Spirit Plus Bed, A-C Powered Hospital Bed
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model Numbers: 5600, Spirit Plus Bed (manufactured between November 1, 2014 January 13, 2016); Serial Numbers: Serial Number 029481 029482 029483 029484 029485 029486 029487 029488 029489 029490 029491 029492 029493 029494 029495 029496 029497 029498 029499 029500 029501 029502 029503 029504 029505 029506 029507 029508 029509 029510 029511 029512 029513 029514 029515 029516 029517 029518 029519 029520 029521 029522 029523 029524 029525 029526 029527 029528 029529 029530 029531 029532 029543 029544 029545 029546 029547 029548 029549 029550 029551 029552 029553 029554 029555 029556 029557 029565 029566 029567 029568 029569 029570 029571 029572 029573 029574 029575 029576 029577 029578 029579 029580 029581 029582 029583 029584 029585 029586 029587 029588 029589 029590 029591 029592 029593 029594 029595 029596 029597 029598 029599 029600 029601 029602 029603 029604 029605 029606 029607 029608 029609 029610 029611 029612 029613 029614 029615 029616 029617 029618 029619 029620 029621 029622 029623 029624 029625 029626 029627 029628 029629 029630 029631 029632 029633 029634 029635 029636 029637 029638 029660 029661 029662 029663 029664 029665 029666 029667 029668 029669 029670 029671 029672 029673 029674 029675 029676 029677 029678 029679 029680 029681 029682 029683 029684 029685 029686 029687 029688 029689 029690 029691 029692 029693 029694 029695 029696 029697 029698 029699 029700 029701 029702 029703 029704 029705 029706 029709 029710 029711 029712 029713 029714 029715 029716 029717 029718 029719 029720 029721 029722 029723 029724 029725 029726 029727 029728 029729 029730 029731 029732 029733 029734 029735 029736 029737 029738 029739 029740 029741 029742 029743 029744 029745 029746 029747 029748 029749 029750 029751 029752 029753 029754 029755 029756 029757 029758 029759 029760 029812 029813 029814 029815 029816 029817 029818 029819 029820 029821 029822 029823 029824 029825 029826 029827 029828 029829 029830 029831 029832 029833 029834 029835 029836 029837 029838 029839 029840 029841 029842 029843 029844 029845 029846 029847 029848 029849 029850 029851 029852 029853 029854 029855 029856 029857 029858 029859 029860 029861 029862 029863 029864 029865 029866 029867 029868 029869 029870 029871 029872 029873 029874 029875 029876 029877 029879 029880 029881 029882 029883 029884 029885 029886 029887 029888 029889 029890 029891 029892 029893 029894 029895 029896 029897 029898 029899 029900 029901 029916 029917 029918 029919 029920 029921 029922 029923 029924 029925 029926 029927 029928 029930 029931 029932 029933 029934 029935 029936 029937 029938 029939 029940 029941 029942 029943 029944 029945 029946 029947 029948 029949 029951 029952 029953 029954 029955 029956 029957 029958 029959 029960 029961 029962 029963 029964 029965 029966 029967 029968 029969 029970 029971 029972 029973 029974 029975 029976 029977 029978 029979 029980 029981 029982 029983 029984 029985 029986 029987 029988 029989 029990 029991 029992 029993 029994 029995 029996 029997 029998 029999 030000 030001 030002 030003 030004 030005 030006 030007 030008 030009 030010 030011 030012 030013 030014 030015 030016 030017 030018 030019 030020 030021 030022 030023 030024 030025 030026 030027 030028 030029 030030 030031 030032 030033 030034 030035 030103 030104 030105 030106 030107 030108 030109 030110 030111 030112 030113 030114 030115 030116 030117 030118 030119 030120 030121 030122 030123 030124 030125 030126 030127 030128 030129 030130 030131 030132 030133 030134 030135 030136 030137 030138 030139 030140 030141 030142 030143 030144 030145 030146 030147 030148 030162 030163 030164 030165 030166 030167 030168 030169 030170 030171 030172 030173 030174 030175 030176 030177 030178 030179 030180 030181 030182 030183 030184 030185 030186 030187 030188 030189 030190 030191 030192 030193 030194 030195 030196 030197 030198 030199 030200 030201 030202 030203 030222 030233 030234 030235 030239 030240 030250 030286 030287 030288 030289 030290 030291 030292 030293 030294 030295 030296 030297 030298 030299 030300 030301 030302 030303 030304 030305 030306 030307 030308 030309 030310 030311 030312 030313 030314 030315 030316 030317 030318 030319 030320 030321 030322 030323 030324 030325 030326 030327 030328 030329 030330 030331 030332 030333 030334 030335 030336 030337 030338 030339 030340 030341 030342 030343 030344 030345 030346 030347 030348 030349 030350 030351 030352 030353 030398 030399 030400 030401 030402 030403 030404 030405 030406 030412 030413 030414 030415 030416 030417 030419 030420 030421 030422 030423 030424 030425 030426 030427 030428 030429 030430 030431 030432 030433 030434 030435 030436 030437 030438 030476 030504 030505 030506 030507 030520 030521 030522 030524 030526 030528 030529 030530 030531 030532 030533 030534 030535 030536 030537 030538 030539 030540 030541 030542 030543 030544 030545 030549 030552 030553 030554 030555 030556 030580 030643 030644 030645 030766 030767 030926 030927 030928 030929 030930 030931 030932 030933 030934 030935 030936 030937 030938 030939 030940 030941 030942 030943 030944 030945 030946 030947 030948 030949 030950 030951 030952 030953 030954 030955 030956 030960 030970 030971 030972 030973 030974 030975 030976 030977 030978 030979 030980 030981 030982 030983 030984 030985 030986 030987 030988 030989 030990 030991 030992 030993 030994 030995 030996 030997 030998 030999 031000 031001 031002 031003 031004 031005 031006 031007 031008 031009 031010 031011 031012 031013 031014 031033 031034 031035 031036 031037 031038 031039 031040 031041 031042 031043 031044 031045 031046 031047 031048 031049 031050 031051 031052 031053 031054 031055 031056 031057 031058 031059 031060 031061 031062 031063 031064 031065 031066 031067 031068 031069 031070 031071 031072 031073 031074 031075 031076 031077 031078 031079 031080 031081 031082 031083 031084 031085 031086 031089 031090 031091 031092 031093 031094 031095 031096 031097 031098 031099 031100 031101 031102 031103 031104 031105 031106 031107 031108 031109 031110 031111 031112 031113 031114 031115 031116 031117 031118 031119 031120 031121 031122 031123 031124 031125 031126 031127 031128 031129 031130 031131 031132 031133 031134 031135 031136 031137 031138 031139 031140 031141 031142 031143 031144 031145 031146 031147 031148 031149 031150 031151 031152 031153 031154 031155 031156 031157 031158 031159 031160 031161 031162 031163 031164 031165 031166 031167 031168 031169 031170 031171 031172 031173 031174 031175 031176 031177 031178 031179 031180 031181 031182 031183 031184 031185 031186 031187 031188 031189 031190 031191 031192 031193 031194 031195 031196 031197 031198 031199 031200 031201 031202 031203 031204 031205 031206 031207 031208 031218 031219 031220 031221 031222 031223 031224 031225 031226 031227 031228 031229 031230 031231 031232 031233 031234 031235 031236 031237 031238 031239 031240 031241 031242 031243 031244 031245 031246 031247 031248 031249 031250 031251 031252 031253 031254 031255 031256 031264 031265 031266 031267 031268 031269 031270 031271 031272 031273 031274 031275 031276 031277 031278 031279 031280 031281 031282 031283 031284 031285 031286 031287 031288 031289 031290 031291 031292 031295 031296 031297 031298 031299 031300 031301 031302 031303 031304 031305 031306 031307 031308 031309 031310 031311 031312 031313 031314 031315 031316 031317 031318 031319 031320 031321 031322 031323 031324 031325 031327 031328 031329 031330 031331 031332 031333 031334 031335 031336 031337 031338 031339 031340 031341 031342 031343 031344 031345 031346 031347 031348 031349 031350 031351 031352 031353 031354 031355 031356 031357 031358 031359 031360 031361 031362 031363 031364 031365 031366 031367 031368 031369 031370 031371 031372 031373 031374 031375 031376 031377 031378 031379 031380 031381 031382 031383 031384 031385 031386 031387 031388 031389 031390 031391 031392 031393 031394 031395 031396 031397 031398 031399 031400 031401 031402 031417 031419 031420 031421 031422 031423 031424 031431 031463 031464 031465 031466 031467 031468 031469 031470 031471 031472 031473 031474 031475 031476 031477 031478 031479 031480 031481 031482 031483 031484 031485 031486 031487 031488 031489 031491 031492 031493 031494 031495 031496 031497 031498 031499 031500 031501 031502 031503 031504 031505 031506 031507 031508 031509 031547 031548 031549 031550 031551 031552 031553 031554 031555 031556 031557 031558 031559 031560 031561 031562 031563 031564 031565 031566 031567 031568 031569 031570 031571 031572 031573 031574 031575 031576 031577 031578 031579 031580 031581 031582 031583 031584 031585 031586 031591 031592 031593 031594 031595 031596 031597 031598 031599 031600 031601 031602 031603 031604 031605 031606 031607 031608 031609 031610 031611 031612 031613 031614 031615 031642 031643 031644 031645 031646 031647 031648 031649 031650 031651 031652 031653 031654 031655 031656 031657 031658 031665 031666 031667 031668 031669 031670 031671 031672 031673 031674 031675 031676 031677 031678 031679 031680 031682 031683 031684 031685 031686 031687 031688 031689 031690 031691 031692 031693 031694 031695 031696 031697 031698 031699 031700 031701 031702 031703 031704 031705 031706 031707 031708 031709 031710 031711 031712 031713 031714 031715 031716 031717 031718 031719 031720 031722 031723 031724 031725 031726 031727 031728 031729 031730 031731 031732 031733 031734 031735 031736 031737 031738 031739 031740 031741 031742 031743 031744 031745 031746 031747 031748 031749 031750 031751 031752 031753 031754 031755 031756 031757 031758 031759 031760 031761 031762 031763 031764 031765 031766 031767 031768 031769 031770 031771 031772 031773 031774 031775 031776 031796 031797 031798 031799 031800 031801 031802 031803 031804 031805 031806 031807 031808 031809 031810 031811 031812 031813 031814 031815 031816 031817 031818 031819 031820 031821 031822 031823 031824 031825 031826 031827 031828 031829 031830 031831 031832 031840 031843 031844 031845 031846 031847 031848 031849 031850 031851 031852 031853 031854 031855 031856 031857 031858 031859 031860 031861 031862 031863 031864 031865 031866 031867 031868 031869 031870 031871 031872 031873 031874 031875 031876 031877 031878 031879 031880 031881 031882 031883 031884 031885 031886 031887 031888 031889 031890 031891 031892 031893 031894 031895 031896 031897 031898 031899 031900 031901 031902 031903 031904 031905 031906 031907 031908 031909 031910 031911 031912 031913 031914 031915 031916 031917 031918 031919 031920 031921 031922 031923 031924 031925 031926 031927 031928 031929 031930 031931 031932 031933 031934 031935 031936 031937 031938 031939 031940 031941 031942 031943 031944 032020 032021 032022 032023 032024 032025 032026 032027 032028 032029 032030 032031 032055 032056 032057 032058 032059 032060 032061 032062 032063 032064 032065 032066 032067 032068 032069 032070 032071 032072 032073 032074 032075 032076 032077 032078 032079 032080 032081 032082 032083 032084 032085 032086 032087 032088 032089 032090 032091 032092 032093 032094 032095 032096
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- CHG Hospital Beds Inc
- Reason for Recall:
- Stryker Medical is initiating a voluntary recall of the Spirit Select and Spirit Plus A-C powered hospital beds due to reports of hi-lo actuators broken at the mount ends, which could cause the bed to unexpectedly lower resulting in patient injury.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Spirit Plus Bed, A-C Powered Hospital Bed
Product Codes/Lot Numbers:
Model Numbers: 5600, Spirit Plus Bed (manufactured between November 1, 2014 January 13, 2016); Serial Numbers: Serial Number 029481 029482 029483 029484 029485 029486 029487 029488 029489 029490 029491 029492 029493 029494 029495 029496 029497 029498 029499 029500 029501 029502 029503 029504 029505 029506 029507 029508 029509 029510 029511 029512 029513 029514 029515 029516 029517 029518 029519 029520 029521 029522 029523 029524 029525 029526 029527 029528 029529 029530 029531 029532 029543 029544 029545 029546 029547 029548 029549 029550 029551 029552 029553 029554 029555 029556 029557 029565 029566 029567 029568 029569 029570 029571 029572 029573 029574 029575 029576 029577 029578 029579 029580 029581 029582 029583 029584 029585 029586 029587 029588 029589 029590 029591 029592 029593 029594 029595 029596 029597 029598 029599 029600 029601 029602 029603 029604 029605 029606 029607 029608 029609 029610 029611 029612 029613 029614 029615 029616 029617 029618 029619 029620 029621 029622 029623 029624 029625 029626 029627 029628 029629 029630 029631 029632 029633 029634 029635 029636 029637 029638 029660 029661 029662 029663 029664 029665 029666 029667 029668 029669 029670 029671 029672 029673 029674 029675 029676 029677 029678 029679 029680 029681 029682 029683 029684 029685 029686 029687 029688 029689 029690 029691 029692 029693 029694 029695 029696 029697 029698 029699 029700 029701 029702 029703 029704 029705 029706 029709 029710 029711 029712 029713 029714 029715 029716 029717 029718 029719 029720 029721 029722 029723 029724 029725 029726 029727 029728 029729 029730 029731 029732 029733 029734 029735 029736 029737 029738 029739 029740 029741 029742 029743 029744 029745 029746 029747 029748 029749 029750 029751 029752 029753 029754 029755 029756 029757 029758 029759 029760 029812 029813 029814 029815 029816 029817 029818 029819 029820 029821 029822 029823 029824 029825 029826 029827 029828 029829 029830 029831 029832 029833 029834 029835 029836 029837 029838 029839 029840 029841 029842 029843 029844 029845 029846 029847 029848 029849 029850 029851 029852 029853 029854 029855 029856 029857 029858 029859 029860 029861 029862 029863 029864 029865 029866 029867 029868 029869 029870 029871 029872 029873 029874 029875 029876 029877 029879 029880 029881 029882 029883 029884 029885 029886 029887 029888 029889 029890 029891 029892 029893 029894 029895 029896 029897 029898 029899 029900 029901 029916 029917 029918 029919 029920 029921 029922 029923 029924 029925 029926 029927 029928 029930 029931 029932 029933 029934 029935 029936 029937 029938 029939 029940 029941 029942 029943 029944 029945 029946 029947 029948 029949 029951 029952 029953 029954 029955 029956 029957 029958 029959 029960 029961 029962 029963 029964 029965 029966 029967 029968 029969 029970 029971 029972 029973 029974 029975 029976 029977 029978 029979 029980 029981 029982 029983 029984 029985 029986 029987 029988 029989 029990 029991 029992 029993 029994 029995 029996 029997 029998 029999 030000 030001 030002 030003 030004 030005 030006 030007 030008 030009 030010 030011 030012 030013 030014 030015 030016 030017 030018 030019 030020 030021 030022 030023 030024 030025 030026 030027 030028 030029 030030 030031 030032 030033 030034 030035 030103 030104 030105 030106 030107 030108 030109 030110 030111 030112 030113 030114 030115 030116 030117 030118 030119 030120 030121 030122 030123 030124 030125 030126 030127 030128 030129 030130 030131 030132 030133 030134 030135 030136 030137 030138 030139 030140 030141 030142 030143 030144 030145 030146 030147 030148 030162 030163 030164 030165 030166 030167 030168 030169 030170 030171 030172 030173 030174 030175 030176 030177 030178 030179 030180 030181 030182 030183 030184 030185 030186 030187 030188 030189 030190 030191 030192 030193 030194 030195 030196 030197 030198 030199 030200 030201 030202 030203 030222 030233 030234 030235 030239 030240 030250 030286 030287 030288 030289 030290 030291 030292 030293 030294 030295 030296 030297 030298 030299 030300 030301 030302 030303 030304 030305 030306 030307 030308 030309 030310 030311 030312 030313 030314 030315 030316 030317 030318 030319 030320 030321 030322 030323 030324 030325 030326 030327 030328 030329 030330 030331 030332 030333 030334 030335 030336 030337 030338 030339 030340 030341 030342 030343 030344 030345 030346 030347 030348 030349 030350 030351 030352 030353 030398 030399 030400 030401 030402 030403 030404 030405 030406 030412 030413 030414 030415 030416 030417 030419 030420 030421 030422 030423 030424 030425 030426 030427 030428 030429 030430 030431 030432 030433 030434 030435 030436 030437 030438 030476 030504 030505 030506 030507 030520 030521 030522 030524 030526 030528 030529 030530 030531 030532 030533 030534 030535 030536 030537 030538 030539 030540 030541 030542 030543 030544 030545 030549 030552 030553 030554 030555 030556 030580 030643 030644 030645 030766 030767 030926 030927 030928 030929 030930 030931 030932 030933 030934 030935 030936 030937 030938 030939 030940 030941 030942 030943 030944 030945 030946 030947 030948 030949 030950 030951 030952 030953 030954 030955 030956 030960 030970 030971 030972 030973 030974 030975 030976 030977 030978 030979 030980 030981 030982 030983 030984 030985 030986 030987 030988 030989 030990 030991 030992 030993 030994 030995 030996 030997 030998 030999 031000 031001 031002 031003 031004 031005 031006 031007 031008 031009 031010 031011 031012 031013 031014 031033 031034 031035 031036 031037 031038 031039 031040 031041 031042 031043 031044 031045 031046 031047 031048 031049 031050 031051 031052 031053 031054 031055 031056 031057 031058 031059 031060 031061 031062 031063 031064 031065 031066 031067 031068 031069 031070 031071 031072 031073 031074 031075 031076 031077 031078 031079 031080 031081 031082 031083 031084 031085 031086 031089 031090 031091 031092 031093 031094 031095 031096 031097 031098 031099 031100 031101 031102 031103 031104 031105 031106 031107 031108 031109 031110 031111 031112 031113 031114 031115 031116 031117 031118 031119 031120 031121 031122 031123 031124 031125 031126 031127 031128 031129 031130 031131 031132 031133 031134 031135 031136 031137 031138 031139 031140 031141 031142 031143 031144 031145 031146 031147 031148 031149 031150 031151 031152 031153 031154 031155 031156 031157 031158 031159 031160 031161 031162 031163 031164 031165 031166 031167 031168 031169 031170 031171 031172 031173 031174 031175 031176 031177 031178 031179 031180 031181 031182 031183 031184 031185 031186 031187 031188 031189 031190 031191 031192 031193 031194 031195 031196 031197 031198 031199 031200 031201 031202 031203 031204 031205 031206 031207 031208 031218 031219 031220 031221 031222 031223 031224 031225 031226 031227 031228 031229 031230 031231 031232 031233 031234 031235 031236 031237 031238 031239 031240 031241 031242 031243 031244 031245 031246 031247 031248 031249 031250 031251 031252 031253 031254 031255 031256 031264 031265 031266 031267 031268 031269 031270 031271 031272 031273 031274 031275 031276 031277 031278 031279 031280 031281 031282 031283 031284 031285 031286 031287 031288 031289 031290 031291 031292 031295 031296 031297 031298 031299 031300 031301 031302 031303 031304 031305 031306 031307 031308 031309 031310 031311 031312 031313 031314 031315 031316 031317 031318 031319 031320 031321 031322 031323 031324 031325 031327 031328 031329 031330 031331 031332 031333 031334 031335 031336 031337 031338 031339 031340 031341 031342 031343 031344 031345 031346 031347 031348 031349 031350 031351 031352 031353 031354 031355 031356 031357 031358 031359 031360 031361 031362 031363 031364 031365 031366 031367 031368 031369 031370 031371 031372 031373 031374 031375 031376 031377 031378 031379 031380 031381 031382 031383 031384 031385 031386 031387 031388 031389 031390 031391 031392 031393 031394 031395 031396 031397 031398 031399 031400 031401 031402 031417 031419 031420 031421 031422 031423 031424 031431 031463 031464 031465 031466 031467 031468 031469 031470 031471 031472 031473 031474 031475 031476 031477 031478 031479 031480 031481 031482 031483 031484 031485 031486 031487 031488 031489 031491 031492 031493 031494 031495 031496 031497 031498 031499 031500 031501 031502 031503 031504 031505 031506 031507 031508 031509 031547 031548 031549 031550 031551 031552 031553 031554 031555 031556 031557 031558 031559 031560 031561 031562 031563 031564 031565 031566 031567 031568 031569 031570 031571 031572 031573 031574 031575 031576 031577 031578 031579 031580 031581 031582 031583 031584 031585 031586 031591 031592 031593 031594 031595 031596 031597 031598 031599 031600 031601 031602 031603 031604 031605 031606 031607 031608 031609 031610 031611 031612 031613 031614 031615 031642 031643 031644 031645 031646 031647 031648 031649 031650 031651 031652 031653 031654 031655 031656 031657 031658 031665 031666 031667 031668 031669 031670 031671 031672 031673 031674 031675 031676 031677 031678 031679 031680 031682 031683 031684 031685 031686 031687 031688 031689 031690 031691 031692 031693 031694 031695 031696 031697 031698 031699 031700 031701 031702 031703 031704 031705 031706 031707 031708 031709 031710 031711 031712 031713 031714 031715 031716 031717 031718 031719 031720 031722 031723 031724 031725 031726 031727 031728 031729 031730 031731 031732 031733 031734 031735 031736 031737 031738 031739 031740 031741 031742 031743 031744 031745 031746 031747 031748 031749 031750 031751 031752 031753 031754 031755 031756 031757 031758 031759 031760 031761 031762 031763 031764 031765 031766 031767 031768 031769 031770 031771 031772 031773 031774 031775 031776 031796 031797 031798 031799 031800 031801 031802 031803 031804 031805 031806 031807 031808 031809 031810 031811 031812 031813 031814 031815 031816 031817 031818 031819 031820 031821 031822 031823 031824 031825 031826 031827 031828 031829 031830 031831 031832 031840 031843 031844 031845 031846 031847 031848 031849 031850 031851 031852 031853 031854 031855 031856 031857 031858 031859 031860 031861 031862 031863 031864 031865 031866 031867 031868 031869 031870 031871 031872 031873 031874 031875 031876 031877 031878 031879 031880 031881 031882 031883 031884 031885 031886 031887 031888 031889 031890 031891 031892 031893 031894 031895 031896 031897 031898 031899 031900 031901 031902 031903 031904 031905 031906 031907 031908 031909 031910 031911 031912 031913 031914 031915 031916 031917 031918 031919 031920 031921 031922 031923 031924 031925 031926 031927 031928 031929 031930 031931 031932 031933 031934 031935 031936 031937 031938 031939 031940 031941 031942 031943 031944 032020 032021 032022 032023 032024 032025 032026 032027 032028 032029 032030 032031 032055 032056 032057 032058 032059 032060 032061 032062 032063 032064 032065 032066 032067 032068 032069 032070 032071 032072 032073 032074 032075 032076 032077 032078 032079 032080 032081 032082 032083 032084 032085 032086 032087 032088 032089 032090 032091 032092 032093 032094 032095 032096
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2331-2016
Related Recalls
Spirit Select Bed, A-C Powered Hospital Bed
CHG Hospital Beds
Stryker Medical is initiating a voluntary recall of the Spirit Select and Spirit Plus A-C powered hospital beds due to reports of hi-lo actuators broken at the mount ends, which could cause the bed to unexpectedly lower resulting in patient injury.
It was identified that in some situations the bed exit alarm may not function as intended. The software code for the bed exit system has the potential to auto-reset erroneously. In some situations, the software code does not allow enough time for the weight value to fall to zero once the patient egresses from the bed, in which case, the bed has the potential to reset the bed exit alarm even without a patient on the bed. The bed exit on the Spirit Select bed may incorrectly reset if the Auto Reset option is selected. A patient fall may result in soft tissue or bone injury which may require medical or surgical intervention to prevent permanent disability. If the bed exit alarm incorrectly resets then medical assistance after a fall or during entanglement may be delayed. Patients with dementia or disorientation are at greatest risk. There have been two reported injuries associated with these issues to date. A Medical Risk Assessment determined that there is a remote probability of occurrence of harm requiring medical intervention.