Brand Name: SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 12 TESTS Product Name: SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 12 TESTS Model/Catalog Number: 1023770 Software Version: N/A Product Description: The SporeTraq Mail-In Spore Testing Service includes 12 bacterial Spore Strips provided by Mesa Labs. The Spore Strip (single species G. stearothermophilus ATCC#7953 or dual species G. stearothermophilus ATCC #7953 and B. atrophaeus ATCC #9372) is intended for use in testing the efficacy of sterilizer systems used in healthcare facilities. The Spore Strips are compatible with Steam sterilization, Chemiclave¿ sterilization, Ethylene Oxide sterilization, or Dry Heat sterilization processes. The test strip is placed in the customer s sterilizer during their sterilization cycle and upon completion the sterilized test strip along with the control strip is returned to Mesa Labs for lab testing in the return envelope. Spore test results are digitally stored in HealthFirst s OnTraq smart automation tracking and reporting system for easy and secure access to a paperless archive from any device with internet access. Component: N/A
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot Code: Please see attached spreadsheet listing products that were not incubated per the following requirements : a. Spore strips used in steam sterilizers are incubated for 24 hours b. Spore strips used in chemiclave sterilizers are incubated for 72 hours c. Spore strips used in dry heat or ethylene oxide sterilizers are incubated for 7 days As noted in the previous recall scope question, HealthFirst's Field Correction scope will be expanded to include notifications to all customers in possession of 1023780 SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 52 TESTS,1023770 SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 12 TESTS or other products serviced by Mesa Labs.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- HF Acquisition Co LLC
- Reason for Recall:
- Due to incorrect incubation process being performed on product.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Brand Name: SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 12 TESTS Product Name: SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 12 TESTS Model/Catalog Number: 1023770 Software Version: N/A Product Description: The SporeTraq Mail-In Spore Testing Service includes 12 bacterial Spore Strips provided by Mesa Labs. The Spore Strip (single species G. stearothermophilus ATCC#7953 or dual species G. stearothermophilus ATCC #7953 and B. atrophaeus ATCC #9372) is intended for use in testing the efficacy of sterilizer systems used in healthcare facilities. The Spore Strips are compatible with Steam sterilization, Chemiclave¿ sterilization, Ethylene Oxide sterilization, or Dry Heat sterilization processes. The test strip is placed in the customer s sterilizer during their sterilization cycle and upon completion the sterilized test strip along with the control strip is returned to Mesa Labs for lab testing in the return envelope. Spore test results are digitally stored in HealthFirst s OnTraq smart automation tracking and reporting system for easy and secure access to a paperless archive from any device with internet access. Component: N/A
Product Codes/Lot Numbers:
Lot Code: Please see attached spreadsheet listing products that were not incubated per the following requirements : a. Spore strips used in steam sterilizers are incubated for 24 hours b. Spore strips used in chemiclave sterilizers are incubated for 72 hours c. Spore strips used in dry heat or ethylene oxide sterilizers are incubated for 7 days As noted in the previous recall scope question, HealthFirst's Field Correction scope will be expanded to include notifications to all customers in possession of 1023780 SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 52 TESTS,1023770 SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 12 TESTS or other products serviced by Mesa Labs.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2332-2026
Related Recalls
Due to incorrect incubation process being performed on product.
Biological, Bio52 HealthFirst Mail In Monitoring System, 52 Tests, Item: 1003600. Biological, Bio12 HealthFirst Mail In Monitoring System, 12 Tests, Item: 1003580
HF Acquisition Co
Spore test strip monitoring system, designed for routine monitoring of sterilization processes, may have a printing error on the return envelope that may cause spore test results to be reported to the incorrect customer's account.
GLYCOPYRROLATE INJECTION, USP 0.2MG/ML, 1mL VIAL, manufactured by HF Acquisition Co. LLC, Mukilteo, WA 98275, NDC 51662-1487-3
HF Acquisition Co
Labeling: Label Mix-up