Ratcheting Cannulated Driver Handle, Catalog #: 210-00-004, Rx Only, Non Sterile
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot: 603600
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Trilliant Surgical Ltd.
- Reason for Recall:
- Firm is voluntarily recalling cannulated driver handles (210-00-003 and 210-00-004) that may demonstrate a degradation of function with the adapter collar sticking, and/or not retaining the mating adapter male shaft.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Ratcheting Cannulated Driver Handle, Catalog #: 210-00-004, Rx Only, Non Sterile
Product Codes/Lot Numbers:
Lot: 603600
Distribution:
Nationwide distribution
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2339-2018
Related Recalls
3.5mm Aresenal Screw Drill Bit , Ref: 330-35-001, Qty. 70, Rx Only, Non-Sterile, UDI: (01)00812926029696 - Product Usage: A drill bit facilitates implant insertion
Trilliant Surgical
Labeling Error; specifically, drill bits from lots # TSL008495A are marked with the lot # of TSL008495. This could lead to parts being misidentified and impact device traceability.
8mm HTR Sterile Hammer Toe Reaming Kit, Material: Stainless Steel, Sterile: R, PX only, UDI: (01)00812926022345(17)231225 (10)TSL007202
Trilliant Surgical
The firm has identified that parts from the kit, lot: TSL007202, have the potential to contain the incorrect contents of the 8mm HTR Sterile Hammer Toe Reaming Kit. The kit should contain (2) reamers total, (1) Concave Reamer and (1) Convex Reamer; there is the potential that kits from the identified lot may have (2) of the same reamer type as opposed to one (1) of each type.
Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.