371105HA SK
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- GE Healthcare Finland Oy
- Reason for Recall:
- When multiple CARESCAPE Monitor B650 units are connected to the same network and a network overload occurs for a prolonged time, the monitors may simultaneously restart as designed. The monitor restart will not be completed until the network issue has been corrected.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
GE Healthcare CARESCAPE Monitor B650
Product Codes/Lot Numbers:
Serial Numbers: SEW13497373HA SEW13497374HA SEW13497375HA SEW13497376HA SEW13508068HA SEW12494591HA SEW14280294HA SEW14280298HA SEW14289769HA SEW14289777HA SK416341018HA SK416341019HA SK416341026HA SEW12494422HA SEW12494425HA SEW12494435HA SEW12494584HA SEW12494585HA SEW12494592HA SEW12494593HA SEW12494596HA SEW12494598HA SEW12494619HA SEW12494621HA SEW12494622HA SEW12494625HA SEW12494626HA SEW12504918HA SEW12504923HA SEW12504940HA SEW12504942HA SEW12504944HA SEW12504954HA SEW14278996HA SEW14279002HA SEW14279003HA SEW14279005HA SK416061047HA SK416061048HA SK416061088HA SK416081102HA SK416461063HA SK416461092HA SEW13109786HA SEW12059577HA SEW12059583HA SEW12059587HA SEW12059589HA SEW12059592HA SEW12059593HA SEW12059594HA SEW12059596HA SEW12059597HA SEW12059598HA SEW12059599HA SEW12059601HA SEW13275956HA SEW13276028HA SEW13276029HA SEW13275967HA SEW13276066HA SEW13276073HA SEW13286164HA SEW13286175HA SEW13286183HA SK415491246HA SK415491249HA SK415491253HA SK415491257HA SK415491259HA SK415491264HA SK415491265HA SQC17490035HA SQC17490038HA SQC17490039HA SQC17490040HA SQC17490042HA SEW15134264HA SEW15134269HA SEW15134270HA SK415211108HA SK415211110HA SK415211116HA SK415221002HA SK415221006HA SK415361118HA SK415361119HA SK415361120HA SK415361121HA SK415361122HA SK415361123HA SK416041059HA SK416051085HA SK416051119HA SK416041053HA SK416041054HA SK416041062HA SK416041063HA SK416041065HA SK415491254HA SK415521029HA SK415531048HA SK416011088HA SK416011113HA SK416501112HA SK416501119HA SK416501122HA SK417051035HA SK417051036HA SK417051038HA SK417051099HA SK417051100HA SK417051118HA SK417051121HA SK417051122HA SK417051125HA SK417051126HA SK417051130HA SK417051131HA SK417051139HA SK417051144HA SK417051145HA SK417051149HA SK417051150HA SK417051151HA SK417051152HA SK417051153HA SK417051154HA SK417051157HA SK417061002HA SK417061023HA SK417061024HA SK417061025HA SEW13508579HA SEW13369186HA SEW13369234HA SEW13369245HA SEW13369246HA SEW13369249HA SEW13369251HA SEW13369252HA SEW13369254HA SEW13369255HA SEW13369256HA SEW14381680HA SEW15166221HA SEW15166222HA SEW15166223HA SEW15166224HA SEW15166226HA SEW15166230HA SEW15166231HA SEW15166235HA SEW15166236HA SEW15166237HA SEW14063178HA SEW14062864HA SEW14062865HA SEW14062927HA SEW14063174HA SEW14063175HA SEW14063176HA SEW14063179HA SEW14062866HA SEW14235956HA SEW14235957HA SEW14235958HA SEW14235993HA SEW14235994HA SEW14235995HA SEW14246223HA SK416051106HA SK416051113HA SK416051114HA SK416051115HA SEW11202291HA SK416121077HA SK416121083HA SEW12245068HA SEW12245071HA SEW12245072HA SEW12245074HA SEW12245075HA SEW12245076HA SEW12482932HA SQC18150040HA SQC18150042HA SK416321071HA SK416321073HA SK416321075HA SK416321076HA SEW13317664HA SQC17500211HA SQC17500212HA SQC17500213HA SQC17500214HA SQC17500215HA SQC17500216HA SQC17500217HA SQC17500218HA SQC17500219HA SQC17500220HA SQC17500221HA SQC17500222HA SQC17500224HA SQC17500225HA SQC17500226HA SQC17500228HA SQC17500230HA SQC17500234HA SQC17500235HA SEW14063243HA SQC18010001HA SK417081005HA SEW13369260HA SEW13369277HA SEW13379768HA SEW13379778HA SEW13379779HA SEW13379780HA SEW13379782HA SEW13379784HA SEW13379790HA SEW13379791HA SEW14031040HA SEW14031041HA SEW14031042HA SEW14031044HA SEW14031045HA SEW14031046HA SEW14031047HA SEW14031060HA SEW14031061HA SEW14031062HA SEW14031066HA SEW14041261HA SEW14041262HA SEW14041263HA SEW14041264HA SEW14041267HA SEW14041268HA SEW14041269HA SEW14041271HA SEW14041282HA SEW14041283HA SEW14041284HA SEW14041286HA SEW14041287HA SEW14041289HA SEW14041290HA SEW14041291HA SEW14063210HA SEW14105484HA SEW14105968HA SEW14105969HA SEW14116795HA SEW14127301HA SEW14127307HA SEW14127308HA SEW14127312HA SEW14127313HA SEW14127314HA SEW14127315HA SEW14127316HA SEW14127317HA SEW14127318HA SEW14127319HA SEW14127320HA SEW14127321HA SEW14127322HA SEW14127323HA SEW14127324HA SEW14127325HA SEW14127326HA SEW14127327HA SEW14127328HA SEW14127329HA SEW14127330HA SEW14127331HA SEW14127332HA SEW14127333HA SEW14127334HA SEW14127335HA SEW14127631HA SEW14127632HA SEW14127633HA SEW14127634HA SEW14127635HA SEW14127636HA SEW14127637HA SEW14127638HA SEW14127639HA SEW14127863HA SEW14127864HA SEW14127865HA SEW14127866HA SEW14127867HA SEW13379789HA SEW13379793HA SQC17450282HA SQC17450283HA SQC17450284HA SQC17450285HA SQC17450287HA SQC17450289HA SQC17450292HA SQC17450293HA SEW12388960HA SEW12388971HA SEW12389025HA SEW12389026HA SEW12494289HA SEW12494291HA SEW12494292HA SEW12494293HA SEW12494294HA SEW12494296HA SEW12494316HA SEW12494319HA SEW12494320HA SEW12494321HA SEW12494323HA SEW12494324HA SEW12494325HA SEW12494328HA SEW12494559HA SEW12494561HA SEW12494562HA SEW12494568HA SEW12494603HA SEW12494605HA SEW12494606HA SEW12494607HA SEW12494608HA SEW12494610HA SEW12494611HA SEW12494612HA SEW12494613HA SEW12494615HA SEW13265672HA SEW13265673HA SEW13265674HA SEW13275965HA SK416071199HA SK416071202HA SK416071205HA SK416071206HA SK416071207HA SK416071208HA SK416071209HA SK416071210HA SK416071211HA SK416071212HA SK416071213HA SK416071214HA SK416071225HA SK416071227HA SK416071228HA SK416071229HA SK416071230HA SK416071231HA SK416071232HA SK416071233HA SEW11243341HA SEW11243343HA SEW11243344HA SEW11243345HA SEW11243350HA SEW11446229HA SEW11446230HA SEW11446231HA SEW11446236HA SEW14127659HA SEW14127663HA SEW14127671HA SEW14127672HA SEW14127673HA SEW15155904HA SEW15166217HA SEW15166225HA SEW15166227HA SK415361034HA SEW12245031HA SK417011071HA SK417011072HA SK417011073HA SK417011077HA SK417011079HA SQC18010097HA SQC18010098HA SQC18010109HA SEW14496326HA SEW14496327HA SEW14496333HA SEW14496334HA SEW14496353HA SK417111097HA SK417111092HA SK417111098HA SK417111100HA SK417111101HA SK417111103HA SK417111104HA SK417111105HA SK417111106HA SK417111110HA SK417111112HA SK417111113HA SEW12131766HA SEW12244579HA SEW12244580HA SEW12357582HA SEW14459881HA SEW15054763HA SEW15054764HA SEW15054765HA SEW15054805HA SK416421111HA SK416421119HA SK416421120HA SK417021019HA SK417021025HA SK417021027HA SQC17520016HA SQC17520017HA SEW12399425HA SEW12399426HA SEW12399427HA SEW12399428HA SEW12399429HA SEW12399430HA SEW12399432HA SEW12399434HA SEW12399435HA SEW12399436HA SEW12399437HA SK415311122HA SK415311131HA SK415311132HA SK416371226HA SK416371230HA SEW14029842HR SQC18110196HA SQC18110198HA SQC18120082HA SQC18120083HA SQC18120084HA SEW11202299HA SK415361010HA SK415361124HA SK415361126HA SK415361127HA SK415361128HA SK415361129HA SK415361130HA SK415361197HA SK415361203HA SK415361206HA SK415361209HA SK415361213HA SK415361216HA SK415361220HA SK415371002HA SK415371006HA SEW13475994HA SEW13476000HA SEW13476001HA SEW13475998HA SEW13476003HA SK416071011HA SK416071020HA SK416071021HA SK416071022HA SK416071023HA SK416071024HA SK416071025HA SK416071026HA SK416071027HA SK416071029HA SEW11243443HA SEW11243450HA SEW11243451HA SEW11243452HA SEW11243456HA SEW11243457HA SEW12059455HA SEW12059456HA SEW12059461HA SEW12059462HA SK415390013HA SK416151168HA SK416151169HA SK416151170HA SK417341201HA SK417021061HA SEW13286262HA SEW13296800HA SEW13296801HA SEW13296803HA SEW13296804HA SEW13296805HA SEW13296806HA SEW13296821HA SEW13296826HA SEW13296827HA SEW13296828HA SEW13296831HA SEW13296843HA SEW13296844HA SEW13296848HA SEW13296849HA SEW13296850HA SK416461126HA SK416461133HA SK416461134HA SK416461136HA SK416461137HA SK416461138HA SK416461139HA SK416461140HA SK416461141HA SK416461142HA SK416461144HA SK416461149HA SEW12018076HA SEW12018094HA SEW13203376HA SQC18120171HA SK417261197HA SK417281074HA SK417281106HA SK417291202HA SK417301011HA SK417301029HA SK417301030HA SK417301031HA SK417301038HA SK417301227HA SK417301230HA SK417301231HA SK417301232HA SK417301233HA SK417301234HA SK417301238HA SK417301239HA SK417311017HA SK417311018HA SK417311019HA SK417311020HA SK417311024HA SK417311025HA SK417311036HA SK417311037HA SK417311038HA SK417311039HA SK417311040HA SK417311041HA SK417311042HA SK417311043HA SK417311044HA SK417311045HA SK417311046HA SK417311053HA SK417311055HA SK417311056HA SK417311057HA SK417311058HA SK417311059HA SK417311060HA SK417311061HA SK417311062HA SK417311063HA SK417311064HA SK417311065HA SK417311066HA SK417311067HA SK417311089HA SK417311090HA SK417311113HA SK417311115HA SK417311116HA SK417311118HA SK417311138HA SK417311139HA SK417311140HA SK417311141HA SK417311142HA SK417311143HA SK417311144HA SK417311145HA SK417311146HA SK417311147HA SK417311148HA SK417311149HA SK417311150HA SK417311151HA SK417311152HA SK417311153HA SK417311154HA SK417311155HA SK417311156HA SK417311157HA SK417311158HA SK417311159HA SK417311161HA SK417311198HA SK417321070HA SK417321114HA SK417321115HA SK417321116HA SK417321117HA SK417321119HA SK417321120HA SK417321121HA SK417321125HA SK417321147HA SK417331003HA SK417331004HA SK417331005HA SK417331008HA SK417331028HA SK417331041HA SK417331042HA SK417331043HA SK417331044HA SK417331157HA SQC17430011HA SEW11202775HA SEW11202778HA SEW11223011HA SEW11223022HA SEW11223123HA SEW11263888HA SEW12234397HA SEW14302254HA SEW14302255HA SEW14302257HA SEW14302263HA SEW14347390HA SEW12028675HA SEW12028676HA SEW12028677HA SEW12028678HA SEW12028680HA SEW12028681HA SEW12028682HA SEW12182716HA SEW12245080HA SEW12245082HA SQC17420031HA SQC17420053HA SQC18100142HA SQC18100143HA SQC18100144HA SQC18100153HA SQC18100154HA SQC18040085HA SEW14473939HA SEW14484145HA SK417111024HA SK417111026HA SK417111027HA SK417111028HA SEW15101852HA SK417341202HA SK417341205HA SK417341206HA SK417341207HA SK417341210HA SK417341212HA SK417341213HA SK417341214HA SK417341229HA SK417381247HA SK417381250HA SQC17420215HA SQC17430002HA SQC17430003HA SQC17430012HA SEW10354778HA SEW12285885HA SK417061148HA SEW13131087HA SEW13131088HA SEW14280584HA SEW14280598HA SEW15271035HA SEW15271038HA SEW15381015HA SEW15381017HA SEW11079419HA SEW11079425HA SEW11079436HA SEW11079441HA SEW11079465HA SEW11079479HA SEW11079496HA SEW11059204HA SEW11079373HA SEW11079378HA SEW11079463HA SEW11079466HA SEW11079469HA SEW11079493HA SEW11079507HA SEW11079509HA SEW11079511HA SEW11079512HA SEW11079534HA SEW13172414HA SEW13182787HA SEW13519440HA SEW13519441HA SEW13519443HA SEW13519444HA SEW14160610HA SEW14160613HA SEW14170856HA SEW14170868HA SEW14171070HA SEW14171071HA SK417081003HA SK417081130HA SK417101089HA SK417121090HA SK417121095HA SK417121096HA SK417121097HA SK417121098HA SK417121099HA SK417121100HA SK417121101HA SK417121102HA SK417121103HA SK417121104HA SK417121105HA SK417121106HA SK417121107HA SK417121108HA SK417121110HA SK417121111HA SK417221043HA SK416331162HA SK416331185HA SK416331186HA SK416341048HA SK416341060HA SK416341065HA SK416341066HA SK416341067HA SK416341068HA SK416341070HA SK416341072HA SK416341073HA SK416341074HA SK416341077HA SK416341078HA SK416341079HA SK416341080HA SK416341081HA SK416341082HA SK416341083HA SK416341084HA SK416341085HA SK416341086HA SK416341087HA SK416341088HA SK416341089HA SK416341090HA SK416341091HA SK416341092HA SK416341093HA SK416341094HA SK416341095HA SK416341096HA SK416341097HA SK416341098HA SK416341099HA SK416341100HA SK416341116HA SEW14392029HA SEW14392037HA SEW14392038HA SEW14392042HA SEW14392044HA SK416371112HA SQC17400053HA SQC17400054HA SQC17400055HA SQC17400056HA SQC17400057HA SQC17400058HA SQC17440152HA SQC17440180HA SQC17440182HA SQC17440198HA SQC17440199HA SQC17450102HA SEW13454551HR SEW15065986HA SEW15065987HA SK415390009HR SK415390012HR SK415390014HR SEW13411759HA SEW13411760HA SEW13411761HA SEW13411762HA SEW13411764HA SEW13411765HA SEW13411766HA SEW13411767HA SEW13411925HA SEW13411928HA SEW13432945HA SEW13432947HA SEW13432948HA SEW13432952HA SEW13432953HA SEW13432954HA SEW13432955HA SEW13432958HA SEW13432959HA SEW13432962HA SEW13464760HA SEW13464764HA SEW13464765HA SEW13464768HA SEW13464769HA SEW13464770HA SEW13464772HA SEW13508138HA SEW13518959HA SEW13518960HA SEW13518962HA SEW14246763HA SEW14246773HA SEW14246774HA SEW14246775HA SEW14247266HA SEW14247287HA SEW14224894HA SK415511077HA SK415511090HA SK415511092HA SK415511094HA SK415511095HA SK415511096HA SK415511098HA SK415511100HA SK415511101HA SK415511102HA SK415511103HA SK415511104HA SK415511123HA SK415511124HA SK415511126HA SK415511127HA SK415511128HA SK415511130HA SK416191169HA SK415511165HA SK415511177HA SK416301151HA SK416311002HA SK416311003HA SK416311004HA SK416311005HA SK416311009HA SK416311050HA SK416311052HA SK416311053HA SK416311054HA SK416311055HA SK416311056HA SK416311057HA SK416311063HA SK416311064HA SK416311065HA SK416311066HA SK416341055HA SK416341056HA SK416341057HA SK416341059HA SK416341062HA SK416341063HA SK416341064HA SK416221117HA SK416221118HA SK416221119HA SK416221120HA SK416221121HA SK416221122HA SK416221123HA SK416221200HA SK416221201HA SK416221205HA SK416221206HA SK416221207HA SK416221209HA SK416231032HA SK416341058HA SK416401057HA SK416401058HA SK416401059HA SK416401062HA SK416401063HA SK416401064HA SK416401066HA SK416401068HA SK416401071HA SK417211179HA SK417221231HA SK417221232HA SK417221234HA SK417221235HA SK417221236HA SK416231214HA SEW12142057HA SEW12152088HA SEW12193058HA SEW12203363HA SEW12203364HA SEW12203365HA SEW12203366HA SEW12203367HA SEW12203368HA SEW12265428HA SEW12265429HA SEW12265430HA SEW12265431HA SEW12265432HA SEW12265592HA SEW12265593HA SEW12265596HA SEW12265597HA SEW12265598HA SEW12265599HA SEW12265600HA SEW12275631HA SEW12275632HA SEW12275633HA SEW12275656HA SEW12275678HA SEW12275679HA SEW12275685HA SEW12275686HA SEW12275687HA SEW12275688HA SEW12275689HA SEW12275690HA SEW12275691HA SEW12275692HA SEW12275694HA SEW12275695HA SEW12275696HA SEW12275698HA SEW12275699HA SEW12275704HA SEW12285973HA SEW12357257HA SEW12368068HA SEW13286647HA SEW13296671HA SEW13296674HA SEW13296681HA SEW13296682HA SEW13296683HA SEW13296685HA SEW13296687HA SEW13296688HA SEW13296689HA SEW13296690HA SEW13296692HA SEW13296697HA SEW13296699HA SEW13296701HA SEW13296702HA SEW13296706HA SEW13296708HA SEW13296710HA SEW13296712HA SK417071135HA SK417071137HA SK417071138HA SK417071140HA SK417231322HA SK417231323HA SK417231324HA SK417231326HA SK417231327HA SK417231330HA SK417231331HA SK417241001HA SK417381100HA SK417381101HA SK417381102HA SK417381103HA SEW12131856HA SEW12131859HA SEW12131860HA SEW12131877HA SEW12131879HA SEW12131886HA SEW12131887HA SEW12142061HA SEW12142062HA SEW12142063HA SEW12244544HA SEW12244545HA SEW12244547HA SEW12244552HA SEW12244554HA SEW12244556HA SEW12244717HA SEW12244718HA SK416431024HA SK416431026HA SK416431027HA SK416431037HA SK417211175HA SK417231206HA SK417231210HA SK417231222HA SK417241042HA SK417241043HA SK417241044HA SK417241045HA SK417241046HA SK417241047HA SK417241048HA SK417241050HA SK417251051HA SK417251052HA SK417251053HA SK417251054HA SK417251056HA SK417251058HA SK417251060HA SK416031042HA SK416031053HA SK416031054HA SEW12131778HA SEW12131779HA SEW12131791HA SEW12131794HA SEW12131797HA SK415260001HA SK416041098HA SEW12296316HA SEW12296317HA SEW12296318HA SEW12306478HA SEW12306479HA SEW12306484HA SEW11477028HA SEW11517992HA SEW11517997HA SK416231013HA SK416231017HA SK416231023HA SK416231025HA SK416231026HA SK416231027HA SK416231028HA SK416231029HA SEW11202381HA SEW14019781HA SK416181031HA SK416181032HA SK416181033HA SK416181034HA SK416181036HA SK416181037HA SK416181038HA SK416181040HA SK416181041HA SK416181042HA SK416181043HA SK416181044HA SK416181045HA SK416181046HA SK416181050HA SK416181051HA SK417141069HA SK417141070HA SEW12494263HA SEW12494269HA SEW12494271HA SEW12494272HA SEW12494275HA SEW12494587HA SEW12494628HA SEW11059217HA SEW11202769HA SEW12245129HA SEW15190043HA SEW15190046HA SEW15190047HA SEW15190048HA SEW15190049HA SEW15200001HA SEW15200002HA SEW15200003HA SEW15200004HA SEW15200005HA SEW15200006HA SEW15200007HA SEW15200008HA SEW15200009HA SEW15200010HA SEW15220001HA SEW15220002HA SEW15220004HA SEW15220005HA SEW15220006HA SEW15220007HA SEW15220008HA SEW15220009HA SEW15220010HA SEW15220011HA SEW15220012HA SEW15220018HA SEW15220020HA SEW15241048HA SEW15241068HA SEW15241097HA SEW15241098HA SEW15241099HA SEW15241100HA SEW15241101HA SEW15241102HA SEW15241103HA SEW15241122HA SEW15241123HA SEW15241124HA SEW15241125HA SEW15241126HA SEW15251001HA SEW15251002HA SEW15251003HA SEW15251004HA SEW15251005HA SEW15251006HA SEW15251007HA SEW15251008HA SEW15251009HA SEW15251011HA SEW15251012HA SEW15251013HA SEW15251014HA SEW15251016HA SEW15251017HA SEW15251022HA SK415141076HA SEW14381659HA SEW14381661HA SEW14381665HA SEW14381666HA SEW14381668HA SEW14381669HA SEW14381670HA SEW14381673HA SEW14381675HA SEW14381676HA SEW14381677HA SEW14381682HA SEW14381683HA SEW14415291HA SEW14415292HA SEW14415298HA SEW14415301HA SEW14415302HA SEW14426253HA SEW14426257HA SEW14426263HA SEW14437098HA SEW14437118HA SEW14437120HA SEW14437125HA SEW14437126HA SEW14437127HA SEW14437129HA SEW14437139HA SEW14437140HA SEW14437141HA SEW14437142HA SEW14437143HA SEW14437149HA SEW14437156HA SEW14437157HA SEW14437158HA SEW14437159HA SEW14148891HA SEW14148892HA SEW14148899HA SK417311199HA SK417311200HA SK417311201HA SK417311204HA SK417311205HA SK417321002HA SK417321004HA SK417321013HA SK415511164HR SK415511167HR SK416021098HR SK416061019HR SQC17470152HA SQC17470153HA SQC17470154HA SQC17470155HA SQC17470161HA SQC17470162HA SQC17470163HA SQC17470164HA SEW13475573HA SEW13475574HA SEW13475576HA SEW14160609HA SEW14170855HA SEW14170864HA SEW14508174HA SEW11263783HA SEW13275963HA SEW13275977HA SEW12378484HA SEW12378485HA SEW12378487HA SEW12378488HA SEW12378489HA SEW12378490HA SEW12378491HA SEW11405568HA SEW11477093HA SEW11487240HA SEW11487251HA SEW11487279HA SEW11487322HA SEW11497477HA SEW12049008HA SEW12244833HA SEW12245017HA SEW12245018HA SEW12245019HA SEW12245029HA SEW12245030HA SEW13464740HA SEW13464742HA SEW13464743HA SK417231204HA SK417231290HA SK417241084HA SK417241086HA SK417241104HA SK417241105HA SK417241106HA SK417241107HA SK417241108HA SK417241109HA SK417241110HA SK417241111HA SK417241112HA SK417241113HA SK417241114HA SK417241115HA SK417241116HA SK417241117HA SK417241118HA SK417241119HA SK417241120HA SK417241121HA SK417241122HA SK417241123HA SK417241124HA SEW14313345HA SEW14313356HA SEW15055425HA SEW15065967HA SEW15065984HA SEW12505570HA SEW13026793HA SEW13026844HA SEW13026861HA SEW13026862HA SEW13026863HA SEW13026865HA SEW13026870HA SEW13026875HA SEW13026936HA SEW13026939HA SEW13026954HA SEW13026955HA SEW12409606HA SEW12483210HA SEW12483214HA SEW13234375HA SEW13234439HA SEW13234440HA SEW13234441HA SEW13234442HA SEW13234444HA SEW13234450HA SK417271092HA SK417271095HA SQC17510134HA SEW12368046HA SEW12389045HA SEW13475604HA SK416461035HA SK417021013HA SK417021022HA SEW11202317HA SEW13519436HA SEW14160612HA SEW14160632HA SEW14170854HA SK415341074HA SK415341075HA SK415341076HA SK415341082HA SK415341091HA SK415341092HA SK415341093HA SK415341108HA SK415341111HA SK415341113HA SK416051018HA SK417291054HA SK417291056HA SK417291082HA SK417291083HA SK417291090HA SK417291091HA SK417291092HA SK417291093HA SK417291097HA SK417291098HA SK417291099HA SK417291100HA SK417291102HA SK417291135HA SK417291136HA SK417291137HA SQC18010091HA SQC18010123HA SQC18020097HA SQC18030119HA SQC18040023HA SQC18040112HA SQC18060132HA SQC18070048HA SQC18070049HA SQC18070055HA SQC18070056HA SQC18070057HA SQC18070058HA SQC18070059HA SQC18070060HA SQC18070061HA SQC18070062HA SQC18070063HA SQC18070064HA SQC18070065HA SQC18070066HA SQC18070077HA SQC18070078HA SQC18070079HA SQC18070080HA SQC18070081HA SQC18070082HA SQC18070084HA SQC18070085HA SQC18070086HA SK416251008HA SK416251011HA SK416511038HA SK416511148HA SK416511149HA SK417031013HA SK416221096HA SK416221097HA SK416221100HA SK416221105HA SK416221173HA SK416221175HA SK416221177HA SK416221192HA SK416221193HA SK416221194HA SK416251004HA SK416251006HA SK416251007HA SK416251012HA SK416251013HA SK416251017HA SK416251018HA SK416251020HA SEW15020439HA SEW15020440HA SEW15020441HA SEW15020442HA SEW15020443HA SEW15020444HA SEW15020445HA SEW15020447HA SEW15020448HA SEW15020449HA SEW14268779HA SEW14268781HA SEW14278924HA SEW14278929HA SEW14278932HA SEW14278936HA SEW14278937HA SEW14278938HA SEW14278942HA SEW14278957HA SEW15020457HA SEW15020458HA SEW15020459HA SEW15020460HA SEW15020462HA SEW15020463HA SEW15020464HA SEW15020465HA SEW15020466HA SEW15020467HA SEW15020469HA SEW15020470HA SEW15020474HA SEW15020478HA SEW15020479HA SEW15020480HA SEW15020481HA SEW15020491HA SEW15020496HA SEW15124003HA SEW12121391HA SEW12121431HA SEW12121432HA SEW12121434HA SEW12121435HA SEW12131743HA SEW12141994HA SEW12142006HA SEW12142010HA SEW12142020HA SEW12142021HA SEW12142022HA SEW12142023HA SEW12049295HA SEW12049297HA SEW14448191HA SEW14448193HA SEW14448196HA SEW15066817HA SEW11436062HA SEW11436070HA SEW11436073HA SEW11436099HA SEW12245028HA SEW14127293HA SK417241067HA SK417251024HA SK417251028HA SK417251029HA SK417251030HA SK417251031HA SK417251033HA SK417251034HA SK417251035HA SK417251036HA SK417251037HA SK417251039HA SK417251041HA SK417251042HA SEW14127870HA SEW14127871HA SEW14127878HA SEW14127879HA SEW14127880HA SEW14127881HA SEW14127882HA SEW14127883HA SEW14127884HA SEW14127885HA SEW14127886HA SEW14127888HA SEW14127889HA SEW14127890HA SEW14127891HA SEW14127892HA SEW14127893HA SEW14127894HA SEW14127895HA SEW14127896HA SEW14127897HA SEW14127898HA SEW14127899HA SEW14127900HA SEW14127907HA SQC18090129HA SQC18090132HA SQC18090133HA SQC18090134HA SQC18090135HA SQC18090136HA SK416221111HA SK416221112HA SEW13327772HA SEW13327802HA SEW13327803HA SEW13327804HA SEW14508178HA SEW12505585HA SEW12505620HA SEW12505772HA SEW12505783HA SEW12505786HA SEW12505789HA SEW12505800HA SEW12505804HA SEW12505805HA SEW12505806HA SEW12505827HA SEW12515638HA SEW12515639HA SK415411123HA SK415411124HA SK415491336HA SK415501020HA SK415501022HA SK415501023HA SK417271132HA SK416211122HA SK416211128HA SK416211174HA SK416221071HA SK416221142HA SK416221143HA SK416221145HA SK416221185HA SK416221186HA SEW14313327HA SEW14313653HA SEW14313663HA SEW14313664HA SEW14313679HA SEW14313680HA SEW14313681HA SEW14313682HA SQC17420061HA SQC17420067HA SQC17420069HA SQC17420075HA SQC18030184HA SQC18040075HA SEW11079353HA SEW11079394HA SEW11079396HA SEW11079526HA SEW11079533HA SEW11109962HA SEW11109977HA SEW11109978HA SEW11202309HA SEW11202322HA SEW14448266HA SK416011112HA SK416011114HA SEW14073459HA SEW14073460HA SEW14149136HA SK415521036HA SK415521039HA SK415521040HA SK415521042HA SK415521043HA SK415521045HA SK415521046HA SK415521048HA SK415521050HA SK417141062HA SK417141063HA SK417141064HA SK417141065HA SK415391019HA SK416051081HA SK416051082HA SK417291094HA SK417291095HA SK417291096HA SK415401027HA SK416011035HA SEW13026364HA SEW13026368HA SEW13026369HA SEW13026377HA SK416261088HA SK416261089HA SK416261090HA SK416261091HA SK416261093HA SK416261095HA SK416261096HA SK416261098HA SK416261099HA SK416261100HA SK416261101HA SK416261104HA SK416261105HA SK416261115HA SK416271171HA SK416271174HA SK416281009HA SK416281017HA SK417381233HA SK416341022HA SEW11079370HA SEW12142002HA SEW13255296HA SEW13327793HA SK415421058HA SK415421063HA SEW14392092HA SK417251040HA SK417251149HA SK417251150HA SK417251151HA SK417251152HA SK417251153HA SK417251154HA SK417251156HA SK417261102HA SK417271146HA SK417271147HA SK417271148HA SK417271153HA SK417291244HA SK417291273HA SK417301006HA SK417361071HA SK417361072HA SK417361073HA SK417121003HA SK417121004HA SK417121005HA SK417121006HA SK417121030HA SK417121031HA SK417121032HA SK417121058HA SK417131011HA SK417131012HA SK417131013HA SK417131014HA SK417131015HA SK417131016HA SK417131017HA SK417131018HA SK417131020HA SK417131021HA SK417131023HA SK417131024HA SK417131025HA SK417131026HA SK417131027HA SK417131028HA SK417131029HA SK417131030HA SK417131031HA SK417131032HA SK417131033HA SK417131034HA SK417131035HA SK417131037HA SK417131038HA SK417131039HA SK417131040HA SK417131041HA SK417131042HA SK417131043HA SK417131044HA SK417131045HA SK417131046HA SK417131047HA SK417131048HA SK417131006HA SK417141072HA SK417141108HA SK417141109HA SK417141110HA SK417141111HA SK417141112HA SK417141113HA SK417141114HA SK417141115HA SK417141116HA SK417141117HA SK417141118HA SK417141119HA SK417141120HA SK417141121HA SK417141122HA SEW13234433HA SEW13244743HA SEW13244760HA SEW13244770HA SEW13244973HA SEW13244974HA SEW13244975HA SEW13244982HA SEW13244983HA SEW13244987HA SEW13244988HA SEW13244989HA SEW13244990HA SEW13244992HA SEW13244993HA SEW13244994HA SEW13244995HA SEW13245019HA SEW13245020HA SEW13245021HA SEW14279006HA SEW14279014HA SEW14279016HA SEW14279017HA SEW14279019HA SEW14279448HA SEW14279451HA SEW14279452HA SEW14279453HA SEW14279454HA SEW14279455HA SEW14279456HA SEW14279457HA SEW14279458HA SEW14279459HA SEW14279460HA SEW14279465HA SEW14291041HA SEW14325257HA SEW14325430HA SEW14325431HA SEW14325432HA SEW14325452HA SEW14325569HA SEW14325579HA SEW14325580HA SEW14325581HA SEW14325582HA SEW14325589HA SK415241033HA SK415251001HA SK415261011HA SK415261012HA SK415261013HA SK415261014HA SK415261021HA SK415261024HA SK415281018HA SK416101070HA SK417201063HA SK417201064HA SK417201065HA SEW11181943HA SEW14127710HA SEW13348789HA SEW13348816HA SEW13348817HA SEW15090492HA SEW15090494HA SEW15090495HA SEW15090496HA SEW15090497HA SEW15101833HA SEW15101834HA SEW15101857HA SK415511060HA SEW14149122HA SEW14138489HA SEW14149117HA SEW14019784HA SK415491115HA SK415491116HA SK415491119HA SK415491172HA SK415491177HA SK415491178HA SEW15134271HA SEW15134272HA SEW15134276HA SEW15134277HA SEW15134278HA SEW15134279HA SEW15134280HA SEW15134281HA SEW15134282HA SEW15134283HA SEW15134284HA SEW15134285HA SEW15134286HA SEW15134287HA SEW15134288HA SEW15134289HA SEW15134290HA SEW15134291HA SEW15134292HA SEW15134293HA SEW15134294HA SEW15134295HA SEW15134296HA SEW15134297HA SEW15134298HA SEW15134299HA SEW15134300HA SEW15134301HA SEW15134302HA SEW15134303HA SEW15134304HA SEW15134305HA SEW15134306HA SEW15134307HA SEW15134308HA SEW15134309HA SEW15134310HA SEW15134311HA SEW15134312HA SEW15134313HA SEW15134314HA SEW15134315HA SEW15134316HA SEW15134317HA SEW15134318HA SEW15134319HA SEW15134320HA SEW15134321HA SK417251104HA SEW13047471HA SEW13099688HA SEW12378574HA SK415280001HA SK415280003HA SK415280004HA SK415280005HA SK415280007HA SEW11365069HA SEW11375254HA SEW11375257HA SEW11405460HA SEW11405461HA SEW11405462HA SEW11405463HA SEW11405464HA SEW11405465HA SEW11405467HA SEW11405472HA SEW11405478HA SEW11405507HA SEW11405509HA SEW11425922HA SEW12090754HA SEW12152084HA SEW12152085HA SEW12152086HA SEW12152093HA SEW12152094HA SEW12152098HA SEW12152099HA SEW12152100HA SEW12152101HA SEW12152102HA SEW12182628HA SEW12182630HA SEW12182702HA SEW12193064HA SEW12193074HA SEW12193076HA SEW12193077HA SEW12193078HA SEW12193079HA SEW12193080HA SEW12193100HA SEW12203354HA SEW12203356HA SEW12203359HA SEW12203360HA SEW12265457HA SEW12275712HA SEW12285848HA SEW12285865HA SEW12285866HA SEW12285867HA SEW12285869HA SEW12285911HA SEW12285915HA SEW12296279HA SEW12296313HA SEW12367718HA SEW12368047HA SEW12368049HA SEW12368051HA SEW12368052HA SEW12368058HA SEW12368060HA SEW12368062HA SEW12368063HA SEW12378573HA SEW12378575HA SEW12378576HA SEW12378577HA SEW12378578HA SEW12378579HA SEW12378580HA SEW12378581HA SEW12378582HA SEW12378585HA SEW12378586HA SEW12378590HA SEW12378591HA SEW12378592HA SEW12378640HA SEW12378643HA SEW12378644HA SEW12378645HA SEW12378646HA SEW12378648HA SEW12378651HA SEW12378652HA SEW12431357HA SEW12472724HA SEW12472725HA SEW12472726HA SEW12482895HA SEW12482946HA SEW12482947HA SEW12483206HA SEW12493488HA SEW12493511HA SEW12493514HA SEW12493595HA SEW12494349HA SEW13026828HA SEW13026829HA SEW13026830HA SEW13026874HA SEW13026880HA SEW13026881HA SEW13026882HA SEW13026916HA SEW13026918HA SEW13026919HA SEW13026920HA SEW13026921HA SEW13026922HA SEW13026923HA SEW13026924HA SEW13026925HA SEW13047425HA SEW13047427HA SEW13047428HA SEW13047431HA SEW13047432HA SEW13047433HA SEW13047434HA SEW13047435HA SEW13047436HA SEW13047437HA SEW13047438HA SEW13047441HA SEW13047443HA SEW13047719HA SEW13057780HA SEW13057786HA SEW13057787HA SEW13057788HA SEW13057790HA SEW13057794HA SEW13057795HA SEW13057796HA SEW13057800HA SEW13057824HA SEW13057827HA SEW13057828HA SEW13057829HA SEW13068514HA SEW13068581HA SEW13068582HA SEW13068587HA SEW13068588HA SEW13068591HA SEW13068607HA SEW13068625HA SEW13068626HA SEW13099402HA SEW13100129HA SEW13100130HA SEW13100132HA SEW13100133HA SEW13100135HA SEW13100136HA SEW13100137HA SEW13100138HA SEW13100140HA SEW13100164HA SEW13100168HA SEW13100171HA SEW13109755HA SEW13109803HA SEW13109804HA SEW13110201HA SEW13110202HA SEW12367703HA SEW12483266HA SEW12483268HA SEW12483270HA SEW12483271HA SEW12483276HA SEW12483277HA SEW12483278HA SEW12483279HA SEW12483280HA SEW12493419HA SEW12493457HA SEW12493458HA SEW12493479HA SEW12493483HA SEW12493487HA SEW12493497HA SEW12493500HA SEW12493510HA SEW12493512HA SEW12493513HA SEW12493515HA SEW12493589HA SEW12493594HA SEW12493596HA SEW12493597HA SEW12493598HA SEW12493611HA SEW12493612HA SEW12493613HA SEW12493614HA SEW12494353HA SEW14461674HA SEW14461677HA SEW14462161HA SEW14462184HA SK417391055HA SK417391147HA SK415501025HA SK415501250HA SK416011022HA SK416021072HA SEW11365014HA SEW11365051HA SEW11365058HA SEW11365065HA SEW13131133HA SEW14073741HA SEW14073742HA SEW13131126HA SEW11364928HA SEW11364973HA SEW11364988HA SEW11364996HA SEW11365005HA SEW11365041HA SEW11365043HA SEW11365045HA SEW11365046HA SEW11365047HA SEW11365048HA SEW11365049HA SEW11365050HA SEW11365052HA SEW11365054HA SEW11365055HA SEW11365056HA SEW11365057HA SEW11365059HA SEW12121539HA SEW12275718HA SEW12285907HA SEW12285908HA SEW12285921HA SEW12285923HA SEW13234418HA SEW12285900HA SEW12234241HA SEW12234484HA SEW14105547HA SEW14106040HA SEW14106041HA SEW14116728HA SEW14116730HA SEW14116731HA SEW14116732HA SEW14116733HA SEW14116740HA SEW14116741HA SEW14116747HA SEW14138452HA SEW14138462HA SEW14346898HA SEW14346899HA SEW14357697HA SEW14357701HA SEW14357704HA SEW14357705HA SEW14357710HA SEW14357712HA SEW14357716HA SEW14357717HA SEW14357734HA SEW14357736HA SEW14357737HA SEW14496309HA SEW15221063HA SEW15221064HA SEW15221065HA SEW15221066HA SEW15221067HA SEW15221069HA SK415261019HA SK415261032HA SK415371001HA SK415371004HA SK415371005HA SK415371007HA SK415371008HA SK415371009HA SK415371010HA SK415371011HA SK415371012HA SK415371013HA SK415371014HA SK415371015HA SK415371017HA SK415371018HA SK415371019HA SK415371022HA SK415371024HA SK416051182HA SK416061003HA SK416061004HA SK416061005HA SK416061007HA SK416061008HA SK416231249HA SK416231284HA SK416241076HA SK416241088HA SK416251176HA SK416251198HA SK416251199HA SK416251200HA SK416251201HA SK416251203HA SK416261034HA SK416261036HA SK417371097HA SK417371098HA SK417371099HA SK417371100HA SK417371101HA SK417371102HA SK417371103HA SK417371104HA SK417371105HA SK
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2340-2018
Related Recalls
Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment.
CARESCAPE ONE, Model no. 2087075-300 - shipped with potentially affected batteries or affected batteries installed as a spare part Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms
GE Healthcare Finland Oy
Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment.
Portrait VSM Model nos. 6660000-666 Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms
GE Healthcare Finland Oy
Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment.